Europace Advance Access originally published online on May 11, 2006
Europace 2006 8(6):443-448; doi:10.1093/europace/eul041
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
RADIATION EXPOSURE IN EP
Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis
1 2nd Department of Radiology, Medical Physics Unit, University of Athens, Athens, Greece; 2 Department of Cardiology, Athens Euroclinic, 9 Athanassiadou Street, 115 21 Athens, Greece; 3 Department of Medical Physics, University of Athens, Athens, Greece; 4 Quality Assurance Reference Centre, Newcastle General Hospital, Newcastle-upon-Tyne, UK
Manuscript submitted 11 July 2005. Accepted after revision 9 March 2006.
* Corresponding author. Tel: +30 210 6416600; fax: +30 210 6819779. E-mail address: dkatritsis{at}euroclinic.gr
| Abstract |
|---|
|
|
|---|
Aims To perform a comprehensive analysis of all aspects of patient and in-room personnel radiation dosimetry in interventional electrophysiology.
Methods and results Measurements were performed during 19 diagnostic electrophysiology studies and 24 catheter ablations. Kerma-area product and exposure time values were 48.7 (6.4230) Gy cm2 and 25.5 (4.479.2) min for ablation, and 12.5 (4.5117.2) Gy cm2 and 4.5 (1.231) min for diagnostic studies, respectively. Patient effective doses were 15.2 (2.159.6) mSv for ablation and 3.2 (1.323.9) mSv for diagnostic procedures. Radiation risk to the patient was estimated to be up to eight cases of fatal cancer in 10 000 procedures. The risk of development of fatal cancer was less than 3x106 per procedure to the primary operator. The risk for the nurse and technician was much lower. The dose per procedure for the primary operator was 7.1 µGy at the eyes, 0.79 µGy at the chest under the lead apron, 13.68 µGy at the chest over the apron, 3.82 µGy at the thyroid, 17.76 µGy at the left hand, and 12.11 µGy at the left knee.
Conclusion As far as radiation exposure is concerned, electrophysiology studies followed by radiofrequency ablation are safe procedures for both patient and personnel when performed in catheterization laboratories with modern equipment, experienced operators, and standard safety precautions.
Key Words: Electrophysiology, Ablation, Radiation, Dosimetry
| Introduction |
|---|
|
|
|---|
Growing use of electrophysiology studies, often followed by radiofrequency (RF) catheter ablation, in young adults has caused increased concern over potential radiation effects. Ionizing radiation is linked to potential injury defined by either stochastic effects (e.g. DNA damage, cancer induction) or deterministic effects (e.g. skin injuries), which require a minimum number of affected cells and a threshold dose value. For electrophysiological and ablation procedures exceeding 1 h of total irradiation time, patient skin doses can be above the threshold for deterministic effects, and skin injuries have been reported.1
The objective of the present study was to consider all radiation aspects of the electrophysiological procedure which may or may not lead to ablation. Thus, it aimed at: (1
) measuring the entrance dose and calculating effective dose for all in-room personnel, (2
) performing an assessment of patient-effective dose and individual dose received by all organs, (3
) investigating the contribution of each projection to kerma-area product (KAP) and irradiation time, and (4
) estimating the risk for fatal cancer to patients and operators. To our knowledge, our report represents the first comprehensive analysis concerning all important issues of patient and in-room personnel dosimetry in interventional electrophysiology.
| Methods |
|---|
|
|
|---|
Patients
Patients subjected to diagnostic electrophysiological procedures with or without subsequent catheter ablation were studied. Diagnostic electrophysiology and RF ablation were accomplished using standard techniques. Electrodes and ablating catheters were positioned through bilateral femoral access as previously described.9
Radiation dose measurements
Procedures were performed on a Philips (Einthoven, The Netherlands) Integris H3000 C-arm X-ray unit with an undercouch X-ray tube/overcouch image intensifier configuration. The X-ray tube had a total filtration of 3 mm Al. Source-to-image intensifier distance was set at 100 cm. A series of data relating to the radiation exposure were recorded throughout each procedure. KAP value (Gy cm2) and exposure time were monitored separately for fluoroscopy and digital cine. Data were separately recorded for each projection used, [i.e. PA (posteroanterior), LAO (left anterior oblique) 60o, and RAO (right anterior oblique) 30o]. Prior to all measurements, the field size corresponding to the field size seen at the monitor was recorded. Furthermore, additional data regarding technical factors (i.e. tube potential kVp, tube current mA exposure, time, and frame rate) were logged.
