Biventricular stimulation to prevent cardiac desynchronization: rationale, design, and endpoints of the Biventricular Pacing for Atrioventricular Block to Prevent Cardiac Desynchronization (BioPace) study
1 Department of Cardiology and Angiology, Clinic of Internal Medicine and Cardiology, University Hospital Philipps-University Marburg, Baldingerstr. 1, D-35033 Marburg, Germany; 2 Centre Hospitalier Régional Universitaire de la Cavale Blanche, Brest, France; 3 Institute of Medical Biometry and Epidemiology (IMBE), Philipps-University Marburg, Germany; 4 Coordinating Centre for Clinical Trials (KKS), Philipps-University Marburg, Germany; 5 SJM International, Inc., Brussels, Belgium
Manuscript submitted 16 June 2005. Accepted after revision 9 April 2006.
* Corresponding author. Tel: +49 6421 28 66462; fax: +49 6421 28 68954. E-mail address: funck{at}mailer.uni-marburg.de
| Abstract |
|---|
|
|
|---|
Despite the deleterious effects of cardiac dyssynchrony and the positive effects of cardiac resynchronization therapy, patients with high-degree atrioventricular block continue to receive desynchronizing right ventricular (RV) pacing systems. Although it is unclear whether the negative effects of RV pacing and left bundle branch block (LBBB) are comparable, and whether they depend on the presence and the degree of structural heart disease, one may hypothesize that RV pacing may have similar effects to LBBB. In the BioPace trial, the long-term effects of RV pacing vs. biventricular pacing will be prospectively compared in 1200 pacemaker patients with high likelihood of mostly paced ventricular events, regardless of whether in sinus rhythm or in atrial fibrillation (AF). After echocardiographic examination of left ventricular (LV) function, patients will be randomly assigned to the implantation of an RV vs. a biventricular pacing system and followed for up to 5 years. Primary study endpoints are survival, quality of life (QoL), and the distance covered in a 6-min hall walk (6-MHW) at 24 months after implantation. Secondary endpoints are QoL and the 6-MHW result at 12 months after implantation, hospitalization rate, LV dimensions, LV ejection fraction, and the development of chronic AF and other adverse events.
Key Words: Cardiac pacing, Biventricular stimulation, Prevention of cardiac desynchronization, Atrioventricular block, Cardiac resynchronization therapy, Left ventricular dysfunction
| Introduction |
|---|
|
|
|---|
Cardiac dyssynchrony has deleterious effects on cardiac function by depressing left ventricular (LV) mechanical performance while increasing myocardial oxygen consumption.1
Although cardiac dyssynchrony is usually associated with spontaneous LBBB, it may also be an iatrogenic adverse effect of right ventricular (RV) pacing. Furthermore, the desynchronizing effects of RV pacing appear similar to those observed with LBBB.1
However, the prognostic implications of RV pacing-induced cardiac desynchronization in patients with and without LV dysfunction are largely unclear. In addition, the conditions in which long-term RV pacing does not induce cardiac remodelling despite obvious electrical desynchronization are not known. We, therefore, designed the Biventricular Pacing for Atrioventricular Block to Prevent Cardiac Desynchronization (BioPace) trial to study prospectively the long-term effects of biventricular vs. conventional RV pacing in patients with an indication for permanent ventricular pacing and any LV size and function.
