Europace Advance Access originally published online on January 9, 2007
Europace 2007 9(2):127-129; doi:10.1093/europace/eul176
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
ELECTROPHYSIOLOGY
Ischaemia modified albumin in radiofrequency catheter ablation
2nd Department of Cardiology, Onassis Cardiac Surgery Center, 356 Syngrou Ave, 17674 Athens, Greece
Manuscript submitted 26 May 2006. Accepted after revision 4 November 2006.
* Corresponding author. Tel: +210 9493 372; fax: +210 9493 373. E-mail address: esbarouni{at}yahoo.gr
| Abstract |
|---|
|
|
|---|
Aim Ischaemia modified albumin (IMA) is considered a marker of myocardial ischaemia, in contrast to the biomarkers of myocardial injury [creatine kinase (CK), the MB isoenzyme of CK, and cardiac troponin I (Tn-I)] that are released when cardiac necrosis occurs. Ischaemia modified albumin has been reported to increase following percutaneous coronary intervention and in acute coronary syndromes. We sought to determine whether IMA increases following radiofrequency (RF) ablation.
Methods and results We studied 40 consecutive patients who underwent RF catheter ablation; 20 were men and 20 women and their age was 47 ± 16 (1677) years. All patients underwent electrophysiological study and subsequent RF ablation. Peripheral venous samples were collected before the procedure (baseline), immediately after the procedure, 2 h post-procedure and the following day (20 h post-procedure) and assayed for CK, the MB isoenzyme of CK, cardiac Tn-I and IMA. Ischaemia-modified albumin plasma levels did not differ significantly at all four time points, baseline, and following ablation (P = 0.5974), whereas CK, CK-MB, and Tn-I increased significantly at all time points compared with baseline (P < 0.0001). Post-ablation, all but three 3 CK measurements were in the normal range; 14 patients had CK-MB plasma levels above the upper limit of normal; all but one patient had Tn-I elevated.
Conclusion The IMA plasma levels do not change significantly following RF ablation, unlike biomarkers of myocardial injury, implying that myocardial necrosis occurs without preceding ischaemia.
Key Words: Ischaemia modified albumin, Radiofrequency ablation, Biomarkers of myocardial necrosis
| Introduction |
|---|
|
|
|---|
Ischaemia-modified albumin (IMA), as assessed by the albumin cobalt binding (ACB) test, is considered a marker of myocardial ischaemia, in contrast to biomarkers of myocardial injury [CK, the MB isoenzyme of CK, and cardiac troponin I (Tn-I)] that are released when cardiac necrosis occurs. Ischaemia, through hypoxia, acidosis, sodium, and calcium pump disruptions and free radical injury, may induce changes in the binding capacity of the NH2 terminus of albumin to bind metals such as cobalt (Co), copper (Cu), and nickel (Ni). Ischaemia modified albumin has been reported to increase following percutaneous coronary intervention (PCI), immediately post-balloon inflation and 30 min after and return to baseline in 612 h.1
| Methods |
|---|
|
|
|---|
We studied 40 consecutive patients who underwent RF catheter ablation between November 2004 and April 2005 in our unit; 20 were men and 20 women and their age was 47 ± 16 (range 1677) years. All patients gave written informed consent and the study protocol was approved by the Ethics Committee of our Hospital.
All patients underwent electrophysiological study and subsequent RF ablation. The indication for intervention was WolffParkinsonWhite syndrome (WPW) in 8 patients, atrioventricular nodal reentrant tachycardia (AVNRT) in 13, atrial tachycardia in 4, atrial flutter in 3, and atrial fibrillation-atrial flutter in 12. All our patients were in a fasting state and were mildly sedated. The number of applications varied greatly and was 30 ± 36 (1151); the delivered energy was 1140 ± 1593 W (305000 W), for a duration of 1160 ± 1625 s (607320 s) and at a temperature of 56 ± 7°C (4570°C).
Peripheral venous samples were collected before the procedure (baseline), immediately after the procedure, 2 h post-procedure and the following day (20 h post-procedure), via an indwelling catheter. The samples were assayed for CK, the MB isoenzyme of CK (CK-MB), cardiac Tn-I and IMA.