Lithium fluoride (LiF) thermoluminescent dosimetry (TLD) chips (Teledyne Isotopes, NJ, USA) were used for personnel and phantom radiation measurements. The TLDs used in this study presented normal sensitivity (detection level 50 µGy) while their variation in light output was less than 7% difference from the mean value. TLDs were calibrated against an ionization chamber and electrometer (Radcal Model 3036) using free in air measurements. TLDs were read out under a constant flow of nitrogen gas on a manual reader (model 2800M-Victoreeen Inc.). In order to avoid errors from different radiation qualities, individual dose-response curves (dose range 0.110 mGy) were obtained for the reference ionization chamber at 50, 70, and 90 kVp for the X-ray tube used in this survey. The calibration factor c was calculated using the formula:
![]() |
For scattered radiation measurements, the dosimeters were also calibrated against a portable dose-rate meter (Berthold LB 133, Bad Wildbad, Germany) and placed on a cylindrical polymethylmethacyclate (PMMA) phantom (r=16 cm, h=30 cm) situated at the most frequent position of the cardiologist during the procedure (50 cm right from the patient pelvis) oriented parallel to the primary beam (broad-beam geometry). The rate meter and the TLDs were placed at a position corresponding to the operator's chest level. A similar cylindrical PMMA phantom was put on the bed at normal patient position (in the primary X-ray beam, 60 cm from focus) to act as the main source of scattered radiation. Calibration was repeated for tube high voltages between 50 and 90 kVp for the three projections (PA, LAO 60°, and RAO 30°).
Kerma-area product measurements were performed with a KAP meter (Gammex-RMI Ltd, Nottingham, UK) built into the X-ray system. The KAP meter was calibrated in situ and had a total uncertainty of 4% at a confidence level of 95%. The calibration was performed according to the manufacturer recommendations, with and without the patient table at the average tube potential used in everyday practice (80 kVp), by using RTI Electronics (Mölndal, Sweden) Solidose dosimeter with an RTI Electronics PMX solid-state detector.
Effective dose calculations of patients data were undertaken using three separate methods: (1
) TLD measurements with the use of an anthropomorphic phantom (the Alderson Rando phantom),12
,13
(2
) the National Radiological Protection Board (NRPB) approach14
with the use of the KAP to effective dose-conversion factor for each projection, and (3
) a commercially available computer programme (WinODS version 1.0a software, 1997, RTI Electronics),15
,16
calculating effective dose by KAP values for each projection, given tube potential kVp, field size, and patient data.
In order to estimate the dose received by the various organs of the patient during a typical examination, the Rando phantom was also used. 177 LiF TLDs (TLD-100 3x3x0.9 mm3
Harshaw) were placed at positions corresponding to the organs and tissues defined by the recommendations of International Commission on Radiological Protection (ICRP) Publication 60. The Rando phantom was irradiated using technique factors and projections representative of a typical procedure. Doses to the skin, bone surfaces, red bone marrow, and lung were estimated according to ICRP publication 60.17
The dose to each organ or tissue in each slice of the phantom was determined from the average of all TLD readings corresponding to that organ in a particular slice. This average dose was then multiplied by a weighting factor appropriate for the organ and slice in the Rando phantom according to:
![]() |
Additionally, organ doses per projection were estimated by using WinODS software Doseguard 100 marketed by RTI Electronics. This software requires knowledge of the average KAP values per projection, the focus patient distance, the X-ray tube voltage, and the weight of the average patient.
In order to estimate the dose to the personnel, TLD chips were attached to each person inside the operating room. Specifically six sets of TLDs (each set containing three chips) were placed on each operator. These dosimeters were attached to the left branch of the eyeglass, the thyroid, the chest over and underneath the lead apron, the left hand and knee. Additionally, the nurse and the technologist wore a TLD each sited on the left side of the chest over the lead apron. All procedures were performed by the same operators. As radiation received by medical personnel is often underestimated due to the small exposure per case, accumulated dose measurements over the total cases performed in this study was used.
