Skip Navigation


Europace Advance Access originally published online on December 24, 2008
Europace 2009 11(2):206-212; doi:10.1093/europace/eun363
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
11/2/206    most recent
eun363v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Sterlinski, M.
Right arrow Articles by Szwed, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sterlinski, M.
Right arrow Articles by Szwed, H.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org


Leads and Lead Extraction

Subacute cardiac perforations associated with active fixation leads

Maciej Sterlinski, Andrzej Przybylski*, Aleksander Maciag, Pawel Syska, Mariusz Pytkowski, Michal Lewandowski, Ilona Kowalik, Bohdan Firek, Piotr Kolsut, Grzegorz Religa, Mariusz Kusmierczyk, Franciszek Walczak and Hanna Szwed

Institute of Cardiology, Spartanska 1, 02-637 Warsaw, Poland

Aims: Having several recently published reports on increased rate of cardiac perforation with some lead models as background, we assess the relation between cardiac perforations and models of leads used.

Methods and results: All pacing and defibrillation leads implantations between 1 January 2007 and 31 March 2008 were analysed retrospectively. There were 2247 leads implanted in 1419 patients aged 67.6 ± 14.1, 1200 (53%) active and 1047 (47%) passive fixation leads. Cardiac perforation occurred in eight patients (0.5%). The number of perforations does not differ significantly between the pacemaker and implantable cardioverter defibrillator implantations (five and three cases, respectively, P = 0.13). All perforations were associated with the active fixation leads implantation (8 vs. 0, P < 0.01). Only four models of leads were associated with perforations, but the risk of their use was not statistically significantly increased, when compared with other active fixation leads placed in the adequate position.

Conclusions: The incidence of cardiac perforation related to pacing and defibrillation leads is low. The use of active fixation leads is associated with an increased risk of cardiac perforation. We did not find any correlation between the perforation rate and any particular model of the implanted lead.

Key Words: Cardiac perforation, Transvenous leads, Pacemaker, Implantable cardioverter-defibrillator, Complications


* Corresponding author. Tel: +48 223434050, Fax: +48 228449510, Email: aprzybylski{at}ikard.pl or aprzybylski{at}adres.pl

Manuscript submitted 17 June 2008. Accepted after revision 2 December 2008.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.