Skip Navigation

Europace 2005 7(6):525-534; doi:10.1016/j.eupc.2005.05.016
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Stagegaard, N.
Right arrow Articles by Svendsen, J. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Stagegaard, N.
Right arrow Articles by Svendsen, J. H.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© 2005 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.

Indication of the radiofrequency induced lesion size by pre-ablation measurements*

Niels Stagegaard, Helen Høgh Petersen*, Xu Chen and Jesper Hastrup Svendsen

Medical Department B 2012, The Heart Centre, Rigshospitalet, National University Hospital Blegdamsvej 9, 2100 Copenhagen, Denmark

Manuscript submitted 10 August 2004. Accepted after revision 25 May 2005.

*Corresponding author. Tel.: +45 35 45 21 42; fax: +45 35 45 27 05. E-mail address: hhp{at}dadlnet.dk


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Clinical implications
 Study limitations
 Acknowledgements
 References
 
BACKGROUND: During radiofrequency ablation of arrhythmias tissue heating and hence lesion size depend on electrode–tissue contact and cooling of the electrode tip caused by cavitary blood flow. These factors are unique and unknown for each catheter placement in the beating heart. A tool for assessing these factors prior to ablation may indicate the lesion size which will be obtained for any given catheter position.

METHODS AND RESULTS: Radiofrequency ablation was performed in vitro on strips of left ventricular porcine myocardium during two different levels of convective cooling (0 or 0.1 m/s), two different contact pressures (10 or 30 g) and parallel or perpendicular electrode–tissue orientation using 7F 4 mm tip catheters. Prior to ablation the impedance rise ({Delta}IMP) caused by the obtained contact and the temperature rise with a 0.6 W 5 s test pulse ({Delta}T) were measured. Subsequently, during unchanged conditions, radiofrequency ablation was performed as either temperature-controlled, power-controlled or irrigated tip ablation and lesion size was determined.

{Delta}IMP increased significantly (P < 0.05) by improved contact, whereas it was not affected by convective cooling. {Delta}T was significantly increased by increasing contact pressure (P < 0.05) and significantly decreased by increased cooling (P < 0.001). {Delta}T was not systematically affected by electrode orientation. The product of {Delta}T and {Delta}IMP showed a significant correlation between the obtained lesion size and power output for temperature-controlled and between lesion size and tip temperature for power-controlled ablation (P < 0.001).

CONCLUSIONS: Pre-ablation measurement of {Delta}IMP and {Delta}T can indicate the lesion size resulting after ablation in temperature-controlled, power-controlled and irrigated ablation in vitro, since {Delta}T reflects cavitary cooling and to a smaller extent electrode–tissue contact, and {Delta}IMP reflects only electrode–tissue contact.

Key Words: catheter ablation, impedance, electrode contact, external cooling, power-controlled ablation, temperature-controlled ablation, irrigated tip, radiofrequency ablation


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Clinical implications
 Study limitations
 Acknowledgements
 References
 
Radiofrequency catheter ablation has proven to be a safe and effective means of treatment of many tachycardias[1–Go5]Go, but during ablation even in the temperature-controlled mode the convective cooling of the electrode tip can cause excess power delivery, thereby causing tissue overheating[6]Go. It is therefore necessary to control the tissue heating in order to obtain the lesion size needed without excess power delivery. This requires knowledge of, and adjustment for, the factors influencing lesion size. Several factors have been shown to affect the obtained tissue temperatures and thereby lesion size during radiofrequency catheter ablation. Some of these factors can be controlled by the choice of catheter and by the setting of the radiofrequency generator, that is: electrode tip size[7–Go11]Go, energy application time[12,Go13]Go, irrigation rate[14–Go16]Go, power or temperature setting[17–Go20]Go. Other factors, which are difficult or impossible to control, such as convective cooling of the electrode tip [6,Go10,Go11,Go21–Go24]Go, electrode–tissue contact pressure [12,Go14,Go24–Go28]Go and electrode tip orientation against the endocardium [10,Go21,Go23,Go24,Go29,Go30]Go have also been shown to have major impact on tissue temperatures and hence lesion size. Since these "uncontrollable" factors are unique for each obtained catheter position in the heart, they are unknown.

