ICD
Patient AlertTM to detect ICD lead failure: efficacy, limitations, and implications for future algorithms
1 Abteilung Kardiologie und Pneumologie, Herzzentrum, Georg-August-Universität Göttingen, Germany; 2 Justus-Liebig-Universität Gießen, Germany; 3 Johannes-Gutenberg-Universität Mainz, Germany; 4 Klinikum Kassel, Germany; 5 Medtronic Bakken Research Center, Maastricht, The Netherlands; 6 Kerckhoff-Klinik Bad Nauheim, Germany
Manuscript submitted 6 August 2005. Accepted after revision 12 February 2006.
* Corresponding author. Tel: +49 551 39 9625; fax: +49 551 39 9628. E-mail address: dirkvollmann2000{at}aol.com
| Abstract |
|---|
|
|
|---|
Aims An algorithm that alerts implantable cardioverter-defibrillator (ICD) patients, in case of abnormal lead impedance (Patient AlertTM, Medtronic), may help to recognize lead dysfunction. We aimed to determine the utility of Patient Alert for ICD lead-failure detection in a prospective study.
Methods and results Three hundred and sixty ICD patients were followed for 22±14 months. Patient Alert was active for pacing impedance <200 and >20003000
, and high-voltage conductor impedance <1020 and >200
. Ten alert events and a total of 29 severe system complications occurred. Patient Alert detected three of 10 ICD lead failures, with a positive predictive value (PPV) of 77.8% for any severe system complication. Retrospective analysis identified 23 patients with a sensing integrity counter (SIC) >300 and revealed an additional four prior undetected lead defects. SIC detected ICD lead failure with 92.9% sensitivity and a PPV of 59.1%. Eight of nine patients with a false-positive SIC had an integrated bipolar lead. Patient Alert combined with SIC detected all ICD lead failures and 71.4% of all severe lead complications.
Conclusions Patient Alert, based on daily lead-impedance measurement, detected one-third of all ICD lead failures. Combined use with continuous lead integrity monitoring (SIC) increased sensitivity to 100%. Integrated bipolar leads may yield a false-positive SIC. Incorporating SIC and automated pace/sense threshold measurement may improve Patient Alert sensitivity for severe lead complications.
Key Words: Implantable cardioverter-defibrillator, Lead failure, Impedance measurement, Oversensing, Complication
| Introduction |
|---|
|
|
|---|
The implantable cardioverter-defibrillator (ICD) has emerged as therapy of choice in patients at risk for ventricular tachyarrhythmias. For appropriate device therapy, the integrity of the implanted system is essential. Unfortunately, lead defects occur in a non-negligible proportion of ICD patients.1
The purpose of this study was to evaluate the utility of Patient Alert for the detection of ICD lead failure in a prospective multicentre study.
In the course of this investigation, the so-called sensing integrity counter (SIC) was introduced. The SIC continuously quantifies very short RR intervals that are typically caused by electrical noise. Recent reports suggest that the SIC may enhance the detection of ICD lead defects.9
Thus, we performed a retrospective analysis after closure of the present study to determine the utility of the SIC for the detection of ICD lead failure.
| Methods |
|---|
|
|
|---|
This prospective observational study was conducted between March 2000 and November 2004 at 17 centres in Germany and Switzerland. The study protocol conformed to the guiding principles of the Declaration of Helsinki and was approved by the local ethics committees or institutional review boards.
Patients
Patients scheduled for ICD replacement were included if they received a Medtronic (Minneapolis, MN, USA) device that incorporates the Patient Alert feature and if the new generator was connected to a chronically implanted ICD lead. All enrolled patients gave informed written consent.
Impedance measurement and Patient Alert
All devices used in this study provide automatic, subthreshold lead-impedance measurement. The daily measurement of pacing impedance has been described elsewhere.9
The HV impedance measurement differs between Medtronic Gem devices (Gem I, II, III, InSync) and more recent Medtronic ICD generations (Marquis, Maximo, Intrinsic, InSync Marquis). For the Gem ICD family, subthreshold HV impedance measurement has been described earlier.9
In more recent Medtronic ICDs, HV impedance is determined with a subthreshold 1 V, 90 µs pulse delivered from right ventricular (RV) coil to can. If connected, the same pulse is also delivered from the superior vena cava coil to the can. In both cases, impedance is calculated by measurements within the same circuit. All impedances are determined once daily at 3.00 AM and are stored in the device memory.