Creatine kinase was measured on a Cobas Integra 800 analyser (Roche, Basel, Switzerland), as was the IMA by the ACB test. Creatine kinase-MB and Troponin-I were measured by immunoassay on the Dimension system (Dade Behring, Newark, NJ, USA). We compared our results using ANOVA with repeated measures; pair comparisons were performed with the Wilcoxon test.
| Results |
|---|
|
|
|---|
Results (Table 1 and Figure 1) are expressed as mean ± SD (minimummaximum). We found, using ANOVA with repeated measures, that IMA plasma levels did not differ significantly at all four time points, baseline and following ablation (P = 0.5974). Ischaemia modified albumin plasma levels were very similar to each other at all time points, with tide distribution; furthermore, they did not differ according to the number of applications or the type of the catheter used (irrigated catheters in 12 patients and non-irrigated in 28). Creatine kinase, CK-MB, and Tn-I increased significantly compared with baseline (P < 0.0001); their levels at all time points post-procedure were compared with baseline by the Wilcoxon test and were found to be significantly (P < 0.0001). Post-ablation, all but three CK measurements were in the normal range (CK < 190 mU/mL); 14 patients had CK-MB plasma levels above the upper limit of normal (CK-MB > 3.6 ng/mL); all but one patient had Tn-I elevated (Tn-I > 0.1 ng/mL).
|
|
| Discussion |
|---|
|
|
|---|
We found that IMA plasma levels did not change immediately after 2 and 20 h after RF ablation compared with baseline values, whereas CK, CK-MB, and Tn-I increased significantly compared with baseline; only 3 patients had, however, CK measurements above the normal range, less than half (14) had CK-MB measurements above the upper limit of normal but all except one patient had Tn-I elevated.
Cardiac Tn-I is a highly specific and sensitive marker of myocardial injury; all but one patient had elevated Tn-I, implying that ablation does cause myocardial cell death. However, this myocardial necrosis is minor as the other markers (CK and CK-MB) remained within normal limits or showed only slight changes. This is in agreement with the previous studies.10
14
We found no significant changes in IMA plasma levels. There is only one published report on IMA in RF ablation, which showed that IMA significantly increases 30 min after the procedure and returns to baseline values in 8 h9
; in our study, blood sampling was performed immediately after 2 and 20 h after the procedure. The different timing may explain the contradictory findings between the latter study and ours, although the effects of ischaemia on N-terminal region of albumin which reduce the binding of cobaltium can be detectable up to 6 h after the ischaemic event.15
Thus, we think that the most likely explanation is that RF ablation induces immediate myocardial necrosis without preceding myocardial ischemia; this is in distinction from PCI and ACS where complete coronary occlusion, usually transient and repetitive, occurs and causes prolonged myocardial ischaemia which subsequently leads to myocardial necrosis. Clearly further studies are required to clarify this.
There is one study reporting that IMA plasma levels increase 1 and 6 h following cardioversion in patients with AF, but whether this reflects myocardial or skeletal ischaemia is not clear;8
CK and CK-MB rise in this setting has been attributed to skeletal muscle damage rather than myocardial necrosis.
In conclusion, IMA plasma levels do not change significantly following RF ablation, whereas biomarkers of myocardial necrosis significantly increase due to minor myocardial injury; this possibly implies that myocardial necrosis, in RF ablation, occurs without preceding ischaemia. In addition, elevated levels of IMA in the peri-ablation period may suggest incidental myocardial ischaemia.
| References |
|---|
|
|
|---|
[1] Bar-Or D, Winkler JV, VanBenthysen K, Harris L, Lau E, Hetzel FW. Reduced albumin-cobalt binding with transient myocardial ischemia after elective percutaneous transluminal coronary angioplasty a preliminary comparison to creatine kinase-MB, myoglobin,: troponin I. Am Heart J 2001; 141: 985991.[CrossRef][Web of Science][Medline]
[2] Sinha MK, Gaze DC, Tippins JR, Collinson PO, Kaski JC. Ischemia modified albumin is a sensitive marker of myocardial ischemia after percutaneous coronary intervention. Circulation 2003; 107: 24032405.
