PACING
Computationally inexpensive methods for intra-cardiac atrial bipolar electrogram compression
1Division of Cardiology, San Martino Hospital Largo Rosanna Benzi 10, 16132, Genoa, Italy; 2Institute of Electronic Circuits, National Research Council Via De Marini 6, 16149, Genoa, Italy; 3Vitatron Medical Italia Via Dell'Arcoveggio 49/5, 40129, Bologna, Italy
Abstract
AIM: This paper reports studies of mathematical algorithms for intra-cardiac atrial bipolar electrogram compression suitable with implementation on implantable devices.
PATIENTS AND METHODS: Bipolar intra-cardiac electrograms (IEGMs) of high right atrium were obtained from 20 patients who underwent electrophysiological studies for arrhythmias. Four thousand seven hundred and eighty-two seconds of IEGM were collected and divided into three rhythm groups: sinus rhythm (SR), atrial fibrillation (AF) and atrial flutter (AFL).
Since mathematical algorithms suitable for use with implantable devices demand low computational cost, we employed piecemeal linear approximation methods (ZOP Zero Order Prediction and SAPA Scan Along Polygonal Approximation), and beat detection method (Peak) both or which need small numbers of operations to perform electrogram compression. Compression ratio (CR) and percent root mean square difference (PRD) were used to compare the three methods, with statistical analyses performed using paired t-test.
RESULTS AND CONCLUSION: The best performance was obtained using the Peak method which reaches an average CR of 10·6 in the case of SR group, 2·8 for AF, and 3·6 for AFL groups, respectively, while PRD lies below 2% for SR and AFL groups and 6% for the AF group. Results show that, for bipolar electrograms, the Peak method reaches statistically significant better performance (P< 0·001) in all cases except for Peak vs SAPA applied to AF (P=0·2). The number of operations necessary to compress the data indicate that time consumption can be reduced to be suitable for real time compression in implantable devices. The Peak method, which was assumed to receive the instant of occurrence of each recognized beat, from the hardware of the device, requires fewer operations than ZOP and SAPA.
Increasing the length of electrograms recorded in pacemakers will enhance the amount of information provided by the implantable device, allowing more detailed characterization of the intra-cardiac activity and leading to new perspectives in arrhythmia diagnosis and therapy. Copyright 2002 The European Society of Cardiology. Published by Elsevier Science Ltd. All rights reserved.
Key Words: Intra-cardiac bipolar electrograms, data compression, pacemakers, mathematical algorithms
Correspondence: Dr Paolo Rossi, Cardiology Division, San Martino Hospital, Largo Rosanna Benzi 10, 16132, Genoa, Italy. Tel: +39 010 555 3448/3453; Fax: +39 010 555 6846; E-mail: paolo.rossi{at}Hsanmartino.liguria.it
References
[1] Warren JA and Nelson JP. Pacemaker and ICD pulse generator circuitry. In Ellenbogen KA, Kay NG, Wilkoff BL (Eds.). Clinical Cardiac Pacing and Defibrillation 2000; pp. 194216.
[2] Moya, A, Syncope of unknown origin after electrophysiologic study and head-up tilt test: how useful is an implantable loop recorder?, Raviele, A, Cardiac Arrhythmias 1999, Italia, Springer-Verlag, 2000, 437, 41.
[3] Saeed M, Links MS, Mahapatra S, Mouded M, Tzeng D, Jung V, Contreras R, Swygman C, Homoud M, Estes NA 3rd, Wang PJ. Analysis of intra-cardiac electrograms showing monomorphic ventricular tachycardia in patients with implantable cardioverter-defibrillators. Am J Cardiol 2000; 85: 580587.[Medline]
[4] Israel CW, Gascon D, Nowak B, Campanale G, Pascotto P, Hartung W, Lellouche D. Pulsar/Selute Picotip VDD Study Group. Diagnostic value of stored electrograms in single-lead VDD systems. Pacing Clin Electrophysiol 2000 Nov; 23: 1801.[Medline]
[5] Kleinert M, Elmqvist H, Strandberg H. Spectral properties of atrial and ventricular endocardial signals. Pacing Clin Electrophysiol 1979; 2: 1119.[Medline]
[6] Cohen A and Zigel Y. Compression of multichannel ECG through multichannel long-term prediction. IEEE Eng Med Biol Mag 1998; 17: 109115.[Medline]
[7] Moody, GB, Mark, RG, Goldberger, AL, Evaluation of the TRIM ECG data compressor, Computers in Cardiology conference 1988, IEEE Computer Society Press, 1989, 167, 70.
[8] Vullings, HJLM, Biomedical Waveform Validation, Technische Universiteit Delft, Delft, Holland ISBN 90-901-2643-0, 1999. [PhD Thesis].
[9] Jalaleddine SMS, Hutchens CG, Strattan RD, Coberly WA. ECG data compression techniques a unified approach. IEEE Trans Biomed Engineering 1990; 37: 329343.
[10] Medtronic Report, Tachyarrhythmias Technical Concept Paper. Vol. 1, N.5 (Autoadjusting sensitivity) pp. 18, Medtronic Inc. 1992.
[11] Rossi P, Sartori G, Casaleggio A, Chiappalone M, Morando M, Corbucci G, Reggiani M, Borgo E. Effects of Data Compression on Intra-cardiac Bipolar Atrial Electrograms. Europace 2001; 2:(suppl. A) A89.
[12] Ishijima M, Shin SB, Hostetter GH, Sklansky J. Scan-along polygonal approximation for data compression of electrocardiograms. IEEE Trans Biomedical Engineering 1983; 30: 723729.
[13] Coggins RJ and Jabri MA. A low complexity intra-cardiac electrogram compression algorithm. IEEE Trans Biomedical Engineering 1999; 46: 8291.
[14] Barold SS. Development of a true pacemaker holter. In Vardas PE (Ed.). Cardiac Arrhythmias, Pacing and Electrophysiology 1998; Kluwer Academic Publishers.
![]()
CiteULike
Connotea
Del.icio.us What's this?
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||