* To whom correspondence should be addressed. In this case report, we present a 57-year-old patient who suffered from incessant slow ventricular tachycardia because of a change in therapy 2 weeks after the implantation of an implantable cardioverter defibrillator (ICD) with Home Monitoring (HM). HM is a fully automatic and wireless system capable of transmitting device data including episode counters. The patient had mild symptoms with only short episodes of dizziness and would not have contacted the primary care centre. Due to HM, the arrhythmia was detected before the next routine follow-up, and the patient was seen the next day. HM made it possible to adjust the dosage of the newly established beta-blocker treatment in an outpatient setting and avoid hospitalization. HM yields the potential for remote detection of arrhythmias in ICD recipients, allowing alterations in device programming or medication.
Received October 25, 2005
Accepted March 18, 2006
Article
Remote detection of incessant slow VT with an ICD capable of home monitoring
Stephanos Siaplaouras 1 *,
Axel Buob 1,
Hans Ruprecht Neuberger 1,
and
Christian Mewis 1
Stephanos Siaplaouras, E-mail: siaplaouras{at}aol.com
![]()
Abstract ![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
G. Andrikopoulos, S. Tzeis, G. Theodorakis, and P. Vardas Monitoring capabilities of cardiac rhythm management devices Europace, October 29, 2009; (2009) eup317v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Masella, P. Zanaboni, F. Di Stasi, S. Gilardi, P. Ponzi, and S. Valsecchi Assessment of a remote monitoring system for implantable cardioverter defibrillators J Telemed Telecare, September 1, 2008; 14(6): 290 - 294. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Heidbuchel, P. Lioen, S. Foulon, W. Huybrechts, J. Ector, R. Willems, and H. Ector Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator Europace, March 1, 2008; 10(3): 351 - 357. [Abstract] [Full Text] [PDF] |
||||