[3] Quiles J, Roy D, Gaze D, Garrido IP, Avanzas P, Sinha M, et al. Relation of ischemia-modified albumin (IMA) levels following elective angioplasty for stable angina pectoris to duration of balloon-induced myocardial ischemia. Am J Cardiol 2003; 92: 322324.[CrossRef][Web of Science][Medline]
[4] Garrido IP, Roy D, Calvino R, Vazquez-Rodriguez JM, Aldama G, Cosin-Sales J, et al. Comparison of ischemia-modified albumin levels in patients undergoing percutaneous coronary intervention for unstable angina pectoris with versus without collaterals. Am J Cardiol 2004; 93: 8890.[CrossRef][Web of Science][Medline]
[5] Christenson RH, Duh SH, Sanhai WR, Wu AH, Holtman V, Painter P, et al. Characteristics of an albumin binding test for assessment of acute coronary syndrome patients: a multicenter study. Clin Chem 2001; 47: 464470.
[6] Bhagavan NV, Lai EM, Rios PA, Yang J, Ortega-Lopez AM, Shinoda H, et al. Evaluation of human albumin cobalt binding assay for the assessment of myocardial ischemia and myocardial infarction. Clin Chem 2003; 49: 581585.
[7] Sinha M, Roy D, Gaze D, Collinson P, Kaski JC. Role of ischemia-modified albumin a new biochemical marker of myocardial ischemia in the early diagnosis of acute coronary syndromes. Emerg Med J 2004; 21: 2934.
[8] Roy D, Quiles J, Sinha M, Aldama G, Gaze D, Kaski JC. Effect of direct-current cardioversion on ischemia-modified albumin in patients with atrial fibrillation. Am J Cardiol 2004; 93: 366368.[CrossRef][Web of Science][Medline]
[9] Roy D, Quiles J, Sinha M, Floros D, Gaze D, Collinson P, et al. Effect of radiofrequency catheter ablation on the biochemical marker ischemia-modified albumin. Am J Cardiol 2004; 94: 234236.[CrossRef][Web of Science][Medline]
[10] del Rey JM, Madrid AH, Valino JM, Rubi J, Mercader J, Moro C, et al. Cardiac troponin I and minor cardiac damage: biochemical markers in a clinical model of myocardial lesions. Clin Chem 1998; 44: 22702276.
[11] Madrid AH, del Rey JM, Rubi J, Ortega J, Gonzalez Rebollo JM, Seara JG, et al. Biochemical markers and cardiac troponin I release after radiofrequency catheter ablation: approach to size of necrosis. Am Heart J 1998; 136: 948955.[CrossRef][Web of Science][Medline]
[12] Manolis AS, Vassilikos V, Maounis T, Melita-Manolis H, Psarros L, Haliasos A, et al. Detection of myocardial injury during radiofrequency catheter ablation by measuring serum cardiac troponin I levels: procedural correlates. J Am Coll Cardiol 1999; 34: 10991105.
[13] Rao SP, Miller S, Rosenbaum R, Lakier JB. Cardiac troponin I cardiac enzymes after electrophysiologic studies, ablations,: defibrillator implantations. Am J Cardiol 1999; 84: 470.
[14] Brueckmann M, Wolpert C, Bertsch T, Sueselbeck T, Liebetrau C, Kaden JJ, et al. Markers of myocardial damage, tissue healing, and inflammation after radiofrequency catheter ablation of atrial tachyarrythmias. J Cardiovasc Electrophysiol 2004; 15: 686691.[CrossRef][Web of Science][Medline]
[15] Berenshtein E, Mayer B, Goldberg C, Kitrossky N, Chevion M. Patterns of mobilization of copper iron following myocardial ischemia: possible predictive criteria for tissue injury. J Mol Cell Cardiol 1997; 29: 30253034.[CrossRef][Web of Science][Medline]
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
P O Collinson and D C Gaze Ischaemia-modified albumin: clinical utility and pitfalls in measurement J. Clin. Pathol., September 1, 2008; 61(9): 1025 - 1028. [Full Text] [PDF] |
||||
![]() |
D. Roy and J. C. Kaski IMA the role of oxidative stress Europace, September 1, 2007; 9(9): 854 - 854. [Full Text] [PDF] |
||||
![]() |
E. Sbarouni IMA the role of oxidative stress: reply Europace, September 1, 2007; 9(9): 854 - 855. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||