Skip Navigation


Europace Advance Access originally published online on August 4, 2006
Europace 2006 8(10):855-858; doi:10.1093/europace/eul091
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
8/10/855    most recent
eul091v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Zupan, I.
Right arrow Articles by Brecelj, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Zupan, I.
Right arrow Articles by Brecelj, A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The European Society of Cardiology 2006. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org


PACING

Retrospective analysis of mode survival, reliability of atrial sensing and incidence of atrial tachyarrhythmias in 307 single-lead VDD pacemaker patients

Igor Zupan1,*, Luka Lipar1, David Zizek1, Wim Boute2, Masa Vidmar3, Tone Gabrijelcic4, Peter Rakovec1 and Ales Brecelj4

1 Department of Cardiology, University Medical Centre Ljubljana, Zaloska 7, 1000 Ljubljana, Slovenia; 2 Vitatron B.V., Meander 1051, 6825 MJ Arnhem, The, Netherlands; 3 Educational Research Institute, Gerbiceva 62, 1000 Ljubljana, Slovenia; 4 Department of Cardiovascular Surgery, University Medical Centre Ljubljana, Zaloska 7, 1000 Ljubljana, Slovenia

Manuscript submitted 19 September 2005. Accepted after revision 16 May 2006.

* Corresponding author: Tel: +386 1 522 28 44; fax: +386 1 522 28 28E-mail address: igor.zupan{at}mf.uni-lj.si


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
Aims The aim of this retrospective analysis was to investigate VDD mode survival, development of atrial tachyarrhythmias (AT), and long-term atrial sensing performance of VDD pacing systems.

Methods and results We implanted single-lead VDD pacemakers in patients with isolated atrioventricular block and performed a retrospective analysis of 307 patients who had their devices implanted between May 1994 and September 2001. In 39 patients (12.7%), the pacing mode had to be reprogrammed to a single-chamber ventricular pacing mode, mostly due to permanent AT. In 16 of these patients, the atrial sensing safety margin was less than 150%. The atrial sensing safety margin was insufficient, i.e. less than 100% in only seven patients. Although only 12 (3.9%) of the patients had a history of paroxysmal AT at the time of pacemaker implantation, 200 (65%) patients presented with AT during follow-up. The mean AT burden at the last follow-up was 2.5%.

Conclusion These data illustrate that single-lead VDD pacemakers can be applied without serious complications in a highly selected group of patients. Our main concern is the development of AT in a large part of our population. Over a 10-year period, two thirds of our patients presented with AT.

Key Words: Single-lead VDD pacemaker, Long-term follow-up, Atrial tachyarrhythmia, Mode survival, Retrospective analysis


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
The importance of the atrial contribution to cardiac output has long been recognized.1Go–3Go Atrial synchronous ventricular pacing has both short and long term advantages over ventricular pacing.4Go In patients with atrioventricular (AV) block and preserved sinus rhythm, a VDD pacing system would be sufficient to eliminate the ventricular bradycardia and to provide AV synchrony and chronotropic response. The presence of paroxysms of atrial fibrillation does not prohibit the use of single-lead VDD pacemakers if the pacemaker features mode switching and a rate-response sensor (mode switch to VDIR pacing mode). Atrial bradycardia or sinus pauses are a contra-indication to VDD pacing systems, because they lack the ability to pace the atrium under these circumstances. Sensing of atrial depolarization in VDD mode is achieved through electrodes that are in physical contact with the atrial wall (dual-lead systems) or through floating electrodes (single-lead systems). Floating electrodes can be placed on the lead body of the ventricular lead at the atrial level. Such a system is referred to as a single-lead VDD pacing system. The single-lead VDD pacemaker concept was proposed as early as 1973 but became commercially available in the late 1980s.5Go–7Go

We have been implanting single-lead VDD pacing systems since the early 1990s according to the strict indication of preserved sinus rhythm in combination with AV block. Primary reasons for this policy are less foreign material in the body, shorter implantation times and radiation exposure, and fewer post implantation complications with lower follow-up costs.8Go,9Go We performed this retrospective analysis to investigate VDD mode survival, development of AT, and long term atrial sensing performance of these pacing systems.


