Europace Advance Access originally published online on March 9, 2007
Europace 2007 9(4):203-207; doi:10.1093/europace/eum007
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
ABLATION
Internal atrial shock delivery by standard diagnostic electrophysiology catheters in goats: effects on atrial electrogram amplitude and tissue architecture
1 Department of Cardiology, Leiden University Medical Centre, PO Box 9600, 2300 RC Leiden, The Netherlands; 2 Department of Cardiology, Cardiology Academic Hospital Maastricht, Maastricht, The Netherlands; 3 Department of Pathology, Academic Hospital Maastricht, Maastricht, The Netherlands; 4 Department of Cardiology, Cardiovascular Institute Maastricht, Maastricht, The Netherlands
Manuscript submitted 20 July 2006. Accepted after revision 9 December 2006.
* Corresponding author. Tel: +31715262020; fax: +31715266809. E-mail address: m.c.e.f.wijffels{at}lumc.nl
| Abstract |
|---|
|
|
|---|
Aims In this study, we evaluated the effects of atrial shock delivered via diagnostic electrophysiology catheters.
Methods and results In 11 anaesthetized goats, decapolar catheters were positioned in the right atrial appendage (RAA) and coronary sinus (CS). Three different catheters and two cardioversion protocols were evaluated. In four goats, 50 J shocks were delivered using catheters with 1 mm electrodes (surface area 70 mm2). In 6 goats, catheters with 2 mm electrodes (area 140 mm2) were used. In three of the six goats, 50 J shocks were given while in the other 3, 10 J shocks were delivered. In 1 goat 50 J shocks were delivered via 5 mm electrode catheters (area 310 mm2). No persisting adverse effects occurred. However, the electrogram amplitude at the RAA and CS decreased by >5098% (P > 0.01). The amount of amplitude decrease was most pronounced at the CS site and for 50 J shocks. Goats were sacrificed after 9 ± 1 days. Macroscopy revealed endocardial lesions at the electrode locations. Microscopy showed endocardial thrombosis, and necrosis with formation of granulation tissue. Changes were most marked with diagnostic catheters and 50 J shocks.
Conclusions Atrial shock delivery via diagnostic catheters causes local ablation lesions. The amount of amplitude decrease, macroscopic and microscopic damages were related to the energy applied and electrode surface area.
Key Words: Goat, Internal atrial shock delivery, Cardiac electrophysiology, Endocardial, Shock tissue damage
| Introduction |
|---|
|
|
|---|
In the early 1960s, external electrical cardioversion was introduced as a highly effective and safe method to treat atrial fibrillation (AF).1
Atrial fibrillation may interfere significantly with electrophysiological studies and/or ablation procedures. If AF is long-lasting, either pharmacological or electrical cardioversion is necessary. A major drawback of pharmacological cardioversion is that antiarrhythmic drugs change the electrical properties of the heart, prohibiting a correct electrophysiological evaluation. For this reason, internal cardioversion under light sedation may be preferred. Although internal cardioversion by specially designed catheters has been shown to be safe and effective in humans,5
9
an alternative would be to deliver shocks via previously introduced diagnostic electrophysiology catheters. However, due to large differences in catheter design [cardioversion catheters typically have multiple large (e.g. 5 mm) electrode poles or incorporated shocking coils]5
7
,13
data from these cardioversion studies cannot be extrapolated to cardioversion with standard diagnostic catheters. Because only very limited data about the efficacy and safety of cardioversion via diagnostic catheters are available,14
,15
it was the aim of our present study to evaluate the safety and the electrophysiological and histological effects of intracardiac shock delivery performed by conventional diagnostic electrophysiology catheters in goats.
| Methods |
|---|
|
|
|---|
Instrumentation and handling of animals
Eleven female goats weighing between 49 and 82 kg (63 ± 11 kg) were used for this study. Animal handling was performed according to the Dutch Law on Animal Experimentation and the European Directive for the Protection of Vertebrate Animals used for experimental and other scientific purposes. The protocol was approved by the Animal Investigation Committee of Maastricht University. Goats were anaesthetized with thiopental (Nesdonal® 1015 mg/kg i.v., Merial, Velserbroek, The Netherlands), intubated and ventilated with halothane (12%) and a 1 : 2 mixture of O2 in N2O. During the experiments, animals were monitored by ECG and expiratory CO2 content. Through the external jugular and hind leg veins 23 sheaths (7 French) were introduced by a modified Seldinger technique. Two 6 French steerable decapolar diagnostic electrophysiology catheters (LivewireTM, St Jude Medical, Daig Division, Minnetonka, MN, USA) were positioned in the right atrial appendage (RAA) and coronary sinus (CS) under fluoroscopic guidance. In the first three animals, a third multipurpose diagnostic catheter was introduced via the jugular vein to visualize the CS by contrast injection. To prevent volume depletion, intravenous fluid (Ringers lactate) was given during the experiment (250 mL/h). In addition, sodiumampicillin (Pentrexyl® 1000 mg, Bristol-Myers Squibb, Woerden, The Netherlands) was given intravenously once directly before and once after surgery. Post-operatively the animals received buprenorfine (Temgesic® 0.015 mg/kg i.m., Reckitt & Colman, Hull, England).
