Skip Navigation


Europace Advance Access originally published online on September 22, 2008
Europace 2008 10(12):1442-1444; doi:10.1093/europace/eun266
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
10/12/1442    most recent
eun266v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in Europace
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Gutleben, K.-J.
Right arrow Articles by Brachmann, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gutleben, K.-J.
Right arrow Articles by Brachmann, J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org


SHORT COMMUNICATIONS

Rescue-stenting of an occluded lateral coronary sinus branch for recanalization after dissection during cardiac resynchronization device implantation

Klaus-Jürgen Gutleben*, Georg Nölker, Harald Marschang, Anil M. Sinha, Martin Schmidt, Guido Ritscher and Johannes Brachmann

Department of Cardiology, Medizinische Klinik, Klinikum Coburg, Ketschendorfer Straße 33, 96450 Coburg, Germany

Cardiac resynchronization therapy (CRT) using left- (LV) or biventricular pacing is widely applied in selected heart failure patients. However, transvenous LV-lead placement into coronary sinus (CS) branches can be challenging. A 77-year-old female patient with New York Heart Association class III symptoms due to dilated cardiomyopathy [LV ejection fraction (LVEF): 10%, QRS-duration: 150 ms], despite optimal medical treatment presented for CRT. Coronary sinus angiograms were performed after transvenous CS cannulation. Within a large posterolateral vein, low phrenic nerve stimulation thresholds were found. The only alternative smaller tortuous lateral branch showed a significant narrowing, making LV-lead advancement impossible. Angioplasty was performed, using a venoplasty balloon. This caused complete branch occlusion. After recanalization of the vessel by implantation of a bare metal stent, the lead could be advanced through the stent. Optimal pacing parameters without phrenic nerve stimulation were established. Angioplasty of CS branches during CRT implantation procedures bears the risk of complete branch occlusion, but recanalization can acutely be achieved by stent implantation. This is the first report on rescue-stenting of a CS branch after angioplasty-related occlusion. Transthoracic lead implantation, accompanied risks, and slower recovery could thus be avoided.

Key Words: Resynchronization therapy, Left ventricular lead implantation, Stenting, Coronary sinus occlusion, Heart failure


* Corresponding author: Tel: +49 9561 22 33211; fax: +49 9561 22 6349. E-mail address: gutleben{at}gmx.de

Manuscript submitted 3 August 2008. Accepted after revision 3 September 2008.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?

Related articles in Europace:

Stenting of coronary veins: a critical comment
Bert Hansky
Europace 2008 10: 1363. [Full Text]  





Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.