PACING
Unusual VDD-pacing
1 Cardiology Division, Internal Medicine University Department, Kantonsspital Bruderholz, H-4101 Bruderholz, Switzerland; 2 Cardiology Division, University Hospital Basel, Switzerland
Manuscript submitted 9 August 2006. Accepted after revision 2 October 2006.
* Corresponding author. Tel: +41 61 436 2220; fax: +41 61 436 3670. E-mail address: peter.rickenbacher{at}ksbh.ch
| Abstract |
|---|
|
|
|---|
The case of an 80-year-old woman who underwent permanent VDD-pacemaker implantation for recurrent syncope in the presence of second-degree type 2 AV-block is reported. During follow-up, low atrial sensing with AV-synchrony of only 5873% was noted. Four years after the pacemaker implantation, the patient was hospitalized for non-cardiac reasons and the chest radiograph showed displacement of the atrial dipole into the right ventricular outflow tract (RVOT). It is hypothesized that AV-synchrony was maintained by left atrial sensing due to the anatomic proximity of the RVOT to the left atrial appendage.
Key Words: VDD pacing, P-wave sensing, Atrial undersensing
| Introduction |
|---|
|
|
|---|
VDD-pacemakers are indicated for patients with AV-block if they have normal sinus node function and no need for atrial pacing.1
| Case report |
|---|
|
|
|---|
A 80-year-old woman with a history of type-2 diabetes and arterial hypertension underwent permanent pacemaker implantation for recurrent syncope in the presence of second-degree type-2 AV-block. An endocardial pacing lead (Vitatron VDD 65/15.5 cm) was placed in the right ventricular apex with the atrial dipole in the right atrium through the right subclavian vein and the pulse generator (Vitatron Saphir 3 VDD(R) Mod. 640; Vitatron BV, Arnhem, NL) was positioned subcutaneously in the right pectoral region (Figure 1). After implantation, the P-wave amplitude was 2.8 mV, the R-wave amplitude 18.8 mV, and the ventricular pacing threshold was 0.4 V at 0.5 ms. At 3 months and during late follow-up, low atrial sensing (0.10.3 mV) with AV-synchrony in the range of 5873% according to the event counter was noted, whereas the ventricular measurements were stable. The P-wave amplitude was not susceptible to breathing manoeuvres or different body positions. Since the patient remained asymptomatic, atrial sensitivity was programmed to 0.1 mV, but no other action was taken.
|
Four years after the pacemaker implantation, the patient was hospitalized for urinary tract infection and dehydration. The chest radiograph (Figure 2) showed displacement of the atrial dipole into the right ventricular outflow tract (RVOT). The atrial dipole position was confirmed by echocardiography. At that time, the P-wave amplitude was 0.2 mV and the ventricular pacing threshold was 1.05 V at 0.4 ms, whereas the R-wave amplitude could not be measured because of complete AV-block with no underlying ventricular escape rhythm. In Figure 3, the intracardiac marker channel of the pacemaker is depicted demonstrating correct atrial sensing and pacing in the VDD mode. Repositioning of the electrode was discussed, but was refused by the patient.
|
|
| Discussion |
|---|
|
|
|---|
Long-term performance of VDD pacing systems has been addressed in a number of studies. Although intermittent loss of P-wave sensing occurs in a considerable percentage of Holter-monitored patients, high mean AV-synchrony rates > 90% have been reported in most of the studies.2
| References |
|---|
|
|
|---|
[1] Gregoratos G, Abrams J, Epstein AE, et al. ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices. Circulation 2002; 106: 214561.
[2] Pitts CJC. European multicenter prospective follow-up study of 1002 implants of a single lead VDD pacing system. The European Multicenter Study Group. Pacing Clin Electrophysiol 1991; 14: 17424.[CrossRef][Medline]
[3] Antonioli GE, Ansani L, Barbieri D, et al. Italian multicenter study on single lead VDD pacing system using a narrow atrial bipole. Pacing Clin Electrophysiol 1992; 15: 18903.[CrossRef][Medline]
[4] Naegeli B, Osswald S, Pfisterer M, et al. VDD(R) pacing: short- and long-term stability of atrial sensing with a single lead system. Pacing Clin Electrophysiol 1996; 19: 45564.[CrossRef][Medline]
[5] Ovsyshcher IE and Crystal E. VDD pacing: under evaluated, undervalued and underused. Pacing Clin Electrophysiol 2004; 27: 13358.[CrossRef][Medline]
[6] Van Campen CMC, de Cock CC, Huijgens J, Visser CA. Clinical relevance of loss of atrial sensing in patients with single lead VDD pacemakers. Pacing Clin Electrophysiol 2001; 24: 8069.[CrossRef][Medline]
[7] Wiegand UKH, Nowak B, Reisp U, et al. Implantation strategy of the atrial dipole impacts atrial sensing performance of single lead VDD pacemakers. Pacing Clin Electrophysiol 2002; 25: 31623.[CrossRef][Medline]
[8] Wiegand UKH, Bode F, Schneider R, et al. Atrial sensing and AV synchrony in single lead VDD pacemakers: a prospective comparison to DDD devices with bipolar atrial leads. J Cardiovasc Electrophysiol 1999; 10: 51320.[Web of Science][Medline]
[9] Ovsyshcher IE, Katz A, Rosenheck S, et al. Single lead VDD pacing: multicenter study. Pacing Clin Electrophysiol 1996; 19: 176871.[CrossRef][Medline]
[10] Mirza I, Holt P, James S. Permanent left atrial pacing: a 2-year follow-up of coronary sinus leads. Pacing Clin Electrophysiol 2004; 27: 3147.[CrossRef][Medline]
[11] De Cock CC, Giudici MC, Twisk JW. Comparison of the haemodynamic effects of right ventricular outflow-tract pacing with right ventricular apex pacing. A quantitative review. Europace 2003; 5: 2758.
![]()
CiteULike
Connotea
Del.icio.us What's this?
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||


