Europace Advance Access originally published online on October 10, 2007
Europace 2007 9(11):1081-1090; doi:10.1093/europace/eum198
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
ELECTROCARDIOLOGY
Incorrect electrode cable connection during electrocardiographic recording
Division of Cardiac and Vascular Sciences, St Georges University of London, Cranmer Terrace, Tooting, London SW17 0RE, UK
Manuscript submitted 5 July 2007. Accepted after revision 22 August 2007.
* Corresponding author. Tel: +44 208 725 5909; fax: +44 208 767 7141.E-mail address: vbatchva{at}sgul.ac.uk
| Abstract |
|---|
|
|
|---|
Incorrect electrode cable connections during electrocardiographic (ECG) recording can simulate rhythm or conduction disturbance, myocardial ischaemia and infarction, as well as other clinically important abnormalities. When only precordial or only limb cables, excluding the neutral cable, have been interchanged the waveforms in the different leads are re-arranged, inverted, or unchanged, whereas the duration of intervals is not changed. The mistake can be recognized by the presence of unusual P–QRS patterns (e.g. negative P–QRS in lead I or II, positive in lead AVR, P–QRS complexes of opposite direction in leads I and V6, etc.), change in the P–QRS axis, or abnormal precordial QRS–T wave progression. Interchange of limb cables with the neutral cable distorts Wilsons terminal and the morphology of all precordial and unipolar limb leads. The telltale sign of the mistake is the presence of (almost) a flat line in lead I, II or III. Interchange of even one of the limb cables, except for the neutral cable, with a precordial cable distorts the morphology of most leads and leaves not more than one lead (I, II, or III) unchanged. Computerized algorithms for detection of lead misplacement, such as those based on artificial neural networks, or on correlation between original and reconstructed leads, have been developed.
Key Words: Electrocardiogram, Electrocardiographic lead, Lead misplacement, Central terminal, Wilsons terminal
| Introduction |
|---|
|
|
|---|
It has been reported that 0.4–4% of all 12-lead electrocardiograms (ECG) acquired in various clinical settings have been recorded with incorrect connection of the electrode cables,1
In order to find out which leads, if any, of the incorrectly recorded ECG are unaffected by the technical error and can still be used to interpret the ECG, it is necessary to decipher exactly which cables have been interchanged. This may sometimes be a challenging problem. Interchange of only precordial cables (e.g. V5 with V6) or only the RA, LA, or left leg (LL) cables (not involving the neutral electrode cable) does not affect the reference electrode for the unipolar* precordial leads (the so-called central or Wilsons terminal*). In such cases, the waveforms in all leads are either unaffected but rearranged or only need to be re-labelled, whereas the duration of intervals and absolute amplitudes are not affected. On the other hand, when a peripheral cable has been interchanged with a precordial or with the right leg (neutral) cable, the shape and amplitude of the waveforms in most or all leads is affected and they cannot be used for precise morphological analysis or measurement of amplitude and intervals.
In the following text, we have tried to summarize the effect of the more common electrode cable misconnections on the normal 12-lead ECG and to present simple visual criteria for their detection. More sophisticated methods, such as those based on computer software algorithms are only briefly mentioned.
Our comments apply to standard ECG equipment that satisfies the common signal acquisition and processing criteria, namely that that resistor and amplifier setting of all three limb leads (i.e. RA, LA, and LL) are the same, and similarly that the resistor and amplifier setting of all precordial leads (i.e. leads V1–V6) are the same. This is not necessarily satisfied in some non-standard equipment and in such cases, lead interchanges may result in other technical abnormalities than those described in this text. We also assumed that the circuit setting of the grounding neutral electrode is such that the position of the electrode on the body is irrelevant.
| Reversal of the left arm, right arm or left leg electrodes |
|---|
|
|
|---|
Table 1 presents the changes resulting from the five possible interchanges of the lead cables not involving the neutral electrode cable.
|
Reversal of the RA and LA cables, which is the most frequent type of cable reversal, is easily recognizable during sinus rhythm by the presence of negative P and QRS waves in lead I, which never occurs in healthy hearts and is a very unusual finding in the presence of cardiac disease (usually chronic obstructive or other pulmonary disease, less frequently congenital heart disease14
|
Reversal of the RA and LL electrode cables can be easily spotted by the inverted P–QRS complex in lead II. This mistake can mimic inferior myocardial infarction as leads AVF and AVR are also interchanged. Unlike inferior infarction in sinus rhythm, however, the P wave in leads AVF and II is also inverted.