The Patient Alert algorithm audibly warns the patient once daily at a programmable time if one of several defined conditions has been met. In this study, the alert was activated in all subjects for the following conditions: pacing impedance <200 or >2000
(Marquis and more recent Medtronic ICDs: <200 or >3000
) and HV impedance <10 or >200
(Marquis and more recent Medtronic ICDs: <20 or >200). Optional alert features were low battery voltage, long charge time, >3 shocks delivered during one episode, and all therapies delivered in one detection zone. The power on reset condition is a non-programmable, obligatory alert feature. All patients were informed about the alert function and were instructed to contact immediately the ICD outpatient clinic, if an alert should occur.
Sensing integrity counter
Lead failure is often associated with oversensing of electrical noise. Intervals between these signals (simulating RR intervals) are typically very short and their occurrence may be intermittent. The SIC continuously quantifies the cumulative number of RR intervals <140 ms between follow-ups. All ICDs used in this study stored SIC data in the device memory. In the Medtronic Gem III and in later Medtronic ICD generations, the SIC was also visible on the programmer at the time of device interrogation.
Data evaluation
Patient characteristics and data on the implanted system were collected upon enrolment and at the time of ICD replacement. Lead integrity and device performance were evaluated at implant, during routine follow-up, at unscheduled visits, after system modification, and in case of a patient death. Unscheduled visits were prompted by Patient Alert or by symptoms thought to be related to the implanted system. Appropriate system performance was evaluated according to the standard follow-up procedure of the participating centres. This included the interrogation of the device with retrieval of all stored events, episode data, and intracardiac electrograms. Included was also the measurement of sensing and pacing threshold and impedance, painless evaluation of the HV impedance, and recording of real-time electrograms. The SIC was evaluated if the corresponding value was visible to the investigator at the time of follow-up. The retrospective SIC analysis (after study closure) considered data that had been retrieved from the device memory at each follow-up in all patients. On the basis of prior reports,9
,10
an increment in the total SIC to >300 events since the last follow-up was considered as an indicator of potential lead failure. During each visit, the initial and final interrogations were documented by save to disk. All data were reviewed and classified by the local investigators and an independent study committee. A system-related severe complication was defined as an undesirable clinical occurrence related to the presence or the performance of the implanted system and resulting in (prolonged) hospitalization, an intervention (e.g. ICD reprogramming and lead replacement), or death. All system-related complications were classified as lead- or device-related. A Patient Alert was classified as appropriate if it detected a system-related severe complication. Lead failure was assumed if the occurrence of abnormal lead impedance, noise oversensing, or other clinical findings (e.g. chest X-ray) suggested conductor fracture or insulation defect, and if these findings resulted in lead replacement. In case of a patient death, physicians were asked to provide a death summary report and a final ICD interrogation. Death was classified as sudden cardiac if it occurred within 1 h of symptoms or if it occurred unwitnessed and unexpected without other apparent cause. Cardiac deaths not classified as sudden cardiac or cardiac deaths of hospitalized patients on inotropic support were classified as non-sudden cardiac death. Non-cardiac deaths were all deaths not classified as cardiac. All cardiac deaths were reviewed for a potential relationship to device dysfunction.
Study endpoints
Primary study endpoint was the proportion of ICD lead failures detected by Patient Alert. Prospectively defined secondary study endpoints were the number of inappropriate alerts and the incidence of system-related severe complications. A retrospective analysis evaluated the SIC sensitivity, specificity, and positive predictive value to detect ICD lead failure.
