Europace Advance Access originally published online on October 25, 2007
Europace 2007 9(12):1203-1208; doi:10.1093/europace/eum228
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
ELECTROCARDIOLOGY
Electrocardiographic markers of structural heart disease and predictors of death in 2332 unselected patients undergoing outpatient Holter recording
1 Department of Cardiology, Policlinico Tor Vergata, Viale Oxford 81, Rome 00133, Italy; 2 Department of Cardiology, Mater Hospital, Dublin, Ireland; 3 Department of Medicine, University Putra Malaysia, Kuala Lumpur, Malaysia
Manuscript submitted 30 April 2007. Accepted after revision 14 September 2007.
* Corresponding author. Tel: +39 03478993059; fax: +39 0620900382.E-mail address: mark_m_gallagher{at}hotmail.com
| Abstract |
|---|
|
|
|---|
Aims: To test the hypothesis that the QS interval of ventricular ectopic beats (VEBs) (ventricular ectopic QS interval, VEQSI) would provide a marker for the presence of structural heart disease and a predictor of mortality.
Methods and results: We interviewed and examined 2332 patients undergoing Holter ECG monitoring for clinical indications. In persons with VEBs, the morphologies were counted and the QS interval was measured for each of these morphologies. The duration of the broadest VEB, measured from the QRS onset in the derivation showing the earliest onset to its end in the derivation showing the latest termination, was taken as that patient's VEQSI. Survival was ascertained from public health records. Of 15 electrocardiographic variables pre-selected as potential prognostic indicators, VEQSI demonstrated the strongest association with the presence of structural heart disease (P = 0.013). Thirty-four persons died in 16 ± 4 months follow-up. Univariate predictors of mortality are age, history of myocardial infarction, maximum heart rate, QS interval, the number of VEB morphologies, and the VEQSI. On multivariate analysis, only age (P < 0.001) and the number of VEB morphologies (P = 0.02) predicted mortality.
Conclusion: VEQSI predicts the presence of structural heart disease. The number of VEB morphologies in a Holter recording predicts all-cause mortality.
Key Words: Risk stratification, Sudden death, Ventricular fibrillation, QRS complex duration, QRS interval, Ventricular ectopic beat, Ventricular premature beat, Heart rate variability, Myocardial infarction
| Introduction |
|---|
|
|
|---|
Of the many electrocardiographic indices proposed as markers of risk of death, only the QS interval has shown a consistent predictive value in large prospective studies.1
Early studies on the use of the ambulatory ECG to measure risk of death concentrated on the frequency of VEBs and on the presence or absence of ventricular tachycardia (VT).4
,5
More recently, heart rate variability (HRV) has received attention as a means of predicting risk of mortality.6
–8
We measured these indices in a population of patients attending for ambulatory ECG monitoring for a variety of clinical indications, and compared them with the VEQSI measured in the same recordings as predictors of mortality. We also considered the number of morphologies of VEB as this seemed likely to constitute a significant confounding variable in analysing the predictive value of VEQSI.
| Methods |
|---|
|
|
|---|
We interviewed unselected patients attending for 24 h Holter ECG monitoring for clinical indications at Policlinico Tor Vergata between February 2003 and October 2004. A clinical history was taken and cardiovascular examination performed by a physician training in cardiology or by a final-year medical student as the monitor was being applied. The selection of patients was based solely on the availability of the interviewing physician or student on the day in question. The patient was asked to bring relevant medical documentation when returning the following day for removal of the recorder.
From the documentation returned by the patient, information gleaned from the patient interview and examination and a search of the computerized archive of our echocardiography laboratory and from hospital admission records, the interviewing physician or student determined whether the patient had structural heart disease defined as a history of acute myocardial infarction (AMI) or valvular or congenital heart disease or heart failure or a reduced left ventricular ejection fraction (LVEF) or evidence of familial heart muscle disease or a history of any form of cardiac surgery. A history of hypertension or arrhythmia alone was not considered to constitute the evidence of structural heart disease.