During sinus rhythm, reversal of the LA and LL electrode cables is frequently difficult or impossible to recognize, because when lead I becomes lead II, AVL and AVF are interchanged and lead III is inverted, the ECG may superficially look even more normal than the correctly recorded one (Figure 2B). On the other hand, in the presence of typical atrial flutter, LA/LL cable interchange can easily be detected by the appearance of the sawtooth flutter waves in leads I, III, and AVL, but not in lead II.14
The observant ECG reader may even notice that the flutter waves in lead III are inverted (e.g. with the acute angles pointing upwards instead of downwards in case of typical (counter-clockwise, CCW) atrial flutter22
). In such case, if the typical pattern was not observed previously, it would not be possible to exclude atypical flutter patterns. The inversion of lead III and interchange of leads AVF and AVL (in which the T wave can normally be inverted) can produce negative T waves in leads III and AVF, simulating inferior myocardial ischaemia.14
In a patient with inferior myocardial infarction, this cable reversal can cause the disappearance of the negative Q waves and T waves in the inferior leads.14
|
Abdollah and Milliken22
Ho and Ho24
advanced this idea much further and developed an elaborate algorithm for detection of the five types of limb lead misplacement not involving the neutral electrode based on analysis of the P wave. Their method is based on the vectorcardiographic premise that the normal frontal plane P loop is (almost) invariably inscribed in the CCW rather than the CW direction. In
94% of all cases, the direction of P loop inscription (CCW or CW) may be discerned from the standard ECG by inspecting the P wave in simultaneously recorded three standard and three unipolar limb leads.24
For example, earlier occurrence of the P wave peak in lead III and AVF relative to lead I and AVL, respectively, signifies CCW inscription of the P loop, whereas biphasic negative–positive P wave in lead III reflects CW inscription, which normally occurs very rarely22
(Figure 2B). Each of the five limb lead misplacements changes the direction of inscription of the P loop and/or the frontal P axis. For example, RA/LA cable reversal reflects the P loop across the vertical line passing through the LL apex of the Einthovens triangle. The resulting mirror image of the P loop has obviously reversed direction of inscription (i.e. CW in most cases) and a frontal axis between 115° and 135° instead of the normal range of 45–65°. The authors showed that each of the six possible peripheral cable connections (1 correct + 5 incorrect) can be characterized by a unique combination of direction of P loop inscription (CW or CCW), frontal P axis and the unipolar limb lead in which the real AVR appears (Table 2).
|
The method requires the presence of P waves and is obviously of limited applicability when P waves are small or abnormal, which renders timing comparisons and hence determination of the P loop direction of inscription difficult or impossible. As mentioned by the authors,24
| Interchange involving the neutral electrode cable |
|---|
|
|
|---|
The neutral electrode, which is, by convention, placed on the right ankle, may be positioned anywhere on the body without affecting the ECG waveforms. On the other hand, interchange of its cable with the LA or RA cables affects the shape of all precordial leads (because the central terminal is changed) and of all unipolar limb leads, whereas only one of the bipolar limb leads is unaffected.14
- Lead I, II or III displays (almost) a straight line because it measures the potential difference between both legs, which is negligible.
- The other two bipolar limb leads look either identical, or one of them is an inverted (upside down) image of the other.
- Two of the unipolar limb leads also look identical, whereas the third looks like one of the bipolar limb leads or its inverted image (Table 3).
|
|
With RA/ground cable interchange, lead II displays a flat line, i.e. II
0 (Figure 3A). Since I + III = II and hence I + III
0, lead I (in this case potential difference between RL and LA) visibly looks like an inverted image of lead III (–III), whereas lead III is the only 1 unchanged of all 12 leads. Lead AVL, which in this case measures LA – (RL + LL)/2 also looks like –III, whereas AVR [RL – (LA + LL)/2] and AVF [LL – (LA + RL)/2] look identical. When the LA cable is interchanged with the ground cable, lead III (RL – LL) is a flat line, whereas lead I (RA – RL) is practically the same as lead II (RA – LL, the only unaffected lead) (Figure 3B). Lead AVR, which in this case measures RA – (RL + LL)/2 looks like inverted lead II, whereas AVL [RL – (RA + LL)/2] and AVF [LL – (RA + RL)/2] look identical.