Electrophysiological study and protocols for shock delivery
After induction of anaesthesia and positioning of the catheters in the RAA and the CS, the electrodes were connected to a custom-built switch box. This switch box allowed connection of the electrodes to either the output of an external defibrillator (Ventitrex HVS-02, division of St Jude Medical, Sunnyvale, CA, USA.) or to a customized acquisition system (PC-EMS 3.0 Cardiovascular Research Institute, Maastricht, The Netherlands) for off-line analysis of electrograms. A total of 10 QRS-synchronized shocks (biphasic, duration 6/6 ms) were administered in all goats. The time interval between the shocks was 1 min.
As can be seen in Table 1, three different decapolar catheters and two cardioversion shock protocols were tested. In the first four goats, diagnostic electrophysiology catheters consisting of 10 parallel 1 mm electrode rings with an electrode spacing of 252 mm and an active surface area of 70 mm2
(LivewireTM, St Jude Medical) were used and 10 shocks of 50 J were given. In another six goats, similar catheters with 2 mm electrode rings and an active surface area of 140 mm2
were used (LivewireTM, St Jude Medical). In 3 of these 6 goats, 10 shocks of 50 J were given while in the other 3, 10 shocks of only 10 J were delivered. In the remaining goats, cardioversion catheters consisting of 10 parallel 5 mm electrode rings with an active surface area of 310 mm2
were used (ElecathTM Electro-Catheter Corp., Rahway, NJ, USA) and 10 shocks of 50 J were given. For each shock, the impedance and effective amount of energy applied was measured (Table 1). The effect of shocks on local electrogram amplitude was analysed off-line, for one of the bipolar electrodes at the middle of the RAA and CS catheters. The average electrogram amplitude of three consecutive beats was measured directly before the first shock and 1 min after the last (10th) shock was given. Catheters and sheets were removed, and a compressive bandage was applied to the puncture site for several hours.
|
Microscopic and macroscopic evaluation
One week (9 ± 1 days, range 710 days) after the experiments, animals were sacrificed by Euthesate® (200 mg/kg i.v., Apharmo, Arnhem, The Netherlands) and the heart was excised by a lateral left thoracotomy. The atrium was opened, and the RAA and CS regions were inspected for macroscopic damage. In all goats, representative tissue samples were taken from areas with an abnormal macroscopic appearance (e.g. haematoma or thrombus). Tissue samples were fixed in 4% zinc-buffered formalin and embedded in paraffin. Sections (4 µm) were cut and processed for haematoxylin and eosin staining and evaluated by light microscopy.
Statistical analysis
Data are presented as mean ± SD. Statistical analysis was performed by either a unpaired Student's t-test (Table 1, comparison between shock voltage for 10 and 50 J shocks and shock impedance for 1 and 2 mm electrode catheters), a Wilcoxon signed-rank test (comparison electrogram amplitude before and after shock delivery) or by a MannWhitney U test (comparison post-shock electrogram amplitude between RA and CS and between 10 and 50 J shocks). A P value of >0.05 was considered to be statistically significant.
| Results |
|---|
|
|
|---|
Safety of internal atrial shocks and effect on electrogram amplitude
The delivery of intracardiac shocks did not result in persistent or serious complications and all animals survived the procedure. Temporary adverse effects were almost exclusively observed during the acute experiments in animals receiving high energy shocks of 50 J. Non-sustained episodes of AF and atrioventricular (AV) block of >5 s occurred immediately after the delivery of a 50 J shock in two of four goats, in which the 1 mm electrode catheters were used, and in 1 of 3 goats in which the 2 mm electrode catheters were used. These effects were not observed in animals receiving lower energy shocks of 10 J. Because asystole (caused by complete AV block) never lasted >5 s, ventricular pacing was not required. Although in none of the animals (sustained) ventricular tachycardia was seen, several ventricular beats with a widened QRS complex were seen immediately after shock delivery in 3 of 4 goats, in which 50 J shocks were applied via the 1 mm electrode catheters, and in 2 of 3 goats receiving 50 J shocks and in 1 of 3 goats who had 10 J shocks via catheters with 2 mm electrodes. At morbid anatomical examination, no signs of perforation were present.
