Skip Navigation


Europace Advance Access originally published online on June 5, 2007
Europace 2007 9(9):711-716; doi:10.1093/europace/eum109
This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
9/9/711    most recent
eum109v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Irnich, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Irnich, W.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The European Society of Cardiology 2007. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org


IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS

Parameters characterizing implantable defibrillator output: a proposal

Werner Irnich

1 Justus-Liebig-University, University Hospital, Friedrichstr. 18, 35392 Giessen, Germany

Aims Recently, a discussion was carried out in Heart Rhythm on the specifications that could characterize implantable defibrillators. It is the intention of this paper to participate in this discussion on defibrillation characteristics and to give recommendations on how this problem could be solved.

Theoretical considerations and results There are different defibrillation theories, all finding that the defibrillation's efficacy depends on the time constant RC which is output capacitance C times load resistance R. Efficacy decreases with increasing RC. This means that (i) the knowledge of C is of paramount importance, (ii) the energy is ‘devalued’ with increasing RC and that those parameter settings such as tilt or pulse duration should be adjusted to the time constant, and (iii) the energy values given without further specification are not meaningful. As there is always a voltage drop across an internal resistance within the ICD, the measured voltage across the output differs from the capacitor voltage and is reduced which determines the efficiency of the device. From the data given by Thammanomai et al., one can determine the parameters maximum voltage, capacitance, internal resistance, and tilt. These parameters are adequate and necessary to describe an ICD device and to derive the effective energy for device comparison.

Discussion The ‘high output devices’ with their high nominal energy are reduced in their effective energies to a degree that they are comparable to the best ‘standard output devices’. They do not offer that superiority which is promised by the nominal energy. Moreover, if the tilt is fixed and larger than optimal, the energy requirements are still higher or the effective energy will further drop. The term ‘delivered energy’ is not used by us because the delivered energy increases with increasing tilt. However, today's tilts are too large as judged by theories, which means that high delivered energies can be worse than lower ones. The delivered energy is, therefore, not a meaningful parameter in judging ICDs.

Conclusion ICD devices should be characterized by: (i) voltage, (ii) capacitance, (iii) tilt or pulse duration (if not programmable), and (iv) internal resistance. All other parameters can be derived from them by simple calculations. Introduction of a ‘devaluation factor’ characterizes the decreasing efficacy with increasing time constant and renders the output characteristics transparent and comparable.

Key Words: Implantable defibrillator, Output specifications, Defibrillation theory, Technical efficiency, Physiological efficacy, Devaluation factor, Optimal truncation


Corresponding author. Tel: +49 641 99 41390; fax: +49 641 99 41399. E-mail address: werner{at}irni.ch or irnich{at}technik.med.uni-giessen.de

Manuscript submitted 28 February 2007. Accepted after revision 27 April 2007.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.