| Background |
|---|
|
|
|---|
Ventricular pacing is the only mode of prevention of profound bradycardia or asystole in patients with complete atrioventricular (AV) block. Conventional RV pacing, however, is associated with abnormal ventricular activation, causing a prominent electromechanical delay in patients with spontaneously narrow QRS complexes.5
There is currently no evidence that cardiac desynchronization by RV pacing is less adverse than spontaneous LBBB, which is usually associated with structural heart disease.4
,16
In patients with LV dilatation and dysfunction, the prognostic content of spontaneous LBBB and RV pacing-induced QRS widening are very likely to be similar.17
This argument is supported by the results of the DAVID study, which showed a higher event-free survival in recipients of implantable cardioverter defibrillators (ICDs) programmed with a VVI backup pacing rate of 40 bpm than in patients whose ICD had been programmed to DDD pacing, such that the RV was nearly incessantly paced.2
This is concordant with the results of other trials where patients paced for AV block at the ventricular level, particularly in the presence of atrial fibrillation (AF), had a less favourable prognosis than patients with sick sinus syndrome.18
Furthermore, in the PAVE study, biventricular pacing was associated with a significantly greater functional capacity than RV pacing after AV nodal ablation for chronic AF.19
Rationale for direct biventricular pacing in AV block
Alternatives proposed to mitigate the negative effects of RV apical pacing have been RVOT, RV septal, dual-site RV, and His-bundle pacing. With RV septal pacing, the QRS duration was reduced in 64% of patients and a negative correlation between the duration of the paced QRS complex and LVEF was observed. However, the method may be time-consuming and RV septal pacing is not predictably associated with a shorter-paced QRS duration than RV apical pacing.11
Regardless of QRS duration, LV stimulation was found to be superior to RV stimulation from the standpoint of LV mechanical performance.20
While the QRS complex was shorter during biventricular pacing, the main haemodynamic benefit seemed attributable to LV pacing. Thus, from an electromechanical point of view, LV-based pacing is probably superior to any RV pacing configuration and should be preferred to dual-site RV pacing. A limitation of His bundle pacing is its inapplicability in patients with infra-Hisian conduction abnormalities, which are common in the clinical setting of CHF.4
This study has, therefore, been designed to compare biventricular with conventional RV pacing. LV pacing only was deliberately not chosen because of the persistent risk of LV lead dislodgement in a potentially pacemaker-dependent patient population.
Rationale for the use of ICDs as pacing devices
In its initial phase, the BioPace study has started without the possibility to implant ICDs. However, meanwhile it became more and more accepted that patients with EFs
35% represent a high-risk population for sudden cardiac death (SCD), even if prior arrhythmic events are lacking, and that ICDs can considerably reduce their risk. This was proven in two major prospective randomized trials, the Multicenter Automatic Defibrillator Implantation Trial (MADIT II)12
and the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) study.21
In the MADIT II study, the effect of ICD therapy on SCD reduction in patients with history of myocardial infarction and LVEFs
30% was investigated. The ICD reduced the risk of SCD by 28%. The SCD-HeFT study included heart failure patients [New York Heart Association (NYHA) class IIIII] with ischaemic and non-ischaemic heart disease and slightly less impaired LV function (EF
35%). The relative risk reduction achieved by the ICD was 23% compared with placebo therapy in this group of patients. These important results on primary SCD prevention could not be ignored by a large study such as BioPace. Therefore, it was only a logical consequence of these published results and a reaction to impending guideline modification22
to add an amendment to the BioPace study protocol so that the patients who fulfil the BioPace inclusion criteria and who, in terms of LV dysfunction, meet the SCD-HeFT and/or the MADIT II criteria can be included in the BioPace study and be implanted with ICDs to be used primarily as pacing devices. This amendment was made in August 2005. It did not alter the inclusion criteria of the study but merely reduced the exclusion criteria. The decision to implant an ICD instead of a pacemaker will only be based on the insight that patients with severely impaired LVEFs are prone to SCD. No other accepted tachyarrhythmia indication for the implantation of an ICD may be present. The decision whether a patient will be scheduled for the implantion of an ICD or a pacemaker system will be left to the discretion of the physician. From the biometrician's point of view, the extension of eligibility criteria will increase the external validity of the trial.
Main objectives of the study
The primary objective of the study is to determine whether synchronous biventricular pacing to prevent iatrogenic ventricular desynchronization confers a clinical benefit in patients with conventional indications for permanent ventricular pacing, regardless of spontaneous QRS duration and morphology or LV size and function. Clinical benefit is defined as a higher survival rate, longer distance covered during the 6-min hall walk (6-MHW) test, and better quality of life (QoL), ascertained by the Minnesota Living with Heart Failure Questionnaire (Table 1). Cardiac structure and function will be investigated by echocardiography.
|
| Methods |
|---|
|
|
|---|
Study design
BioPace has been designed as a controlled, randomized, single-blind, parallel group trial. At least 80 international medical centres will enrol 1200 patients over an estimated recruitment period of 40 months. A biventricular pacing group will be compared with a group of recipients of conventional RV pacing systems. The study design is depicted in Figure 1.
|
Inclusion and exclusion criteria
The study inclusion and exclusion criteria are detailed in Tables 2 and 3. All patients have a class I indication for permanent ventricular pacing23
35% will be allowed in accordance with the actual guidelines for the implantation of arrhythmia devices.23
|
|
Stratification and balanced randomization
Major prognostic factors that may influence the primary objectives are LVEF24
Randomization results will be distributed by facsimile from the independent Coordinating Centre for Clinical Trials at Philipps-University Marburg, Germany. A balanced randomization is used according to the following stratifying scheme:
- LVEF: (i)
35%, (ii) between 36 and 50%, (iii) >50%
- AF: present or absent
- ICD implant or implantation of a low-voltage (pacemaker) system.