When the 2 arm and the respective 2 leg cables are interchanged, lead I displays a flat line, as it records the voltage difference between the two legs instead of the two arms; lead II (RL – LA) is practically the same as lead III (LL–LA), whereas lead III is inverted because its two electrodes are reversed. Lead AVR [RL – (LA + LL)/2] looks identical to AVL [LL – (RL + LA)/2], whereas AVF [LA – (RL + LL)/2] looks like lead III. Unlike the previous 2 cases of cable interchange, in this case none of the 12 leads is unaffected (Figure 3C).
Interchange of the left leg and the ground (i.e. right leg) cables does not visibly affect the ECG, because the potentials of the legs are almost equal.
When a precordial cable is interchanged with the neutral cable, the respective precordial lead displays essentially the VF lead (AVF without augmentation*), because the potential of the right leg (which is approximately equal to that of the left leg) is referenced against the Wilsons terminal instead of the average of the RA and LA potentials. The peripheral and the rest of the precordial leads are unchanged (Figure 4).
|
All the above considerations about misplacement of the ground electrode cable refer to the cases when the leg electrodes are placed on their standard places, i.e. the ankles. In clinical practice, however, the peripheral electrodes are often placed on the torso in accordance with the so-called Mason–Likar system28
|
Ho et al.30
|
From a practical point of view, it is important to remember that whilst in most cases, limb lead misplacement shifts the frontal QRS axis significantly (and may simulate a fascicular block), it may sometimes change the frontal QRS axis only negligibly as, for example, RA/LA reversal in a patient with vertical QRS axis (Figure 1) or LA/LL reversal in a patient with QRS axis of 30°.
| Interchange of precordial electrodes |
|---|
|
|
|---|
Interchange of precordial electrode cables is a much less common technical mistake than incorrect placement of the precordial electrodes31
Other methods for detection of incorrect cable connections
Hedén et al.35
applied artificial neural networks (ANN) to detect RA/LA cable reversal. The authors took the detection criteria incorporated in two popular computer programs for ECG interpretation (based on QRS and P wave axis) and used them to train the ANNs. Following that, they compared the performance of the ANNs with the two programs. The study was based on 11 009 ECG without lead misplacement and an equal number of computer-derived ECGs with RA/LA cable interchange, divided into a training set and a test set. In the presence of P wave, ANN further increased the high sensitivity of the two programs up to an average of 99.11%. In the absence of P waves, the low sensitivity of the two programs (average of 31 and 39%, respectively) was increased by the ANNs to 63 and 95%, respectively. Both computer programs and ANNs showed very high specificity (close to 100%), both in the presence as well as absence of P waves.
In a second study,3
the authors applied the same method to the same database to test the ability of the ANNs to detect LA/LL and five cable reversal of adjacent precordial cables (V1/V2, V2/V3,..., V5/V6). The sensitivities of the networks ranged between 45 and 83% and were higher than that of the conventional programmes, which were between 0.1 and 10%.
Kors and van Herpen2
reported an ingenious method for automatic detection of lead misplacement in digital 12-lead ECGs. It is based on the fact that the 8 independent leads of the 12-lead ECG (i.e. any 2 peripheral and the 6 precordial leads) contain both independent as well as redundant information and therefore each of them (and hence each of the 12 leads) can be reconstructed from the 7 others in reasonable approximation.36
For a given ECG, each of the 12-leads can be reconstructed from the rest and the correlation coefficient between each original and the corresponding reconstructed lead may be calculated. If the ECG has been recorded with correct cable connections the correlations between each original and the respective reconstructed lead will be high. However, if the ECG has been recorded with some cable interchange, these correlations will generally be lower.
