Skip Navigation

Europace 2006 8(11):962-965; doi:10.1093/europace/eul100
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Nilsson, B.
Right arrow Articles by Svendsen, J. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nilsson, B.
Right arrow Articles by Svendsen, J. H.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The European Society of Cardiology 2006. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org


ELECTROPHYSIOLOGY

The effectiveness of a high output/short duration radiofrequency current application technique in segmental pulmonary vein isolation for atrial fibrillation

Brian Nilsson*, Xu Chen, Steen Pehrson and Jesper Hastrup Svendsen

Medical Department B, Department of Cardiology, The Heart Centre, Cardiac Catheterization Laboratory, Rigshospitalet, Copenhagen University Hospital, 9 Blegdamsvej, DK-2100 Copenhagen, Denmark

Manuscript submitted 25 July 2006. Accepted after revision 6 June 2006.

* Corresponding author. Tel: +45 3545 8449; fax: +45 3545 2705. E-mail address: brni{at}dadlnet.dk


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Acknowledgement
 References
 
Aims Segmental pulmonary vein (PV) isolation by radiofrequency (RF) catheter ablation has become a curative therapy for atrial fibrillation (AF). However, the long procedure time limits the wide application of this procedure. The aim of the current study was to compare a novel ablation technique with a high power output and short application time vs. a conventional technique using a low power output and long application time.

Methods and results The study included 90 consecutive patients (age 53±10 years; 66 men). Segmental PV isolation was performed by irrigated RF catheter ablation in both groups. In the conventional group (Group 1, 45 patients), the power output was limited to 30 W with a target temperature of 50°C and an RF preset duration of 120 s. In the novel group (Group 2, 45 patients), the maximum power output was preset to 45 W, with a target temperature of 55°C and duration of 20 s. In Group 2, a significant reduction in the PV isolation time (127±57 vs. 94±33 min, P<0.02), mean fluoroscopy time (73±23 vs. 55±16 min, P<0.001), and radiation dose was observed. According to the application time and number, Group 2 showed a reduction in RF application time, but a higher number of RF applications were required for creation of complete PV isolation. During a mean follow-up of 15±7 months, a total of 74% of patients in Group 1 and 76% of patients in Group 2 demonstrated stable SR.

Conclusion Segmental PV isolation using a high power output and short application time is safe and effective in PV isolation in patients with AF. This technique can significantly reduce the procedure and fluoroscopy time compared with a low-power output technique.

Key Words: Atrial fibrillation, Catheter ablation, Pulmonary vein isolation


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Acknowledgement
 References
 
Radiofrequency (RF) catheter ablation to isolate the pulmonary veins (PVs) has become an increasingly common treatment option for patients with atrial fibrillation (AF). However, this technique includes ablation of an extensive area in the left atrium and ostia of the PVs. It is time-consuming with a substantial radiation exposure even in experienced centres. In addition, lack of experienced operators and the long procedure time preclude the widespread application of this procedure, overall increasing the potential for complications like thrombo-embolic events. Thus, significant reduction in the procedure time becomes a major challenge in performance of the catheter isolation of PVs. Currently, in most centres, segmental ostial RF application is applied in a temperature-controlled mode with low power output and a long application time with the intention to reduce the risk of PV stenosis and thrombus formation.1Go–3Go However, a few centres are now using a high output technique during extra-ostial ablation, apparently without increased complication rates.2Go,4Go,5Go

In the present study, we demonstrate a new RF application technique using a high power output and short RF application time. The aim of this retrospective study was to compare the conventional technique using a low power output and a long pulse duration (Group 1) with this new RF application technique (Group 2) in terms of procedure time, RF time, radiation exposure time, and complications in patients referred for segmental ostial PV isolation of AF.


    Methods
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Acknowledgement
 References
 
Study patients
The study included 90 consecutive patients (age 53±10 years, 66 men) who had undergone one segmental PV isolation of drug-refractory paroxysmal (59 patients) or persistent AF (31 patients) at the Rigshospitalet, Copenhagen University Hospital. Patients fulfilling at least one of the following criteria were excluded from the study: congenital heart disease, younger than 18 years, significant valve disease, left ventricular ejection fraction<20 %, NYHA class IV, and prior ablation for AF.