- Unpaced QRS duration: (i)
120 ms, (ii) between 121 and 150 ms, (iii) >150 ms
- LBBB during spontaneous rhythm: present or absent
- Enrolling centre
35% into LVEF
35% with ICD implant and LVEF
35% with implantation of a low-voltage (pacemaker) system. The previously existing stratum LVEF
35% will be pooled with the stratum LVEF
35% without ICD implant. | Pacing systems |
|---|
|
|
|---|
Pulse generators
Patients assigned to biventricular pacing receive the Model 5510 Frontier 3x2TM (St Jude Medical Inc., Sylmar, CA, USA) triple-chamber pacemaker or newer models. Depending on whether sinus rhythm or chronic AF is present, patients randomized to the RV pacing group receive conventional single or dual-chamber pulse generators manufactured by St Jude Medical Inc.
Implantable cardioverter defibrillators
The EpicTM HF (Model V-339) or any other in terms of study relevant functions comparable with St Jude Medical CRT device with ICD backup will be used as the triple-chamber stimulation device in patients randomized to biventricular pacing who have been selected for the implantation of an ICD as an antibradycardia pacing device. No atrial lead will be implanted in patients with permanent AF in order to avoid procedural differences compared with the control group.
Depending on the presence or absence of sinus rhythm the Epic DR (Model V-233) or the Epic +VR (Model V-196) or any other in terms of study relevant functions comparable with St Jude Medical ICD will be used as stimulation device in patients randomized to RV pacing who have been selected for the implantation of an ICD. All ICDs will be programmed as shock-only devices so that shocks will be delivered for ventricular fibrillation and fast, haemodynamically not tolerated, malignant ventricular tachycardias. The use of antitachycardia pacing therapies is not allowed under this protocol. This is crucial for an additional secondary data analysis in which every appropriate shock will be counted as an event of death.
AV synchronization
The individual AV delay is optimized in both study groups, for VDD and for DDD pacing, using preferably an electrocardiogram-based or echocardiogram-based27
method. The dual- and triple-chamber pulse generators as well as the corresponding ICD devices allow the programming of different AV intervals for atrial-triggered ventricular stimulation (PV delay) and for AV sequential pacing (AV delay).
Study-protocol for upgrade from RV to biventricular pacing
In the RV-paced group, a patient may be upgraded to a biventricular system if one or both of the following criteria are fulfilled: (i) the patient has been hospitalized three times for management of CHF progression to NYHA functional class IV since the beginning of the trial, (ii) the patient has become dependent on intravenous inotropic support and does not tolerate its withdrawal. With respect to the primary survival endpoint, patients who need an upgrade are classified as deceased, as they were refractory to standard CHF therapy and had access to no other therapeutic option.
Lead system
The Model 1055 K AesculaTM or Model 1056 K QuickSiteTM Left Heart Lead (St Jude Medical Inc.) are the LV leads used for this study. The model 1056T or newer St Jude Medical LV leads may be used when they become available. The use of bipolar leads is recommended. The RV lead is preferably placed at the RV apex, though other implantation sites are allowed. The LV lead is implanted transvenously via the coronary sinus, with a view to reach a lateral or postero-lateral LV segment.27
Should these sites be unattainable, the lead may be implanted elsewhere, as long as biventricular pacing is associated with a narrower QRS complex than RV pacing. An epicardial LV lead implantation is allowed if the transvenous approach is unsuccessful. Any commercially available ICD lead may be used, although the use of leads manufactured by St Jude Medical Inc. is encouraged.
Randomization
Screened patients who met the inclusion criteria and have given their informed consent are randomly assigned within 24 h to the implantation of a biventricular pacing or ICD system (treatment group) vs. implantation of a pacemaker or ICD for conventional RV pacing (control group) and undergo the implantation procedure within
5 workdays. The decision whether a pacemaker or an ICD device will be implanted is left to the discretion of the physician. Follow-up visits are scheduled at 1, 3 (optional depending on new devices), 6, and 12 months of follow-up and every 12 months thereafter.