Effective dose to the personnel was calculated by using two calculation methods already reported in literature.18
,19
The first, published by Niklason et al.18
calculates effective dose Deff as Deff=0.02(HosHu)+Hu, where Hos is the dose outside the lead apron and Hu is the dose recorded under the lead apron. The second method proposed by Faulkner and Marshall19
calculates Deff as Ho/32, where Ho is the total dose.
Radiation risk to the patient
An analysis quantifying the risk of fatal cancer over patient's lifetime has been proposed by the ICRP in publication 60.17
A risk factor of 5.6x102 Sv1 applies to the whole population, whereas corresponding value of 4.6x102 Sv1 has been used to calculate lifetime probability of a fatal cancer.
| Results |
|---|
|
|
|---|
Patients
Measurements were performed during 43 electrophysiological procedures conducted on 30 male and 13 female patients. Twenty four of the procedures were diagnostic electrophysiological studies followed by RF catheter ablation. In the remaining patients, catheter ablation was deferred or radiation measurements were obtained only during the diagnostic study. Clinical characteristics are provided in Table 1.
|
Radiation dose measurements
If all procedures are considered together, an average total KAP value equal to 22.3 (4.5230) Gy cm2
|
In Figure 1 the contribution of each projection to KAP values for all patients is demonstrated. The anteroposterior projection, the preferred projection of the operators, contributed to the highest KAP values.
|
Phantom radiation dose measurements
The measured organ doses at the Rando phantom as well as the number of TLDs used per organ are presented in Table 3. For comparison, the corresponding organ doses calculated with WinODS are given in Table 3 as well. The effective dose deduced from measurements with the Rando Phantom is 7.5 mSv. This value is comparable with the one obtained using the NRPB approach which was 7.1 mSv. The phantom effective doses do not agree with the corresponding values estimated by using the WinODS software which were 12.1 mSv. The corresponding total risk for fatal cancer, for a typical procedure for the patients is 4.2x104.
|
Personnel radiation dose measurements
In Table 4, the effective dose per procedure for the primary operator, the assistant operator, the technician radiographer, and the nurse calculated from the TLD measurements using the two different methods presented in the literature, is demonstrated. If the higher calculated effective dose is considered, then the total risk per procedure of developing a fatal cancer estimated by multiplying the effective dose by the risk factor for radiation workers, (4.6x102 Sv1), was 3.0x106 for the primary operator, 2.5x106 for the assistant operator, 2.3x107 for the technician, and 2.85x107 for the nurse.
|
Radiation risk to the patient
The corresponding total risk for fatal cancer for the patients due to irradiation during the procedures can be calculated if the radiation risk factor 5.6x102 Sv1 for the general population is multiplied by the calculated effective doses. If the lowest and highest values are considered, the total risk is estimated between 3.4x105 and 2x103 for ablation procedures, and between 3.9x105 and 103 for electrophysiological procedures.
| Discussion |
|---|
|
|
|---|
The aim of this study was to assess all radiation dosimetric aspects of cardiac electrophysiological procedures. Thus, we measured patient skin doses and assessed patient effective doses, and we also performed dose measurements in an anthropomorphic phantom in order to compare them with effective dose calculations. We calculated radiation risk to the patient during these procedures. We also performed personnel radiation dose measurements in order to calculate the risk arising from these procedures. Although several reports have dealt with radiation issues during ablation procedures, no such comprehensive study exists.
The estimated average effective dose values measured for electrophysiological procedures are among the lowest that have been reported in the literature for cardiac catheterization procedures,20
27
despite our rather long exposure times. This can be attributed to the fact that exposure time is mainly due to fluoroscopy, whereas as we have previously shown,28
,29
the main reason for increased patient radiation doses is digital cine exposure time. Considering the effective dose calculated for ablation procedures, McFadden et al.4
have reported 17 mSv, Kovoor et al.5
6.34 mSv per 60 min of fluoroscopy, while Lickfett et al.6
have reported effective dose values in the range of 1.4849.75 mSv. In this study, the estimated patient effective dose values when calculated separately for ablations and diagnostic electrophysiological procedures were found to be (median value and range) 15.2 (2.159.6) mSv and 3.2 (1.323.9) mSv, respectively when the NRPB method was used. Important differences in effective dose values were found when both NRPB and WinODS methods were used and could be attributed to the different input data that the two methods use in calculating effective dose. The fact that the effective dose measured in the phantom for a typical procedure is in good agreement with the value calculated by using the NRPB method, supports the use of this rather than the WinODS method for such purposes.