In order to detect cable interchange, each digital ECG is subjected to a series of computationally simulated cable interchanges, and the correlation coefficients between each recorded and the corresponding reconstructed lead are calculated before as well as after each simulated cable interchange.2
If an ECG has originally been recorded with correct cable connections, a computational simulation of a cable interchange will reduce the correlation coefficients. On the other hand, if the same ECG has originally been recorded with a particular cable interchange, the computational simulation of the same interchange will increase the correlation coefficients (because it will remove the cable interchange). The method was tested on a set of more than 3000 ECGs (QRS complex analysis only) and demonstrated specificities of
99.5% with sensitivities of
93% for detection of 14 common cable interchanges, except LA/LL cable interchange (sensitivity of 17.9%).2
Interchange of precordial and peripheral cables and other cable interchanges
It is impossible to review all 3 628 800 (=1 x 2 x...x 10) possible interchanges of the 10 ECG cables, some of which are extremely unlikely. Interchange of even one peripheral with a precordial cable will distort the Wilsons terminal and hence the morphology of the precordial leads may augment the amplitude of some limb leads and will leave not more than 12 (lead I, II, or III) of all 12 leads unchanged. Such bizarre looking ECGs occasionally occur in clinical practice6
(Figure 6) and can be misinterpreted even by experienced electrocardiographers.11
|
| Conclusions |
|---|
|
|
|---|
Some simple yet frequently overlooked rules can help detecting electrode cable reversal in the majority of cases. When a previously recorded tracing is available for comparison, the ECG reader should be aware of any inexplicable change in the axis, polarity or morphology of the P–QRS complex, while bearing in mind that sometime peripheral cable reversal may change only slightly (or normalize) the axis of the QRS complex or P wave, the latter may be too small for the tell-tale signs of lead misplacement to be seen, and the changes in the precordial leads may be due to variable (i.e. incorrect) placement of the precordial electrodes.
The experienced electrocardiographer should immediately notice unusual (in the normal heart during sinus rhythm) ECG patterns (e.g. negative P–QRS complexes in lead I or II or positive in lead AVR, abnormal precordial R, S, or T wave progression) or physiologically improbable patterns (e.g. an almost flat line in a peripheral lead, opposite polarity of P–QRS in lead I and lead V6, incompatibility between the frontal and horizontal P–QRS–T axis,11
etc.).
Vectorial analysis of the P wave from simultaneously recorded and properly displayed peripheral leads, as described by Ho and Ho,24
can be very helpful for detection of peripheral cable reversal. Computerized or computer-based ECG recorders, which allow the ECG to be displayed in various formats and amplitude gain, or directly display the vectorcardiographic loops, can obviously facilitate such analysis. Generally speaking, the study of temporal relations between corresponding parts of the P wave or QRS complex in different simultaneously recorded leads can render very valuable information not only for detection of lead misplacement.37
,38
All electrode cable interchanges, except those involving only peripheral cables (excluding the ground electrode cable) or only precordial cables, affect the reference electrodes for the unipolar limb leads and the precordial leads (Wilsons central terminal). As a result, in such cases the morphology of all leads (expect, possibly, not more than one of leads I, II, or III) is affected and they should not be used for morphological analysis or precise measurement of intervals (e.g. the QT interval). However, the degree to which different cable interchanges affect the duration of the QT and other intervals has never been systematically investigated.
Computer algorithms capable of reliable recognition of all possible cable interchanges (and, eventually, of reconstruction of the ECG with correct cable connections from the one recorded with lead misplacement) will eventually be introduced. Nevertheless, these should not cause the clinician to forget about the possible clinical errors due to poorly recorded ECGs.
| Acknowledgements |
|---|
|
|
|---|
We thank Prof. Ivaylo Christov from the Centre of Biomedical Engineering at the Bulgarian Academy of Sciences, for the valuable advice, and Mrs Slaveia Petrova, MSc, from St. Georges University of London, for the important help with the preparation of the figures.
Conflict of interest: none declared.
