Skip Navigation


Europace Advance Access originally published online on January 20, 2008
Europace 2008 10(3):334-335; doi:10.1093/europace/eum299
This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
10/3/334    most recent
eum299v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Di Pino, A.
Right arrow Articles by Bianca, I.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Di Pino, A.
Right arrow Articles by Bianca, I.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org


PACING

Efficacy of closed-loop stimulation with epicardial leads in an infant with congenital atrioventricular block

Alfredo Di Pino*, Salvatore Agati and Innocenzo Bianca

Paediatric Cardiology and Cardiac Surgery Unit, S. Vincenzo Hospital, Contrada Sirina, 98039 Taormina, Messina, Italy

We present the case of an infant with a congenital atrioventricular block and sinus node dysfunction that was implanted with a pacemaker with physiological closed-loop stimulation (DDD-CLS) pacing mode with two epicardial leads. We have observed the response of DDD-CLS system with respect to conventional rate responsive accelerometer-based mode (DDDR): the atrial pacing percentage and the 24 h rate trend showed higher values during DDD-CLS compared with DDDR. In this patient, rate responsive feature seems preserved in DDD-CLS also using epicardial leads.

Key Words: Epicardial, Closed-loop stimulations, Infant


* Corresponding author. Tel: +39 0942579550. E-mail address: dipinoa{at}yahoo.it

Manuscript submitted 19 October 2007. Accepted after revision 19 December 2007.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.