To control lesion size, one must adjust power or temperature setting according to the actual obtained electrode–tissue contact and actual convective cooling of the electrode tip at the given site. Therefore, pre-ablation evaluation of parameters reflecting these factors might be a valuable tool.

The purpose of the present in vitro study was to evaluate if measurement of the impedance rise caused by the variable electrode–tissue contact correlates with the lesion size obtained and to evaluate whether this impedance change can add information to the correlation between lesion size and the temperature rise from a pre-ablation low power test pulse, which in other studies has been shown to correlate with lesion size[23,Go26]Go.


    Methods
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Clinical implications
 Study limitations
 Acknowledgements
 References
 
The model
Freshly excised pig hearts, used less than 4 h postmortem, were dissected and strips of left ventricular myocardium were suspended in a bath of isotonic saline at 37 °C.

The electrode was mounted in a holder maintaining the tip either perpendicular or parallel to the tissue. The electrode–tissue contact pressure was controlled by a strain gauge dynamometer adjusted to maintain a pressure of either 10 or 30 g. A pump circulating the saline in the bath releasing it through a laminator close to the electrode tip created external convective cooling around the electrode tip, simulating cavitary blood flow. The pump was adjusted to create a flow velocity of either 0 or 0.1 m/s around the electrode tip measured by pulsed wave Doppler (Toshiba Sonolayer SSA-270A) using an oesophageal probe (Hewlet Packard Omniplane 5 mHz PEF 511-SA).

Ablation equipment
Two radiofrequency generators were used. For pre-ablation impedance measurement a generator capable of measuring impedance at 50 kHz prior to initiating ablation was used (EP-SHUTTLE, Stockert GMBH, Freiburg, Germany).

For test pulse delivery and unipolar ablation a generator delivering radiofrequency current with a frequency of 500 kHz and maximal power output of 100 W (ATAKR II, Medtronic CardioRhythm, Minneapolis, MN, USA) was used. This generator is capable of delivering a short non-destructive output of 0.6 W for 30 s. (LEMTM mode)[26]Go. The change in electrode-tip temperature during LEMTM is displayed continuously on the generator with one digit. The accuracy of the temperature measurement is 0.1 °C[31]Go.

Thermocouple-type 7F, 4 mm tip catheters, either conventional non-cooled (ABLATR, Medtronic CardioRhythm, San Calif., USA) or irrigated showerhead-type (SPRINKLR, Medtronic CardioRhythm, San Calif., USA) were used. During ablation the electrode-tip temperature, power output and impedance were continuously monitored. Power output was automatically discontinued if impedance was outside the range of 40–250 {Omega} or if tip temperature exceeded 90 °C. A Harvard "44" syringe pump was used to control infusion rate through the irrigation catheters.

Measurement of pre-ablation parameters
Prior to each ablation impedance at 50 kHz was measured with the convective cooling adjusted according to the protocol (see below) and the tip placed approximately 2 mm above the tissue and then measured again when electrode–tissue contact was obtained with the desired orientation and pressure (see below). The difference between these two impedance measurements, i.e. the increase in impedance obtained by electrode–tissue contact ({Delta}IMP) was registered.

Then with unchanged electrode position and unchanged cooling in the tissue bath the low power test pulse (0.6 W) was delivered and the increase in tip temperature ({Delta}T) at 5 s was registered. After approximately 1 min, to allow tip temperature to normalize, ablation was performed with unchanged contact and convective cooling.

Ablation protocols
Electrode orientation was either perpendicular or parallel. Electrode–tissue contact pressure was either 10 or 30 g and convective cooling was either 0 or 0.1 m/s.

Pre-ablation measurements were performed as described above.

Ablation with non-cooled catheters was performed in two different modes: temperature-controlled ablation with a target temperature of 70 °C and power-controlled ablation with a power setting of 15 W.

Ablation with irrigated catheters was performed in the power-controlled mode with a power setting of 40 W and an irrigation rate of 20 ml/min using isotonic saline at 37 °C.

Energy application time was 60 s for all modes of ablation.

After each ablation the average power output and average tip temperature were registered.

Tissue preparation and lesion volume determination
The lesions were cut by two incisions through the centre perpendicular to the surface and perpendicular to each other. The tissue blocks were then incubated with 0.5 mg nitroblue tetrazolium/ml 0.2 M Soerenson's buffer for 10 min at 37 °C. The maximum lesion depth and width in both dimensions (widtha and widthb) were then measured with calipers.

