Europace, doi:10.1093/europace/eum123
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Do small (6.6 Fr.) active and passive fixation defibrillation leads perform as well as larger sized leads? a multi-centre analysis
1 Schüchtermann Klinik, Bad Rothenfelde, Germany; 2 London Health Sciences Centre, London, Ontario, Canada; 3 Hôpital Laval, Sainte-Foy, Québec, Canada; 4 Kingston General Hospital, Kingston, Ontario, Canada; 5 Centro Cardiologico Monzino, Milan, Italy; 6 Medtronic, Bakken Research Center BV, Maastricht, The, Netherlands; 7 Medtronic Inc., Minneapolis, MN, USA; 8 St. Johannes Hospital, Dortmund, Germany
Aims The 6.6 Fr. Sprint Fidelis® lead family may allow multiple lead implantation procedures with reduced risk of venous obstruction.
Methods and results Two prospective, historically controlled, multi-centre studies were conducted in Europe (80 patients) and Canada (79 patients). The purpose was to assess the ventricular lead-related adverse events (LRAEs) and performance of the small Models 6948 and 6949 defibrillation leads, respectively, in patients with a standard indication for an ICD implant. Safety was assessed by demonstrating equivalence of the LRAE free rate at 1 month to comparable but larger leads (Models 6942, 6943, 6944, 6947and 4074). Seventy-five of 80 patients with a 6948 lead (93.8%) remained free of LRAEs. Seventy-four out of 79 patients (93.7%) with the 6949 lead remained free of LRAEs. The 95% lower confidence bounds were above the critical difference limits. Thus, safety of the Sprint Fidelis® leads is similar to that of larger leads. Electrical performance through 1-month follow-up proved to be acceptable in comparison with other established leads.
Conclusion These multi-centre studies confirm that smaller defibrillation leads offer similar safety and efficacy features to widely used larger leads; they have low LRAE rates and defibrillation thresholds, while providing the advantage of a smaller introducer size and reduced venous obstruction.
Key Words: Defibrillation lead, Dual coil, ICD, Lead handling, Safety, Electrical performance
* Corresponding author: Abteilung Kardiologie/Angiologie, Elisabeth-Krankenhaus, Klara-Kopp-Weg 1, D-45138 Essen, Germany. Tel: +49 201 897 3201; fax: +49 201 288 525. E-mail address: b.kuepper{at}elisabeth-essen.de
Manuscript submitted 12 December 2006. Accepted after revision 24 May 2007.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
C. J. W. Borleffs, L. van Erven, R. J. van Bommel, E. T. van der Velde, E. E. van der Wall, J. J. Bax, F. R. Rosendaal, and M. J. Schalij Risk of Failure of Transvenous Implantable Cardioverter-Defibrillator Leads Circ Arrhythm Electrophysiol, August 1, 2009; 2(4): 411 - 416. [Abstract] [Full Text] [PDF] |
||||