Skip Navigation

Europace 2003 5(1):83-89; doi:10.1053/eupc.2002.0279
© 2003 by European Society of Cardiology
This Article
Right arrow FREE Full Text (PDF) Freely available
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (5)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Sperzel, J.
Right arrow Articles by Neuzner, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sperzel, J.
Right arrow Articles by Neuzner, J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Automatic capture verification in ICD lead systems using intracardiac ventricular evoked response and reduced coupling capacitance

J. Sperzel1, H. F. Pitschner1, T. Schwarz2, A. König3, Q. Zhu4 and J. Neuzner1

1Kerckhoff-Klinik, Department of Cardiology and Electrophysiology Benekestrasse 2-8, 61231, Bad Nauheim, Germany; 2Department of Heart Surgery Benekestrasse 2-8, 61231, Bad Nauheim, Germany; 3Guidant Europe NV/SA, Research Department Park Lane, Culliganlaan 2B, 1831, Diegem, Belgium; 4Guidant, Research and Development Mailstop E 211, 4100 Hamline Avenue North, 55112-5798, St Paul, MN, U.S.A.

AIM: Intracardiac ventricular evoked-response (ER) signals detected by implanted cardioverter defibrillator (ICD) lead systems were investigated for automatic capture verification (AC).

METHOD: ER signals were evaluated with an external pacing system equipped with a reduced coupling capacitance (CC=2·2 #x03BC;F) in the pacing output circuit during ventricular step-down threshold testing at 0·4 ms pulse duration. Real-time pacing markers, surface ECG and intracardiac electrograms pre- and post-filtering were digitally recorded.

RESULTS: Twenty consecutive patients, age 61±12 years, with leads from two different manufacturers were tested — 10 were implanted with acute leads (AL) and 10 with chronic leads (CL). The analysis was based on the ER amplitude during capture and on the ER-to-afterpotential ratio (SAR), with SAR>2 as the criterion for successful capture detection. ER amplitudes (median and range) were 8·1 mV (2·1–19·5 mV) for AL and 8·3 mV (3·7–14·2 mV) for CL. SAR values (median and range) were 48·0 (2·5–682·6) for AL and 13·2 (6·3–35·9) for CL, indicating that AC could successfully be applied in all patients.

CONCLUSIONS: Reducing the pacing CC allows adequate ER detection for automatic capture verification on non-selected ICD lead systems. The effect of high-voltage shock treatment on ER signal detection requires further investigation.

Key Words: ICD, ventricular autocapture, evoked response, afterpotential


Correspondence: Dr med. Johannes Sperzel, Kerckhoff-Klink, Department of Cardiology and Electrophysiology, Benekestraße 2-8, 61231 Bad-Nauheim, Germany. Tel.: +49 6032 9960; Fax: +49 6032 996 2236; E-mail:Johannes.Sperzel{at}kerckhoff.med.uni-giessen.de


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
EuropaceHome page
J. Sperzel, B. Nowak, E. Himmrich, G. Zhang, A. König, R. Willems, C. Reister, A. Sathaye, and G. Fröhlig
Acute performance evaluation of a new ventricular automatic capture algorithm.
Europace, January 1, 2006; 8(1): 65 - 69.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
M. Biffi, G. Boriani, M. Bertini, P. Silvestri, C. Martignani, and A. Branzi
Pacing with capture verification in candidates for resynchronisation therapy: A feasibility study
Europace, January 1, 2005; 7(3): 255 - 265.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.