VENTRICULAR ARRHYTHMIA
Unusual cause for an increase of the sensing integrity counter in a patient with inappropriate implantable cardioverter-defibrillator therapy
Abteilung Kardiologie und Pneumologie, Herzzentrum, Klinikum der Georg-August Universität Göttingen, Robert-Koch-Strasse 40, 37075 Göttingen, Germany
Manuscript submitted 16 October 2006. Accepted after revision 2 February 2007.
* Corresponding author. Tel: +49 551 39 9625; fax: +49 551 39 9628. E-mail address: dirkvollmann2000{at}aol.com
| Abstract |
|---|
|
|
|---|
We describe the case of a patient who presented with multiple implantable cardioverter-defibrillator (ICD) shock discharges 12 months after device implantation. Upon device interrogation, intermittent oversensing of electrical noise and potential ICD lead failure were suggested by a significant increase in the sensing integrity counter (SIC), a cumulative count of very short ventricular sensed intervals. Analysis of stored episodes, however, revealed that inappropriate ICD therapy had been caused by intermittent T-wave oversensing (TWO), and that the increase of the SIC resulted from the coincidence of TWO and premature ventricular complexes (PVCs). T-wave oversensing resolved and the SIC did not increase any more during follow-up after adjustment of ventricular sensitivity. The coincidence of TWO and PVCs should therefore be considered as an uncommon cause for short ventricular sensed intervals in ICD patients presenting with a suspect increase in the SIC.
Key Words: Implantable-cardioverter defibrillator, Oversensing, Inappropriate therapy, Sensing integrity counter
| Case report |
|---|
|
|
|---|
A 69-year-old man presented for regular follow-up 12 months after prophylactic implantation of a dual-chamber implantable cardioverter-defibrillator (ICD) (Medtronic EnTrustTM, Medtronic Inc. Minneapolis, MN, USA). Underlying cardiac diseases were ischaemic cardiomyopathy with a left ventricular ejection fraction of 30% and sick sinus syndrome with a history of paroxysmal atrial fibrillation. At presentation, the patient reported that he had experienced multiple shock discharges several weeks earlier, and that, 4 weeks ago, these events had triggered an unscheduled follow-up at another institution, where the ICD had been interrogated. Thereafter, the device had delivered one additional shock. The patient denied loss of consciousness, and there was no evidence of myocardial ischaemia or cardiac decompensation. The surface ECG showed, similar to prior recordings, an atrial paced rhythm with first-degree atrioventricular block and right bundle branch block.
Several device alerts were noticed upon interrogation. The sensing integrity counter (SIC), which cumulatively counts short (120130 ms) VV intervals and thereby typically indicates intermittent oversensing of electrical noise, had increased to 3471 since last interrogation. Furthermore, all therapies including shocks had been delivered without success for one VT/VF episode. Last, patient activity had progressively decreased during the last months to currently <2 h per day. The device had been programmed to AAIR
DDDR pacing at a rate of 50120 bpm, with two ventricular tachyarrhythmia detection zones (VT >158 bpm, VF >231 bpm). In both detection windows, antitachy pacing (ATP) and shock therapy were active (ATP during charging in the VF zone). Ventricular sensitivity was programmed to the nominal value of 0.3 mV.
The episode counter showed that 8 VT, 117 non-sustained VT, and no AF/AT episodes had been detected since the last device interrogation. One high-voltage shock and multiple ATP sequences had been delivered. Analysis of stored data also revealed that 5 VF, 65 VT, and 621 non-sustained VT episodes had been documented at the previous follow-up 4 weeks earlier. Also, 12 shocks had been delivered in the corresponding time interval. Some of these older episodes could be retrieved from the device memory. Of note, no reprogramming and no additional diagnostic or therapeutic procedures had been performed during the follow-up at the other institution.
At the present follow-up, analysis of stored episode data revealed intermittent T-wave oversensing (TWO). Furthermore, combined occurrence of TWO and premature ventricular complexes (PVCs) was found to cause short VV intervals (Figure 1), thereby explaining the significant increase in the SIC as well as the inappropriate detection of VF. Analysis of lead trend data and acute measurements of the sensed R-wave amplitude (19 mV) and ventricular pacing impedance (584
) provided no evidence of dysfunction of the implanted ICD lead (Medtronic model 6947). Comparison of ventricular electrograms, recorded 3 days, 3 months and 12 months post-implant, revealed discrete changes in the RT-interval and T-wave amplitude over time (Figure 2).
|
|
To prevent additional episodes of TWO, ventricular sensitivity was reprogrammed to 0.6 mV after ensuring that appropriate VF detection had intra-operatively been tested with a sensitivity of 1.2 mV. Four weeks later, the SIC had not increased any more, and no more atrial or ventricular episodes had been detected by the device.
