Skip Navigation

Europace 2007 9(Supplement 6):vi119-vi123; doi:10.1093/europace/eum216
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by van Dam, P. M.
Right arrow Articles by van Oosterom, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by van Dam, P. M.
Right arrow Articles by van Oosterom, A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The European Society of Cardiology 2007. All rights reserved. For permissions, please e-mail: journals.permissions@oxfordjournals.org

Analysing the potential of Reveal® for monitoring cardiac potentials

Peter M. van Dam1,* and Adriaan van Oosterom2

1 Medtronic SQDM, Meander 1051, 6825 MJ Arnhem, The Netherlands; 2 Department of Cardiology, CHUV, Lausanne, Switzerland

* Corresponding author. Tel: +31 26 3767292; fax: +31 26 3767305. E-mail address: peter.van.dam{at}medtronic.com


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
Aims: The implantable loop recorder (ILR) continuously monitors the heart’s electric activity by means of a subcutaneous bipolar electrogram (Elg). Currently, the relationship between the Elg and the surface electrocardiogram (ECG) has been poorly documented. This model-based study aimed at investigating the differences between the bipolar surface and subcutaneous signals, as well as the effect of the insulating boundary of the ILR on these signals. Additionally, the model is used for determining the optimal implant location of the device.

Methods and results: Sinus rhythm ECG of a complete heart cycle was simulated by means of a biophysical model. Different volume conductors were created to investigate the effect of the insulating boundary of the ILR. The Elg closely matched the nearby bipolar ECG, both in morphology and in amplitude. The optimal localization and orientation of the ILR was found to depend on the Elg signal feature of interest, e.g. PQ, QRS, or STT waveforms.

Conclusion: The differences between the bipolar ECG on the surface and the subcutaneous electrogram are negligible. The optimal implant location may be based on nearby surface recordings. The simulation model is an eligible tool for determining the optimal implant location for the ILR, for all signal features of interest.

Key Words: Volume conductor models, Subcutaneous ECG, Implantable loop recorder


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
In clinical practice, the 12-lead electrocardiogram (ECG) is commonly used for diagnosing specific electric malfunctioning of the heart. The normal recording practice provides snapshots of the electrical activity of the heart only. Rhythm disorders or the onset of acute myocardial infarction frequently remain unobserved. To capture such events, recording time can be prolonged, up to days, using a Holter system. Yet, despite the longer recording time, relevant clinical events are still frequently missed. In cases such as these, the heart may be monitored by an even more extended period (months) by an implantable loop recorder (ILR).1Go

The Reveal, a commercially available ILR, is such a monitor. It is often implanted just below the subcutaneous fat in the left shoulder region. It monitors the heart rhythm by means of a bipolar electrogram with an electrode spacing of ~45 mm. To reduce the influence of myo-potentials, the electrodes are introduced facing the skin. Currently, the experience obtained from using the subcutaneous signals is still limited and the relationship between subcutaneous bipolar electrogram (Elg) and the nearby bipolar surface ECG is insufficiently documented. In this study, a computer model was used to quantify the differences between the two types of bipolar signals.

To simulate the type of signals involved, the electrical activities of both the atria and the ventricles need to be modelled. On the basis of a realistic source model and an MRI-derived volume conductor model, the bipolar ECG wave forms generated during sinus rhythm are compared with those of the IRL.

The boundary of the Reveal is a perfect insulator (with the exception of two small electrode contacts). The presence of such a non-conducting object might significantly influence the current density distribution within the torso, and hence also the potential field and in particular the potential difference between the electrodes of the ILR. This particular effect was studied by comparing the potential differences observed for the bipolar signals while either including or excluding the isolating properties of the device.

Finally, the suitability of the model for determining an optimal implant location was tested. At present, the optimal implant location in terms of signal amplitude remains to be determined.


    Methods
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
Electrocardiogram simulation
Any description of the genesis of electrocardiographic potentials requires the specification of the current sources generated by the biochemical reactions at the cell membrane: the source model. In addition, the conductive properties (extent and conductivity values) of the medium in which the potential field is set up need to be specified: the volume conductor model. The relevant geometry of the heart and the volume conductor as used in this study were derived from MRI images of a healthy 22-year-old human male. These were retrieved from images acquired during breath hold, half way expiration. The geometries of the atria and ventricles were reconstructed at the beginning of the P-wave, ~165 ms before the QRS onset.