RF ablation
The same experienced operator (XC) performed all catheter ablations. The PVs were mapped after transseptal catheterization with two catheters: a circular mapping catheter (Lasso, Biosense Webster, Diamond Bar, CA, USA) and a 5 mm tip quadripolar saline-irrigated mapping/ablation catheter (Celsius thermo-cool, Biosense Webster). PV angiograms were performed before and after PV isolation. RF energy was applied at ostial sites with the earliest PV activation guided by the circular catheter. The RF energy was delivered from an RF generator (Cordis, Stockert) in temperature-controlled mode with an irrigation rate of 2 mL/min. In the first 45 patients (Group 1), the RF power output was limited to 30 W, with a target temperature of maximal 50°C, and a preset duration of 120 s. In the last 45 patients (Group 2), the maximum power output was preset to 45 W, a target temperature of 55°C, and duration of 20 s. The procedure endpoint was complete elimination of electrical conduction into the PV. Thirty minutes following the final RF application, PV isolation was confirmed by recording no potentials by the Lasso catheter.

Post-ablation care and follow-up
Patients were discharged the day after the ablation procedure unless complications occurred. Follow-up was obtained by outpatient clinical visits or by telephone interview. Arrhythmia events within the first month following the ablation were not included in the final results. All antiarrhythmic medication was stopped 1 month after the ablation. However, in the case of symptomatic recurrent AF, the patients were offered another (and final) ablation or resumption of antiarrhythmic drug therapy. At study initiation, patients with low risk of thrombo-embolic complications were only treated with aspirin. However, for safety reasons (discussed subsequently), this was changed to comprise anticoagulant therapy for the last 35 patients, starting 2 months prior to the ablation and discontinued 3 months following the ablation in patients without recurrent AF.

Statistical analysis
Continuous variables are expressed as mean±SD and analysed using the unpaired Student's t-test. In the case of data not normally distributed, the Mann–Whitney U test was used. Categorical data were analysed by Fisher's exact test. Statistical significance was established at P<0.05.


    Results
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Acknowledgement
 References
 
The patient characteristics and clinical history, including antiarrhythmic medication, are summarized in Table 1. No significant difference was found between the groups with regard to age, gender, structural heart disease, AF type, and duration. However, more antiarrhythmic drugs had been tried in Group 2, but no difference was observed with respect to the use of amiodarone. In addition, a trend towards older patients and patients with known persistent AF were seen in Group 2 compared with Group 1.


View this table:
[in this window]
[in a new window]

 
Table 1 Clinical characteristics

 
Procedure results
Complete isolation was achieved in all targeted PVs except in four patients, three patients in Group 1 and one in Group 2 (P=0.355, Table 2). Significantly more PVs were isolated in Group 2 (3.0±0.8 vs. 3.4±0.7, P=0.002). According to the application time and energy used, Group 2 showed a reduction in RF time (19±14 vs. 36±17 min, P<0.001) and total amount of energy (51.1±37.2 vs. 65.5±29.9 kJ, P=0.06). However, a higher number of RF applications (53±20 vs. 28±12, P<0.001) were required for creation of complete PV isolation. No audible pop occurred in any of the groups.


View this table:
[in this window]
[in a new window]

 
Table 2 Results of RF PV isolation

 
In comparison of the low and high output ablation techniques, Group 2 showed a 26% reduction in the ablation procedure time, defined as the time between the first and last RF application (94±33 vs. 127±57 min, P<0.019) (Figure 1). In addition, Group 2 showed a 25% reduction in mean fluoroscopic time (55±16 vs. 73±23 min, P<0.001) (Figure 2) and a 33% reduction in radiation dose (67±49 vs. 100±51 Gy/cm2, P=0.004).


Figure 1001
View larger version (15K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 1 Comparison of the procedure times in Groups 1 and 2 (time between the first and last RF applications).

 


Figure 1002
View larger version (17K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 2 Comparison of the fluoroscopy times in Groups 1 and 2.

 
Clinical outcome
During a mean follow-up of 15±7 months (range 5–25 months), a total of 74% of patients in Group 1 and 76% of patients in Group 2 demonstrated stable SR (ns) without symptomatic recurrent AF. However, 46 and 44% of patients required additional antiarrhythmic drug therapy in Groups 1 and 2, respectively (ns). Two patients experienced a transient cerebral ischaemic episode, one in each group. In the first patient, the symptoms of left-sided hemiparesis appeared just following the ablation. In the other patient, symptoms of reduced and blurred vision appeared within a few hours following the ablation. Both patients recovered within 24 h. No other major complications, especially no PV stenosis, were observed acutely or during follow-up.