| Discussion |
|---|
|
|
|---|
Inappropriate ICD therapy occurs in 1520% of the patients and is the most common device-related complication in this population.1
Prior studies showed that TWO is a common cause of inappropriate device therapy.8
Implantable cardioverter-defibrillator devices have to ensure reliable detection of low-amplitude VF and at the same time avoid oversensing. To reduce the likelihood of TWO, most ICDs automatically adjust sensitivity in relation to the amplitude of the preceding R-wave.10
In Medtronic devices, sensitivity is adjusted after each sensed ventricular event to an initial starting value that is related to the R-wave amplitude and the programmed sensitivity. Sensitivity then decreases exponentially until the programmed (maximum) sensitivity value is reached. Therefore, ventricular sensitivity and the consecutive risk for TWO is higher if the time interval between R- and T-wave is longer. In our patient, the occurrence of intermittent TWO was related to a progressive prolongation of the RT-interval and an increase in T-wave amplitude. Considering the algorithm for auto-adjusted sensitivity, both changes probably contributed to the development of TWO in the present case.
Several reversible causes have been associated with the occurrence of TWO (e.g. hyperglycaemia, hyperkalaemia, and use of QT-prolonging drugs). In our patient, we excluded that TWO and the underlying electrographic changes were related to any of these causes. T-wave oversensing and the associated complications resolved after reprogramming ventricular sensitivity from 0.3 to 0.6 mV. Reliable detection of VF always has to be ensured when reprogramming sensitivity to higher values. In the present case, reprogramming was safe because adequate detection had been tested intraoperatively with a sensitivity of 1.2 mV. Furthermore, the sensed R-wave had constantly been
10 mV during the follow-up. In the presence of a such a relatively large R-wave amplitude, changes in programmed sensitivity are known to alter substantially the entire auto-adjusting sensitivity curve.10
The device had been interrogated at another institution after the delivery of 12 shock discharges, but no corrective action had been taken to prevent additional inappropriate therapies. This suggests that inappropriate device therapy and the underlying cause had not been identified. Earlier re-programming of ventricular sensitivity could have prevented TWO and inappropriate therapy in our patient.
In summary, this case illustrates several problems that may be encountered during the ICD follow-up. First, the coincidence of TWO and PVC is identified as an uncommon cause of short VV intervals and a resulting increase in the SIC. Second, we found that slow and progressive changes in the ventricular electrogram may cause TWO and inappropriate ICD therapy late after device implantation. Last, our findings suggest that tailored programming of ventricular sensitivity should be considered to prevent TWO.
| Acknowledgements |
|---|
|
|
|---|
The authors thank Bruce D. Gunderson for his helpful comments.
| References |
|---|
|
|
|---|
[1] Rosenqvist M, Beyer T, Block M, den Dulk K, Minten J, Lindemans F. Adverse events with transvenous implantable cardioverter-defibrillators: a prospective multicenter study. European 7219 Jewel ICD investigators. Circulation 1998; 98: 66370.
[2] Vollmann D, Luthje L, Vonhof S, Unterberg C. Inappropriate therapy and fatal proarrhythmia by an implantable cardioverter-defibrillator. Heart Rhythm 2005; 2: 3079.[CrossRef][Web of Science][Medline]
[3] Vollmann D, Erdogan A, Himmrich E, Neuzner J, Becker D, Unterberg-Buchwald C, et al. Patient alert to detect ICD lead failure: efficacy, limitations, and implications for future algorithms. Europace 2006; 8: 3716.
[4] Ellenbogen KA, Wood MA, Shepard RK, Clemo HF, Vaughn T, Holloman K, et al. Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications. J Am Coll Cardiol 2003; 41: 7380.
[5] Becker R, Ruf-Richter J, Senges-Becker JC, Bauer A, Weretka S, Voss F, et al. Patient alert in implantable cardioverter defibrillators: toy or tool? J Am Coll Cardiol 2004; 44: 958.
[6] Gunderson BD, Patel AS, Bounds CA, Shepard RK, Wood MA, Ellenbogen KA. An algorithm to predict implantable cardioverter-defibrillator lead failure. J Am Coll Cardiol 2004; 44: 1898902.
[7] Gunderson BD, Wood NM, Pearson AM, Wang WX, Olson WH, Investigators at GDCW. A new sensing integrity indicator for ventricular pace/sense leads. Pacing Clin Electrophysiol 2000; 23: 591.
[8] Weretka S, Michaelsen J, Becker R, Karle CA, Voss F, Hilbel T, et al. Ventricular oversensing: a study of 101 patients implanted with dual chamber defibrillators and two different lead systems. Pacing Clin Electrophysiol 2003; 26: 6570.[Medline]
[9] Schulte B, Sperzel J, Carlsson J, Dursch M, Erdogan A, Pitschner HF, et al. Inappropriate arrhythmia detection in implantable defibrillator therapy due to oversensing of diaphragmatic myopotentials. J Interv Card Electrophysiol 2001; 5: 48793.[CrossRef][Web of Science][Medline]
[10] Swerdlow CD and Friedman PA. Advanced ICD troubleshooting: Part I. Pacing Clin Electrophysiol 2005; 28: 132246.[CrossRef][Medline]
![]()