Volume was calculated as {pi} × 6–1 × depth × widtha × widthb.

Statistical analysis
Values are expressed as mean ± SD. Assessments of relations were performed using linear or parabolic regression, the least squares method was used for curve fitting. Individual groups were compared by two-sample t-test. Values of P < 0.05 were considered statistically significant. Statistical software Excel 2000 was used.


    Results
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Clinical implications
 Study limitations
 Acknowledgements
 References
 
A total of 213 lesions were produced. Eighty-four lesions were performed in the temperature-controlled mode, 86 lesions in the power-controlled mode and 43 lesions were performed with irrigated tip ablation. All applications were continued for the planned 60 s duration.

During four applications popping occurred. In two of these lesions a crater was seen. All of these four applications were performed in the temperature-controlled mode with 10 g pressure and external cooling of 0.1 m/s. Compared with lesions with no popping, these four applications had significantly lower {Delta}T and {Delta}IMP (P < 0.05) and mean power output and lesion volume were significantly higher compared with lesions where popping did not occur (P < 0.05).

IA. Pre-ablation measurements: relation to electrode–tissue contact and external cooling, non-irrigated ablation (n = 170)

  1. {Delta}T: When electrode–tissue contact pressure was increased from 10 to 30 g a significant increase in {Delta}T was seen for both electrode orientations and for both flow situations (P < 0.05), whereas {Delta}T was not significantly altered by changing electrode orientation except in the group of applications with 10 g contact pressure in the flow situation, where {Delta}T was slightly, but significantly higher for the parallel orientation compared with the perpendicular orientation (P < 0.05). Increasing external cooling by changing the flow in the tissue bath from 0 to 0.1 m/s significantly decreased {Delta}T for all settings (P < 0.001) (Table 1).


View this table:
[in this window]
[in a new window]
 
Table 1 Pre-ablation measurements for standard and irrigated tip ablation

 
{Delta}IMP: When electrode–tissue contact pressure was increased from 10 to 30 g a significant increase in {Delta}IMP was seen for both electrode orientations and for both flow situations (P < 0.05). Changing electrode orientation from parallel to perpendicular caused an increase in {Delta}IMP for both flow situations, which reached statistical significance for a contact pressure of 30 g (P < 0.001), whereas the increase was not statistically significant for a contact pressure of 10 g. Changing external cooling by changing the flow in the tissue bath did not cause any changes in {Delta}IMP, except for a contact pressure of 10 g in the perpendicular orientation, where {Delta}IMP was slightly, but significantly higher (P < 0.05) in the no-flow situation.

IB. Pre-ablation measurements: relation to electrode–tissue pressure and external cooling, irrigated ablation (n = 43), Table 1
When electrode–tissue contact pressure was increased from 10 to 30 g a significant increase in {Delta}IMP was seen for both electrode orientations and for both flow situations (P < 0.001). Changing external cooling by changing the flow in the tissue bath did not change {Delta}IMP, except for a contact pressure of 10 g where {Delta}IMP was significantly decreased, when flow was increased (P < 0.05).

{Delta}T was not relevant for irrigated ablation as the high irrigation rate cools the catheter tip to an extent, where the applied 0.6 W test pulse cannot heat the tissue sufficiently to heat the tip to any measurable extent.

II. Pre-ablation measurements: relation to subsequent ablation parameters and lesion size
As the impact of changing the electrode orientation, contact pressure and flow around the electrode varies with the applied ablation mode, the relations between the two pre-ablation parameters ({Delta}T and {Delta}IMP) and the tip temperature and power output reached during ablation, as well as the lesion size obtained are considered separately for each ablation mode given below.

IIA. Power-controlled ablation
Combining all groups of electrode orientation, electrode contact pressure and flow around the electrode tip (N = 86), simulating the clinical situation, where neither of these parameters are known, showed a modest, but significant positive correlation (r = 0.49, P < 0.001) between {Delta}T and lesion volume, and between {Delta}T and average tip temperature achieved (r = 0.90, P < 0.001) (Table 2A and Fig. 1).