    Discussion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Acknowledgement
 References
 
To avoid PV stenosis, a low power output has been used to create PV isolation in patients with AF.2Go,3Go,6Go However, even though irrigated catheter ablation is used, a long RF time is still required to achieve PV block, particularly in areas with thick myocardial tissue. In the present study, we demonstrate the use of a novel RF application technique with a high power output and short application time in PV isolation. To the best of our knowledge, this is the first study comparing a high output/short application time with a low output/long application time. In comparison with the conventional method, this technique significantly reduces the procedure and radiation exposure time, radiation dose, and RF time. In addition, no significant difference was found in overall success (possibly due to identical procedure endpoints) and complication rate, without occurrence of PV stenosis in either groups.

Even though high output was used, the total amount of energy was far lower compared with the conventional ablation method before achieving PV isolation. An explanation for this could be that a major part of the energy during the low output/long duration ablation technique will be absorbed in the tissue surface and surroundings (blood), before creating a lesion deep enough (unless long duration RF is used) to achieve PV entrance block. Even though, irrigated-tip ablation is known to create lesions, which are deeper and wider beneath the surface with little damage to the intima,7Go,8Go this greater amount of energy may eventually increase the risk of complications, especially thrombo-embolism. Another explanation could be that an application for 2 min at the wrong place would not be efficient. More applications (with a shorter duration) may increase the likelihood of destroying the principal conduction sites between the left atrium and the PVs.

In a few reports, a similar ablation technique (high output/short duration) has been used to the one described in the present study;4Go,5Go however, this was only used in circumferential (extra-ostial) ablation with a non-irrigated catheter technique, thus the current study is the first describing this technique in segmental PV isolation. In a study by Mansour et al.,4Go a total of 80 consecutive patients underwent circumferential or segmental PV isolation using irrigated-tip ablation. In the circumferential approach, RF energy was delivered with a maximum temperature of 60°C and 50 W power limit for 12–15 s at each site. In the segmental approach, RF energy was delivered with a target temperature of 50°C and a power limit of 25–30 W for 30 s at each site. Cardiac tamponade occurred in two patients in the circumferential ablation group (5%) and one patient in the segmental ablation group (2.5%). One patient in each group (2.5%) had a cerebrovascular accident. In another study by Pappone et al.,5Go 560 patients with AF underwent circumferential extra-ostial catheter ablation of the PVs. RF energy was applied for up to 30 s with a limit of 100 W and a target temperature of 60°C. During repeat procedures, RF energy was delivered with a limit of 80 W and a target temperature of 60°C at each site. In 563 procedures, only two cases of pericardial tamponade occurred, and there were no thrombo-embolic events or cases of PV stenosis. As opposed to the technique described in the present study, usually the catheter was dragged around the PVs while continuously applying RF energy. Presumably, if the catheter stayed for a long time at the same site, this could increase the risk of atrial perforation with subsequent life-threatening complications such as atrio-oesophageal fistula,9Go especially in view of the close anatomical relationship between PVs and the oesophagus.10Go In contrast to the above-mentioned studies using a high output technique,4Go,5Go we did not experience any perforations or PV stenosis, probably due to the limited application time of 20 s. Finally, some other laboratories using irrigated catheters have reported application times of 30–60 s,11Go–13Go longer times than reported in the current study.

In conclusion, this new ablation technique with a high power and short application time seems safe in segmental PV isolation and has now been used in our laboratory for ~2.5 years with good results. If these initial findings were confirmed in larger randomized clinical trials, this technique may be attractive worldwide, due to the potential advantages of reduced procedure and fluoroscopy times.

Limitations
The present study is a retrospective analysis. Although our results are highly significant, they need to be further validated in experimental and clinical prospective randomized studies. The influence of a learning curve on the results cannot be definitely excluded. However, the same operator, who had performed more than 150 PV isolation procedures before this study performed all procedures.

In Group 2, a trend was seen towards including older patients and those with persistent AF. In addition, more PVs were isolated compared with Group 1. Overall, we believe that these differences reflect development in ablation strategy extending the indications including older patients with higher AF burden and going from isolation of arrhythmogenic PVs only to that of today, with isolation of all PVs. Overall, this suggets that our results may be an underestimation of the real difference.

No systematic Holter monitoring of the patients was performed, thus the reported recurrence rate may be an underestimation. However, in the current study, a successful ablation procedure was defined as a complete entrance block into the PVs. This was achieved in all but four patients with no difference between Groups 1 and 2, indicating no major difference in clinical success rates between the two ablation techniques. Apart from venography, we did not perform any systematic imaging procedures of the PVs prior to and following catheter ablation. However, no patient complained of pulmonary symptoms necessitating further investigation of PV stenosis.