    Methods
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
All patients were implanted with a single-lead VDD pacing system, consisting of a Saphir or Saphir II VDDR pacemaker and a Brilliant or Brilliant+ quadripolar lead (Vitatron, Arnhem, Netherlands). Between May 1994 and September 2001, we implanted and followed 441 patients with such pacing systems. Patient follow-ups were included in this analysis until September 2004. We included 307 patients in our analysis (mean age 75±13 years; 56% male; mean follow-up 5.98±1.59 years). We excluded patients from the analysis that were considered ‘lost to follow-up’ either because they did not have any follow-up visits in the last 18 months or they died and did not have a single follow-up visit.

We assessed sinus node function prior to pacemaker implant according to our local practice (long-term ECG, carotid sinus massage, adenosine, stress testing). For the mode survival analysis, we accepted only the VDD or VDDR mode. Patients were considered to have atrial tachyarrhythmia (AT), if this was noted in the patient's file by the physician and/or if the AT burden indicated 1% or higher during one or more of the follow-up visits. AT burden was defined as the percentage of the time that the atrial rhythm was classified as AT by the implanted pacemaker.

The Saphir VDD pacemakers include a P-wave amplitude histogram, showing the distribution of P-waves and AT signals over the entire follow-up period. The atrial sensing safety margin was defined as a quotient of the smallest amplitude in the P-wave amplitude histogram and the programmed atrial sensitivity. Values above 100% indicate reliable sensing.

Statistical analysis
Descriptive statistics (mean±standard deviation) were calculated for most of the data sets. VDD mode survival and incidence of AT were evaluated using Kaplan–Meier cumulative survival curve. Independent variables were dichotomized and one or two-tailed Mann–Whitney was applied to assess differences between the groups for dependent variables. A P value of less than 0.05 was considered statistically significant.


    Results
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
A total of 249 patients had at least one follow-up visit in the last 18 months and 58 patients died and had at least one follow-up visit.

The indication for pacing was AV block (91.7%), bundle branch block (6.3%), sick sinus syndrome (SSS) (1%), and other (1%). The atrial rhythm at the time of implantation was sinus rhythm (94.8%), paroxysmal atrial fibrillation (AF) (3.9%), SSS (1%), and missing (0.3%). The AV conduction status at the time of implantation was normal (3.6%), second-degree AV block (29%), complete (third degree) AV block (65.5%), and missing (2%).

VDD mode survival
Figure 1 shows the Kaplan–Meier cumulative survival curve for the VDD pacing mode. In the cohort of 307 patients, 39 (12.7%) patients had their pacing mode reprogrammed to a single-chamber ventricular pacing mode. In almost all (36) these patients, AT was the primary reason for mode reprogramming and atrial undersensing accounted for the remainder. The atrial sensing safety margin was less than 150% in 16 of the patients in whom the mode was reprogrammed. The atrial sensitivity was programmed to its lowest value (0.1 mV) in 14 of these patients (0.18 mV and 0.25 mV in the two other patients). The mean atrial sensing safety margin was higher in patients who remained in the VDD pacing mode compared with patients in whom the pacing mode was reprogrammed (223±156% vs. 165±106%; P<0.05).


Figure 0911
View larger version (17K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 1 Kaplan–Meier cumulative survival curve for the VDD pacing mode. Only 12.7% of the patients had their pacing mode reprogrammed to a single-chamber ventricular pacing mode.

 
Incidence and burden of atrial tachyarrhythmias
Figure 2 shows the Kaplan–Meier cumulative survival curve for AT free patients. Only 3.9% of the patients had a history of paroxysmal AT at the time of pacemaker implantation. Figure 3 illustrates the AT burden at the last follow-up visit (n=191). The mean AT burden was 2.5±8.6% and 104 (54.5%) of the patients were free of AT burden.


Figure 0912
View larger version (17K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 2 The cumulative AT-free survival curve shows that only one-third of patients remain AT free during follow-up.

 


Figure 0913
View larger version (25K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 3 Distribution of the AT burden at the last follow-up visit.

 
Aspects related to AT
In our patient cohort, the incidence of right ventricular apical pacing was high (92.1±15.1%) as the vast majority of patients had their pacemaker implanted because of high-degree AV block. The incidence of ventricular pacing did not differ significantly between the AT-free patients and patients with AT.

