SHORT SERIES REPORT
Runaway pacemaker: A forgotten phenomenon?
Instituto Cardiovascular de Buenos Aires, Electrofisiología, Blanco Encalada 1543 DCO 1428 Buenos Aires, Argentina
Manuscript submitted 21 January 2005. Accepted after revision 1 June 2005.
*Corresponding author. Tel.: +54 1148635770; fax: +54 1147877500. E-mail address: vicsam1604{at}yahoo.com.ar
Abstract
Runaway is an uncommon pacemaker dysfunction, characterized by fast and erratic spikes at non-physiological rates.
This infrequent but potentially lethal failure mode may be related to low battery voltage.
Four single chamber pacemaker patients were analyzed (Medtronic Minix ST 8330, Minneapolis, MN, had been implanted in two patients and two CPI Triumph VR 1124, St Paul, MN, in the other two). They had been admitted because of presyncopal episodes. Typical high rate stimuli at 2000 ppm alternating with pacing at 6065 ppm were recorded in all ECGs. Lead system tests were normal. The pulse generators had to be replaced.
Key Words: failure of capture, battery depletion, runaway
Introduction
Pacemaker patient follow-up is recommended to be more frequent near replacement time to avoid generator failure due to low or depleted battery voltage.
Runaway is a rare but dangerous phenomenon due to generator dysfunction which may be related to its late replacement and battery depletion.
Low battery voltage can provoke low rate stimuli, capture and sensing failures, mode changes and less frequently the runaway phenomenon.
This event shows a typical ECG with captured beats alternating with non-captured high rate spikes. Fortunately, in these cases the patients displayed spontaneous rhythm.
Runaway could have been avoided by timely pacemaker replacement[1]
.
Case 1
A Medtronic model Minix ST 8330 (Minneapolis, MN) was implanted in a 67-year-old patient in another institution in 1994 because of complete atrioventricular (AV) block, with no recent follow-up.
He was admitted to our Intensive Care Unit because of presyncopal episodes for 3 days.
He was alert, with an average heart rate of 40 bpm, and his blood pressure was 130/80 mmHg.
The ECG showed a pacemaker rhythm at 60 ppm in VOO mode interrupted by repeated spike bursts at 2000 ppm of 3 s duration and progressively decreasing amplitude.
During bursts, atrial flutter was observed with an escape ventricular rhythm at 30 bpm.
Low blood pressure was observed during runaway episodes (Fig. 1).
|
Magnet manoeuvres did not provoke any change.
The generator was interrogated by telemetry (Fig. 2) and showed VOO mode programmed at 70 ppm. Battery depletion was detected: 1.83 V, an extremely low level, well below elective replacement level[2,
3]
. A Medtronic model Thera SR 8942 (Minneapolis, MN) was implanted to replace the exhausted device. The lead system was retained because threshold and resistance values were in acceptable ranges.
|
Case 2
A VVIR, CPI model Triumph 1124 (St Paul, MN), was implanted in a 78-year-old patient in another institution in 1994 because of complete AV block, programmed at a low rate of 60 ppm.
This patient was admitted to our Intensive Care Unit because of presyncopal episodes, normotensive with an average heart rate of 40 bpm.
The ECG showed 2000 ppm spikes with decreasing amplitude without any ventricular capture, and an escape rhythm recorded at 40 bpm followed by isolated captured beats at 60 ppm (Fig. 3).
|
Magnetic manoeuvres did not provoke any change.
Telemetric reports were not obtained due to lack of telemetric communication. The patient had not attended follow-up for 2 years.
Case 3
This case was referred by another institution and was admitted to the Intensive Care Unit because of syncopal episodes and stroke. A Medtronic ST 8330 pacemaker (Minneapolis, MN) had been implanted 8 years earlier because of complete AV block.
A typical high rate spike ECG alternating with captured beats at 60 ppm was recorded (Fig. 4).
|
Only the pacemaker was replaced as threshold and resistance of the ventricular lead were normal.
Case 4
A 55-year-old woman with a history of aortic valve replacement 24 years earlier, due to severe aortic stenosis, required a VVIR pacemaker because of complete AV block.
She had undergone five generator replacements due to depleted battery, the last in 1996.
She was admitted to our institution for dizziness, presyncopal episodes, and fatigue. Her ECG (Fig. 5) showed a typical runaway phenomenon. The pacemaker was replaced by a Medtronic VVI. During her hospital stay, she presented a "Torsade de Pointes" episode that was successfully defibrillated (Fig. 6).
|
|
Discussion
A typical runaway ECG (Fig. 1, 35) shows intermittent ventricular capture at a rate slower than normal, numerous spikes at a very high rate with different voltages and often without capture beats, simulating an ECG artifact.
