Europace 2007 9(2):105-107; doi:10.1093/europace/eul143
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Sharif, A. Y.
Right arrow Articles by Sadeghian, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sharif, A. Y.
Right arrow Articles by Sadeghian, S.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The European Society of Cardiology 2007. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org


PACING

New technique: repositioning of dislodged atrial pacing lead with a specially designed urological basket

Ahmad Yamini Sharif*, Gholamreza Davoodi, Ali Kazemi Saeed and Saeed Sadeghian

Tehran Heart Center, Tehran University of Medical Sciences, Jalal Al Ahmad and North Kargar Cross, PO Box 1411713138, Tehran, Islamic Republic of Iran

Manuscript submitted 10 March 2006. Accepted after revision 3 October 2006.

* Corresponding author. Tel: +0098 21 88029600 69; fax: +0098 21 88029731. E-mail address: yaminisharif.a{at}doctor.com


    Abstract
 Top
 Abstract
 Introduction
 Case 1
 Methods
 Case 2
 Discussion
 References
 
The rate of dislodgement of atrial pacing leads is ~3%. To solve this problem, reoperation and repositioning of these leads is one of the solutions. Some operators have reported repositioning these leads with snare systems or deflectable catheters. In this communication, we present a new method using a specially designed urological basket to solve this problem.

Key Words: Atrial pacing lead, Dislodgement, Urological basket


    Introduction
 Top
 Abstract
 Introduction
 Case 1
 Methods
 Case 2
 Discussion
 References
 
Lead dislodgement requiring reoperation is a complication that raises the costs of pacemaker implantation surgery, while adding to patient discomfort. Migration of a dislodged lead can cause complications such as undersensing, loss of atrial capture, and loss of atrial function in patients who need AV synchrony. Once detected, the dislodged lead should immediately be repositioned.1Go,2Go

In recent years, percutaneous transcatheter repositioning of displaced permanent pacemaker leads has been advocated before consideration of standard repositioning by reopening the generator pocket. The procedure is easy and safe, allowing a reduction of the need for surgical lead revision and the associated morbidity and cost.3Go–5Go In this report, we introduce a method with a specially designed urological basket to solve this problem in two patients.


    Case 1
 Top
 Abstract
 Introduction
 Case 1
 Methods
 Case 2
 Discussion
 References
 
A 68-year-old man was admitted in January 2006 with dyspnoea, mild dizziness, and evidence of pacemaker malfunction due to atrial lead dislodgement. He had a history of coronary bypass surgery in 1985 and angioplasty (implantation of seven stents in two stenotic saphenous vein grafts) in 2003. In 2004, he experienced an inferior MI, after which he had transient complete heart block accompanied by respiratory arrest and cardiogenic shock, which was treated medically and the block resolved. In the next admission (4 months later) he had first-degree atrioventricular block and also had episodes of complete heart block accompanied by dizziness and exacerbation of dyspnoea. Electrocardiogram showed narrow complexes and old inferior MI. In the latest echocardiographic study ejection fraction was 40% and there was no evidence of significant dyssynchrony; we therefore decided to implant a dual chamber pacemaker.

The implantation was successful, and the patient was discharged in good condition in November 2005. At follow-up, pacemaker analysis showed no atrial sensing and pacing. Programming the pacemaker to VVIR mode did not relieve symptoms, and the patient was scheduled for repositioning of atrial lead, which was performed with this specially designed urological basket in February 2006.

Further description of specially designed urological basket (cardiac pacing lead hook)
This device is specially designed based on the non-metallic urological stone basket known as the Dormia basket that is used for removal of stones that are located in the ‘lower ureter’.

By making changes, its function is altered so that instead of forming a basket, the operator can form a hook around the cardiac pacing lead by pulling the steerable inner line of the device. (Figure 1A–C).


Figure 1431
View larger version (94K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 1 (AC) These show the specially designed urological basket and how it makes a hook. (By inactivating three lines of basket (arrow) only one line is active (arrow head) So by dragging the lines, the head bends towards the active line and makes a hook.

 
In the next step the operator is able to reposition the lead by traction on the device.

To remove the hook, the operator releases the inner line; this will allow the hook to be straightened as it enters again into the delivery system (Mullins sheath).


    Methods
 Top
 Abstract
 Introduction
 Case 1
 Methods
 Case 2
 Discussion
 References
 
To perform repositioning of the atrial pacing lead, the patient was brought to the catheterization laboratory in the post-absorptive non-sedated state. The Mullins sheath was placed around the atrial lead of pacemaker and by using deflectable ablation catheter. The sheath was then positioned over the loop of the atrial pacing lead (Figure 2).