The percentage of the time during which the device operated in an AV synchronous mode was significantly higher in patients that were free of AT (87.2±18.0% for patients free of AT vs. 75.0±22.6% for patients with AT; P<0.001).

P-wave sensing
The P-wave amplitude was measured during a follow-up visit using the pacemaker programmer. Figure 4 shows the mean value of these measurements, illustrating that in most patients a sufficient atrial sensing safety margin is available with respect to the highest programmable sensitivity of 0.1 mV.


Figure 0914
View larger version (17K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 4 Mean value of the P-wave amplitude measured during a follow-up visit using the pacemaker programmer. The highest atrial sensitivity (0.1 mV) is also shown, illustrating that in most patients a sufficient atrial sensing safety margin is available.

 
The lowest P-waves in the P-wave amplitude histogram were 0.32±0.21 mV for patients free of AT and 0.27±0.28 mV for patients with AT (P=0.005). Using the lowest amplitude in the P-wave amplitude histogram, we found that the atrial sensing safety margin was less than 100% in seven patients. In these patients, the atrial sensitivity was programmed to 0.1 mV (n=2) and 0.18 mV or 0.25 mV (n=5) and therefore could have been improved in the majority of these patients. Atrial undersensing can cause intermittent AV asynchrony and more ventricular events may be classified as premature ventricular complex (PVC) by the pacemaker. The incidence of PVCs in the seven patients with a sensing safety margin of less than 100% was 8.0±18.8% vs. 1.2±4.2% in patients with a sensing safety margin of more than 100% (P<0.05). AV synchrony was significantly higher in patients with a sensing safety margin of more than 100% (79.2±21.4% vs. 66.5±25.7%; P<0.05).

Other results
Atrioventricular delay programming had no significant influence on the incidence of ventricular pacing because of the high-degree AV block in most patients.


    Discussion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
One of the appealing characteristics of single-lead VDD pacemakers is their simplicity: the implantation of a single lead allows the physician to obtain dual-chamber function. Several studies have shown that the rate of complications requiring surgical intervention, the implantation time, and the fluoroscopic time of these systems are comparable with that of VVI pacemakers and significantly lower than that of DDD pacing systems.10Go,11Go

There are several mechanisms in patients with single-lead VDD pacemakers that potentially affect the development of AT. A disadvantage of VDD pacing systems is the inability to pace the atrium and thus to prevent long atrial intervals e.g. as a result of sino-atrial block or sinus bradycardia.12Go In contrast, single-lead VDD pacemakers do not injure the atrial wall nor do the atrial electrodes pose a mechanical stress on the atrial wall, factors that may facilitate development of AT.

In our study, we have found that 65.1% of our patients revealed AT during follow-up. We used the pacemaker arrhythmia counters to access the presence of AT. Because the leads that we used all had an atrial inter-electrode distance of 10 mm, we feel that far field R-wave sensing, and thus false positive AT counting and over-estimation of AT burden, is not significantly affecting the results.13Go

Our results are not consonant with the results published by Kong et al.14Go and Folino et al.15Go An important difference between their methodology and ours is the determination whether AT is present in a given patient. Their retrospective analysis relied on documented AT in the ECG during follow-up or in 24 h Holter ECG recordings. This may have influenced their results as AF is a rather random phenomenon that may occur in ‘storms’.16Go Using the pacemaker diagnostics to diagnose the presence of AT avoids under-reporting whereas over-reporting, e.g. due to far field sensing, is expected to be limited. This may explain why our results show a higher proportion of patients with AT. Schuchert et al. used pacemaker diagnostic data to study the incidence of AT in patients with no prior history of them.17Go They concluded that AT was detected within one year in 50% of patients with a single-lead VDD pacemaker. Other studies that used pacemaker diagnostics to monitor the incidence of AT have shown similar results.18Go These results are more in line with ours.

Paroxysms of AT are usually not a contraindication for single-lead VDD pacemakers if these devices feature mode switching to a non-tracking pacing mode upon detection of AT.

If patients with single-lead VDD pacemakers require replacement of their pacemaker and they have AT, then upgrading to a DDDR system should be seriously considered, as atrial pacing may reduce the AT burden.19Go

Several recent studies have shown that a high percentage of right ventricular apical pacing is associated with a higher incidence of AF.20Go,21Go However, the incidence of ventricular pacing in our group did not differ significantly between the AT-free patients and patients with AT.

