Skip Navigation


Europace Advance Access originally published online on March 23, 2006
Europace 2006 8(5):377-384; doi:10.1093/europace/eul014
This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
8/5/377    most recent
eul014v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Irnich, W.
Right arrow Articles by Bernstein, A. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Irnich, W.
Right arrow Articles by Bernstein, A. D.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The European Society of Cardiology 2006. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org


PACING

Do induction cooktops interfere with cardiac pacemakers?

Werner Irnich1,* and Alan D. Bernstein2

1 University Hospital, Friedrichstr. 18, 35392 Giessen, Germany; 2 CardioScript International, Salt Lake City, UT, USA

Aims In induction cooktops, coils produce time-varying magnetic fields that induce eddy currents in the ferromagnetic bottom of a pot or pan, thereby heating it, while the cooktop itself remains cool. Interference with pacemaker sensing could conceivably be produced by voltages induced directly by induction or indirectly by leakage currents.

Methods and results A worst-case pacemaker-patient (PP) model representing left-sided implantation of a unipolar pacemaker was used for measurement of induced voltages, to judge whether induction cooktops could interfere with pacemaker sensing. Eleven induction cooktops of European manufacture were tested using the PP model. The pacemaker sensitivity with respect to 24 kHz voltages, amplitude-modulated at 100 Hz, was investigated in 244 devices. The current passing through the body of a grounded patient touching a metal pot was determined by measuring the voltage from hand to hand and between electrodes placed on the thorax to simulate an implanted unipolar pacing system underneath. The results obtained were complex. If the pot is positioned concentrically with the induction coil, the smallest pot produced the largest stray field, but the induced voltage always remained below the critical value of 100 mV. With eccentrically positioned large pots, voltages of up to 800 mV could be induced. The induced voltage could always be reduced to ≤60 mV by maintaining a distance of 35 cm. The most sensitive pacemaker reacted at 90.5 mV. Because of leakage current, ~2% of the voltage between pot and ground appears across the pacemaker's sensing input.

Conclusion Patients are at risk if the implant is unipolar and left-sided, if they stand as close as possible to the induction cooktop, and if the pot is not concentric with the induction coil. Unipolar pacing systems can sense interference generated by leakage currents if the patient touches the pot for a long period of time. The most likely response to interference is switching to an asynchronous interference mode. Patients with unipolar pacemakers are at risk only if they are not pacemaker-dependent.

Key Words: Cardiac pacemaker, Induction cooktop, Electromagnetic interference, Induced voltage, Leakage current


* Corresponding author. E-mail address: werner{at}irni.ch


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.