View this table:
[in this window]
[in a new window]
 
Table 2 Ablation parameters and lesion size for standard ablation

 



View larger version (16K):
[in this window]
[in a new window]
 
Figure 1 For all power-controlled applications, independent of the electrode–tissue contact (pressure and orientation) and independent of external cooling, both of which, in a clinical setting, are unknown, the relations between the product of the two pre-ablation parameters and the lesion volume obtained and the average tip temperature reached are shown. For abbreviations see text.

 
We found a significant positive correlation between {Delta}IMP and lesion volume (r = 0.64, P < 0.001), and a poor, but significant positive correlation between {Delta}IMP and average tip temperature achieved (r = 0.35, P < 0.01). Combining the two pre-ablation parameters improved the correlations for both lesion volume and tip temperature: A significant positive correlation (r = 0.67, P < 0.001) between the product {Delta}T × {Delta}IMP and lesion volume (Fig. 1A), and between the product {Delta}T × {Delta}IMP and lesion depth (r = 0.57, P < 0.001) and also between the product {Delta}T × {Delta}IMP and average tip temperature was achieved (r = 0.90, P < 0.001) (Fig. 1B).

The impact of pressure, orientation and cooling on lesion size and tip temperature achieved in the present study is consistent with the findings from earlier studies.

IIB. Temperature-controlled ablation
Combining all groups of electrode orientation, electrode contact pressure and flow around the electrode tip (N = 84), simulating the clinical situation, where neither of these parameters are known, showed a significant negative correlation between {Delta}T and lesion volume (r = –0.87, P < 0.001) and a negative correlation between {Delta}T and average power output (r = –0.91, P < 0.001). We found no correlation between {Delta}IMP and lesion volume (r < 0.1, P > 0.05), but a significantly negative correlation between {Delta}IMP and average power output (r = –0.61, P < 0.001) (Table 2B and Fig. 2).



View larger version (16K):
[in this window]
[in a new window]
 
Figure 2 For all temperature-controlled applications, independent of the electrode–tissue contact (pressure and orientation) and independent of external cooling, both of which, in a clinical setting, are unknown, the relations between the product of the two pre-ablation parameters and the lesion volume obtained and the average power applied are shown. For abbreviations see text.

 
Combining the two pre-ablation parameters did not change the correlation for lesion volume, but improved the correlation for average power output: a significant negative correlation was found between the product {Delta}T × {Delta}IMP and lesion volume (r = –0.67, P < 0.001) (Fig. 2A), and between the product {Delta}T × {Delta}IMP and lesion depth (r = –0.58, P < 0.001), and a significant negative correlation was found between the product {Delta}T × {Delta}IMP and average power output (r = –0.82, P < 0.001) (Fig. 2B).

The impact of pressure, orientation and cooling on lesion size and average power output in the present study is consistent with the findings from earlier studies.

IIC. Irrigated ablation
Combining the two levels of external cooling and the two contact pressures (N = 43) we found a significant positive correlation between {Delta}IMP and lesion volume (r = 0.82, P < 0.001) (Fig. 3) and between {Delta}IMP and lesion depth (r = 0.71, P < 0.001) (Table 3).


View this table:
[in this window]
[in a new window]
 
Table 3 Irrigated tip ablation

 
Consistent with earlier studies[14]Go we found that increasing electrode–tissue contact pressure increased lesion dimensions significantly (P < 0.05), whereas increasing external cooling did not change lesion dimensions significantly.


    Discussion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Clinical implications
 Study limitations
 Acknowledgements
 References
 
Lesion size is determined by the amount of energy delivered by the generator and by the fraction of this delivered energy that actually enters the tissue. These two parameters, in turn, are determined by the electrode-to-tissue contact and by the external cooling of the electrode tip, which are two unknown factors, since they are unique for each electrode position. This can be explained as follows:

The lesion size obtained during radiofrequency ablation is correlated with the tissue temperature reached during energy application [10,Go21,Go22,Go32–Go34]Go. This, in turn, is determined by the amount of energy entering the tissue. Since the electrode tip is not completely embedded in the tissue, only part of the energy delivered by the radiofrequency generator is entering the tissue, the rest of the energy is "wasted" into the blood. How large the fraction of the energy entering the tissue is, depends almost entirely on the degree of tissue contact obtained, which, as demonstrated in the present study can be estimated by pre-ablation measurement of {Delta}IMP, since the conductivity of blood (or saline) and tissue differs (0.95 and 0.61 1/{Omega} m at 37 °C, respectively)[35]Go. The impedance increase which occurs when the electrode tip is moved from "no contact" to "contact", will therefore reflect the fraction of the tip touching tissue.

