Europace Advance Access originally published online on January 5, 2006
Europace 2006 8(3):175-177; doi:10.1093/europace/euj001
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
8/3/175    most recent
euj001v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (1)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Al Khadra, A. S.
Right arrow Articles by Al Shuhri, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Al Khadra, A. S.
Right arrow Articles by Al Shuhri, S.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The European Society of Cardiology 2006. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org


ICDS

Detection of refrigerator-associated 60 Hz alternating current as ventricular fibrillation by an implantable defibrillator

Ayman S. Al Khadra*, Abdulaziz Al Jutaily and Salem Al Shuhri

Department of Adult CardiologyPrince Sultan Cardiac CenterPO Box 7897 (X-982), Riyadh 11159 Saudi Arabia

Manuscript submitted 11 May 2004. Accepted after revision 26 January 2005.

* Corresponding author. Tel: +966 5 561 6245; fax: +966 1 474 9995. E-mail address: avnrt{at}hotmail.com


    Abstract
 Top
 Abstract
 Introduction
 Case report
 Discussion
 Conclusions
 References
 
This report describes a patient with an implantable defibrillator who suffered an inappropriate defibrillation shock upon retrieving some food items from his inadequately earthed refrigerator. Noise typical of electrical interference can be observed in the stored electrogram of the episode. The patient was instructed to earth his home appliances, but he decided to avoid his refrigerator altogether, and has had no subsequent shocks.

Key Words: Implantable defibrillator, Electromagnetic interference, Inappropriate shock


    Introduction
 Top
 Abstract
 Introduction
 Case report
 Discussion
 Conclusions
 References
 
Electromagnetic interference (EMI), such as that generated by handheld radiofrequency remote controls,1Go electric razors,2Go electronic article surveillance systems,3Go–5Go power drills and washing machines,6Go or slot machines,7Go may rarely result in false triggering of implantable cardioverter defibrillators (ICDs) discharge, triggering of noise-reversion mode, inadvertent reprogramming of the device, or damage to the circuitry. This type of interference may result in delivery of inappropriate detection and therapy discharge to the patient. A common source of EMI is electrical leakage from unearthed electrical appliances at home or work.8Go Usually such interference is easy to recognize in the stored intracardiac electrograms (EGMs) of ICDs because of the typical pattern of electrical noise.


    Case report
 Top
 Abstract
 Introduction
 Case report
 Discussion
 Conclusions
 References
 
The patient is a 77-year-old male with a history of coronary artery disease and previous myocardial infarction, with depressed left ventricular function, who suffered from recurrent symptomatic ventricular tachycardia (VT). He had a Medtronic single chamber ICD (GEM II VR, model 7229) and bipolar Sprint lead implanted. ICD detection and treatment were programmed for two zones, ventricular fibrillation (VF) zone [290 ms, number of intervals detected (NID) 18/24] and VT zone (400 ms, NID 16). Antitachycardia pacing (ATP) algorithms were programmed on. Ventricular sensitivity was 0.3 mV. During the 20 months follow-up after ICD implant, the patient was followed regularly and had several appropriate ATP treatments due to VT and shocks due to VF. The episodes were typically characterized by dizziness, followed by palpitations and then resolution of symptoms.

The patient presented to the pacemaker clinic complaining of a single shock without warning symptoms of any kind. The patient explained that the shock was received when he was at home taking food items from his refrigerator. He felt the shock at the moment he was touching the inside of the appliance, and the shock threw him to the floor. No further shocks were felt subsequently. Interrogation of his ICD showed a single episode of detected VF, for which a 19.4 J (20 J stored) shock was delivered (Figure 1). This is the first programmed shock in the VF zone. Impedance of the high voltage lead was within the normal range (49 ohm) and the episode lasted for 9 s. Examination of the stored EGM in Figure 1 shows regular and normal appearance of R-waves at an interval of ~840 ms. However, there is undulating high-frequency noise distorting the baseline. This noise disappeared after delivery of the shock. Closer examination of the detected noise reveals that it is being detected in a regular pattern, but following every normal R-wave, there is cessation of detection of the noise for just over 400 ms. Arm movement as well as positional changes did not reproduce this noise on the EGM. Combined with delivered shock impedance of 49 ohm, this is evidence inconsistent with conductor fracture.


Figure 0011
View larger version (22K):
[in this window]
[in a new window]
 
Figure 1 Histogram from the episode report showing inappropriate VF detection and cessation following the delivery of the shock.

 
Given the history, the fact that the refrigerator was not earthed, the appearance of EMI on the intracardiac EGM, and the characteristic grouped ventricular sensed events, an electrical leak from the refrigerator causing 60 Hz interference was inferred (alternating current in Saudi Arabia has a voltage of 120 or 220 V at 60 Hz).