Statistics
Based on data described in prior product performance reports, the expectation was that in a time period of 3 years, 4.2% of the patients would experience ICD lead failure because of conductor fracture or an insulation defect. The null-hypothesis was that the proportion of ICD lead failures detected by Patient Alert was
50%. It was calculated that 300 patients and a follow-up period of 3 years were necessary to reach 80% power.
| Results |
|---|
|
|
|---|
A total of 360 patients were enrolled and followed for 22±14 (range 256) months. Patient characteristics and implanted ICD systems are summarized in Table 1. None of the patients had a separate pacing system. Seventeen subjects had an abandoned RV lead. Complete follow-up information was available from 324 (90%) of the patients. Nine patients relocated to another hospital for follow-up and nine patients left the study for other reasons. Twenty-five deaths were reported and classified as sudden cardiac (n=2), non-sudden cardiac (n=14), and non-cardiac (n=8). In one patient, the mode of death remained unclassified but ICD interrogation provided no evidence of device or lead malfunction. In one case of sudden cardiac death, dysfunction of the implanted ICD lead (Endotak DSP, Guidant, St Paul, MN, USA) could not be excluded, because no final ICD interrogation was available. In all remaining cases, no evidence was found for a potential relationship between patient death and implanted system.
|
ICD lead failures and system-related complications
A total of 29 system-related severe complications occurred in 8% of the patients (Figure 1). Ten ICD lead failures were observed in 2.8% of all patients. Detailed information on clinical presentation, implanted lead model, and outcome of each event is given in Table 2. ICD lead failure caused inappropriate detection of ventricular fibrillation in four subjects and resulted in shock delivery in two patients. The majority of ICD lead failures (7/10) affected the Medtronic Transvene model 6936. On the basis of mode of failure and prior reports,3
|
|
Efficacy and limitations of Patient Alert
Ten Patient Alert events occurred in 10 patients. Event classification is illustrated in Figure 2. As shown, 70% (7/10) of all alert events were appropriate. Patient Alert detected three ICD lead failures (Table 2). Patient Alert also revealed one case of atrial lead failure and three other severe device-related complications (Figure 2).
|
Two inappropriate Patient Alert events were observed (Figure 2). In one patient, the alert was triggered because the algorithm was not capable of measuring atrial impedance during atrial fibrillation. This algorithm limitation has been resolved in newer devices. Another alert occurred after the lower threshold for RV pacing impedance had been programmed to <300
(nominal: <200
). Impedance ranged between 200 and 300
in this patient, but no evidence for system malfunction was found. The issue resolved by re-programming the nominal threshold value. One Patient Alert remained unclassified because of missing follow-up information. The alert was triggered by a pacing impedance >2000
and resulted in surgical revision. Intraoperative lead values were normal. The ICD was replaced but no defect was found upon technical examination, suggesting that a loose connector screw or an incomplete lead fracture may have caused the alert. Table 3 summarizes the utility of Patient Alert for the detection of ICD lead failure and system-related severe complications. Seven ICD lead failures were not detected by the Patient Alert. These cases presented with episodes of noise oversensing and/or an increased SIC (Table 2). Also undetected was one case of increased ICD lead pacing threshold, a reduction in ICD lead R-wave amplitude sensing and one atrial lead dislodgement. These cases were all revealed by altered pacing and sensing thresholds during the follow-up testing. One case of phrenic nerve stimulation and two cases of lead dislodgement were associated with a left ventricular lead (Patient Alert not applicable).
|
Of note, two patients did not hear the alarm tone. In patients who heard the alarm, the average time from the first alert sounding until ICD interrogation was 5.3 days. Five patients reported having heard an alarm tone without a documented event in the device memory.
Sensing integrity counter
The individual SIC increased to a total of >300 (7508±1783) events since the last follow-up in 23 patients. In 11 patients with a Medtronic Gem/Gem II device, this increase was not visible to the investigator at the time of device interrogation. Retrospective analysis showed that the SIC had increased in 9 of 10 cases with confirmed ICD lead failure. In four patients, ICD lead failure was detected during routine follow-up, only because of an increase in the SIC (Table 2). In three of these patients (no. 57, no. 147, and no. 261), a decrease in ring-coil impedance confirmed an insulation breach of the Medtronic 6936 lead. The change in ring-coil impedance only became evident during retrospective analysis. In another patient, lead failure was confirmed at the time of follow-up by a loss of sensing during arm movement. Two patients with ICD lead failure in which the SIC had not been visible on the programmer presented with episodes of ventricular fibrillation (VF) detection (oversensing), causing inappropriate shock delivery in one patient. One case of lead failure within the HV circuit was not detected by the SIC (Patient no. 12). Of note, retrospective analysis of all data revealed the early onset of ICD lead failure in four additional patients with a Medtronic Transvene model 6936. Early onset of lead failure was indicated by a combined increase in SIC and change in ring-coil impedance. These patients were scheduled for lead replacement after study closure.