The amount of energy delivered by the generator is again determined, either by the operator during power-controlled ablation or by automatic regulation during temperature-controlled ablation up to the power limit set by the operator. In both modes the amount of energy that can be or is delivered is highly dependent on the cooling of the electrode for the particular electrode position obtained, which as demonstrated in the present study, can be estimated by pre-ablation measurement of {Delta}T.

Therefore, the combination of evaluating electrode-to-tissue contact (determines the fraction of delivered energy that will enter the tissue) and cooling (determines power output during temperature-controlled ablation, and heat-loss during power-controlled ablation) can give an indication of tissue heating.

{Delta}T and {Delta}IMP have to be interpreted with consideration of the planned ablation mode, since the impact on lesion size caused by contact and cooling differs for the different ablation modes. For example, the effect of increasing external convective cooling is directly opposite in the two modes: in the temperature-controlled mode increasing cooling will increase power delivery and, therefore, lesion size, whereas during power-controlled ablation power delivery is constant and increased cooling will cause a slight decrease in lesion size, due to increased heat transfer away from the electrode tip and from the tissue [6,Go10,Go11,Go21–Go24]Go. {Delta}T is reflecting cooling and this opposite effect of cooling between the two modes is demonstrated by the negative correlation between {Delta}T and lesion size for temperature-controlled ablation and by the positive correlation between {Delta}T and lesion size for power-controlled ablation.

Since excessively high temperatures, above approximately 100 °C, of the myocardial tissue entail a risk of pop and crater formation[6,Go36]Go, pre-ablation measurement evaluating the factors (tissue contact and external cooling) determining the amount of energy entering the tissue and thereby tissue heating, can allow for adjustment of target temperature/power setting to reduce the risk of overheating. Our group has recently demonstrated this for left ventricular ablation in the beating heart in an in vivo study[37]Go. As shown in the present study, even during temperature-controlled ablation, where the electrode-tip temperature is constant, pop and crater formation occurred in some lesions. These applications had significantly lower {Delta}T and {Delta}IMP compared with the applications without popping, indicating high power delivery and increased risk of popping. Adjustment of either target temperature or power limit for these applications could probably have reduced this risk.

The present study demonstrates for the first time that the temperature rise induced by a pre-ablation low power test pulse ({Delta}T) and the impedance rise caused by the obtained electrode–tissue contact ({Delta}IMP) can give an indication of the cavitary cooling of the electrode tip and of the electrode–tissue contact in vitro. For irrigated tip ablation, where cavitary cooling has little or no impact on lesion size, since its effect is overcome by the high irrigation rate, and {Delta}T is not measurable, we demonstrated for the first time that {Delta}IMP can reflect electrode tip contact pressure and is positively correlated with the obtained lesion size. In recent in vitro studies[23,Go26]Go it has been shown that the temperature rise induced by a pre-ablation low power test pulse mainly reflects the external cooling of the electrode tip and to a smaller extent the electrode–tissue contact. The electrode–tissue contact can be evaluated more precisely by measuring the difference in impedance before and after tissue contact[38–Go41]Go, as illustrated in the present study.


    Clinical implications
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Clinical implications
 Study limitations
 Acknowledgements
 References
 
Assuming these findings can be applied to the human heart, they imply a possible means of controlling power delivery to the tissue during radiofrequency catheter ablation with consideration of each catheter position's unique electrode–tissue contact and external cooling. This should in theory allow for a more controllable power delivery into the tissue and hence a more controllable lesion size, thereby reducing the risk of tissue overheating, popping and crater formation.