    Discussion
 Top
 Abstract
 Introduction
 Case report
 Discussion
 Conclusions
 References
 
This case illustrates a typical example of potential environmental hazards to patients who have implantable defibrillators, particularly when electrical appliances and instruments are not sufficiently earthed. Because household appliances are typically not earthed in Saudi Arabia, the potential for electrical leak, particularly for refrigerators and other kitchen appliances is very high.

The history and examination of stored EGMs are typical of electrical current leak with detection by the implanted defibrillator. The ICD detected this current leak while the patient was touching the inside of the refrigerator and responded with a defibrillation shock. Review of the intracardiac EGM (Figure 2) reveals the onset of high frequency noise (120 ms) detected by the ICD as VF, superimposed on normal R-waves marching through the noise. The noise seen on the stored EGM is typical of this type of EMI, and the intermittent lack of detection following a normal sensed R-wave is due to automatic adjustment of sensitivity. For this particular Medtronic device, the sensitivity is reduced to 75% of the EGM peak following the sensed (but not paced) event. The ICD recognized the noise as EMI and responded with a pacing spike coupled at 500 ms to the last detected impulse. Pacing terminates the under-sensing period 500 ms following the native R-wave. Following the paced beat, however, detection of noise as VF resumes. After VF detection criteria are satisfied, and during capacitor charge, ventricular pacing ceases and typical grouped detection of the noise is evident (three sensed and three missed). Under-sensing of R-waves is due to the auto-adjusting sensitivity threshold mechanism, discussed earlier. This pattern is not seen with actual VF, where nearly all the waves are sensed because the difference between the waves is not as significant as it is in this example. Following shock delivery, which resulted in the patient moving away from the source of EMI, the EGM no longer shows the EMI noise.


Figure 0012
View larger version (43K):
[in this window]
[in a new window]
 
Figure 2 There is evidence of EMI, which is sensed by the ICD in a characteristic pattern before and during charging of the device (see text). The noise disappears after delivery of the shock.

 
The patient was reassured about the function of the device and was educated thoroughly on EMI and ways to avoid it. He was instructed on appropriate earthing of his household appliances and has not returned with any further inappropriate shocks.


    Conclusions
 Top
 Abstract
 Introduction
 Case report
 Discussion
 Conclusions
 References
 
This patient has suffered an inappropriate ICD discharge due to an electrical leak from the refrigerator. The most important protection from such incidents is patient awareness including education about the possibility of EMI-triggered ICD discharge and avoidance of situations where electrical leaks may exist. They should be urged to have their electrical appliances and instruments appropriately earthed.


    References
 Top
 Abstract
 Introduction
 Case report
 Discussion
 Conclusions
 References
 
[1] Man KC, Davidson T, Langberg JJ, Morady F, Kalbfleisch SJ. Interference from a hand held radiofrequency remote control causing discharge of an implantable defibrillator. Pacing Clin Electrophysiol 1993; 16: 1756–1758.[CrossRef][Medline]

[2] Seifert T, Block M, Borggrefe M, Breithardt G. Erroneous discharge of an implantable cardioverter defibrillator caused by an electric razor. Pacing Clin Electrophysiol 1995; 18: 1592–1594.[CrossRef][Medline]

[3] McIvor ME. Environmental electromagnetic interference from electronic article surveillance devices: interactions with an ICD. Pacing Clin Electrophysiol 1995; 18: 2229–2230.[Medline]

[4] Mathew P, Lewis C, Neglia J, Krol RB, Saksena S. Interaction between electronic article surveillance systems and implantable defibrillators: insights from a fourth generation ICD. Pacing Clin Electrophysiol 1997; 20: 2857–2859.[Medline]

[5] Santucci PA, Haw J, Trohman RG, Pinski SL. Interference with an implantable defibrillator by an electronic antitheft-surveillance device. N Engl J Med 1998; 339: 1371–1374.[Free Full Text]

[6] Chan NY and Ho LW-L. Inappropriate implantable cardioverter-defibrillator shock due to alternating current leak: report of two cases. Europace 2005; 7: 193–196.[Abstract/Free Full Text]

[7] Madrid A, Sanchez A, Bosch E, Fernandez E, Moro Serrano C. Dysfunction of implantable defibrillators caused by slot machines. Pacing Clin Electrophysiol 1997; 20: 212–214.[CrossRef][Medline]

[8] Sabate X, Moure C, Nicolas J, Sedo M, Navarro X. Washing machine associated 50 Hz detected as ventricular fibrillation by an implanted cardioverter defibrillator. Pacing Clin Electrophysiol 2001; 24: 1281–1283.[Medline]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?



This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
8/3/175    most recent
euj001v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (1)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Al Khadra, A. S.
Right arrow Articles by Al Shuhri, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Al Khadra, A. S.
Right arrow Articles by Al Shuhri, S.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?