As a comparison group, we identified those patients whose data suggested an absence of any cardiac pathology. The interviewing physician or medical student designated a patient as apparently normal if they did not have any evidence of structural heart disease, and had no history of hypertension or sustained arrhythmia and were not receiving any medications suggestive of a history of a cardiovascular disorder including beta blocking agents, anti-arrhythmic drugs or aspirin.
Holter recording and analysis
Dedicated nurses applied Holter recording systems with meticulous skin preparation. A seven electrode three-channel system was connected to give ECG derivations approximating to limb lead II and chest leads V1 and V5. All recordings were performed at a sampling frequency of 1 kHz using a three-channel device with solid-state memory (Rozinn Cardio ID RZ153, Rozinn Electronics Inc., Glendale, NY, USA).
Analysis was performed on a workstation running commercial Holter analysis software (Rozinn Holter for Windows, Rozinn Electronics Inc, Glendale, NY, USA). All recordings were analysed by the same physician, an experienced cardiac electrophysiologist who performed careful manual over-reading to eliminate artefacts and to correct the automated identification of VEBs and their classification by morphology. The interpreting physician was not present during the patient interview. At the time of Holter analysis, the physician was unaware of the full clinical characteristics of the patient, having only a single line summary of information necessary to generate a clinical report. Any Holter containing <18 h of good quality recording was excluded from analysis.
We prospectively selected 15 Holter ECG variables for evaluation including VEQSI, QS interval, three measures of heart rate (mean, minimum and maximum rate), four variables relating to ventricular ectopy (VEB frequency, number of VEB morphologies, number of episodes of VT and maximum frequency during VT), and three time domain (standard deviation of NN intervals, SDNN; root meam square standard deviation, RMSSD and HRV triangular index) and three frequency domain indices of HRV (high, low and very low frequency power). The indices were selected on the basis of published studies regarding Holter predictors of sudden cardiac death1
–8
or as potential confounding variables in evaluating the prognostic importance of VEQSI.
All VEBs in each recording were inspected. VEBs were considered to be of the same morphology if they had the same sequence of deflections (expressed as Q, q, R, r, S, s, R', r', S, s) in all three derivations with the amplitude of each deflection differing by <25% when expressed as a proportion of the amplitude of the conducted QRS complex preceding the VEB.
The number of morphologies of VEB in each recording was counted, up to a maximum of six morphologies. For each VEB morphology, regardless of the number of morphologies, the QS interval was measured on a single QRS complex representative of that morphology and chosen for the clarity of its onset and termination. Measurements were made using electronic callipers on a simultaneous three-derivation ECG segment represented on screen at 3.5 s per screen and at the maximum gain setting of the system. We measured from the start of the QRS in the derivation showing the earliest onset to the end of the QRS in the derivation showing the latest termination. The duration of the broadest VEB was considered to be the VEQSI of that patient. The QS interval of a conducted QRS complex was measured in the same way, choosing a broad QRS complex in patients with intermittent bundle branch block, a spontaneous beat in patients with periods of pacing, and always using a complex with a good clarity of onset and termination.
HRV was characterized using the same commercial Holter analysis software (Rozinn Holter for Windows, Rozinn Electronics Inc, Glendale, NY, USA) after careful editing to assure accurate identification of all normal beats and freedom from artifacts. Recordings of imperfect quality and patients in persistent atrial fibrillation or atrial tachycardia or with a persistently high frequency of ectopic beats were excluded from HRV analysis, as were recordings with persistent pacing of any modality. Recordings containing periods of pacing or periods of arrhythmia were accepted if all such episodes had been identified by the system and thus excluded from HRV analysis.
Follow-up
Survival was ascertained from the computerized records of the Comune di Roma, a regional archive of data derived from death certification. In the case of patients who had died during the period of follow-up, the date of death was noted. At the time of our search, the records contained information on deaths occurring before the end of 2004.