According to our data, the projection that contributed more to KAP values was the PA projection. This can be attributed to the fact that this was the preferred projection of the operators. Actually, as we have previously shown, the LAO projection is responsible for more radiation dose per unit time of exposure.29
The risk for fatal cancer calculated here for the patients is low, particularly when considered in the context of the benefits of a successful electrophysiological and ablation procedure. Regarding medical personnel, the dose received per procedure was also found to be relatively small. Significant reductions in staff doses may be achieved using protective shielding.
Study limitations
The main limitation of our study is that all cases were performed in a modern electrophysiological laboratory by an experienced operator with a high workload and exposure to all kinds of ablation procedures. Secondly, our study has a relatively small number of patients. Thirdly, the majority of cases was not pulmonary vein isolation that currently represents the most demanding ablation procedure. This might explain differences from other reports. Thus, our results may not be valid for other electrophysiological laboratories and for operators being at the steep part of their learning curves.
Acknowledging these limitations, we conclude that as far as radiation exposure is concerned, electrophysiological studies followed by RF ablation should be considered safe procedures for both patient and personnel when performed in catheterization laboratories with modern equipment, experienced operators, and standard safety precautions. Still, optimization of protection should be constantly sought through continuous review of procedures, conducting regular measurements that make the operators aware of their performance and producing data that allow comparison with practice elsewhere.
| References |
|---|
|
|
|---|
[1] Vano E, Arranz L, Sastre JM, et al. Dosimetric and radiation protection considerations based on some cases of patient skin injuries in interventional cardiology. Br J Radiol 1998; 71: 5106.[Abstract]
[2] Damilakis J, Theocharopoulos N, Perisinakis K, et al. Coceptus radiation dose and risk from cardiac catheter ablation procedures. Circulation 2001; 104: 8937.
[3] Perisinakis K, Damilakis J, Theocharopoulos N, Manios E, Vardas P, Gourtsoyiannis N. Accurate assessment of patient effective radiation dose and associated detriment risk from radiofrequency catheter ablation procedures. Circulation 2001; 104: 5862.
[4] Kovoor P, Ricciardello M, Collins L, Uther JB, Ross DL. Risk to patients from radiation associated with radiofrequency ablation for supraventricular tachycardia. Circulation 1998; 98: 153440.
[5] Lickfett L, Mahesh M, Vasamreddy C, et al. Radiation exposure during catheter ablation of atrial fibrillation. Circulation 2004; 110: 300310.
[6] Rosenthal LS, Mahesh M, Beck TJ, et al. Predictors of fluoroscopy time and estimated radiation exposure during radiofrequency catheter ablation procedures. Am J Cardiol 1998; 82: 4518.[CrossRef][Web of Science][Medline]
[7] Kadish AH, Mayuga KA, Yablon Z, et al. Effectiveness of shielding for patients during cardiac catheterization or electrophysiologic testing. Am J Cardiol 2001; 88: 13203.[Medline]
[8] McFadden SL, Mooney RB, Shepherd PH. X-ray dose and associated risks from radiofrequency catheter ablation procedures. Br J Radiol 2002; 75: 25365.
[9] Katritsis DG, Ellenbogen KA, Panagiotakos DB, et al. Ablation of superior pulmonary veins compared to ablation of all four pulmonary veins: a randomized clinical trial. J Cardiovasc Electrophysiol 2004; 15: 6415.[CrossRef][Medline]
[10] Katritsis D, Giazitzoglou E, Korovesis S, Zambartas C. Comparison of the transseptal approach to the transaortic approach for ablation of left-sided accessory pathways in patients with WolffParkinsonWhite syndrome. Am J Cardiol 2003; 91: 6103.[Medline]
[11] Katritsis D, Becker A, Ellenbogen K, Giazitzoglou E, Korovesis S, Camm AJ. Effect of slow-pathway ablation in atrioventricular nodal reentrant tachycardia on the electrophysiologic characteristics of the inferior atrial inputs to the human atrioventricular node. Am J Cardiol 2006; 97: 8605.[CrossRef][Web of Science][Medline]
[12] Alderson SW, Lanzl LH, Rollins M, et al. An instrumented phantom system for analog computation of treatment plans. Am J Roentgenol 1962; 87: 185.