| Study organization |
|---|
|
|
|---|
Echocardiographic recordings and core analyses
Echocardiographic examinations are performed at each study site prior to pacemaker implantation at 1, 12, and 24 months after implantation and at the end of the study. All recordings are stored on super-VHS videotapes or on DVD/CD-ROM. After the on-site analysis of the pre-implant echocardiogram, the data are forwarded to the echocardiographic core laboratory for central analysis (appendix).
| Independent Safety Review Committee |
|---|
|
|
|---|
The Independent Safety Review Committee (ISRC) is composed of three experts who are not investigators in this study (appendix). The ISRC may request interim analyses if an excessive rate of adverse events is suspected in one or the other study group. This committee is responsible for classifying in a blinded fashion all events that may constitute an endpoint or adverse event. The ISRC may recommend the premature termination of the trial to the sponsor. So far, the ISRC has not made any specific comments.
| Statistical considerations |
|---|
|
|
|---|
Confirmatory statistical analysis
The analysis of the effectiveness of biventricular stimulation will be based on (i) patient survival, (ii) distance covered during the 6-MHW at 24 months, and (iii) QoL at 24 months, ascertained by the Minnesota© Living with Heart Failure Questionnaire score. The log-rank and WilcoxonMannWhitney test will be used with a view to establish the superiority of biventricular stimulation over conventional RV pacing. Two-sided stratified tests with eight strata will be carried out, considering the main prognostic factors LVEF in combination with ICD implant (
35% with ICD implant,
35% without ICD implant, >35% and
50%, and >50%) and AF (PRESENT/ABSENT). The intention-to-treat principle will be applied for the primary analysis.
Strongly controlling for multiplicity at an overall, two-sided significance level of
=5%, the closed testing principle by Marcus and Peritz will be used.28
The global intersection hypothesis will be tested by the three elementary hypotheses at a level of (i) 2.5% (survival), (ii) 1.25% (6-MHW at 24 months), and (iii) 1.25% (Minnesota at 24 months). In a second stage, the three intersection hypotheses will be tested by the elementary hypotheses at a level of 2.5% each. Finally, the elementary hypotheses will be tested in a third stage at a level of 5% each. In the case of a significant improvement of functional capacity or QoL, the superiority of biventricular stimulation will only be claimed if there is no statistically significant or trend towards decrease in survival.
The study was initially planned as a fixed sample design. However, the data will be monitored statistically. Interim analyses and design adaptations based on grouped data may be performed at any time during the course of the trial, according to the conditional rejection probability principle.29
32
Sample size and power considerations
Sample size was calculated on the basis of expected survival time, a two-sided significance level of 2.5%, the log-rank test assuming proportional hazards, and the method of Schoenfeld and Richter32
assuming nearly exponential survival curves and a uniform recruitment of patients. We expect a median survival time of 5 years for the RV-paced group and, furthermore, we considered an increase in survival time from 5 to 7 years by biventricular pacing to be clinically relevant. With a recruitment period of 2 years and a follow-up period of 3 years, at least 382 events are required to achieve a power of 80%. The numbers have been adjusted for 5% of randomized patients not fulfilling the criterion of
66% ventricular pacing at 1 month after implantation and for an assumed 15% missing survival information due to loss of follow-up. Consequently, we have planned for an overall recruitment of 1200 patients, 600 assigned to each group. The assumption of a median survival time of 5 years in this study corresponds to a hazard ratio of 5/7=0.714, which is not unrealistic if the data of, for example, the Care-HF study (HR=0.64)33
are taken into account. In contrast, this appears as a rather conservative estimation, even if the effect of biventricular pacing in patients with LBBB was considered to be greater than the effect of biventricular pacing compared with RV pacing.34
,35
Real median survival time of the recruited patients may turn out to be either shorter or longer than expected. This will determine the follow-up period that is needed to achieve the required number of deaths to be observed. In this context, it is noteworthy that not only the sample size but also the number of events to be observed can be adapted. This is the advantage of the possibility to perform interim analyses with design adaptations based on grouped data at any time during the course of the trial.30
,31
| Diary |
|---|
|
|
|---|
The study started in year 2003 and it was initially expected to complete patient enrolment by the end of 2005. Patient recruitment as well as to some extent also centre recruitment, however, appeared to be slower than expected. The main reasons were as follows.
- The implantation of a biventricular device as an ad hoc procedure in patients with AV block turned out not to be per se comparable with scheduled CRT system implants in stable patients with chronic heart failure and LBBB. Urgent biventricular implants can sometime be somewhat challenging, especially in terms of hospital logistics.