Capture beat rate is low because this phenomenon occurs at the end of pacemaker life with depleted battery voltage.
High rate spikes fail to capture because of low amplitude and due to the fact that at such very high rates the ventricle is refractory to many stimuli.
Two clinical aspects were important:
- The patients survived because they were not pacemaker-dependent.
- Runaway did not provoke any potentially lethal arrhythmias. The "torsade de pointes" episode in one patient was related to bradyarrhythmia and not due to overstimulation.
This phenomenon is observed when elective replacement is overdue and the battery had low voltage, well below the safety margin[4
6]
.
The runaway phenomenon was first described in external and internal pacemakers in 1970 and it is one of the causes of sudden death due to pacemaker failure[1]
. Interestingly, these findings had already been attributed to an electric circuit failure, early in pacemaker implantation in Argentina[7,
8]
.
However, in the last decade only isolated cases have been published [9,
10
].
The most frequent causes of runaway are associated with:
- Sudden failure in one of the pacemaker circuits[3]
.
- Battery depletion.
- Circuit destruction or deterioration due to electric diathermy or therapeutic radiation[5]
.
Runaway was also described in implantable cardioverter defibrillator patients due to crystal oscillator failure[2]
.
Today, pacemakers and defibrillators have security designed to prevent runaway: the presence of a maximum rate, hermetic sealing, and a decrease in pulse amplitude at high rates with concomitant loss of capture. Despite these measures, correct performance of the pacemaker below the security battery voltage is not guaranteed[5]
.
Even though the potential runaway hazard is induction of ventricular fibrillation or other ventricular arrhythmias, loss of capture and bradyarrhythmias or asystole in pacemaker-dependent patients must also be considered.
The patients in this report did not have frequent follow-up visits after implantation. In fact, it was impossible to determine the presumed replacement period.
Patients were almost asymptomatic at rest so no active measure was taken at day 1. Pacemakers were replaced on the day after admission. The most appropriate behaviour should have been urgent pacemaker removal because, as mentioned earlier, during runaway periods ventricular arrhythmias might occur. Furthermore, intermittent capture could induce subsidiary foci suppression by overdrive pacing and the bradycardia in the complete heart block can provoke "torsade de pointes".
This potentially lethal phenomenon can be avoided by having appropriate follow-up. We presume that there might have been other cases which were not identified because their first episode was lethal.
Conclusions
The runaway phenomenon is associated with very low battery voltage together with late pacemaker replacement.
It is a potentially lethal clinical entity.
Few reports do not mean an absence of this phenomenon.
Appropriate follow-up programmes will prevent future pacemaker failures.
References
[1] Furman S. and Escher D. Principles and techniques of cardiac pacing. 1970; New York, NY Harper and Row 1012.
[2] Mickley H., Anderson C., Nielson H. Runaway pacemaker: a still existing complication and therapeutic guidelines. Clin Cardiol 1989; 12: 412414.[Medline]
[3] Solow E., Bacharach B., Chung E.K. Runaway pacemaker. Unpredictable pacemaker failure. Arch Intern Med 1979; 139: 11901191.
[4] Carpenter C.M., Galvin J., Guy M., et al. Runaway pacemaker in an implantable cardioverter defibrillator. J Cardiovasc Electrophysiol 1998; 9: 10081011.[CrossRef][Web of Science][Medline]
[5] Katzenburg C.A., Marcus F., Heusinkveld R.S., et al. Pacemaker failure due to radiation therapy. Pacing Clin Electrophysiol 1982; 5: 156159.[CrossRef][Medline]
[6] Vijayaraman P., Vaidya K., Kim S.G., Gross J.N., Ferrick K.J., Palma E., et al. Runaway pulse generator malfunction resulting from undetected battery depletion. Pacing Clin Electrophysiol 2002; 25: 220222.[Medline]
[7] Posse R., Bein I.J., D'Aiutolo R. Arritmias cardíacas en pacientes con marcapaso electrónico implantado. Rev Arg Cardiol 1967; 34: 4749.
[8] D'Aiutolo R. and Posse R. Tratamiento de las arritmias cardíacas. 1968; BsAs Argentina Cabrera y Rey editores 6971.
[9] Bohm A., Hajdu L., Pinter A., Duray G., Preda L. Runaway syndrome and intermittent nonoutput as manifestations of end of life of a VVI pacemaker. Pacing Clin Electrophysiol 2000; 23: 21432144.[Medline]
[10] Griffin J., Smithline H., Cook J. Runaway pacemaker: a case report and review. J Emerg Med 2000; 19: 177181.[Medline]
![]()