In Figure 1, the effects of internal atrial shocks on the RAA and CS electrogram amplitude are shown according to the type of decapolar catheter and shock protocol used. In all the experiments and at both locations, the electrogram amplitude decreased markedly after atrial shock delivery (RAA, P = 0.003; CS, P = 0.005). The decrease in electrogram amplitude was larger for the CS (92%) compared with the RAA site (76%) (P = 0.024) and for 50 J shocks (89%) compared with 10 J shocks (67%) (P = 0.023). Figure 2 shows an example of the effect of 10 intracardiac 50 J shocks on the amplitude of local RAA and CS electrograms. Importantly, we observed that a single shock of 1050 J reduced electrogram amplitude to >1030% of baseline amplitude in most goats. Again, this decrease was less marked at the RAA compared with the CS (data not shown).
|
|
Macroscopic and microscopic evaluation
After 9 ± 1 (range 710) days, goats were sacrificed and the heart excised. Opening of the right atrium showed several endocardial lesions in all but one goat (number 2). These lesions, found at the predicted location of the RAA and CS electrodes, had a typical size of 12 mm and a typical separation between lesions of about 5 mm (skip lesions; Figure 3), which corresponded with the design of the catheters used. Although there were no clear differences in the number of endocardial lesions found between the goats tested with the different electrodes (1 vs. 2 mm) or different shock protocols (50 vs. 10 J), they were less pronounced in animals in which only shocks of 10 J were applied. In the single goat, in which the ElecathTM catheters were used the lesions were more elongated and superficial. Additionally, in several animals endocardial haematomas were observed, often in the CS or close to the ostium of the CS, which may have been caused by catheter manipulation rather than by the effect of the shocks applied.
|
Microscopy of the lesions showed an endocardial, partly organized thrombus and necrosis of the underlying adipose tissue in atrial and even more deeply in ventricular myocardium (Figure 4). Formation of granulation tissue was present, compatible with myocardial necrosis caused 1 week earlier. These changes were most marked in goats in which diagnostic catheters and high energy shocks of 50 J were used, whereas they were less prominent for the 5 mm ElecathTM catheters and almost absent in the goats in which shocks of only 10 J were applied.
|
| Discussion |
|---|
|
|
|---|
Several studies have shown that internal cardioversion of AF in humans by specialized catheters is effective and safe.5
Adverse events and effect on electrogram amplitude
Although we did not observe that DC shocks via standard decapolar electrophysiology catheters caused serious (persistent) adverse effects, our results indicate that intracardiac shocks at 50 J can cause local damage to the endocardium, (atrial) myocardium and fat tissue. In all the experiments, using different electrode designs (1 and 2 mm electrode rings) and different shock energies (1050 J), we observed a marked decrease in electrogram amplitude. However, this effect was most obvious with high energy shocks of 50 J and for the CS site (Figure 1). The most likely explanation for the latter observation is that the position of the catheter in the CS was much more fixed compared with the RAA. Although the tip of this catheter had its position in the RAA, the tissue contact of the more proximal electrodes may have been less stable due to movements caused by contraction of the heart, respiration and shocks. Interestingly, also in the animals in which the specialized cardioversion catheters were used (ElecathTM), a marked decrease in local electrogram amplitude was found (RAA 50% and CS 98%). As far as we know, no study on internal cardioversion using these catheters in animals or humans has ever reported such an effect. Although it can be argued that this is caused by the use of lower energy shocks, it is more likely that it has never previously been systematically evaluated. A decrease in electrogram amplitude probably remained unnoticed, because atrial amplitudes are in general smaller during AF (before cardioversion) than during sinus rhythm (after cardioversion).
