Long-term results of endocardial pacing with AutocaptureTM threshold tracking pacemakers in children
lucaHacettepe University School of Medicine Ankara, Turkey; bAtatürk University School of Medicine Erzurum, Turkey; cÇukurova University School of Medicine Adana, Turkey
AIM: We aimed to evaluate the long-term results of endocardial pacing with AutocaptureTM threshold tracking pacemakers in children.
METHODS AND RESULTS: Implantation and follow-up data of 20 children with these pacemakers were retrospectively evaluated. The pacemakers were implanted subpectorally in five and subcutaneously in 15 patients. The indication for pacing was high-grade atrioventricular block in 18 cases. The mean age at implantation was 7 ± 4.8 years. Four patients were pacemaker dependant (heart rate < 30 bpm). At implantation, the mean pacing threshold was 0.5 V at 0.5 ms. The mean evoked response (ER) signal was 8.5 ± 3.6 mV, and the polarisation signal (PS) was <1 mV in 15 patients and 12 mV in five patients. During the mean follow-up period of 60 months, mean ER signal decreased significantly to 7.7 ± 6.3 mV at 24 months and 6.5 ± 2.5 mV at 60 months (P < 0.05). In four of 15 patients (26.6%), with a predischarge PS value of <1 mV, it increased between 1 and 2 mV over time. During follow-up, autocapture function was deactivated in six (30%) patients; due to inappropriate ER/PS values in four and due to severe muscle twitching in two with subpectoral implants. These problems occurred during a median period of 21 months after implantation. Generators were replaced in three patients with Microny pacemakers because of battery depletion at 54, 66 and 78 months. In two of them autocapture function had been working since implantation. In seven of 10 patients, who completed
60 months of follow-up, battery impedances were still at the predischarge level.
CONCLUSIONS: Autocapture function works well in most children at implantation. Mean ER signal significantly decreases over time despite stable pacing parameters. Autocapture function may become nonoperational due to decreased ER signal in some patients. Muscle twitching may be an important problem that may result in discontinuation of autocapture function in children with subpectoral implants.
Key Words: pacemakers, AutocaptureTM, children, transvenous application
*Corresponding author. Atatürk Üniversitesi Lojmanlari 59. Blok, Kat 4, No 20, 25170 Erzurum, Turkey. Tel.: +90 442 2361212/1064/1209; fax: +90 442 2361301. E-mail address: nceviz{at}atauni.edu.tr
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
M. S. Silvetti, A. De Santis, S. Marcora, T. De Santo, N. Grovale, and F. Drago Circadian pattern of atrial pacing threshold in the young Europace, February 1, 2008; 10(2): 147 - 150. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Tomaske, P. Harpes, R. Pretre, A. Dodge-Khatami, and U. Bauersfeld Long-term experience with AutoCapture(R)-controlled epicardial pacing in children Europace, August 1, 2007; 9(8): 645 - 650. [Abstract] [Full Text] [PDF] |
||||