Skip Navigation



Europace Advance Access published online on July 9, 2008

Europace, doi:10.1093/europace/eun176
This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
10/10/1238    most recent
eun176v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Hayashi, M.
Right arrow Articles by Leenhardt, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hayashi, M.
Right arrow Articles by Leenhardt, A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org


CASE REPORT

Short–long–short sequence caused by ventricular safety pacing inducing ventricular tachycardia in a patient with a dual-chamber implantable cardioverter defibrillator

Meiso Hayashi, Seiji Takatsuki, Anne Messali, Paul Milliez, Fabrice Extramiana and Antoine Leenhardt*

Cardiology Department, Lariboisière Hospital, Paris 7 University, 2 Rue Ambroise Paré, 75475 Paris Cedex 10, France

Ventricular safety pacing (VSP) is an algorithm used to prevent crosstalk inhibition and ventricular capture during the vulnerable period. We report a 78-year-old man with implantable dual-chamber defibrillator, in whom clusters of ventricular tachycardias (VTs) were provoked by the VSP. During rapid DDDR pacing, the delivery of the VSP after every other atrial-paced beat resulted in short–long–short ventricular sequences and induced VTs. An atrial-based lower rate timing, long atrioventricular pacing interval, and automatic gain control also accounted for this arrhythmogenic ventricular sequence. The VSP and the subsequent VT were eliminated by decreasing the pacing rate.


* Corresponding author. Tel: +33 1 49 95 82 23; fax: +33 1 49 95 82 06. E-mail address: antoine.leenhardt{at}lrb.aphp.fr

Manuscript submitted 25 March 2008. Accepted after revision 11 June 2008.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.