In nine patients with a SIC>300, no other evidence of lead failure was found. The average SIC increase per day was somewhat lower in these subjects (86±36) than in those with confirmed lead failure (266±120, P=0.12). Eight of the nine patients with a false-positive SIC had an ICD lead for integrated bipolar sensing (Guidant model 0125: n=4; Ventritex SPL: n=2; Guidant Endotak C: n=1; Guidant model 0144: n=1; Medtronic model 6932: n=1). In three of these patients, intermittent T-wave oversensing had been documented. A false-positive SIC increase affected 4.7% of the patients with integrated bipolar leads and 0.6% of the patients with true bipolar leads. None of these patients had an abandoned RV lead. In one patient with a Medtronic Transvene model 6936 lead, the SIC increased 2 days after the ICD had been replaced. Classification was not possible because of missing follow-up information.
In summary, an SIC>300 identified ICD lead failure with 92.9% sensitivity (13/14), 97.1% specificity (336/346), and a positive predictive value of 59.1% (13/22). The combined use of Patient Alert and SIC identified ICD lead failure with 100% sensitivity and detected 71.4% (15/21) of all lead-related severe complications. These data include those lead failures that were detected after retrospective analysis of the SIC.
| Discussion |
|---|
|
|
|---|
Patient Alert, based on daily measurement of lead impedance, identified 30% of all ICD lead failures in our prospective observational study. This sensitivity is below the anticipated value of 50% and lower than the sensitivity of 69% reported by Becker et al.4
), impedance measurement identified 6.9% of the lead failures. Using thresholds defined in relation to the baseline (e.g. <50% of the minimum baseline), they found that abnormal impedance detected 41.4% of the lead failures with 99.7% specificity.
In contrast to the daily measurement of lead impedance, the SIC continuously quantifies the cumulative amount of short RR intervals (<140 ms). This also allows the detection of very short, intermittent, and clinically silent episodes of electrical noise oversensing. Accordingly, we found that continuous surveillance of lead integrity by the SIC was a more sensitive tool for the detection of ICD lead failure than the present Patient Alert algorithm. Compared with our results, Becker et al.4
reported a higher Patient Alert sensitivity for the detection of ICD lead failure (69 vs. 30%). In their analysis, however, older device generations than in the present investigation were studied, and the SIC was probably visible in a lower proportion of patients during follow-up. Thus, more lead defects with sporadic dysfunction may have remained undetected in the Becker study. This could explain why their relative number of ICD lead failures detected by Patient Alert was larger than in our study.
Retrospective analysis of SIC data from all our patients revealed four additional cases of lead failure in the Medtronic Transvene model 6936. The resulting cumulative failure rate of this coaxial, polyurethane lead design was 24%. Dorwarth et al.2
and Ellenbogen et al.3
reported comparable high failure rates and found that metal ion oxidation and polyurethane breakdown are common causes for insulation breach in this lead design.
Several authors have suggested that integration of the SIC may enhance the sensitivity of Patient Alert.4
,9
Gunderson et al.9
found that a combined algorithm of abnormal impedance trend and increase in SIC detected 86% (25/29) of all ICD lead failures. Our analysis supports these findings, but also elucidates potential limitations of the SIC. In
40% of the patients with a total SIC >300 since the last interrogation, system dysfunction could not be confirmed. Almost all false-positive cases occurred with leads designed for integrated bipolar sensing. It has been shown that integrated bipolar leads are more susceptible than true bipolar electrodes for oversensing of P- and T-waves,11
and diaphragmatic myopotentials.12
Furthermore, double-sensing of premature ventricular complexes has been reported as a cause of false-positive SIC.10
This event also occurred in a patient with an integrated bipolar lead (Medtronic model 6945, B. Gunderson, personal communication). A causal relationship between oversensing and increase in SIC could not be established in our study, but we observed intermittent T-wave oversensing in three patients with integrated bipolar leads and false-positive SIC. Thus, the SIC may be a sensitive indicator for intermittent oversensing, but one should be aware that the specificity for lead failure is limited in patients with integrated bipolar sensing electrodes.