    Conclusions
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Acknowledgement
 References
 
This present study demonstrates that PV isolation using a high power output and short RF pulse duration approach significantly reduces the procedure and radiation exposure times compared with the conventional low power output technique in segmental PV isolation for AF. If prospective randomized studies are able to confirm these findings, this application technique may become attractive to perform worldwide.


    Acknowledgement
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Acknowledgement
 References
 
The authors wish to thank the technical staff of Medical Department B, Sections 3141 and 3142, and the Cardiac Catheterization Laboratory for their help. This study was supported by grants from ‘Lægernes Forsikringsforening af 1891’, Biosense Webster Scandinavia, Medtronic-ViCare A/S, St Jude Medical Denmark A/S, and Guidant Denmark.


    References
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 Acknowledgement
 References
 
[1] Haïssaguerre M, Shah DC, Jaïs P, et al. Electrophysiological breakthroughs from the left atrium to the pulmonary veins. Circulation 2000; 102: 2463–5.[Abstract/Free Full Text]

[2] Oral H, Scharf C, Chugh A, et al. Catheter ablation for paroxysmal atrial fibrillation: segmental pulmonary vein ostial ablation versus left atrial ablation. Circulation 2003; 108: 2355–60.[Abstract/Free Full Text]

[3] Pappone C, Oreto G, Rosanio S, et al. Atrial electroanatomic remodeling after circumferential radiofrequency pulmonary vein ablation: efficacy of an anatomic approach in a large cohort of patients with atrial fibrillation. Circulation 2001; 104: 2539–44.[Abstract/Free Full Text]

[4] Mansour M, Ruskin J, Keane D. Efficacy and safety of segmental ostial versus circumferential extra-ostial pulmonary vein isolation for atrial fibrillation. J Cardiovasc Electrophysiol 2004; 15: 532–7.[Web of Science][Medline]

[5] Pappone C, Manguso F, Vicedomini G, et al. Prevention of iatrogenic atrial tachycardia after ablation of atrial fibrillation: a prospective randomized study comparing circumferential pulmonary vein ablation with a modified approach. Circulation 2004; 110: 3036–42.[Abstract/Free Full Text]

[6] Pappone C, Rosanio S, Oreto G, et al. Circumferential radiofrequency ablation of pulmonary vein ostia: a new anatomic approach for curing atrial fibrillation. Circulation 2000; 102: 2619–28.[Abstract/Free Full Text]

[7] Nakagawa H, Yamanashi WS, Pitha JV, et al. Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation 1995; 91: 2264–73.[Abstract/Free Full Text]

[8] Petersen HH, Chen X, Pietersen A, Svendsen JH, Haunso S. Tissue temperatures and lesion size during irrigated tip catheter radiofrequency ablation: an in vitro comparison of temperature-controlled irrigated tip ablation, power-controlled irrigated tip ablation, and standard temperature-controlled ablation. Pacing Clin Electrophysiol 2000; 23: 8–17.[CrossRef][Medline]

[9] Scanavacca MI, D'avila A, Parga J, Sosa E. Left atrial-esophageal fistula following radiofrequency catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol 2004; 15: 960–2.[Web of Science][Medline]

[10] Mönnig G, Wessling W, Juergens KU, Milberg P, Ribbing M, Fishbach R, et al. Further evidence of a close anatomical relation between the oesophagus and pulmonary veins. Europace 2005; 7: 540–5.[Abstract/Free Full Text]

[11] Macle L, Jaïs P, Weerasooriya R, et al. Irrigated-tip catheter ablation of pulmonary veins for treatment of atrial fibrillation. J Cardiovasc Electrophysiol 2002; 13: 1067–73.[CrossRef][Web of Science][Medline]

[12] Horlitz M, Schley P, Shin DI, et al. Circumferential pulmonary vein ablation for treatment of atrial fibrillation using an irrigated-tip catheter. Am J Cardiol 2004; 94: 945–7.[CrossRef][Web of Science][Medline]

[13] Taylor GW, Kay GN, Zheng X, Bishop S, Ideker RE. Pathological effects of extensive radiofrequency energy applications in the pulmonary veins in dogs. Circulation 2000; 101: 1736–42.[Abstract/Free Full Text]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?



This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Nilsson, B.
Right arrow Articles by Svendsen, J. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nilsson, B.
Right arrow Articles by Svendsen, J. H.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?