CASE REPORT
Avoiding inappropriate ventricular tachycardia detection due to T-wave oversensing in an implantable cardioverter defibrillator: a novel application of the electrogram width criterion
Cardiac Arrhythmia Unit, University Hospital of Zurich Zurich, Switzerland
Appropriate sensing is an essential function of an implantable cardioverter defibrillator (ICD). T-wave oversensing by an ICD can be a serious problem in some patients, causing overestimation of the heart rate, inappropriate tachyarrhythmia detection and therapy delivery. Decreasing the sensitivity or programming longer refractory periods can sometimes overcome T-wave oversensing, but these measures may interfere with the ability of the ICD to correctly detect tachyarrhythmias. This report proposes a novel application of the electrogram (EGM) width criterion using a recently introduced detection enhancement algorithm intended to improve the specificity of ventricular tachycardia detection, to avoid T-wave oversensing. Based on the course of a case with persistent T-wave oversensing and review of previously published reports, oversensing problems in ICDs and management strategies are discussed.
Key Words: T-wave oversensing, implantable cardioverter defibrillator, ventricular tachycardia, detection enhancement, electrogram width criterion
Correspondence: F
rat Duru, Cardiac Arrhythmia Unit, University Hospital of Zurich, Raemistr. 100, CH-8091, Zurich, Switzerland. E-mail: firat.duru{at}dim.usz.ch
[1] Barold SS, Falkoff MD, Ong LS, Heinle RA. Oversensing by single-chamber pacemakers: mechanisms, diagnosis, and treatment. Cardiol Clin 1985; 3: 565585.[Medline]
[2] Singer I, de Borde R, Veltri EP, et al. The automatic implantable cardioverter defibrillator: T wave sensing in the newest generation. Pacing Clin Electrophysiol 1988; 11: 15841591.[CrossRef][Medline]
[3] Grimm W, Flores BF, Marchlinski FE. Electrocardiographically documented unnecessary, spontaneous shocks in 241 patients with implantable cardioverter defibrillators. Pacing Clin Electrophysiol 1992; 15: 16671673.[CrossRef][Medline]
[4] Grimm W, Flores BF, Marchlinski FE. Complications of implantable cardioverter defibrillator therapy: follow-up of 241 patients. Pacing Clin Electrophysiol 1993; 16: 218222.[CrossRef][Medline]
[5] Hook BG, Callans DJ, Kleiman RB, Flores BT, Marchlinski FE. Implantable cardioverter-defibrillator therapy in the absence of significant symptoms. Rhythm diagnosis and management aided by stored electrogram analysis. Circulation 1993; 87: 18971906.
[6] Kelly PA, Mann DE, Damle RS, Reiter MJ. Oversensing during ventricular pacing in patients with a third-generation implantable cardioverter-defibrillator. J Am Coll Cardiol 1994; 23: 15311534.[Abstract]
[7] Reiter MJ and Mann DE. Sensing and tachyarrhythmia detection problems in implantable cardioverter defibrillators. J Cardiovasc Electrophysiol 1996; 7: 542558.[Web of Science][Medline]
[8] Deshmukh P and Anderson K. Myopotential sensing by a dual chamber implantable cardioverter defibrillator: two case reports. J Cardiovasc Electrophysiol 1998; 9: 767772.[Web of Science][Medline]
[9] Calkins H, Brinker J, Veltri EP, Guarnieri T, Levine JH. Clinical interactions between pacemakers and automatic implantable cardioverter-defibrillators. J Am Coll Cardiol 1990; 16: 666673.[Abstract]
[10] Sticherling C, Klingenheben T, Skupin M, Li YG, Hohnloser SH. Kombinierte Therapie mit transvenosen Kardioverter/Defibrillator- und antibradykarden Schrittmachersystemen. Z Kardiol 1997; 86: 105112.[Medline]
[11] Geiger MJ, O'Neill P, Sharma A, et al. Interactions between transvenous nonthoracotomy cardioverter defibrillator systems and permanent transvenous endocardial pacemakers. Pacing Clin Electrophysiol 1997; 20: 624630.[CrossRef][Medline]
[12] Curwin JH, Roelke M, Ruskin JN. Inhibition of bradycardia pacing caused by far-field atrial sensing in a third-generation cardioverter defibrillator with an automatic gain feature. Pacing Clin Electrophysiol 1996; 19: 124126.[CrossRef][Medline]
[13] Yokoyama M, Wada J, Barold SS. Transient early T wave sensing by implanted programmable demand pulse generator. Pacing Clin Electrophysiol 1981; 4: 6874.[CrossRef][Medline]
[14] Perry GY and Kosar EM. Problems in managing patients with long QT syndrome and implantable cardioverter defibrillators: a report of two cases. Pacing Clin Electrophysiol 1996; 19: 863867.[CrossRef][Medline]
[15] Frazier DW and Stanton MS. Pseudo-oversensing of the T wave by an implantable cardioverter defibrillator: a nonclinical problem. Pacing Clin Electrophysiol 1994; 17: 13111315.[Medline]
[16] Mann DE, Damle RS, Kelly P, Landers M, Otto L, Reiter MJ. Comparison of oversensing during bradycardia pacing in two types of implantable cardioverter-defibrillator systems. Am Heart J 1998; 136: 658663.[Medline]
[17] Minten J, Cuypers A, van Binsbergen E. Medtronic Model 7219 Jewel PCD European Clinical Report.
[18] Use of the EGM Width Criterion to discriminate VT from SVT.Tachyarrhythmia Technical Concept Paper (Medtronic). Volume 2, Number7.
[19] Klingenheben T, Sticherling C, Skupin M, Hohnloser SH. Intracardiac QRS electrogram width an arrhythmia detection feature for implantable cardioverter defibrillators: exercise induced variation as a base for device programming. Pacing Clin Electrophysiol 1998; 21: 16091617.[Medline]
[20] Duru F, Schönbeck M, Lüscher TF, Candinas R. The potential for inappropriate ventricular tachycardia confirmation using the intracardiac electrogram width criterion. Pacing Clin Electrophysiol 1999; 22: 10391046.[CrossRef][Medline]
[21] Duru F, Schönbeck M, Candinas R. The need for exercise testing for optimal programming of the EGM width criterion in patients with an implantable defibrillator (abstract). J Am Coll Cardiol 1999; 33: 133.
[22] Duru F, Junga G, Schönbeck M, Candinas R. Effect of slew threshold on electrogram width during ventricular tachycardia: Recommendations for optimal function of the EGM width criterion in an implantable cardioverter-defibrillator (abstract). Kardiovaskuläre Medizin 1999; 2: 51.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
K. Srivathsan, L. R. Scott, and G. T. Altemose T-wave oversensing and inappropriate shocks: a case report Europace, May 1, 2008; 10(5): 552 - 555. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Rauwolf, M. Guenther, N. Hass, A. Schnabel, M. Bock, M.U. Braun, and R.H. Strasser Ventricular oversensing in 518 patients with implanted cardiac defibrillators: incidence, complications, and solutions Europace, November 1, 2007; 9(11): 1041 - 1047. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||