| Footnotes |
|---|
* The term unipolar lead is imprecise.16
Initially, the unipolar limb leads were obtained by pairing the central terminal with the electrode of the right arm (lead VR), left arm (lead VL) and left leg (lead VF), respectively. In 1942, Goldberger18
found a way to augment their voltage. When recording lead VR, he disconnected the RA electrode from the central terminal. Thus was obtained the AVR lead which measures the potential difference between the RA and the average of the LA and LL electrodes. In the same way, lead AVL (difference between LA and the average of RA and LL electrodes) and AVF (difference between LL and the average of RA and LA) were modified. ![]()
As an example, consider an ECG recorded with RA/LA cable reversal (i.e. with inverted lead I and interchanged leads II and III). Lead I of this ECG will be reconstructed from lead II (which is actually lead III) and leads V1–V6 (unchanged) using a universal set of coefficients developed for reconstruction of lead I from lead II and leads V1–V6. The error resulting from using in the reconstruction procedure lead III instead of lead II will be different with different ECGs (e.g. it will be small if lead II and lead III are similar—for example, when the frontal QRS axis is 90°). The resemblance (correlation) between the reconstructed and original (i.e. inverted) lead I will vary with different ECGs, but generally it will be smaller than the correlation between a reconstructed and original (not inverted) lead I of the same ECG recorded without RA/LA cable reversal. ![]()
| References |
|---|
|
|
|---|
[1] Rudiger A, Hellermann JP, Mukherjeec R, Follath F, Turina J. Electrocardiographic artifacts due to electrode misplacement and their frequency in different clinical settings. Am J Emer Med (2007) 25:174–8.[CrossRef]
[2] Kors JA, van Herpen G. Accurate automatic detection of electrode interchange in the electrocardiogram. Am J Cardiol (2001) 88:396–9.[CrossRef][Web of Science][Medline]
[3] Heden B, Ohlsson M, Holst H, Mjöman M, Rittner R, Pahlm O, et al. Detection of frequently overlooked electrocardiographic lead reversals using artificial neural networks. Am J Cardiol (1996) 78:600–4.[CrossRef][Web of Science][Medline]
[4] Criley JM, Nelson P. Virtual tools for teaching electrocardiographic rhythm analysis. J Electrocardiol (2006) 39:113–9.[CrossRef][Web of Science][Medline]
[5] Castellanos A, Pastor JA, Zambrano JP, Myerburg RJ. Left bundle-branch block with technical right-axis deviation. Circulation (2002) 106:2288–9.
[6] Dhingra RC, Wyndham C, Ehsani AA, Rosen KM. Left anterior hemiblock concealing diaphragmatic infarction and simulating anteroseptal infarction. Chest (1975) 67:713715.
[7] Green J, Palma E. Variable pattern of preexcitation: what is the mechanism? J Cardiovasc Electrophysiol (2003) 14:435–6.[CrossRef][Web of Science][Medline]
[8] Chanarin N, Caplin J, Peacock A. Pseudo reinfarction: a consequence of electrocardiogram lead transposition following myocardial infarction. Clin Cardiol (1991) 4:1–2.
[9] Marafioti V, Variola A. Pseudoinfarction pattern by misplacement of electrocardiographic precordial leads. Am J Emerg Med (2004) 22:62–3.[CrossRef][Web of Science][Medline]
[10] Akel R, Saeed M, Ware DL, Chamoun AJ, Birnbaum Y. Unusual evolution of ST elevation acute myocardial infarction. Ann Noninvas Electrocardiol (2004) 9:410–4.[CrossRef][Web of Science][Medline]
[11] Horwitz S, Medrano G. Left anterior hemiblock or inadvertent lead misplacement? Chest (1976) 69:449.[Web of Science][Medline]
[12] Rudiger A, Schoöb L, Follath F. Influence of electrode misplacement on the electrocardiographic signs of inferior myocardial ischemia. Am J Emerg Med (2003) 21:574–7.[CrossRef][Web of Science][Medline]
[13] Guglin ME, Thatai D. Common errors in computer electrocardiogram interpretation. Int J Cardiol (2006) 106:232–7.[CrossRef][Web of Science][Medline]
[14] Surawicz B, Knilans TK. Misplacement of leads and electrocardiographic artifacts. In: Chows Electrocardiography in Clinical Practice—Surawicz B, Knilans TK, eds. (2001) Philadelphia: WB Saunders. 569–82.
[15] Wagner GS. Marriotts Practical Electrocardiography (2001) Philadelphia: Lippincott. 33–4.
[16] Kligfield P, Gettes LS, Bailey JJ, Childers R, Deal BJ, Hancock EW, et al. Recommendations for the standardization and interpretation of the electrocardiogram. Part I: the electrocardiogram and its technology. A scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American College of Cardiology Foundation; and the Heart Rhythm Society. J Am Coll Cardiol (2007) 49:1109–27.
[17] Wilson FN, Johnston FD, Macleod AG, Barker PS. Electrocardiograms that represent the potential variations of a single electrode. Am Heart J (1934) 9:447–58.[CrossRef]
[18] Goldberger E. A simple, indifferent, electrocardiographic electrode of zero potential and a technique of obtaining augmented, unipolar, extremity leads. Am Heart J (1942) 23:483–92.[CrossRef][Web of Science]
[19] Zieli
ski J. Negative P wave in lead I due to right atrial enlargement and displacement of the heart in the chest. Lung (1976) 153:197–201.