It should also be noted that both the impedance measurement and the SIC failed to detect complications resulting from a decrease in R-wave amplitude sensing or an increase in pacing threshold. In addition, our study highlights the fact that present algorithms do not monitor the function of the left ventricular lead. In the light of the increasing number of cardiac resynchronization devices being implanted, this problem could become more relevant in the near future. Monitoring of left ventricular pacing impedance and automated measurements of atrial, right, and left ventricular sensing and pacing thresholds could help to overcome the limitations of the present Patient Alert feature. Performing multiple measurements per day may increase the chance to detect intermittent system dysfunction. Last, a significant proportion of patients do not hear the alert signal, and the time between alert and ICD interrogation may be substantial, even in subjects who immediately notice the alarm. Therefore, remote monitoring systems combined with advanced algorithms could be a promising way to enhance early detection of ICD system dysfunction and to prevent severe complications in the future.
Limitations
The effect of Patient Alert on ICD-related morbidity and mortality cannot be concluded from our findings and has to be evaluated in a prospective, randomized study. The main objective of the present investigation is to determine the efficacy and potential limitations of the Patient Alert algorithm for the detection of ICD lead failure. The development and use of novel lead models may yield other failure modes and thus provide different results. We addressed only one lead-monitoring algorithm from a single manufacturer, because no comparable algorithm existed in devices from other companies at the time the study was initiated. Last, this investigation was sponsored by industry, but the manufacturer had no role in the interpretation of the results.
| Conclusions |
|---|
|
|
|---|
Patient Alert, based on daily measurement of lead impedance, detects approximately one-third of all ICD lead failures. Combined use with continuous lead-integrity monitoring by the SIC increased sensitivity to 100%. Leads for integrated bipolar sensing may yield a false-positive SIC, most likely due to intermittent far-field oversensing of cardiac or diaphragmatic potentials. Integration of the SIC and automated pacing and sensing threshold measurements could provide high Patient Alert sensitivity for all lead-related complications.
| Appendix |
|---|
|
|
|---|
The following investigators and institutions participated in this study (listed in descending order of the number of enrolled patients): Georg-August-Universität Göttingen: C. Unterberg-Buchwald, D. Zenker, D. Vollmann; Kerckhoff Klinik Bad Nauheim: J. Sperzel; Justus-Liebig-Universität Gießen: A. Erdogan; Medizinische Hochschule Hannover: T. Korte; Universitätsklinik Ulm: L. Binner, S. Stiller; Johann-Gutenberg-Universität Mainz: E. Himmrich; Klinikum Wuppertal: M. Horlitz; Universitätsspital Zürich: F. Duru; Westfälische Wilhelms-Universität Münster: D. Böcker; Phillips-Universitat Marburg: W. Grimm; Kardiologische Gemeinschaftspraxis Riesa: H.-H. Ebert, G. Stenzel; Klinikum Lippe-Detmold: J. Carlsson; Klinikum Lüdenscheid: B. Lemke; Klinikum Kassel: J. Neuzner, C. Hansen; Universitätskrankenhaus Eppendorf Hamburg: A. Schuchert; Städtisches Klinikum Brandenburg a.d. Havel: Dr R. Schäfer.
| Acknowledgements |
|---|
|
|
|---|
This study was supported by Medtronic.
| References |
|---|
|
|
|---|
[1] Schulte B, Schwarz T, Sperzel J, Pitschner HF, Klovekorn WP, Neuzner J. Dysfunctions of transvenous cardioverter-defibrillator electrode systems: clinical significance of system integrated diagnosis and measurement functionpossibilities of partially automated system control. Z Kardiol 1998; 87: 6309.[CrossRef][Web of Science][Medline]
[2] Dorwarth U, Frey B, Dugas M, et al. Transvenous defibrillation leads: high incidence of failure during long-term follow-up. J Cardiovasc Electrophysiol 2003; 14: 3843.[CrossRef][Web of Science][Medline]
[3] Ellenbogen KA, Wood MA, Shepard RK, et al. Detection and management of an implantable cardioverter-defibrillator lead failure: incidence and clinical implications. J Am Coll Cardiol 2003; 41: 7380.