Statistical analysis
Continuous variables are shown as average ± SD, categorical ones as absolute and relative frequency. Comparison of variables was assessed by a t-test or Mann–Whitney non-parametric test, as appropriate. Correlations between distributions were performed by the Pearson method, showing the correlation coefficient (R) and its significance. Multivariate predictors of the presence of known structural heart diseases were identified using logistic regression analysis. Predictors of Death were analysed using Cox Regression analysis; potential predictors were inserted into the multivariable model, where P < 0.1 in univariate analysis.
Analyses were performed with SPSS 12.0 statistical package (SPSS Inc, Chicago, IL, USA). Significance was set at P < 0.05.
| Results |
|---|
|
|
|---|
During the study period, we performed Holter ECG monitoring on over 4000 outpatients of whom 2811 were recruited to the study, others were excluded because the interviewing researcher was not available on the day of their attendance. Of these, 479 were subsequently excluded because the recording was not of satisfactory quality or duration.
Clinical characteristics andelectrocardiographic findings
The demographic details of our Holter ECG population have previously been described.9
Patients were 49% male, aged 60 ± 16 years. The majority of the study subjects were healthy persons referred for the evaluation of mild symptoms such as palpitation or dizzy spells, but 479 (21%) had evidence of structural heart disease (Table 1). Persons with structural heart disease were significantly older than the remainder of the group (67 ± 11 vs. 59 ± 17 years, P < 0.001).
|
VEBs were present in 79% of Holter recordings including 91% of those performed in patients with structural heart disease. Of the 15 electrocardiographic features prospectively investigated as possible markers of structural heart disease, logistic regression analysis identified VEQSI and maximum heart rate as the only variables that independently distinguished between persons with structural heart disease and those with apparently normal hearts, of which VEQSI was the more powerful predictor (Table 2, Figure 1). Compared to apparently normal subjects those with structural heart disease had a higher VEQSI (163 ± 19 vs. 151 ± 17) and lower maximum heart rate (114 ± 19 vs. 122 ± 23). Using a dichotomy limit of VEQSI
170 ms, the index had a sensitivity of 0.41 and a specificity of 0.83 in identifying the presence of structural heart disease in our population.
|
|
Left ventricular function
Of the 2332 patients whose data were analysed, a measurement of LVEF was available from an imaging study performed within 1 year before the Holter recording in 913 cases (39%) including 279/446 (63%) of patients with structural heart disease. In none of these cases was there evidence of a substantial event such as AMI or cardiac surgery between the time of LVEF measurement and the time of the Holter recording. In almost all cases, LVEF was measured by echocardiography. Of the 16 electrocardiographic variables evaluated as potential markers of LV dysfunction (Table 2), the number of VEB morphologies (R = –0.318, P < 0.001) and VEQSI (R = –0.302, P < 0.001) correlated most closely with LVEF (Figure 2).
|
Correlations between electrocardiographic variables
There was a strong correlation between VEQSI and the number of VEB morphologies (R = 0.42, P < 0.001). There was a modest correlation between VEQSI and conducted QS interval (R = 0.27, P < 0.001), and between the number of VEBs and the number of VT episodes (R = 0.26, P < 0.001). Other than these and correlations within the set of heart rate variables (mean, minimum and maximum heart rate and rate of fastest VT) and within the set of HRV variables, no other correlation achieved a Pearson coefficient of >0.2.
Predictors of mortality
In 16 ± 4 months subsequent to the qualifying Holter recording, 34 persons died. On univariate analysis, predictors of all cause mortality were age, history of MI, maximum heart rate, QRS index, the number of VEB morphologies, and the VEQSI. On multivariate analysis, only age and the number of VEB morphologies were predictive of mortality (Table 3).
|
| Discussion |
|---|
|
|
|---|
Our study demonstrates that the number of morphologies of VEB in a Holter recording is a predictor of all-cause mortality. Of the other 14 electrocardiographic variables tested, only VEQSI, conducted QS interval, and maximum heart rate show a trend towards association with mortality. Of these, VEQSI shows most promise. Our data show that VEQSI is a marker for the presence of structural heart disease and is correlated with left ventricular (LV) function. In our series, it is the most powerful independent electrocardiographic predictor of the presence of structural heart disease.