    Study limitations
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Clinical implications
 Study limitations
 Acknowledgements
 References
 
This study was performed in non-perfused porcine myocardium in an in vitro model. The movements of the beating heart are likely to increase the variation of both pre-ablation parameters. Further, in the present model the tissue was suspended in saline, which has a lower impedance than blood, and the {Delta}IMP in the present study might therefore not be quite comparable with the clinical situation. But we have yet unpublished data concerning {Delta}IMP from an in vivo study of left ventricular ablation[37]Go, where a mean {Delta}IMP of 31.3 {Omega} was found, which is comparable with the high pressure situations in the present study. Another factor which might have impact is the perfusion of the cardiac tissue in vivo. The present findings should, therefore, be tested in an in vivo model.



View larger version (8K):
[in this window]
[in a new window]
 
Figure 3 For irrigated tip ablation all applications, independent of the electrode–tissue contact pressure and independent of external cooling, both of which, in a clinical setting, are unknown, the relations between pre-ablation {Delta}IMP and the lesion volume obtained are shown. For abbreviations see text.

 

    Acknowledgements
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Clinical implications
 Study limitations
 Acknowledgements
 References
 
The radiofrequency generators were kindly placed at our disposal by Medtronic CardioRhythm, Minneapolis, MN, USA and by Stockert GMBH, Freiburg, Germany.


    Footnotes
 
*The study has been financially supported by: The Danish Medical Research Council, Copenhagen, Denmark, The Danish Heart Foundation, Copenhagen, Denmark, Eva and Henry Frænkels Mindefond, Copenhagen, Denmark, and Grosserer L.F. Foght Foundation, Copenhagen, Denmark. Back


    References
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Clinical implications
 Study limitations
 Acknowledgements
 References
 
[1] Morady F, Harvey M, Kalbfleisch SJ, El-Atassi R, Calkins H, Langberg JJ. Radiofrequency catheter ablation of ventricular tachycardia in patients with coronary artery disease. Circulation 1993; 87: 363–372.[Abstract/Free Full Text]

[2] Klein LS, Shih HT, Hackett FK, Zipes DP, Miles WM. Radiofrequency catheter ablation of ventricular tachycardia in patients without structural heart disease. Circulation 1992; 85: 1666–1674.[Abstract/Free Full Text]

[3] Fontaine G. The ablative techniques from surgery to catheter ablation in the treatment of cardiac arrhythmias: a 20 year experience. Acta Cardiol 1995; 50: 467–468.[Web of Science][Medline]

[4] Choi YS, Sohn KS, Sohn DW, Oh BH, Lee MM, Park YB, et al. Temperature-guided radiofreqeuncy catheter ablation of slow pathway in atrioventricular nodal reentrant tachycardia. Am Heart J 1995; 129: 392–394.[CrossRef][Web of Science][Medline]

[5] Willems S, Chen X, Kottkamp H, Hindricks G, Haverkamp W, Rotman B, et al. Temperature-controlled radiofrequency catheter ablation of manifest accessory pathways. Eur Heart J 1996; 17: 445–452.[Abstract/Free Full Text]

[6] Eick O, Gerritse B, Schumacher B. Popping phenomena in temperature-controlled radiofrequency ablation: when and why do they occur? Pacing Clin Electrophysiol 2000; 23: 253–258.[CrossRef][Medline]

[7] Haines DE, Watson DD, Verow AF. Electrode radius predicts lesion radius during radiofrequency energy heating. Validation of a proposed thermodynamic model. Circ Res 1990; 67: 124–129.[Abstract/Free Full Text]

[8] Langberg JJ, Lee MA, Chin MC, Rosenqvist M. Radiofrequency catheter ablation: the effect of electrode size on lesion volume in vivo. Pacing Clin Electrophysiol 1990; 13: 1242–1248.[CrossRef][Medline]

[9] Langberg JJ, Gallagher M, Strickberger SA, Amirana O. Temperature-guided radiofrequency catheter ablation with very large distal electrodes. Circulation 1993; 88: 245–249.[Abstract/Free Full Text]

[10] Otomo K, Yamanashi WS, Tondo C, Antz M, Bussey J, Pitha JV, et al. Why a large tip electrode makes a deeper radiofrequency lesion: effects of increase in electrode cooling and electrode–tissue interface area. J Cardiovasc Electrophysiol 1998; 9: 47–54.[Web of Science][Medline]

[11] Petersen HH, Chen X, Pietersen A, Svendsen JH, Haunsø S. Lesion dimensions during temperature-controlled radiofrequency catheter ablation of left ventricular porcine myocardium. Impact of ablation site, electrode size and convective cooling. Circulation 1999; 99: 319–325.[Abstract/Free Full Text]

[12] Haines DE. Determinants of lesion size during radiofrequency catheter ablation: the role of electrode–tissue contact pressure and duration of energy delivery. J Cardiovasc Electrophysiol 1991; 2: 509–515.