[4] Becker R, Ruf-Richter J, Senges-Becker JC, et al. Patient alert in implantable cardioverter-defibrillators: toy or tool? J Am Coll Cardiol 2004; 44: 958.
[5] Pinski SL and Fahy GJ. The proarrhythmic potential of implantable cardioverter-defibrillators. Circulation 1995; 92: 165164.
[6] Duru F, Luechinger R, Scharf C, Brunckhorst C. Automatic impedance monitoring and patient alert feature in implantable cardioverter-defibrillators: being alert for the unexpected!. J Cardiovasc Electrophysiol 2005; 16: 4448.[Medline]
[7] Vollmann D, Luthje L, Vonhof S, Unterberg C. Inappropriate therapy and fatal proarrhythmia by an implantable cardioverter-defibrillator. Heart Rhythm 2005; 2: 3079.[CrossRef][Web of Science][Medline]
[8] Stevens J, Buchwald AB, Krieglstein H, Unterberg C. Early detection of lead fracture by painless high voltage lead impedance measurement in a transvenous ICD lead system. J Interv Card Electrophysiol 2000; 4: 26972.[CrossRef][Web of Science][Medline]
[9] Gunderson BD, Patel AS, Bounds CA, Shepard RK, Wood MA, Ellenbogen KA. An algorithm to predict implantable cardioverter-defibrillator lead failure. J Am Coll Cardiol 2004; 44: 1898902.
[10] Gunderson BD, Wood NM, Pearson AM, Wang WX, Olson WH. Investigators at GDCW: a new sensing integrity indicator for ventricular pace/sense leads. Pacing Clin Electrophysiol 2000; 23: 591.
[11] Weretka S, Michaelsen J, Becker R, et al. Ventricular oversensing: a study of 101 patients implanted with dual chamber defibrillators and two different lead systems. Pacing Clin Electrophysiol 2003; 26: 6570.[CrossRef][Medline]
[12] Schulte B, Sperzel J, Carlsson J, et al. Inappropriate arrhythmia detection in implantable defibrillator therapy due to oversensing of diaphragmatic myopotentials. J Interv Card Electrophysiol 2001; 5: 48793.[CrossRef][Web of Science][Medline]
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
H. Burri and D. Senouf Remote monitoring and follow-up of pacemakers and implantable cardioverter defibrillators Europace, June 1, 2009; 11(6): 701 - 709. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Marine Remote monitoring for prevention of inappropriate implantable cardioverter defibrillator shocks: is there no place like home? Europace, April 1, 2009; 11(4): 409 - 411. [Full Text] [PDF] |
||||
![]() |
A. Goette, F. Cantu, L. van Erven, P. Geelen, F. Halimi, J. L. Merino, J. M. Morgan, and on behalf of the Scientific Initiative Committee o Performance and survival of transvenous defibrillation leads: need for a European data registry Europace, January 1, 2009; 11(1): 31 - 34. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. D. Swerdlow, B. D. Gunderson, K. T. Ousdigian, A. Abeyratne, R. W. Stadler, J. M. Gillberg, A. S. Patel, and K. A. Ellenbogen Downloadable Algorithm to Reduce Inappropriate Shocks Caused by Fractures of Implantable Cardioverter-Defibrillator Leads Circulation, November 18, 2008; 118(21): 2122 - 2129. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Vollmann, L. Luthje, and M. Zabel Far-field oversensing of atrial signals: an unusual cause for very short V-V intervals and inappropriate implantable cardioverter defibrillator therapy Europace, August 1, 2008; 10(8): 1009 - 1011. [Full Text] [PDF] |
||||
![]() |
L. van Erven and M. J Schalij Troubleshooting implantable cardioverter-defibrillator related problems Heart, May 1, 2008; 94(5): 649 - 660. [Full Text] [PDF] |
||||
![]() |
D. Vollmann, L. Luthje, and M. Zabel Unusual cause for an increase of the sensing integrity counter in a patient with inappropriate implantable cardioverter-defibrillator therapy Europace, May 1, 2007; 9(5): 275 - 277. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||