- Because BioPace patients have a routine antibradycardia pacing indication and in order to let the study take place under real-world conditions, also centres with up to now lacking biventricular experience have been included: especially during the initial phase of the study, some of these centres had a lower recruitment rate than expected.
- The recruitment of centres was probably delayed because some of the potential investigators while ignoring potential negative sequelae from biventricular implantations under urgent conditions might have tended to anticipate positive prognostic effects of biventricular compared with RV pacing from positive mechanistic effects that have been found in smaller studies with symptomatic or mechanistic endpoints such as the PAVE study.19
- During the period when the BioPace study was launched, the scientific context of the study with respect to the important role of ICDs in primary prevention of SCD21
,22
was not only heterogeneous but also changing in the participating countries, when the initial study protocol did not offer the possibility to implant ICDs as pacing devices. This turned out to be slowing down not only the recruitment of new centres but also the inclusion of patients in already collaborating centres, because some investigators already accepted the results of the MADIT II study21
and the SCD-Heft studies22
to implant ICDs instead of pacemakers in patients with need for chronic ventricular stimulation and severely impaired LV function. Therefore, in August 2005, the ICD amendment to the original study protocol was made to allow them to be implanted as pacing devices in patients who fulfil the inclusion criteria of the BioPace study and who present with severe LV dysfunction.
| Appendix |
|---|
|
|
|---|
Study Steering Committee: Jean-Jacques Blanc, MD (Chair), Luc De Roy, MD, Reinhard C. Funck, MD (Principal Investigator), Maurizio Lunati, MD, Vince Paul, MD, Christophe Bailleul, PhD (Sponsor Representative)
Independent Safety Review Committee: Christophe Leclercq, MD, Cecilia Linde, MD (Chair), Hans J. Trampisch, PhD
Echocardiographic Core Laboratory: Alison Duncan, MD, Michael Y. Henein (Director), MD, Wei Li, MD
| References |
|---|
|
|
|---|
[1] Heyndrickx GR, Vilaine JP, Knight DR, et al. Effects of altered site of electrical activation on myocardial performance during inotropic stimulation. Circulation 1985; 71: 10106.
[2] Wilkoff BL, Cook JR, Epstein AE, et al. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA 2002; 288: 311523.
[3] Barold SS. Adverse effects of ventricular desynchronization induced by long-term right ventricular pacing. J Am Coll Cardiol 2003; 42: 6246.
[4] Baldasseroni S, Opasich C, Gorini M, et al. Left bundle-branch block is associated with increased 1-year sudden and total mortality rate in 5517 outpatients with congestive heart failure: a report from the Italian network on congestive heart failure. Am Heart J 2002; 143: 398405.[CrossRef][ISI][Medline]
[5] Garrigue S, Barold SS, Valli N, et al. Effect of right ventricular pacing in patients with complete left bundle branch block. Am J Cardiol 1999; 83: A84.
[6] Rosenqvist M, Bergfeldt L, Haga Y, et al. The effect of ventricular activation sequence on cardiac performance during pacing. Pacing Clin Electrophysiol 1996; 19: 127986.[CrossRef][Medline]
[7] Lee MA, Dae MW, Langberg JJ, et al. Effects of long-term right ventricular apical pacing on left ventricular perfusion, innervation, function and histology. J Am Coll Cardiol 1994; 24: 22532.[Abstract]
[8] Frias PA, Corvera JS, Schmarkey L, et al. Evaluation of myocardial performance with conventional single-site ventricular pacing and biventricular pacing in a canine model of atrioventricular block. J Cardiovasc Electrophysiol 2003; 14: 9961000.[CrossRef][Medline]
[9] Karpawich PP, Justice CD, Cavitt DL, et al. Developmental sequelae of fixed-rate ventricular pacing in the immature canine heart: an electrophysiologic, hemodynamic, and histopathologic evaluation. Am Heart J 1990; 119: 107783.[ISI][Medline]
[10] Karpawich PP and Mital S. Comparative left ventricular function following atrial, septal, and apical single chamber heart pacing in the young. Pacing Clin Electrophysiol 1997; 20: 19838.[CrossRef][Medline]
[11] Schwaab B, Fröhlig G, Alexander C, et al. Influence of right ventricular stimulation site on left ventricular function in atrial synchronous ventricular pacing. J Am Coll Cardiol 1999; 33: 31723.
[12] Tse HF, Yu C, Wong KK, et al. Functional abnormalities in patients with permanent right ventricular pacing: the effect of sites of electrical stimulation. J Am Coll Cardiol 2002; 40: 14518.
