OUP user menu

Left and right ventricular lead positions are imprecisely determined by fluoroscopy in cardiac resynchronization therapy: a comparison with cardiac computed tomography

Anders Sommer , Mads Brix Kronborg , Bjarne Linde Nørgaard , Christian Gerdes , Peter Thomas Mortensen , Jens Cosedis Nielsen
DOI: http://dx.doi.org/10.1093/europace/euu056 1334-1341 First published online: 31 March 2014

Abstract

Aims Fluoroscopy is the routine method for localizing left ventricular (LV) and right ventricular (RV) lead positions in cardiac resynchronization therapy (CRT). However, the ability of fluoroscopy to determine lead positions in a standard ventricular segmentation is unknown. We aimed to evaluate the accuracy and reproducibility of fluoroscopy to determine LV and RV lead positions in CRT when compared with cardiac computed tomography (CT).

Methods and results Fifty-nine patients undergoing CRT were included. Bi-plane fluoroscopy and cardiac CT were evaluated in all patients. Pacing lead positions were assessed in a standard LV 16-segment model and in a simplistic RV 8-segment model. Four patients with LV lead displacement were excluded from the agreement analysis of LV lead position. Agreement of LV lead position between fluoroscopy and cardiac CT was observed in 19 (35%) patients with fluoroscopy demonstrating a 1-segment and ≥2-segment error in 30 (55%) and 6 (11%) patients, respectively. Agreement of RV lead position was found in 13 (22%) patients with fluoroscopy showing a 1-segment and ≥ 2-segment error in 28 (47%) and 18 (31%) patients, respectively. The interobserver agreement on LV and RV lead positions was poor for fluoroscopy (kappa 0.20 and 0.23, respectively) and excellent for cardiac CT (kappa 0.87 and 0.85, respectively).

Conclusion Fluoroscopy is inaccurate and modestly reproducible when assessing LV and RV lead positions in a standard ventricular segmentation when compared with cardiac CT. Cardiac CT should be applied to determine the exact pacing site in future research evaluating the optimal pacing lead position in CRT.

  • Cardiac resynchronization therapy
  • Fluoroscopy
  • Cardiac computed tomography
  • Left ventricular lead position
  • Right ventricular lead position
View Full Text

Sign in

ESC members Follow the 'ESC Member and Congress Delegate Sign In' link below for free online access if your subscription to this journal is via the European Society of Cardiology, either as a member or an ESC Congress delegate. Discover if you are an ESC member here.

Otherwise, if your subscription is via OUP, enter your OUP username and password, or select an alternative sign in option below.

Log in through your institution