© 2001 by European Society of Cardiology
PACING
Immediate and long-term atrial sensing stability in single-lead VDD pacing depends on right atrial dimensions
Department of Cardiology, S. Filippo Neri Hospital Rome; 1Aosta Hospital Aosta; 2Macerata Hospital Macerata-Italy
Abstract
AIMS: Single-pass VDD is a physiological stimulation mode which requires a single-lead with a floating dipole to detect the atrial signal. We investigated the impact of right heart dimensions on immediate and long-term atrial sensing stability in VDD systems to draw guidelines for optimal atrial sensitivity programming.
METHODS: Forty-one patients (23 males, mean age 73±11 years) with II or III degree AV block and normal sinus node function received a Thera VDD Medtronic 8948 pulse generator with Medtronic 5032 lead. Atrial sensing was evaluated at pre-discharge and during 12-months follow-up in the supine and upright positions, during normal and forced breathing. Atrial sensing variability, quantified by a D value (mV), which represents the difference between the maximum and the minimum atrial electrogram amplitude obtained during the various activities, was analyzed and related to right heart dimensions measured at echocardiography.
RESULTS: Long-term appropriate atrial sensing was obtained in 40/41 patients. P wave mean amplitude at pre-discharge, during every specific activity, was predictive of 12-month values. No atrial oversensing was observed. According to the right atrial dimension (RAD), patients with RAD >38 mm had a sensing variability significantly higher than those with RAD
34 mm (D=1±1 Vs D=0·4±0·5;P< 0·05). A linear relationship between D and RAD dimensions was observed (r=0·47;P< 0·005). According to the distance from the roof of the right atrium to the right ventricular apex (RAVD), patients with RAVD <93 mm had D=1·1±1, while patients with RAVD >93 mm had D=0·5±0·4;P< 0·05); a inverse correlation between RAVD and D was highlighted (r=0·43;P< 0·01). Using a cluster analysis, the combination of RAD and RAVD identified a subgroup (RAD >30 mm and RAVD <80 mm) at high risk of loss of AV synchrony with 67% sensitivity and 80% specificity. Nevertheless, apart from heart dimensions, no patient showed an atrial signal amplitude lower than 0·3 mV.
CONCLUSIONS: According to our data, in VDD single-lead systems the amplitude of the atrial signal is stable over time in every physical activity. High RAD and low RAVD values may select patients with poor atrial sensing stability. Anyway, taking into account that no atrial oversensing was observed, atrial sensitivity setting at the highest value should be recommended.
Key Words: VDD pacing mode, physiological pacing, echocardiography, right heart dimensions, atrial sensing
Correspondence: Prof Massimo Santini, Department of Cardiology-Ospedale S. Filippo Neri Via Martinotti 20-00135 Roma, Italy. E-mail: m.santini{at}rmnet.it
References
[1] Rediker DE, Eagle KA, Homma S, et al. Clinical and hemodynamic comparison of WI versus DDD pacing in patients with DDD pacemakers. Am J Cardiol 1988; 61: 323329.[CrossRef][Web of Science][Medline]
[2] Crick JCP. European multicenter prospective follow-up study of 1002 implants of a single lead VDD pacing system. Pacing Clin Electrophysiol 1991; 14: 17421744.[CrossRef][Medline]
[3] Parravicini U, Zanetta M, Zenone F, et al. Complications of a single lead VDD pacing in 35 patients with AV block. Pacing Clin Electrophysiol 1992; 15: 19011902.[CrossRef][Medline]
[4] Sermasi S and Marconi M. VDD single pass lead pacing: Sustained mediated tachycardias unrelated to retrograde atrial activation. Pacing Clin Electrophysiol 1992; 15: 19031907.[Medline]
[5] De Cock CC, Van Campen LCMC, Huygens J, et al. Usefulness of echocardiography to predict inappropriate atrial sensing in single-lead VDD pacing. Pacing Clin Electrophysiol 1999; 22: 13441347.[Medline]
[6] Dhingra RC. Sinus node dysfunction. Pacing Clin Electrophysiol 1983; 6: 10621069.[CrossRef][Medline]
[7] Gregoratos G. ACC/AHA Guidelines for implantation of cardiac pacemakers and antiarrhythmia devices. A report of the American College of Cardiology/American Heart Association task force on practice guidelines (Committee on pacemaker implantation). J Am Coll Cardiol 1998; 31: 11751209.
[8] Antonioli GE. Single lead atrial synchronous ventricular pacing: a dream come true. Pacing Clin Electrophysiol 1994; 17: 15311547.[Medline]
[9] Ansani L, Percoco GF, Guardigli G., et al. Long-term reliability of single lead atrial synchronous pacing systems using closely spaced atrial dipoles: five-year experience. Pacing Clin Electrophysiol 1994; 17: 18651869.[CrossRef][Medline]
[10] Naegeli B, Osswald S, Pfisterer M, et al. VDD(R) pacing: short- and long-term stability of atrial sensing with a single lead system. Pacing Clin Electrophysiol 1996; 19: 455464.[CrossRef][Medline]
[11] Faerestrand S and Ohm OJ. Atrial synchronous ventricular pacing with a single lead: reliability of atrial sensing during physical activities, and long-term stability of atrial sensing. Pacing Clin Electrophysiol 1998; 21: 271276.[Medline]
[12] Lau CP, Leung SK, Lee ISF. Comparative evaluation of acute and long-term clinical performance of two single lead atrial synchronous ventricular (CDD) pacemakers: diagonally arranged bipolar versus closely spaced bipolar ring electrodes. Pacing Clin Electrophysiol 1996; 19: 15741581.[Medline]
[13] Toivonen L and Lommi J. Dependence of atrial sensing function on posture in a single-lead atrial triggered ventricular (VDD) pacemaker. Pacing Clin Electrophysiol 1996; 19: 309313.[CrossRef][Medline]
[14] Sun ZH, Stjernvall J, Laine P, et al. Extensive variation in the signal amplitude of the atrial floating VDD pacing electrode. Pacing Clin Electrophysiol 1998; 21: 17601765.[Medline]
[15] Frøhlig G, Blank W, Schwerdt H, et al. Atrial sensing performance of AV universal pacemakers during exericse. Pacing Clin Electrophysiol 1988; 11: 4760.[Medline]
[16] Ross BA, Zeigler V, Zinner A, et al. The effect of exercise on atrial electrogram voltage in young patients. Pacing Clin Electrophysiol 1991; 14: 10401048.[Medline]
[17] Lekven J, Chatterjee K, Tyberg JV, et al. Reduction in ventricular endocardial and epicardial potentials during acute increments in left ventricular dimensions. Am Heart J 1979; 98: 200206.[Medline]
![]()