With respect to mode survival, our results are in line with other long-term follow-up studies with single-lead VDD pacemakers. Sassara et al.8Go reported a mode survival of 94.6% for a follow-up period of almost 4 years (90.6% after 4 years in our study) and Rey et al.22Go reported 95%, whereas Chamberlain-Webber et al.23Go reported 88.6% for a 2 year follow-up period (94.8% after 2 years in our study).

The initial concern about single-lead VDD pacemakers was the reliability of atrial sensing using a floating dipole.24Go–26Go Long-term atrial sensing of single-lead VDD pacing systems is reliable although the floating atrial electrodes in VDD pacing systems yield smaller P-waves when compared with dual-lead pacing systems.11Go The P-wave amplitude decreases during the immediate post-implantation phase and afterwards remains relatively constant.27Go The high atrial sensitivity that can be programmed in these devices allows the programming of an adequate sensing safety margin.


    Conclusions
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
Our study shows that single-lead VDD pacemakers can be applied without major technical difficulties in a highly selected group of patients. Over a follow-up period of 10 years, only 12.7% of the patients required mode reprogramming to a single-chamber ventricular pacing mode (Kaplan–Meier), the primary reason being the development of AT.

Long-term atrial sensing is reliable and, in the vast majority of patients (97.7%), an adequate sensing safety margin was sustained over a long period of time, even for patients with AT.

Our main concern is the development of AT in a large part of our population. During the follow-up period, two-thirds of our patients presented with AT.

Therefore, we suggest that single-lead VDD pacemakers should be implanted in carefully selected patients and that patients with these devices should be monitored closely for the development of AT.


    References
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusions
 References
 
[1] Carleton RA, Passovoy M, Graettinger JS. The importance of the contribution and timing of left atrial systole. Clin Sci 1966; 30: 151–9.[ISI][Medline]

[2] Fananapazir L, Bennett DH, Monks PH. Atrial synchronized pacing: Contribution of the chronotropic response to improved exercise performance. Pacing Clin Electrophysiol 1983; 6: 601–8.[CrossRef][Medline]

[3] Kristensson B, Arnman K, Rydén L. The haemodynamic importance of atrioventricular synchrony and rate increase at rest and during exercise. Eur Heart J 1985; 6: 773–78.[Abstract/Free Full Text]

[4] Kruse I, Arnman K, Conradson TB, Rydén L. A comparison of the acute and long term hemodynamic effects of ventricular inhibited and atrial synchronous ventricular inhibited pacing. Circulation 1982; 65: 846–55.[Abstract/Free Full Text]

[5] Chamberlain DA, Woollons DJ, White NM, English MJ. Synchronous AV pacing with a single pervenous electrode. Br Heart J 1973; 35: 559.[Medline]

[6] Cornacchia D, Fabbri M, Maresta A, Grassi G, Vaiani P. Clinical evaluation of VDD pacing with a unipolar single- pass lead. Pacing Clin Electrophysiol 1989; 12: 604–18.[CrossRef][Medline]

[7] Varriale P, Pilla AG, Tekriwal M. Single lead VDD pacing system. Pacing Clin Electrophysiol 1990; 13: 757–66.[CrossRef][Medline]

[8] Sassara M, Achilli A, Guerra R, et al. Long-term clinical assessment of single-lead VDD electric stimulation. Ital Heart J. 2000; 1:(Suppl.), 777–82.

[9] Wiegand UK, Potratz J, Bode F, Schreiber R, Bonnemeier H, Peters W, Katus HA. Cost-effectiveness of dual-chamber pacemaker therapy: does single lead VDD pacing reduce treatment costs of atrioventricular block? Eur Heart J 2001; 22: 174–80.[Abstract/Free Full Text]

[10] Wiegand UK, Bode F, Bonnemeier H, Eberhard F, Schlei M, Peters W. Long-term complication rates in ventricular, single lead VDD, and dual chamber pacing. Pacing Clin Electrophysiol 2003; 26: 1961–9.[CrossRef][Medline]