Ventricular ectopy as a predictor of cardiac mortality
The presence of multiple VEB morphologies has traditionally been regarded as a predictor of adverse outcome in coronary artery disease. In the classification of ventricular ectopy proposed by Lown and Wolf,10
the finding of multiform ectopic beats placed the patient in the third of five grades. Only the presence of couplets, of VT or of VEBs of great prematurity was considered of greater importance. Although this grading scheme was widely used for many years, no clear confirmation has emerged for the importance of the number of VEB morphologies, and some evidence suggests that it is of little importance.11
Of the criteria used in the Lown classification, only the presence of non-sustained VT is still widely considered to be important.12
Counting the morphologies of VEBs in a Holter recording is tedious and to some extent subjective. The QRS axis changes with posture and in a 24 h recording changes in position are inevitable.VEBs arising from the same point in the heart may appear quite different when the patient is supine and when upright. In this study, we sought to minimize this effect by judging the similarity of VEBs with reference to an adjacent conducted QRS complex. We expect that it will be difficult for any automated Holter analysis system to measure the number of VEB morphologies in a reproducible way.
VEB morphology and structural heart disease
A correlation between VEB morphology and structural cardiac disease was documented by Moulton et al.13
who used 12 lead recordings of short duration recorded at the time of cardiac catheterization. They categorized patients on the basis of VEB morphology into Group 1, in whom the ventricular ectopic QRS was smooth or displayed notching of
40 ms, and Group 2 in whom the ventricular ectopic QRS displayed notching of >40 ms. They noted that VEBs of Group 2 were associated with greater LV dilatation and that VEBs of this type were generally of greater duration than those of Type 1.
Moulton et al. noted that VEBs of persons with structural heart disease were of greater duration. They suggested a dichotomy limit of 160 ms in the identification of patients with LV dilatation, but found that a morphological classification based on the presence of notching of the ectopic QRS complex was more sensitive than a classification based on duration. We quantified the duration but not the notching of the VEB because of the practical difficulty in measuring notching in a reproducible way.
Conducted QS interval
In populations with a high prevalence of left bundle branch block, the QS interval has emerged as an important predictor of sudden cardiac death and of benefit from implantable cardioverter (ICD) therapy, possibly because a broader QRS reflects the greater distance traversed by a wave of depolarization in a dilated left ventricle or slowed conduction through diseased myocardium. In persons with an intact conduction system, the QRS complex is narrow irrespective of the state of the myocardium. In persons without bundle branch block and in patients with right bundle branch block, the QS interval of conducted beats does not predict cardiac mortality.3
In our series, conducted QS interval showed a limited correlation with the presence of structural heart disease, and no ability to predict death. This may reflect the fact that this was a relatively healthy ambulant population with a low prevalence of left bundle branch block.
Ventricular ectopic QS interval
In most persons suspected of having an elevated risk of sudden cardiac death, 24 h Holter ECG monitoring reveals the presence of VEBs. As these arise in the ventricles and are conducted in a manner largely independent of the specialized conduction tissues, the duration of the resulting QRS should depend only on the site of origin, the degree of ventricular dilatation and the velocity of conduction in the ventricular myocardium. It should therefore reflect the state of the myocardium more closely than the QS interval of conducted beats, which is influenced more by the state of the His-Purkinje system. Our data support this hypothesis.
As VEB morphology depends on the site of origin of the VEB, we would expect VEQSI to be linked to the number of VEB morphologies present in a recording as persons with multiple sites of origin are more likely to have one site of origin located far from the midline of the heart and capable of producing a broader QRS complex. This effect could be augmented by the error involved in any biological measurement. The more morphologies are measured, the more likelihood there is that a high value will be returned in error. Our data suggest that there is a correlation between VEQSI and the number of VEB morphologies in a recording. Of these two measurements, our data suggest that the VEB morphology count has greater prognostic power but we found VEQSI the easier to perform as little subjectivity was involved. We expect that the automation of measurement will also prove easier for VEQSI.