Source model
The model of the cardiac current generator used was the equivalent double layer (EDL), located at the closed surface Sh bounding the myocardium, comprising the epicardium and the endocardium. The double layer may be viewed as a sheet of current dipoles directed along the local surface normal of Sh. This source model has a direct link with electrophysiology and has previously been shown to be very effective in the simulation of the potentials during depolarization and repolarization of the atria2Go,3Go and ventricles.4Go,5Go The MRI-derived geometries of atria and ventricles are shown in Figure 1D. The endocardial segments of Sh bordering the cavities are shown alongside.


Figure 1
View larger version (43K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 1 The three volume conductor models used. (A) Anterior view of the standard volume conductor model with lungs, subcutaneous fat (not shown), and heart. The position of the precordial electrodes as indicated. (B) As in (A), but now a horizontally placed ILR is incorporated. (C) As in (A), but now incorporating a vertically placed ILR. (D) top: the endocardial surface segments of Sh and a part of the connecting vessels bordering the left (in red) and right cavities (in blue). (D) bottom: surfaces Sh of the atria and the ventricles as used in the study, shown separated to enhance clarity. (E) the ILR: Reveal. The Elg is the potential measured at electrode a minus the potential at electrode b.

 
The time course of the local strength of the double layer is that of a stylized version of the trans-membrane potential (TMP) of healthy myocytes. It is specified by the local magnitude of the upstroke of the TMP at node n of the surface Sh, its timing of depolarization ({tau}n) and its timing of repolarization ({rho}n). The latter is defined as the timing of the inflection point of the repolarization sequence (Figure 2). Accordingly, the source strength at node n at time instant t can be denoted as sn(t,{tau}n,{rho}n). For discrete time steps, the entire source may be denoted by S, the columns of this matrix represent the instantaneous strengths of the source elements:


Formula 216M1

(1)


Figure 2
View larger version (22K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 2 Transmembrane potentials (TMPs) of some of the atrial (red) and ventricular nodes acting as the sources for simulating the signals during a sinus beat. The TMP at each node n is specified by the timing of depolarization ({tau}n) and repolarization ({rho}n). For one atrial node {tau}n and {rho}n are indicated by black dotted lines. Note that: (i) the specified timing influences TMP wave form and (ii) atrial repolarization at the nodes starts during the PQ interval, and ends, at least for some, during the QRS complex.

 
Volume conductor model
To quantify the differences between the bipolar surface ECG and the Elg, three volume conductor models were created: a basic model that does not take into account the insulating properties of the ILR, and two variants in which the insulating effect was incorporated.

The basic volume conductor model used consisted of triangulated versions of the atria, the ventricles, and the surfaces bounding the major volumes that influence the morphology of the ECG signals, e.g. the torso, lungs, and blood cavities (Figure 1). Additionally, a local subcutaneous fat layer was included under the complete body surface. The assigned thickness of the layer was 8 mm, the approximate value of the subcutaneous fat in the left shoulder region of the lean subject studied. The conductivity values assigned to the individual compartments were: thorax, atrial, and ventricular muscle: 0.2 Sm–1, lungs: 0.04 Sm–1, fat: 0.07 Sm–1, and blood cavities: 0.6 Sm–1.3Go The triangulated atria consisted of 1504 nodes, and the triangulated ventricles consisted of 1500 nodes. For the torso, 501 nodes were used; for the left and right lung, 486 and 259 nodes were used, respectively.

Additionally, two variant volume conductor models were considered in which the insolating properties of the ILR were incorporated, placed in either a horizontal (slightly oblique) or a vertical orientation (Figure 1). The size of the ILR model was taken from the Reveal, length: 60 mm, width: 20 mm, and thickness 8 mm, with an electrode spacing of 45 mm. In both cases, ILRs were positioned ~10 mm under the skin, i.e. just below the subcutaneous fat layer.

Forward computation
From the source and volume conductor descriptions as specified above, the potential field throughout the volume conductor can be computed by means of the boundary element method.6Go The transfer from all source elements on the nodes, n, to any set of field points, {ell}, considered can be expressed by a transfer matrix A, with elements a({ell},n).

In this study, three different transfer matrices were computed, specific for different volume conductor models and field points considered. For each of the transfer matrices, the potential {phi}({ell},t) at field point {ell} was then computed as


Formula 216M2

(2)

Simulated electrocardiogram during sinus rhythm
The source elements are located at all nodes of the triangulated surfaces specifying the geometry of the atria and the ventricles. Time was discretized at 1 ms interval. The timings of local depolarization and repolarization for the atria and ventricles were estimated by a dedicated inverse procedure.7Go The resulting isochrones for the depolarization are depicted in Figure 3.