[13] Heyndrickx GR and Paulus WJ. Effect of asynchrony on left ventricular relaxation. Circulation 1990; 81: III417.
[14] Tse HF and Lau CP. Long-term effect of right ventricular pacing on myocardial perfusion and function. J Am Coll Cardiol 1997; 29: 7449.[Abstract]
[15] Stambler BS, Ellenbogen K, Zhang X, et al. Right ventricular outflow versus apical pacing in pacemaker patients with congestive heart failure and atrial fibrillation. J Cardiovasc Electrophysiol 2003; 14: 11806.[CrossRef][ISI][Medline]
[16] Iuliano S, Fisher SG, Karasik PE, et al. QRS duration and mortality in patients with congestive heart failure. Am Heart J 2002; 143: 108591.[CrossRef][ISI][Medline]
[17] Owen CH, Esposito DJ, Davis JW, et al. The effects of ventricular pacing on left ventricular geometry, function, myocardial oxygen consumption, and efficiency of contraction in conscious dogs. Pacing Clin Electrophysiol 1998; 21: 141729.[CrossRef][Medline]
[18] Brunner M, Olschewski M, Geibel A, et al. Long-term survival after pacemaker implantation. Prognostic importance of gender and baseline patient characteristics. Eur Heart J 2004; 25: 8895.
[19] Doshi RN, Daoud EG, Fellows C, et al. Left ventricular-based cardiac stimulation post AV nodal ablation evaluation (the PAVE study). J Cardiovasc Electrophysiol 2005; 16: 11605.[CrossRef][ISI][Medline]
[20] Prinzen FW, Van Oosterhout MF, Vanagt WY, et al. Optimization of ventricular function by improving the activation sequence during ventricular pacing. Pacing Clin Electrophysiol 1998; 21: 225660.[CrossRef][Medline]
[21] Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002; 346: 87783.
[22] Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005; 352: 22537.
[23] Gregoratos G, Abrams J, Epstein AE, et al. ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devicessummary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). J Am Coll Cardiol 2002; 40: 170319.
[24] Gradman A, Deedwania P, Cody R, et al. Predictors of total mortality and sudden death in mild to moderate heart failure. Captopril-Digoxin Study Group. J Am Coll Cardiol 1989; 14: 56470.[Abstract]
[25] Lip GY. Atrial fibrillation and mortality. Eur Heart J 1999; 20: 15257.
[26] Mahoney P, Kimmel S, DeNofrio D, et al. Prognostic significance of atrial fibrillation in patients at a tertiary medical center referred for heart transplantation because of severe heart failure. Am J Cardiol 1999; 83: 15447.[CrossRef][ISI][Medline]
[27] Butter C, Auricchio A, Stellbrink C, et al. Effect of resynchronization therapy stimulation site on the systolic function of heart failure patients. Circulation 2001; 104: 30269.
[28] Marcus R, Peritz E, Gabriel KR. On closed testing procedures with special reference to ordered analysis of variance. Biometrika 1976; 63: 6605.
[29] Muller HH and Schafer H. Adaptive group sequential designs for clinical trials: combining the advantages of adaptive and of classical group sequential approaches. Biometrics 2001; 57: 88691.[CrossRef][ISI][Medline]
[30] Schafer H and Muller HH. Modification of the sample size and the schedule of interim analyses in survival trials based on data inspections. Stat Med 2001; 20: 374151.[CrossRef][ISI][Medline]
[31] Muller HH and Schafer H. A general statistical principle for changing a design any time during the course of a trial. Stat Med 2004; 23: 2497508.[CrossRef][ISI][Medline]
[32] Schoenfeld DA and Richter JR. Nomograms for calculating the number of patients needed for a clinical trial with survival as an endpoint. Biometrics 1982; 38: 16370.[CrossRef][ISI][Medline]
[33] Cleland JG, Daubert JC, Erdmann E, et al. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 2005; 352: 153949.
[34] Baldasseroni S, De Biase L, Fresco C, et al. Cumulative effect of complete left bundle-branch block and chronic atrial fibrillation on 1-year mortality and hospitalization in patients with congestive heart failure. A report from the Italian network on congestive heart failure (in-CHF database). Eur Heart J 2002; 23: 16928.
[35] Fisher W and Ritter P. Cardiac Pacing in Clinical Practice 1998; Berlin-Heidelberg Springer pp. 3569.
![]()
CiteULike
Connotea
Del.icio.us What's this?
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||