[13] Skrumeda L and Mehra R. Comparison of standard and irrigated radiofrequency ablation in the canine ventricle. J Cardiovasc Electrophysiol 1998; 9: 1196–1205.[Web of Science][Medline]

[14] Weiss C, Antz M, Eick O, Eshagzaiy K, Meinertz T, Willems S. Radiofrequency catheter ablation using cooled electrode: impact of irrigation flow rate and catheter contact pressure on lesion dimensions. Pacing Clin Electrophysiol 2002; 25: 463–469.[CrossRef][Medline]

[15] Ruffy R, Imran MA, Santel DJ, Wharton JM. Radiofrequency delivery through a cooled catheter tip allows the creation of larger endomyocardial lesions in the ovine heart. J Cardiovasc Electrophysiol 1995; 6: 1089–1096.[Web of Science][Medline]

[16] Petersen HH, Chen X, Pietersen A, Svendsen JH, Haunsø S. Temperature-controlled irrigated tip radiofrequency catheter ablation: comparison of in vivo and in vitro lesion dimensions for standard catheter and irrigated tip catheter with minimal infusion rate. J Cardiovasc Electrophysiol 1998; 9: 409–414.[Web of Science][Medline]

[17] Haverkamp W, Hindricks G, Gülker H, Rissel U, Pfennings W, Borggrefe M, et al. Coagulation of ventricular myocardium using radiofrequency alternating current: bio-physical aspects and experimental findings. Pacing Clin Electrophysiol 1989; 12: 187–195.[CrossRef][Medline]

[18] Haines DE. The biophysics of radiofrequency catheter ablation in the heart: the importance of temperature monitoring. Pacing Clin Electrophysiol 1993; 16: 586–591.[CrossRef][Medline]

[19] Hindricks G, Haverkamp W, Gülker H, Rissel U, Budde T, Richter KD, et al. Radiofrequency coagulation of ventricular myocardium: improved prediction of lesion size by monitoring catheter tip temperature. Eur Heart J 1989; 10: 972–984.[Abstract/Free Full Text]

[20] Petersen HH, Roman-Gonzalez J, Johnson SB, Svendsen JH, Haunsø S, Packer DL. Mechanisms for enlarging lesion size during irrigated tip radiofrequency ablation: is there a virtual electrode effect? J Interv Cardiol 2004; 17: 171–177.[CrossRef][Medline]

[21] Kongsgaard E, Steen T, Jensen Ø., Aass H, Amlie JP. Temperature guided radiofrequency catheter ablation of myocardium: comparison of catheter tip and tissue temperatures in vitro. Pacing Clin Electrophysiol 1997; 20: 1252–1260.[CrossRef][Medline]

[22] Anfinsen OG, Kongsgaard E, Aass H, Amlie JP. Radiofrequency catheter ablation in vitro: the difference between tissue and catheter tip temperature depends on location of the temperature sensor. Eur J Cardiac Pacing Electrophysiol 1996; 6: 195–203.

[23] Petersen HH and Svendsen JH. Can lesion size during radiofrequency ablation be predicted by the temperature rise to a low power test pulse in vitro. Pacing Clin Electrophysiol 2003; 26: 1653–1659.[CrossRef][Medline]

[24] Petersen HH, Chen X, Pietersen A, Svendsen JH, Haunsø S. Temperature-controlled radiofrequency ablation of cardiac tissue: an in vitro study of the impact of electrode orientation, electrode tissue contact pressure and external convective cooling. J Interv Card Electrophysiol 2000; 3: 257–262.