Figure 1432
View larger version (113K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 2 This figure shows how to position the Mullins sheath (single arrow head) over the loop of dislodged atrial lead of pacemaker (double arrow head).

 
After removal of the ablation catheter, the specially designed urological basket was passed over the loop of the J-shaped atrial lead. The deflectable part of this basket was passed through the tip of the Mullins sheath. By deflecting the tip of the basket, a hook was formed around the lead (Figure 3A–C). By pulling back the basket and Mullins sheath (Figure 3D), the tip of the atrial lead could be put in a new position in the atrium. Immediately, pacing analysis was performed, atrial pacing and sensing parameters were acceptable. By pulling back the basket catheter into the Mullins sheath (Figure 3F), all devices were removed.


Figure 1433
View larger version (54K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 3 (AF) These show how the basket (arrow) makes a hook around the pacemaker lead (single arrow head) and after traction of the lead, it can be removed.

 

    Case 2
 Top
 Abstract
 Introduction
 Case 1
 Methods
 Case 2
 Discussion
 References
 
A 68-year-old man was admitted in January 2003 with dyspnoea and dizzy spells. His ECG showed complete heart block. Echocardiographic ejection fraction was 40%. Coronary angiography showed borderline lesions (about 50%) in the left anterior descending and circumflex arteries. A dual chamber pacemaker was implanted and the patient was discharged. Four weeks later he came to the pacemaker clinic because of diaphragmatic stimulation. He reported that this symptom began after a severe bout of sneezing. His chest X-ray revealed migration of the atrial lead of his pacemaker to the superior vena cava (SVC). By changing the pacemaker mode to VVIR diaphragmatic stimulation ceased, but the patient showed signs of pacemaker syndrome. We therefore decided to perform repositioning of the dislodged atrial lead using the specially designed urological basket. The dislodged atrial lead was repositioned into right atrium, but one day later this J-shaped passive fixation atrial lead migrated again into the SVC and the patient preferred his pacemaker problem to be corrected by reoperation. In the second operation session another atrial lead (an active fixation lead) was inserted into right atrial appendage.


    Discussion
 Top
 Abstract
 Introduction
 Case 1
 Methods
 Case 2
 Discussion
 References
 
This experience highlights the feasibility of repositioning of pacemaker atrial leads without reopening the generator pocket by using a new device. The precondition for success is the presence of a passive fixation J-shaped atrial lead. There are reports of repositioning of dislodged atrial pacing leads by snare systems or deflectable catheters.3Go–5Go

To reposition the dislodged atrial pacing lead by a snare system, the tip of the dislodged atrial pacing lead should move freely in the right atrium or SVC.

There are some devices such as Needle's Eye®snare that is designed to grasp objects without a free end, but these devices are used for percutaneous retrieval of cardiac leads and, to our knowledge, they have not been used for lead repositioning.

When the tip of a dislodged atrial pacing lead is attached to the wall of the atrium or SVC, deflectable catheters could be used; but if the tip of the displaced lead is tightly fixed, it may be impossible to detach the lead.

By using this new device (cardiac pacing lead hook), the operator was able to reposition the lead.

The results of this experience encourage us to use ‘the cardiac pacing lead hook’ in more cases and evaluate the results.


    References
 Top
 Abstract
 Introduction
 Case 1
 Methods
 Case 2
 Discussion
 References
 
[1] Byrd CL. Management of implant complications. In Ellenboger KA, Kay GN, Wilkoff BI (Eds.). Clinical Cardiac Pacing 1995; Philadelphia WB Saunders pp. 491–522.

[2] Kiviniemi MS, Pirnes MA, Eranen HJK, et al. Complications related to permanent pacemaker therapy. Pacing Clin Electrophysiol 1999; 22: 711–20.[CrossRef][Medline]

[3] Favale S and Nacci F. Percutaneous transcatheter repositioning of displaced permanent pacemaker lead. Pacing Clin Electrophysiol 1999; 22: 1817–9.[CrossRef][Medline]

[4] Kursaklioglu H, Barcin C, Iyiosoy A, Kose S, Isik E. Percutaneous transcatheter repositioning of displaced permanent pacemaker atrial lead with a simple system. Int J Cardiovasc Intervent 2004; 6: 88–90.[CrossRef][Medline]

[5] Sochman J, Lefflerova K, Vrbska J, Kovac J. Catheter aided repositioning of displaced permanent pacemaker lead. Pacing Clin Electrophysiol 1995; 18: 1964–5.[CrossRef][Medline]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?



This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Sharif, A. Y.
Right arrow Articles by Sadeghian, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sharif, A. Y.
Right arrow Articles by Sadeghian, S.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?