Pacemaker troubleshooting: improved efficacy using a new diagnostic digital Holter recording system
Department of Cardiology, Children's National Medical Centre Washington, DC, U.S.A.; 1Medtronic Inc, Fridley Minnesota, U.S.A.
Asymptomatic dysfunction of cardiac pacing systems is not uncommonly detected using long-term ambulatory monitoring techniques. We report two patients with atrial and ventricular sensing abnormalities noted only on Holter monitoring. Multiple empirical attempts at pacemaker reprogramming based on deductive analysis of the pacing anomaly were unsuccessful. Through the use of a new digital recording system that allowed collection of surface electrocardiographic data, intracardiac electrograms, and Marker ChannelTM data a correct diagnosis was made (pacing lead insulation failure). This new recording system has the advantage of significantly improving diagnostic sensitivity and resulting in cost savings.
Key Words: Pacemaker troubleshooting, sensing abnormalities, Holter monitoring, pacing lead fracture
Correspondence: Dr Jeffrey P. Moak, Department of Cardiology, Children's National Medical Centre, Washington, DC 20010, U.S.A. E-mail: jmoak{at}CNMC.org
[1] Brandes A, Bethge KP, Gonska BD, Diederich KW. Transient pacemaker dysfunction and spontaneous arrhythmias in symptomatic and asymptomatic patients with demand pacemakers. Z Kardiol 1996; 85: 237247.[Medline]
[2] Secemsky SI, Hauser RG, Denes P, Edwards LM. Unipolar sensing abnormalities: incidence and clinical significance of skeletal muscle interference and undersensing in 228 patients. Pacing Clin Electrophysiol 1982; 5: 1019.[Medline]
[3] Wiegand UK, Schier H, Bode F, Brandes A, Potratz J. Should unipolar leads be implanted in the atrium? A Holter electrocardiographic comparison of threshold adapted unipolar and high sensitive bipolar sensing. Pacing Clin Electrophysiol 1998; 21: 16011608.[Medline]
[4] Gillis AM, Hillier KR, Rothschild JM, McDonald M, Simpson CA, Wyse DG. Ambulatory electrocardiography for the detection of pacemaker lead failure. Pacing Clin Electrophysiol 1997; 20: 12741282.[Medline]
[5] Bohm A, Pinter A, Preda I. Pacemaker induced torsade-de-pointe tachycardia. Heart 1998; 80: 466467.
[6] Wiegand UK, Bode F, Schneider R, et al. Diagnosis of atrial undersensing in dual chamber pacemakers: impact of autodiagnostic features. Pacing Clin Electrophysiol 1999; 22: 894902.[Medline]
[7] Kelen GJ, Bloomfield DA, Hardage M, et al. A clinical evaluation of an improved Holter monitoring technique for artificial pacemaker function. Pacing Clin Electrophysiol 1980; 3: 192197.[Medline]
[8] Nowak B, Middeldorf T, Housworth CM, et al. Holter recordings with continuous marker annotations: a new tool in pacemaker diagnostics. Pacing Clin Electrophysiol 1996; 19: 17911795.[CrossRef][Medline]
![]()
CiteULike
Connotea
Del.icio.us What's this?
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||