Skip Navigation


Europace Advance Access originally published online on February 1, 2008
Europace 2008 10(3):351-357; doi:10.1093/europace/eun010
This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrowOA All Versions of this Article:
10/3/351    most recent
eun010v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Google Scholar
Right arrow Articles by Heidbüchel, H.
Right arrow Articles by Ector, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Heidbüchel, H.
Right arrow Articles by Ector, H.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org
The online version of this article has been published under an open access model. Users are entitled to use, reproduce, disseminate, or display the open access version of this article for non-commercial purposes provided that the original authorship is properly and fully attributed; the Journal, Learned Society and Oxford University Press are attributed as the original place of publication with correct citation details given; if an article is subsequently reproduced or disseminated not in its entirety but only in part or as a derivative work this must be clearly indicated. For commercial re-use, please contact journals.permissions©oxfordjournals.org.


ICDS

Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator{dagger}

Hein Heidbüchel*, Pieter Lioen, Stefaan Foulon, Wim Huybrechts, Joris Ector, Rik Willems and Hugo Ector

Department of Cardiology–Electrophysiology, University Hospital Gasthuisberg, University of Leuven, Herestraat 49, B-3000 Leuven, Belgium

Aims: Follow-up of implantable cardioverter defibrillator (ICD) patients, with regular in-office visits every 3–6 months, puts a significant burden on specialized electrophysiology clinics. New technology allows for remote monitoring of device function. We wanted to investigate its potential reliability and to which extent its use can reduce in-office visits.

Methods and results: We retrospectively analysed data from 1739 prospectively coded ICD visits in a random set of 169 patients (followed between 2 month and 10.4 year in an academic centre). We defined (i) whether the visit was planned or not, (ii) what were the reasons for unplanned visits, (iii) whether any relevant finding was made, (iv) whether a remote monitoring system with the ability or not to detect pacing threshold would have been able to capture the problem, and (v) what actions were taken. The standard follow-up scheme consisted of in-office visits 1 month after implantation and then every 6 months, unless approaching battery depletion. From the 1739 visits, 1530 were performed according to clinical schedule (88%) and in 1197 of those (78.2%), no relevant finding was made. In 0.52% (n = 8) early post-implant pacing threshold increases would not have been detected by remote monitoring without the ability to determine thresholds (although two patients showed a high impedance). Moreover, in 6% of the planned visits, reprogramming would require a consecutive in-office visit (4%) or hospitalization (2%). Only 175 visits (9.6% of all) were conducted prior to the planned follow-up date due to patient symptoms [another 42 (2.4%) were due to planned surgery or safety warnings]. The proportion of relevant findings during unscheduled visits was significantly higher than during scheduled visits (80.6 vs. 21.8%; P < 0.0001) and a higher proportion of those was arrhythmia- and/or device-related (85.1 vs. 55.3%, P < 0.0001). Reprogramming was required more often (33.1 vs. 4%; P < 0.0001) and hospitalization rate was higher (18.3 vs. 2%; P < 0.0001), so that 51.4% of unscheduled visits would require in-office evaluation. Overall, remote follow-up would correctly exclude device function abnormalities or arrhythmic problems in 1402 (82.2%), identify an arrhythmic problem in 262 (15.3%), correctly identify a device-related problem in 35 (2.1%), but potentially miss an isolated pacing problem in 6 (0.46%). Clinical evaluation would diagnose a relevant clinical problem in the absence of any device interrogation abnormality in 170 patients (10%).

Conclusion: ICD remote monitoring can potentially diagnose >99.5% of arrhythmia- or device-related problems if combined with clinical follow-up by the local general practitioner and/or referring cardiologist. It may provide a way to significantly reduce in-office follow-up visits that are a burden for both hospitals and patients.

Key Words: Telemedicine, Remote visit, Implantable defibrillators, Pacemaker, Telemetry


* Corresponding author. Tel: +32 16 34 42 48; fax: +32 16 34 42 40.E-mail address: hein.heidbuchel{at}uz.kuleuven.ac.be

{dagger} Presented in part at the Europace meeting, Lisbon, June 2007.

Manuscript submitted 29 August 2007. Accepted after revision 3 January 2008.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer:
Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.