Figure 3
View larger version (28K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 3 Isochrones of the timing of depolarization during sinus rhythm. (A) Atrial depolarization (left: natural anterior view; right natural posterior view). (B) Ventricular depolarization (left: natural anterior view; right: posterior basal view). The isochrones are drawn at 10 ms intervals.

 
The simulated 12-lead ECG based on the standard volume conductor model is shown in Figure 4 (blue traces). The quality of this simulation can be judged by comparing the simulated signals (red traces) with the ones measured on the subject involved.


Figure 4
View larger version (13K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 4 Measured (blue) and simulated (red) standard 12- lead ECG. All signals use the so-called zero-mean as a reference, i.e. the mean of all measured or simulated signals. Extremity signals are shown un-augmented.

 

    Results
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
For all three volume conductors described, the resulting electric signals were simulated to investigate the differences between surface and subcutaneous signals as well as to study the influence insulating properties of the ILR. The subcutaneous electrograms were computed at the locations of the electrodes of the ILR (Figure 1), both while including or ignoring the insulating effect of the ILR (Figure 5). The red lines relate to the simulated Elgs: the solid lines when incorporating the insulating effect, the dotted lines when disregarding it. In addition, Figure 5 includes the corresponding signals for field points on the surface of the thorax closest to the electrode positions of the ILR (blue lines).


Figure 5
View larger version (11K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 5 Influence of the ILR on the electrical signals. Simulated signals during sinus rhythm. Red lines show the subcutaneous bipolar EGC; solid lines: the signals when accounting for the non-conductive properties of the SubQ device; dotted lines: the signals when ignoring these. Blue traces: bipolar ECG at the body surface locations closest to the electrodes of the SubQ device. (A) Signals of the vertically orientated ILR (Figure1). (B) Signals of the horizontally orientated ILR (Figure 1).

 
The signals in Figure 5 demonstrate that the effect of the insulating properties of the ILR is very small, certainly compared with the inter-subject variability. At the locations of the ILR, the subcutaneous ECGs were found to be highly similar to the bipolar ECGs on the most proximal part of the body surface. Hence, the study of the optimal placement of the ILR may be carried out using the standard volume conductor, as reported in the remainder of this paper.

Three positions in the left side of the torso were selected to construct subcutaneous amplitude maps of the P-wave, the R-wave, and the T-wave: the first map was located above the heart, the next was located on top of the heart, and the last was located below the heart (see torso insets Figure 6). For each position within the map region, an Elg was generated between that point and the center of the map, which served as the potential reference for the Elg. All map positions were located 10 mm under the surface of the torso. The maps in Figure 6 show that the maximum R-wave amplitudes were found in the region directly overlying the heart. Above and below the heart, the amplitudes were about half that size. For positions below the diaphragm, the P- and the T-wave disappeared almost completely. In all amplitude maps, a line could found for which the amplitude is approximately zero. The difference in amplitude of the Elg features for different positions can best be observed in the signals shown in Figure 6.


Figure 6
View larger version (64K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 6 Amplitude maps of P-, R-, and T-wave at three different locations (see torso insets). The potential reference is at the center of the maps. Radius of disk: 50 mm. Iso-potential lines are drawn at 0.01 mV for the P-wave amplitude, 0.1 mV for the R-wave amplitude, and 0.05 mV for the T-wave amplitude. Positive values are drawn with solid lines and negative values are drawn with a dashed line. The panels on the right show the subcutaneous EGC signals at the border of the disc, 10 mm below the skin’s surface, at four orientations (0, 90, 180, and 270°).

 

    Discussion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
For the locations of the ILR selected, the subcutaneous Elg’s signals were found to be highly similar to those on the most proximal part of the body surface, both in morphology and amplitude. The effect of the insulating properties of the ILR on the Elg is negligible (Figure 5), certainly compared with the inter-subject variability. As a consequence, the study of the optimal placement of the ILR and the selection of its signal features may be carried out by using the surface signals.

In most mapping areas explored (Figure 6), the maximum amplitudes of the P-wave were found to be at least a factor 10 smaller than those of the R-wave, whereas those of the T-wave were about a factor less than four. From a practical implantation point of view, as well as by taking into consideration signal magnitude, the left shoulder region seems to be an appropriate implantation site. In order to determine the optimal orientation, the pattern of the local surface potentials has to be taken into account, e.g. both the maximum and minimum amplitude play a role.