Screening for persons at risk of sudden death
Randomised clinical trials have established that ICD therapy can improve survival in persons at high risk of sudden cardiac death.1
The existence of effective preventative therapy makes it imperative that patients at risk of sudden death be identified. Current guidelines on the use of ICD therapy in the primary prevention of sudden death base their selection criteria almost exclusively on LVEF, in accordance with the entry criteria of the relevant clinical trials. Unfortunately, the use of a single dichotomy limit for LVEF is unsatisfactory regardless of the cut-off value used.14
There is evidence that the conducted QS interval1
and T wave alternans15
can provide additional information about the likelihood that an individual will receive benefit from the implantation of an ICD. Other electrocardiographic indices have been proposed as markers of risk of death, including the dispersion of the QT interval, HRV and the presence of late potentials in the signal averaged ECG. Although each of these was shown to predict death in initial studies, none has shown a consistent and powerful predictive value in multivariate analysis across larger prospective studies.16
–18
The failure of traditional HRV measurements is confirmed in the current study in which the Holter recordings were edited meticulously to assure a standard of measurement higher than that obtained in routine practice. HRV analysis in the form of heart rate turbulence requires further evaluation but this also has demonstrated only limited independent predictive value, at least in recordings of short duration.19
The field of electrocardiographic prediction of death is littered with failed methodologies and conflicting evidence. Although no single method has proved satisfactory, there is evidence that risk factors can be combined to give an estimate of risk more accurate than any one index can provide.12
We believe that VEQSI may have a role in a combined scoring scheme of this type.
| Conclusions |
|---|
|
|
|---|
The current study suggests that VEQSI correlates with the presence and severity of structural heart disease and exhibits a trend towards association with the risk of death. The number of VEB morphologies is also a marker of reduced LVEF, and correlates with all-cause mortality.
The number of VEB morphologies found in a nominal 24 h Holter recording and the VEQSI show promise in the assessment of an individual's risk of death in addition to prognostic information provided by measurement of LVEF.
| Limitations |
|---|
|
|
|---|
The population studied was unselected and incompletely characterized. The non-electrocardiographic data were limited to that performed by the patients' physicians, and available to the patient or present on our hospital computer system at the time of the Holter recording. These data may have been incomplete in some cases. Cardiac imaging data were not available for all subjects and the data available were derived from different imaging methods performed in different centers without a uniform protocol.
The physician performing the measurement of VEQSI and other electrocardiographic variables could not be blinded to all of the clinical information because of the need to generate a clinical report for the Holter recording in parallel with the research study. This could have introduced interpretation bias.
Mortality figures and cause of death were derived from death certification. The cause of death listed on the death certificate could not be verified independently nor could we determine how the cause of death was ascertained by the certifying physician. The coding system did not permit the identification of the mode of death, only the underlying pathological processes, so we do not know which if any of the deaths occurred suddenly.
| Acknowledgements |
|---|
|
|
|---|
We are indebted to Tiziana De Santo and Raffaele Pulvirenti of Medtronic Italy for statistical help and to the Department of Epidemiology of ASL Roma E for access to their archives.
Conflict of interest: none declared.
| References |
|---|
|
|
|---|
[1] Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med (2002) 346:877–83.
[2] Kalahasti V, Nambi V, Martin DO, Lam CT, Yamada D, Wilkoff BL, et al. QRS duration and prediction of mortality in patients undergoing risk stratification for ventricular arrhythmias. Am J Cardiol (2003) 92:798–803.[CrossRef][Web of Science][Medline]
[3] Baldasseroni S, Gentile A, Gorini M, Marchionni N, Marini M, Masotti G, et al. Intraventricular conduction defects in patients with congestive heart failure: left but not right bundle branch block is an independent predictor of prognosis. Ital Heart J (2003) 4:607–13.[Medline]
[4] Ruberman W, Weinblatt E, Goldberg JD, Frank CW, Chaudhary BS, Shapiro S. Ventricular premature complexes and sudden death after myocardial infarction. Circulation (1981) 64:297–305.
