Skip Navigation


Europace Advance Access originally published online on January 12, 2008
Europace 2008 10(3):364-365; doi:10.1093/europace/eum292
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
10/3/364    most recent
eum292v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Osman, F.
Right arrow Articles by Morley-Davies, A.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Osman, F.
Right arrow Articles by Morley-Davies, A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org


ICDS

Twiddler's syndrome in a patient with a biventricular-defibrillator device

Faizel Osman*, Ross Ward and Adrian Morley-Davies

Department of Cardiology, City General Hospital, University Hospital of North Staffordshire NHS Trust, Newcastle Road, Stoke on Trent ST4 6QG, UK

A 69-year-old man with previous myocardial infarction and proximal three-vessel coronary artery disease underwent coronary bypass grafting, with an epicardial lead placed on the lateral left ventricular wall during surgery. A cardiac resynchronization therapy-defibrillator (CRT-D) device was subsequently implanted using active right atrial and right ventricular leads, with the pulse generator placed in a pre-pectoral pocket. Four weeks later, the right atrial lead was failing to sense or capture, and chest X-ray revealed it had pulled out of the myocardium and coiled up behind the device; a diagnosis of Twiddler's syndrome was made. Twiddler's syndrome is unusual in patients with CRT-D devices and may cause symptoms such as inappropriate shocks and hiccups. Placement of the pulse generator in a sub-pectoral position may help prevent it.

Key Words: Twiddler's syndrome, Cardiac resynchronization therapy, Implantable defibrillator


* Corresponding author. Tel: +44 1782 55305. E-mail address: f.osman{at}bham.ac.uk

Manuscript submitted 1 November 2007. Accepted after revision 10 December 2007.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer:
Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.