[11] Huang M, Krahn AD, Yee R, Klein GJ, Skanes AC. Optimal pacing for symptomatic AV block: a comparison of VDD and DDD pacing. Pacing Clin Electrophysiol 2004; 27: 19–23.[CrossRef][Medline]

[12] Israel CW and Hohnloser SH. Pacing to prevent atrial fibrillation. J Cardiovasc Electrophysiol 2003; 14: S20–6.[CrossRef][ISI][Medline]

[13] Nash A, Frohlig G, Taborsky M, et al. Rejection of atrial sensing artifacts by a pacing lead with short tip-to-ring spacing. Europace 2005; 7: 67–72.[Abstract/Free Full Text]

[14] Kong CW, Yu WC, Chen SA, Lin YJ, Huang CY, Chung SL. Development of atrial fibrillation in patients with atrioventricular block after atrioventricular synchronized pacing. Pacing Clin Electrophysiol 2004; 27: 352–7.[CrossRef][Medline]

[15] Folino AF, Buja G, Corso L Dal, Nava A. Incidence of atrial fibrillation in patients with different mode of pacing. Long-term follow-up. Pacing Clin Electrophysiol 1998; 21: 260–3.[CrossRef][Medline]

[16] Patel AM, Westveer DC, Man KC, Stewart JR, Frumin HI. Treatment of underlying atrial fibrillation. Paced rhythm obscures recognition. J Am Coll Cardiol 2000; 36: 784–7.[Abstract/Free Full Text]

[17] Schuchert A, Lepage S, Ostrander JJ, et al. Automatic analysis of pacemaker diagnostic data in the identification of atrial tachyarrhythmias in patients with no prior history of them. Europace 2005; 7: 242–7.[Abstract/Free Full Text]

[18] Pignalberi C, Ricci R, Canale G, et al. Paroxysmal atrial fibrillation in patients paced for atrioventricular block. Comparison between DDD and VDD single-lead pacing system. Ital Heart J 2001; 2: 772–7.[Medline]

[19] Hoffmann E, Janko S, Steinbeck G, Edvardsson N, Camm J. Onset scenarios of paroxysmal atrial fibrillation using new diagnostic pacemaker functions. (Abstract). Pacing Clin Electrophysiol 2000; 23-II: 415A.

[20] Sweeney MO, Hellkamp AS, Ellenbogen KA, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation 2003; 107: 2932–7.[Abstract/Free Full Text]

[21] Kristensen L, Nielsen JC, Mortensen PT, Pedersen OL, Pedersen AK, Andersen HR. Incidence of atrial fibrillation and thromboembolism in a randomised trial of atrial versus dual chamber pacing in 177 patients with sick sinus syndrome. Heart 2004; 90: 661–6.[Abstract/Free Full Text]

[22] Rey JL, Tribouilloy C, Elghelbazouri F, Otmani A. Single-lead VDD pacing: long-term experience with four different systems. Am Heart J 1998; 135: 1036–9.[CrossRef][ISI][Medline]

[23] Chamberlain-Webber R, Barnes E, Papouchado M, Crick JP. Long-term survival of VDD pacing. Pacing Clin Electrophysiol 1998; 21: 2246–8.[CrossRef][Medline]

[24] Toivonen L and Lommi J. Dependence of atrial sensing function on posture in a single lead atrial triggered ventricular (VDD) pacemaker. Pacing Clin Electrophysiol 1996; 19: 309–13.[CrossRef][Medline]

[25] Taborsky M, Neuzil P, Vopálka R, Rezak Z, Bicik V. Acute and long term stability of P-wave sensing in a new VDDR single pass lead pacing system. (Abstract). Pacing Clin Electrophysiol 1993; 16: 106A.

[26] Toivonen L, Laine P, Stjemvall J. High safety margin requirements for atrial floating electrode sensing in single-lead VDD pacemakers. (Abstract). Pacing Clin Electrophysiol 1996; 19: 464A.

[27] Liebold A, Merk J, Keyl C, Aebert H, Birnbaum DE. Clinical results with single lead VDD pacing. Eur J Cardiothorac Surg 1997; 11: 722–7.[Abstract]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?



This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
8/10/855    most recent
eul091v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Zupan, I.
Right arrow Articles by Brecelj, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Zupan, I.
Right arrow Articles by Brecelj, A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?