[5] Denes P, Gillis AM, Pawitan Y, Kammerling JM, Wilhelmsen L, Salerno DM. Prevalence, characteristics and significance of ventricular premature complexes and ventricular tachycardia detected by 24-hour continuous electrocardiographic recording in the Cardiac Arrhythmia Suppression Trial. Am J Cardiol (1991) 68:887–96.[CrossRef][Web of Science][Medline]
[6] Cripps TR, Malik M, Farrell TG, Camm AJ. Prognostic value of reduced heart rate variability after myocardial infarction: clinical evaluation of a new analysis method. Br Heart J (1991) 65:14–9.
[7] Nolan J, Batin PD, Andrews R, Lindsay SJ, Brooksby P, Mullen M, et al. Prospective study of heart rate variability and mortality in chronic heart failure: results of the United Kingdom heart failure evaluation and assessment of risk trial (UK-heart). Circulation (1998) 98:1510–6.
[8] Mäkikallio TH, Huikuri HV, Hintze U, Videbaek J, Mitrani RD, Castellanos A, et al. Fractal analysis and time- and frequency-domain measures of heart rate variability as predictors of mortality in patients with heart failure. Am J Cardiol (2001) 87:178–82.[CrossRef][Web of Science][Medline]
[9] Piccirilli S, Gallagher MM, Vellini M, Padula M, Voumvourakis A, Santini L, et al. Appropriateness of ECG Holter requests in an outpatient service: a prospective study. J Cardiovasc Med (2007) 8:517–20.[Web of Science]
[10] Lown B, Wolf M. Approaches to sudden death from coronary heart disease. Circulation (1971) 44:130–42.
[11] Ruberman W, Weinblatt E, Goldberg JD, Frank CW, Chaudhary BS, Shapiro S. Ventricular premature complexes and sudden death after myocardial infarction. Circulation (1981) 64:297–305.
[12] Church TR, Hodges M, Bailey JJ, Mongin SJ. Risk stratification applied to CAST registry data: combining 9 predictors. J Electrocardiol (2002) 35:117–22.[CrossRef][Web of Science][Medline]
[13] Moulton KP, Medcalf T, Lazzara R. Premature ventricular complex morphology—a marker for left ventricular structure and function. Circulation (1990) 81:1245–51.
[14] Yap YG, Duong T, Bland JM, Malik M, Torp-Pedersen C, Køber L, et al. Optimising the dichotomy limit for left ventricular ejection fraction in selecting patients for defibrillator therapy after myocardial infarction. Heart (2007) 93:832–6.
[15] Bloomfield DM, Steinman RC, Namerow PB, Parides M, Davidenko J, Kaufman ES, et al. Microvolt T-Wave alternans distinguishes between patients likely and patients not likely to benefit from implanted cardiac defibrillator therapy—a solution to the multicenter automatic defibrillator implantation trial (MADIT) II conundrum. Circulation (2004) 110:1885–9.
[16] Gang Y, Ono T, Hnatkova K, Hashimoto K, Camm AJ, Pitt B, et al. QT dispersion has no prognostic value in patients with symptomatic heart failure: an ELITE II substudy. Pacing Clin Electrophysiol (2003) 26:394–400.[CrossRef][Medline]
[17] Grimm W, Christ M, Bach J, Müller HH, Maisch B. Noninvasive arrhythmia risk stratification in idiopathic dilated cardiomyopathy: results of the Marburg Cardiomyopathy Study. Circulation (2003) 108:2883–91.
[18] Abildstrom SZ, Jensen BT, Agner E, Torp-Pedersen C, Nyvad O, Wachtell K, et al. Heart rate versus heart rate variability in risk prediction after myocardial infarction. J Cardiovasc Electrophysiol (2003) 14:168–73.[Web of Science][Medline]
[19] Berkowitsch A, Zareba W, Neumann T, Erdogan A, Nitt SM, Moss AJ, et al. Risk stratification using heart rate turbulence and ventricular arrhythmia in MADIT II: usefulness and limitations of a 10-minute Holter recording. Ann Noninvasive Electrocardiol (2004) 9:270–9.[CrossRef][Web of Science][Medline]
![]()