[20] Macfarlane PW, Coleman EN. Resting 12-Lead ECG electrode placement and associated problems. http://www.scst.org.uk/coleman/resting.htm (14 August 2007).
[21] Friedman WF. Congenital heart disease in infancy and childhood. In: Brawnwald E, ed. In: Heart Disease. A Textbook of Cardiovascular Medicine. (1980) Philadelphia: WB Saunders. 1035.
[22] Abdollah H, Milliken JA. Recognition of electrocardiographic left arm/left leg lead reversal. Am J Cardiol (1997) 80:1247–9.[CrossRef][Web of Science][Medline]
[23] Heden B. Electrocardiographic lead reversal. Am J Cardiol (2001) 87:126–7.[Web of Science][Medline]
[24] Ho KKL, Ho SK. Use of the sinus P wave in diagnosing electrocardiographic limb lead misplacement not involving the right leg (ground) electrode. J Electrocardiol (2001) 34:161–71.[CrossRef][Web of Science][Medline]
[25] Castellanos A, Saoudi NC, Schwartz A, Sodi-Pallares D. Electrocardiographic patterns resulting from improper connections of the right leg (ground) cable. Pacing Clin Electrophysiol (1985) 8:364–8.[CrossRef][Medline]
[26] Haisty WK, Pahlm O, Edenbrandt, Newman K. Recognition of electrocardiographic electrode misplacements involving the ground (right leg) electrode. Am J Cardiol (1993) 71:1490–5.[CrossRef][Web of Science][Medline]
[27] Hoffman I. A flatline electrocardiogram in lead II is a marker for right arm/right leg electrode switch. J Electrocardiol (2007) 40:226–7.[CrossRef][Web of Science][Medline]
[28] Mason RE, Likar I. A new system of multiple-lead exercise electrocardiography. Am Heart J (1966) 71:196–205.[CrossRef][Web of Science][Medline]
[29] Papouchado M, Walker PR, James MA, Clarke LM. Fundamental differences between the standard 12-lead electrocardiography and the modified (Mason-Likar) exercise lead system. Eur Heart J (1987) 8:725–33.
[30] Ho RT, Mukherji L, Evans T Jr. Simple diagnosis of limb-lead reversals by predictable changes in QRS axis. Pacing Clin Electrophysiol (2006) 29:272–7.[CrossRef][Medline]
[31] Wenger W, Kligfield P. Variability of precordial electrode placement during routine electrocardiography. J Electrocardiol (1996) 29:179–84.[Web of Science][Medline]
[32] Childers R. Teaching electrocardiogram interpretation. J Electrocardiol (2006) 39:426–9.[CrossRef][Web of Science][Medline]
[33] Wagner GS. Marriotts Practical Electrocardiography (2001) 10th ed. Philadelphia, USA: Lippincott. 51.
[34] Kumar S, Spodick DH. Specificity of RV6>RV5 in the electrocardiographic recognition of left ventricular hypertrophy. Chest (1966) 50:519–22.
[35] Hedén B, Ohlsson M, Edenbrandt L, Rittver R, Pahlm O, Peterson C. Artificial neural networks for recognition of electrocardiographic lead reversal. Am J Cardiol (1995) 75:929–33.[CrossRef][Web of Science][Medline]
[36] Nelwan SP, Kors JA, Meij SH, van Bemmel JH, Simoons ML. Reconstruction of the 12-lead electrocardiogram from reduced lead sets. J Electrocardiology (2004) 37:11–8.[CrossRef][Web of Science][Medline]
[37] Warner RA, Hill NE, Mookherjee S, Smulyan H. Electrocardiographic criteria for the diagnosis of combined inferior myocardial infarction and left anterior hemiblock. Am J Cardiol (1983) 51:718–22.[CrossRef][Web of Science][Medline]
[38] Warner RA, Battaglia J, Hill NE, Mookherjee S, Smulyan H. Importance of the terminal portion of the QRS in the electrocardiographic diagnosis of inferior myocardial infarction. Am J Cardiol (1985) 55:896–9.[CrossRef][Web of Science][Medline]
![]()