[13] Huda W and Sandison GA. Estimation of mean organ doses in diagnostic radiology from rando phantom measurements. Health Phys 1984; 47: 4637.[Web of Science][Medline]
[14] Hart D, Jones D, Wall B. Estimation of effective dose in diagnostic radiology from entrance surface dose-area product measurements. 1994; Chilton, London National Radiological Protection Board NRPB-R262.
[15] Rannikko S, Ermakov I, Lampinen JS, Toivonen M, Karila KT, Chervjakov A. Computing patient doses of X-ray examinations using a patient size- and sex-adjustable phantom. Br J Radiol 1997; 70: 70818.[Abstract]
[16] Servomaa A, Rannikko S, Ermakov I. Assessment of effective dose in x-ray imaging in view of the proposed ICRP risk factors. Radiat Prot Dosimetry 1992; 43: 2257.[Abstract]
[17] .ICRP. Recommendations of the International Commission on Radiological Protection. ICRP Publication 60. Ann ICRP 1990; 60: 11437.
[18] Niklason LT, Marx MV, Chan HP. The estimation of occupational effective dose in diagnostic radiology with two dosimeters. Health Phys 1994; 67: 6115.[Web of Science][Medline]
[19] Faulkner K and Marshall NW. The relationship of effective dose to personnel and monitor reading for simulated fluoroscopic irradiation conditions. Health Phys 1993; 64: 5028.[Medline]
[20] Broadhead DA, Chapple CL, Faulkner K, Davies ML, McCallum H. The impact of cardiology on the collective effective dose in the North of England. Br J Radiol 1997; 70: 4927.[Abstract]
[21] Efstathopoulos EP, Makrygiannis SS, Kottou S, et al. Medical personnel and patient dosimetry during coronary angiography and intervention. Phys Med Biol 2003; 48: 305968.[CrossRef][Medline]
[22] Faulkner K, Love HG, Sweeney JK, Bardsley RA. Radiation doses and somatic risk to patients during cardiac radiological procedures. Br J Radiol 1986; 59: 35963.
[23] Kuon E, Empen K, Rohde D, Dahm JB. Radiation exposure to patients undergoing percutaneous coronary interventions: are current reference values too high? Herz 2004; 29: 20817.[CrossRef][Web of Science][Medline]
[24] Kuon E, Schmitt M, Dahm JB. Significant reduction of radiation exposure to operator and staff during cardiac interventions by analysis of radiation leakage and improved lead shielding. Am J Cardiol 2002; 89: 449.[CrossRef][Web of Science][Medline]
[25] Kuon E, Dorn C, Schmitt M, Dahm JB. Radiation dose reduction in invasive cardiology by restriction to adequate instead of optimized picture quality. Health Phys 2003; 84: 62631.[Web of Science][Medline]
[26] Vano E, Gonzalez L, Ten JI, Fernandez JM, Guibelalde E, Macaya C. Skin dose and dose-area product values for interventional cardiology procedures. Br J Radiol 2001; 74: 4855.
[27] Neofotistou V. Review of patient dosimetry in cardiology. Radiat Prot Dosim 2001; 94: 17782.[Abstract]
[28] Betsou S, Efstathopoulos EP, Katritsis D, Faulkner K, Panayiotakis G. Patient radiation doses during cardiac catheterization procedures. Br J Radiol 1998; 71: 6349.[Abstract]
[29] Efstathopoulos E, Karvouni E, Kottou S, et al. Patient dosimetry during coronary interventions: a comprehensive analysis. Am Heart J 2004; 147: 46875.[CrossRef][Web of Science][Medline]
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
H. Niinuma, R. T. George, A. Arbab-Zadeh, J. A.C. Lima, and C. A. Henrikson Imaging of pulmonary veins during catheter ablation for atrial fibrillation: the role of multi-slice computed tomography Europace, November 1, 2008; 10(suppl_3): iii14 - iii21. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ector, O. Dragusin, B. Adriaenssens, W. Huybrechts, R. Willems, H. Ector, and H. Heidbuchel Obesity Is a Major Determinant of Radiation Dose in Patients Undergoing Pulmonary Vein Isolation for Atrial Fibrillation J. Am. Coll. Cardiol., July 17, 2007; 50(3): 234 - 242. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||