[25] Avitall B, Mughal K, Hare J, Helms R, Krum D. The effects of electrode–tissue contact on radiofrequency lesion generation. Pacing Clin Electrophysiol 1997; 20: 2899–2910.[CrossRef][Medline]

[26] Eick O, Wittkampf FH, Bronneberg T, Schumacher B. The LETR-principle: a novel method to assess electrode–tissue contact in radiofrequency ablation. J Cardiovasc Electrophysiol 1998; 9: 1180–1185.[Web of Science][Medline]

[27] Hoyt RH, Huang SKS., Marcus FI, Odell RC. Factors influencing trans-catheter radiofrequency ablation of the myocardium. J Appl Cardiol 1986; 1: 469–486.

[28] Vanderbrink B, Gu Z, Rodriguez V, Link M, Homoud M, Estes Md, et al. Microwave ablation using a spiral antenna in a porcine thigh muscle preparation: in vivo assessment of temperature profile and lesion geometry. J Cardiovasc Electrophysiol 2000; 11: 193–198.[Web of Science][Medline]

[29] Chugh S, Chan RC, Johnson SB, Packer DL. Catheter tip orientation affects radiofrequency ablation lesion size in the canine left ventricle. Pacing Clin Electrophysiol 1999; 22: 413–420.[CrossRef][Medline]

[30] Chan RC, Johnson SB, Seward JB, Packer DL. The effect of ablation electrode length and catheter tip to endocardial orientation on radiofrequency lesion size in the canine right atrium. Pacing Clin Electrophysiol 2002; 25: 4–13.[CrossRef][Medline]

[31] Schumacher B, Eick O, Wittkampf FH, Pezold C, Tebbenjohanns J, Jung W, et al. Temperature response following nontraumatic low power radiofrequency application. Pacing Clin Electrophysiol 1999; 22: 339–343.[CrossRef][Medline]

[32] Petersen HH, Chen X, Pietersen A, Svendsen JH, Haunsø S. Tissue temperatures and lesion size during irrigated tip catheter radiofrequency ablation: an in vitro comparison of temperature-controlled irrigated tip ablation, power-controlled irrigated tip ablation and standard temperature-controlled ablation. Pacing Clin Electrophysiol 2000; 23: 8–17.[CrossRef][Medline]

[33] Eick O and Bierbaum D. Tissue temperature-controlled radiofrequency ablation. Pacing Clin Electrophysiol 2003; 26: 725–730.[CrossRef][Medline]

[34] Nakagawa H, Yamanashi WS, Pitha JV, Arruda M, Wang X, Ohtomo K, et al. Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation 1995; 91: 2264–2273.[Abstract/Free Full Text]

[35] Incropera FP and DeWittt DP. In Incropera FP and DeWitt DP (Eds.). Introduction to heat transfer 1991; New York John Wiley and Sons.

[36] Petersen HH, Chen X, Pietersen A, Svendsen JH, Haunsø S. Tissue temperatures and lesion size during irrigated tip ablation in the temperature controlled and the power controlled mode. Arch Mal Coeur 1998; 91: 99 [Abstract].

[37] Stagegaard N, Petersen HH, Ramlose L, Chen X, Svendsen JH. Control of lesion size by setting a target temperature determined from a pre-ablation low power test pulse during in vivo radiofrequency ablation. Eur Heart J 2003; 24: 592 [Abstract].

[38] Zheng X, Walcott G, Hall J, Rollins D, Smith W, Kay GN, et al. Electrode impedance: an indicator of electrode–tissue contact and lesion dimensions during linear ablation. J Interv Card Electrophysiol 2000; 4: 645–654.[CrossRef][Web of Science][Medline]

[39] Cao H, Tungjitkusolmun S, Choy Y, Tsai J, Webster J. Using electrical impedance to predict catheter–endocardial contact during RF cardiac ablation. IEEE Trans Biomed Eng 2002; 49: 247–253.[CrossRef][Web of Science][Medline]

[40] Wang D, Hulse JE, Walsh EP, Saul JP. Factors influencing impedance during radiofrequency ablation in humans. Chin Med J 1995; 108: 450–455.[Medline]

[41] Strickberger SA, Weiss R, Knight BP, Bahu M, Bogun F, Brinkman K, et al. Randomized comparison of two techniques for titrating power during radiofrequency ablation of accessory pathways. J Cardiovasc Electrophysiol 1996; 7: 795–801.[Web of Science][Medline]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?



This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Stagegaard, N.
Right arrow Articles by Svendsen, J. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Stagegaard, N.
Right arrow Articles by Svendsen, J. H.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?