As stated, the differences between the subcutaneous Elg and the bipolar surface ECG are very small. Initially, we compared both signals without the subcutaneous fat layer included in the volume conductor model (results are not shown). Even than the differences were very small, although slightly more compared with the signals with a subcutaneous fat layer included. This even smaller difference between both bipolar signals can, up to a certain extent, be explained by the fact that subcutaneous fat has a lower conductivity than the tissue underneath, thereby limiting the current flow across the layer of fat. As a consequence, the potentials on both sides of the fat layer are approximately the same. It is to be expected that for subjects with more subcutaneous fat the differences between the surface ECG and the subcutaneous Elg will, similarly, be small.

In clinical practice, the ILR sometimes fails to sense the R-wave,8Go thereby failing to record the Elg during clinically relevant events. As the simulation results show, this can occur when the relative orientation of the ILR and the heart is such that the R-wave amplitude is ~0. For an optimal implant orientation, the line of minimal amplitude for certain Elg feature, e.g. PQ, QRS, or STT interval, should serve as guideline. The angle between the minimal amplitude line and the ILR should be large enough (~60°) in order to enable the recording of the selected feature despite changes in heart orientation because of posture changes.9Go–11Go Unfortunately, the minimal amplitude line differs per Elg feature as shown in Figure 6. This may lead to conflicting implantation orientations for the ILR in cases where more than one feature requires to be monitored.


    Conclusion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
The bipolar EGCs sensed by the ILR closely match with the bipolar surface ECG, both in morphology and in amplitude. The optimal orientation of the ILR depends on the focus of interest, i.e. PQ, QRS, or STT interval. For optimal R-wave sensing, both ILRs location and its orientation for which the minimum R-wave amplitudes occur have to be considered.

Conflict of interest: none declared.


    References
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
[1] Krahn A, Klein GJ, Skanes A, Yee R. Use of the implantable loop recorder in evaluation of patients with unexplained syncope. J Cardiovasc Electrophysiol (2003) 14:S70–3.[CrossRef][Web of Science][Medline]

[2] van Oosterom A. Solidifying the solid angle. J Electrocardiol (2002) 35(Part 2):181–92.[Web of Science][Medline]

[3] van Dam PM, van Oosterom A. Volume conductor effects involved in the genesis of the P wave. Europace (2005) 7:S30–8.[Abstract/Free Full Text]

[4] van Oosterom A. Genesis of the T wave as based on an equivalent surface source model. J Electrocardiography (2001) 34(Suppl):217–27.[CrossRef]

[5] van Oosterom A, Oostendorp TF. ECGSIM: an interactive tool for studying the genesis of QRST waveforms. Heart (2004) 90:165–8.[Abstract/Free Full Text]

[6] Gulrajani RM, Roberge FA, Mailloux GE. The forward problem of electrocardiography. In: Comprehensive Electrocardiology.—Macfarlane PW, Lawrie TTV, eds. (1989) Oxford: Pergamon Press. 197–236.

[7] van Oosterom A, van Dam PM. The intra-myocardial distance function as used in inverse computations of the timing of depolarization and repolarization. Comput Cardiol—Murray A, ed. (2005) Lyon, France. 567–70.

[8] Chrysostomakis SI, Klapsinos NC, Simantirakis EN, Marketou ME, Kambouraki DC, Vardas PE. Sensing issues related to the clinical use of implantable loop recorders. Europace (2003) 5:143–8.[Abstract/Free Full Text]

[9] Adams MG, Drew BJ. Body position effects on the ECG. J Electrocardiol (1997) 30:285–91.[CrossRef][Web of Science][Medline]

[10] Engblom H, Foster JE, Martin TN, Groenning B, Pahlm O, Dargie HJ, et al. The relationship between electrical axis by 12-lead electrocardiogram and anatomical axis of the heart by cardiac magnetic resonance in healthy subjects. Am Heart J (2005) 150:507–12.[CrossRef][Web of Science][Medline]

[11] Sutherland DJ, McPherson DD, Spencer CA, Armstrong CS, Horácek MB, Montague TJ. Effects of posture and respiration on body surface electrocardiogram. Am J Cardiol (1983) 52:595–600.[CrossRef][Web of Science][Medline]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?



This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by van Dam, P. M.
Right arrow Articles by van Oosterom, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by van Dam, P. M.
Right arrow Articles by van Oosterom, A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?