Effects on macroscopy and microscopy
Macroscopic evaluation 710 days after the acute experiments showed several endocardial lesions corresponding to the position of the electrodes on the decapolar catheters. Although the sample size of our study was too small to allow precise quantification of these lesions, we observed the most pronounced effects in goats in which small electrodes rings (1 vs. 2 or 5 mm) associated with higher shock impedances and higher shock intensities (50 vs. 10 J) were used. Microscopy of these lesions showed endocardial thrombus formation and necrosis of the superficial atrial myocardium, the underlying adipose tissue and even of the deeper ventricular myocardium. In animals in which shocks of 10 J were given (2 mm electrode rings), only superficial macroscopic abnormalities were seen (discolouration) and microscopic signs of necrosis were minimal. For this reason, the decrease in electrogram amplitude observed in these animals is most likely caused by (reversible) oedema rather than by ablation of the underlying tissue. Although, as far as we know, no previous study has ever evaluated macroscopic and/or microscopic damage caused by internal atrial cardioversion shocks, our findings are not very unexpected for several reasons. Before the introduction of radiofrequency ablation, intracardiac DC shocks have been used for ablation of the His bundle and accessory pathways,16
,17
indicating that these shocks may cause considerable myocardial damage. This was confirmed in a study in dogs, showing that high energy DC shocks (50400 J) applied to the left ventricle resulted in haemorrhagic areas and contraction band necrosis.18
In addition, several studies in patients showed that internal cardioversion with relatively low energies (up to 16 J) and specialized catheters resulted in a slight elevation of cardiac biomarkers suggesting minor myocardial injury.19
,20
Clinical implications
Previous studies on cardioversion with diagnostic catheters are limited and did not evaluate the effects of internal shocks on electrogram amplitude or (macroscopic or microscopy) damage to underlying tissues.14
,15
We believe that our findings have important clinical implications. The smaller the electrode surface area which is related to a higher shock impedance (see Table 1) and the higher the shock intensity used, the higher the current density will be and the more likely that significant local damage of the endocardial, myocardial, and adipose tissues will occur. Further supported by the finding of thrombus formation on the lesions we strongly feel that cardioversion with regular diagnostic catheters should bot be performed with catheters have an even smaller active surface area (e.g. quadripolar catheters) and should be limited to lower energies only (e.g. >15 J). However, even with the use of such a conservative approach, and illustrated by our observation that even after a single shock of 10 J electrogram amplitude could be reduced markedly, it should be realized that information on local electrogram characteristics may be modified and partially lost. For instance, in patients with a left-sided accessory pathway, cardioversion with a catheter in the CS may cause loss of electrogram amplitude which may make it more difficult to localize the optimal site for ablation.
Limitations of the study
It was the aim of this study to evaluate the safety and the electrophysiological and histological effects of internal shock delivery via diagnostic electrophysiology catheters. For this reason, we used a protocol with several shocks of a fixed intensity and did not evaluate the feasibility of internal cardioversion to restore sinus rhythm. The feasibility to cardiovert AF with diagnostic electrophysiology catheters has been described by others and was not the purpose of our study.14
,15
Because our histological analysis was limited to RAA and CS areas showing macroscopic damage, we cannot exclude that less obvious histological abnormalities (limited to the microscopic level) were missed in some goats.
Finally, we are aware of the fact that the number of animals used is quite limited. However, our findings are consistent and logical in such a way that high energy shocks, a small electrode size, and high shock impedance resulted in the most prominent tissue damage. We were primarily not interested in quantifying these effects in more detail and believed that increasing the number of experiments would not change our conclusions.
| Conclusions |
|---|
|
|
|---|
Our study indicates that intracardiac shock delivery via diagnostic decapolar electrophysiology catheters did not result in serious side effects and for this reason appears to be safe. However, the smaller the size of electrodes used for shock delivery and the higher the shock impedance and intensity, the more likely that cardioversion results in a marked decrease in electrogram amplitude and ablation of underlying (myocardial) tissues. These findings may have implications for clinical practice, suggesting that internal cardioversion via regular diagnostic electrophysiology catheters should preferably be limited to delivery of a limited amount of low energy shocks.
| Acknowledgements |
|---|
|
|
|---|
The authors would like to thank S. Philippens, E. van der Veen, and T. van der Nagel for their support during the experiments and St Jude Medical (The Netherlands) for their material and financial support.
| References |
|---|
|
|
|---|
[1] Lown B, Perlroth MG, Kaidbey S, Abe T, Harken DE. Cardioversion of atrial fibrillation. A report on the treatment of 65 episodes in 50 patients. N Engl J Med 1963; 269: 32531.[Web of Science][Medline]
[2] Morris JJ Jr, Kong Y, North WC, McIntosh HD. Experience with Cardioversion of atrial fibrillation and flutter. Am J Cardiol 1964; 14: 94100.[CrossRef][Web of Science][Medline]
[3] Powell AC, Garan H, McGovern BA, Fallon JT, Krishnan SC, Ruskin JN. Low energy conversion of atrial fibrillation in the sheep. J Am Coll Cardiol 1992; 20: 70711.[Abstract]
[4] Cooper RA, Alferness CA, Smith WM, Ideker RE. Internal cardioversion of atrial fibrillation in sheep. Circulation 1993; 87: 167386.
