Skip Navigation



Europace Advance Access published online on October 11, 2007

Europace, doi:10.1093/europace/eum225
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
9/12/1177    most recent
eum225v2
eum225v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Biffi, M.
Right arrow Articles by Giuseppe, B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Biffi, M.
Right arrow Articles by Giuseppe, B.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2007. For permissions please email: journals.permissions@oxfordjournals.org

Left ventricular pacing by automatic capture verification

Mauro Biffi*, Matteo Bertini, Matteo Ziacchi and Boriani Giuseppe

Institute of Cardiology, University of Bologna, Via Massarenti 9, 40138 Bologna, Italy

Aims: To investigate the feasibility of transvenous left ventricular (LV) pacing by autocapture at long term. A reliable measurement of LV evoked response (ER) is the pivotal requirement for beat-to-beat detection of ventricular capture and automatic output adjustment.

Methods and results: Seven patients with accepted class I indication to permanent cardiac pacing received a DDDR pacemaker with automatic output adjustment based on beat-to-beat capture verification (Insignia Ultra 1290, Guidant), whose ventricular port was connected to a LV lead placed in a branch of the coronary sinus. The device allows LV threshold trending, performing a threshold test every 21 h, and diagnoses acute and non-acute issues of ER detection during follow up. Average follow up after implantation was 34 ± 6 months (range 28–45, median 34). Left ventricular pacing threshold showed an increase from implantation (1.2 ± 0.4 V at 0.4 ms) that peaked at week 4 (1.6 ± 0.7 V at 0.4 ms), and returned to baseline (1.1 ± 0.5 at 0.4 ms) by the end of the 7th week. Autocapture performance at long term, as assessed by the trend of LV threshold and of ER diagnostic issues, did not show any pitfall.

Conclusions: Our observations support the feasibility and safety of capture verification during LV pacing alone. A possible application of this pacing technology could be biventricular stimulation.

Key Words: Left ventricular pacing, Automatic capture verification, Pacing threshold, Evoked response


* Corresponding author. Tel: +39 051 345898; fax: +39 051 344859.E-mail address: mauro.biffi{at}aosp.bo.it, mbiffi{at}aosp.bo.it

Manuscript submitted 7 May 2007. Accepted after revision 18 September 2007.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer:
Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.