Latest generation of unipolar and bipolar steroid eluting leads
Long-term comparison of electrical performance in atrium and ventricles
1Cardiology Dept, Degli Infermi Hospital Feanza (RA), Italy; 2Cardiology Dept, Fatebenefratelli Hospital Rome, Italy; 3Cardiology Dept, S. Agostino Hospital Modena, Italy; 4Cardiology Dept, S. Carlo Borromeo Hospital Milan, Italy; 5Cardiology Dept, Casa di Cura Villabianca Bari, Italy; 6Cardiology Dept, S. Maria Nuova Hospital Reggio Emilia, Italy; 7Cardiology Dept, Della Val Camonica Hospital Esine, Italy; 8Medtronic Italy Italy
One hundred and seventy-three patients, mean age 74 years permanently paced with 123 atrial (53 unipolar, 70 bipolar) and 143 ventricular (73 unipolar, 70 bipolar) pacing leads were included in this study. The pacing leads were recent generation low surface area steroid eluting leads from one manufacturer: leads with silicone and polyurethane insulation were studied, and they were combined with generations of one pacemaker family from the same manufacturer permitting identical measurements to be made over a follow-up of 2 years. Pacing threshold was measured using pulse duration at a fixed voltage of 1·5 V: peak to peak P and R wave amplitude and pacing impedance at 2·5 V and 0·5 ms were all measured using the manufacturer's standard programmer. Although many significant differences, in the parameters measures, existed between atrium and ventricle and unipolar and bipolar configurations, none was felt to be of clinical significance. These data permit the physician to choose the lead type with regard to sensing performance and long-term lead integrity.
Key Words: Unipolar, bipolar atrial and ventricular pacing leads, steroid elution, stimulation threshold, sensing
Correspondence: Dr Daniele Cornacchia, Via Orto Caporal, 5, 48018, Faenza (RA) Italy.
[1] Møller M and Arnsbo P. Appraisal of pacing lead performance from the Danish pacemaker register. PACE 1996; 19: 13271336.
[2] Furman S and Benedek ZM. Survival of implantable pacemaker leads. PACE 1990; 13: 19101914.
[3] Mond H, Strathmore N, Hunt P, Hunt D. Bipolar and unipolar permanent pacing leads which is superior? (Abstr). PACE 1989; 12: 678.
[4] Masterson M, Tuzcu EM, Maloney JD, Morant V, Castle L. Atrial pacing and sensing: unipolar vs bipolar. Seven different leads compared. (Abstr). PACE 1987; 10: 435.
[5] Breivik K, Engedal H, Ohm O. Electrophysiological properties of a new permanent endocardial lead for uni- and bipolar pacing. PACE 1982; 5: 268274.
[6] Czernin J, Kaliman J, Laczkovics A, et al. Unipolar vs bipolar pacing a prospective intrapatient, long-term study. (Abstr). PACE 1987; 10: 663.
[7] Joglar J, Welch P, Wilkinson W, Hamdan M, Page L. Initial experience with a high-impedance tined endocardial pacemaker lead: Evidence for increased lead failure. Am Heart J 1997; 134: 161164.[Medline]
[8] Schuchert A, Cappato R, Kuck K-H, Meinertz T. Programmable polarity: effects on pacing and sensing of bipolar steroid-eluting leads. PACE 1996; 19: 20992102.
![]()
CiteULike
Connotea
Del.icio.us What's this?
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||