[5] Cooper RA, Johnson EE, Wharton JM. Internal atrial defibrillation in humans. Improved efficacy of biphasic waveforms and the importance of phase duration. Circulation 1997; 95: 148796.
[6] Levy S, Ricard P, Gueunoun M, Yapo F, Trigano J, Mansouri C, et al. Low-energy cardioversion of spontaneous atrial fibrillation. Immediate and long-term results. Circulation 1997; 96: 2539.
[7] Timmermans C, Rodriguez LM, Ayers GM, Lambert H, Smeets J, Wellens HJ. Effect of electrode length on atrial defibrillation thresholds. J Cardiovasc Electrophysiol 1998; 9: 5827.[Web of Science][Medline]
[8] Taramasco V, Socas A, Ricard P, Levy S. Internal low-energy cardioversion: a therapeutic option for restoring sinus rhythm in chronic atrial fibrillation after failure of external cardioversion. Europace 1999; 1: 17982.
[9] Karch MR, Schmieder S, Ndrepepa G, Schneider MA, Zrenner B, Schmitt C. Internal atrial defibrillation during electrophysiological studies and focal atrial fibrillation ablation procedures. Pacing Clin Electrophysiol 2001; 24: 14649.[CrossRef][Medline]
[10] Levy S, Lauribe P, Dolla E, Kou W, Kadish A, Calkins H, et al. A randomized comparison of external and internal cardioversion of chronic atrial fibrillation. Circulation 1992; 86: 141520.
[11] Schmitt C, Alt E, Plewan A, Ammer R, Leibig M, Karch M, et al. Low energy intracardiac cardioversion after failed conventional external cardioversion of atrial fibrillation. J Am Coll Cardiol 1996; 28: 9949.[Abstract]
[12] Levy S. Internal defibrillation: where we have been and where we should be going? J Interv Card Electrophysiol 2005; 13: 616.
[13] Boriani G, Biffi M, Camanini C, Sammali A, Bacchi L, Accorti P, et al. Transvenous internal cardioversion for atrial fibrillation: A randomized comparison between catheters with different coil length. Am Heart J 2002; 144: 8517.[CrossRef][Web of Science][Medline]
[14] Raitt MH, Kusumoto W, Giraud GD, McAnulty JH. Electrophysiologic predictors of the recurrence of persistent atrial fibrillation within 30 days of cardioversion. Am J Cardiol 2004; 93: 10710.[CrossRef][Web of Science][Medline]
[15] Kusumoto FM. Internal atrial and ventricular defibrillation during electrophysiology procedures. J Interv Card Electrophysiol 2005; 13:Suppl. 1, 718.[Web of Science][Medline]
[16] Scheinman MM, Morady F, Hess DS, Gonzalez R. Catheter-induced ablation of the atrioventricular junction to control refractory supraventricular arrhythmias. JAMA 1982; 248: 8515.
[17] Davis MJ, Mews GC, Cope GD. Transvenous ablation of atrioventricular conduction for refractory or malignant supraventricular arrhythmias. Aust N Z J Med 1984; 14: 47986.[Web of Science][Medline]
[18] Lerman BB, Weiss JL, Bulkley BH, Becker LC, Weisfeldt ML. Myocardial injury and induction of arrhythmia by direct current shock delivered via endocardial catheters in dogs. Circulation 1984; 69: 100612.
[19] Boriani G, Biffi M, Cervi V, Bronzetti G, Magagnoli G, Zannoli R, et al. Evaluation of myocardial injury following repeated internal atrial shocks by monitoring serum cardiac troponin I levels. Chest 2000; 118: 3427.
[20] Gorenek B, Kudaiberdieva G, Goktekin O, Cavusoglu Y, Birdane A, Ata N, et al. Detection of myocardial injury after internal cardioversion for atrial fibrillation. Can J Cardiol 2004; 20: 1658.[Web of Science][Medline]
![]()
CiteULike
Connotea
Del.icio.us What's this?
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||