© 2001 by European Society of Cardiology
PACING
Autocapture compatibility in patients with the MembraneEX lead and Affinity pulse generators
Department of Cardiology, University-Hospital Hamburg-Eppendorf Germany; 1Mustamäe Hospital Estonia; 2St Jude Medical AB Sweden; 3University Hospital of Tartu Estonia; 4District Hospital Leer Germany; 5Hospital Xeral Cies Spain
Abstract
The first Autocapture generation worked well with all recommended leads. The newer Autocapture generation provides a more sensitive resolution for evoked response testing and its implementation in a dual-chamber device. The purpose of the study was to evaluate the performance of the Affinity SR/DR pacemaker with the new Autocapture algorithm in combination with the small surface area pacing lead MembraneEX in 129 patients. Autocapture ventricular threshold, sensing threshold, lead impedance, evoked response (ER) and polarization signals were determined at implantation and discharge, as well as after 1 and 3 months. Autocapture recommendation rate was based on the ER sensitivity test. The median pacing threshold was 0·38, 0·50, 0·75, 0·75 V at implant, discharge, 1 and 3 months post-implant, respectively. The respective data for median lead impedance were 744, 605, 649 and 691 ohms; median sensing threshold was 12·5 mV at all visits. The median ER amplitude was 9·0, 10·1, 9·9 and 10·1 mV and the median polarization signal 0·39 mV at all visits. The frequency of recommended Autocapture activation was 98·3%, 99·2%, 98·3% and 96·2% of all patients at implant, at discharge, 1 and 3 months post-implant respectively. In conclusion, the studied pulse generator enabled, in combination with this pacing lead, in >95% of all patients activation of Autocapture.
Key Words: Ventricular Autocapture activation, high-impedance small-surface area electrode, pacing threshold, lead polarization, evoked response amplitude
Correspondence: Andreas Schuchert, MD, Medical Clinic, Department of Cardiology, University-Hospital Hamburg-Eppendorf, Martinistr. 52, D 20246 Hamburg, Germany. E-mail: schuchert{at}uke.uni-hamburg.de
References
[1] Clarke M, Liu B, Schüller H, et al. Automatic adjustment of pacemaker stimulation output correlated with continuously monitored capture thresholds: a multicenter study. Pacing Clin Electrophysiol 1998; 21: 15671575.[CrossRef][Medline]
[2] Schüller H, Kruse IB, Svennsson O, Mortensen P, et al. Long-term benefits of autocapture four year of follow-up. Pacing Clin Electrophysiol 1999; 22: 807 (abstract).
[3] Schuchert A, Ventura R, Meinertz T. Autocapture Activation without theIntraoperative Evaluation of the Evoked Response Amplitude. Pacing Clin Electrophysiol 2000; 23: 321324.[CrossRef][Medline]
[4] Kacet S, Jarwe M, Klug D. Evaluation of the compatibility of various bipolar pacing leads with Autocapture. Pacing Clin Electrophysiol 1998; 21: 825 (abstract).
[5] Lau C, Nishimura SC, Philippon F, et al. Polarization signal and evoked response characteristics in current endocardial electrodes. Pacing Clin Electrophysiol 1998; 21: 841 (abstract).
[6] Aggarwal RK, Connelly DT, Ray SG, Ball J, Charles RG. Early complication of permanent pacemaker implantation: no difference between dual and single chamber systems. Br Heart J 1995; 73: 571575.
[7] Kiviniemi MS, Pirnes MA, Eramen HJ, Kettumen RV, Hartikainen JE. Complications related to permanent pacemaker therapy. Pacing Clin Electrophysiol 1999; 22: 711720.[CrossRef][Medline]
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
M. Biffi, J. Sperzel, C. Martignani, A. Branzi, and G. Boriani Evolution of pacing for bradycardia: Autocapture Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I23 - I32. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Boriani, L. Rusconi, M. Biffi, L. Pavia, M. Sassara, D. Malfitano, M. G. Bongiorni, L. Padeletti, I. Filice, D. Sanfelici, et al. Role of ventricular autocapture function in increasing longevity of DDDR pacemakers: a prospective study. Europace, March 1, 2006; 8(3): 216 - 220. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Sperzel, B. Nowak, E. Himmrich, G. Zhang, A. König, R. Willems, C. Reister, A. Sathaye, and G. Fröhlig Acute performance evaluation of a new ventricular automatic capture algorithm. Europace, January 1, 2006; 8(1): 65 - 69. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Çeliker, N. Ceviz, and O. Küçükosmanoglu Long-term results of endocardial pacing with AutocaptureTM threshold tracking pacemakers in children Europace, January 1, 2005; 7(6): 569 - 575. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||

