Brady-tachy syndrome: Rapid atrial pacing efficacy in preventing atrial fibrillation recurrence assessed by reliable electrograms: the prefib pilot study
aDepartment of Cardiology, Angoulême Hospital 16470 Angoulême, France; bDepartment of Clinical Research Guidant France Rueil-Malmaison, France
Manuscript submitted 10 January 2005. Revision received 25 July 2005. Accepted after revision 11 May 2005.
*Corresponding author. Fax: +33 545 244 105. E-mail address: daniel.flammang{at}ch-angouleme.rss.fr (D. Flammang).
| Abstract |
|---|
|
|
|---|
AIMS: Recent studies have tested different atrial pacing rates, modes, and sites for preventing atrial fibrillation (AF) recurrence. Present generation pacemakers offer reliable electrograms (EGMs) storage for optimizing the arrhythmia diagnosis. Based on these EGMs, the study objective was to assess the rate of AF recurrence at two different pacing rates.
METHODS: Thirty patients suffering exclusively from symptomatic brady-tachy syndrome (BTS) resistant to
2 drugs, were implanted with a DDDR pacemaker. After a 5-days observation period, the DDD pacing rate was randomly programmed at 60 bpm (15 bpm hysteresis) or at 80 bpm for 12 weeks. The two sequences were crossed over at the end of this fixed period or when earlier symptomatic AF recurred. Antiarrhythmics were maintained. Stored EGMs of
4 s duration identified all AF recurrence.
RESULTS: Thirty patients (17 males, 77.2±8.1 years old) were included. One patient withdrew prematurely for severe heart failure associated with AF recurrence and rapid ventricular response. For the remainder of the 29 patients, fast atrial pacing neither provoked symptoms nor haemodynamic change. AF recurred in 16 patients paced at 60 (15) bpm (mean time: 29 days; range 161) and in 9 patients paced at 80 bpm (mean time: 55 days; range 583) (P<0.05). AF recurrence was asymptomatic in 50% of patients.
CONCLUSIONS: These results confirm that rapid atrial pacing is 1) significantly effective for preventing AF recurrence in symptomatic BTS patients, and 2) haemodynamically well tolerated.
Key Words: brady-tachy syndrome, atrial pacing, AF prevention
| Introduction |
|---|
|
|
|---|
The incidence of atrial fibrillation, its haemodynamic and vascular complications associated with a decrease in the quality of life have shaped our therapeutic approach [1]
The objective of this study was to re-assess the benefit of rapid atrial pacing on atrial fibrillation recurrence validated by reliable stored EGMs, in patients implanted exclusively for brady-tachy syndrome.
| Patients and methods |
|---|
|
|
|---|
The study included 30 patients, 17 males, with a mean age of 77.2±8.1 years from December 1999 to December 2001. The inclusion criterion was symptomatic brady-tachy syndrome resistant to at least two antiarrhythmics (class 1 indication for permanent pacing [15]
Low impedance bipolar atrial leads with an interelectrode distance of 11 mm (Vitatron, Crystalin®, Dieren, NL) were positioned in the right atrial appendage using passive fixation. All ventricular leads were low impedance, bipolar and steroid eluting leads (Medtronic 5092®, Minneapolis, MN, USA), and were positioned at the right ventricular apex. There were no lead dislodgements during follow-up.
Standard measures of sensing, pacing threshold, and lead impedance were taken in both chambers, in bipolar and unipolar settings at implant, discharge, and at the end of both study periods. No abnormal values were observed.
The DDDR pacemakers were the Pulsar Max 972 in 21 patients and the Pulsar Max II 1280 in 9 patients (Guidant Inc., St Paul, MN, USA), all implanted in the pre-pectoral region. These pacemakers store 40 s of high quality EGMs from both chambers simultaneously; this limited memory time may be organized in various ways with a maximum of 10 episodes of 4 s each. The difference between the two pacemakers is that, in the 1280 model, every recording includes the triggering arrhythmia EGMs with onset and simultaneous channel markers (Fig. 1). For long-term follow-up, in order to acquire the maximum information from the 40 s memory capability, a storage of 8 records of 5 s each was selected for the Pulsar 972; a storage of 5 records of 8 s, divided into two periods of 4 s, one before and one after the trigger onset was chosen for the Pulsar 1280.
|
The primary endpoint of this single centre, single blind, and randomized cross-over trial was to compare the time to AF recurrence at two pacing rates during two treatment periods of three months. AF recurrence was defined as any episode of at least 4 s stored in the pacemaker memory. The second endpoint was to estimate the AF burden in both periods, evaluated by the contents of the pacemaker memory. The AF burden is defined by the amount of time spent in arrhythmia and by the number of mode switching episodes, estimated by the number of atrial tachycardia responses (ATR) and the total time spent in ATR. Considering the pilot aspect of this trial and the singular inclusion criterion, a limited cohort of 30 patients was estimated to be sufficient to draw valuable conclusions.
After the implantation, the various algorithms offered by the Pulsar pacemaker were monitored during the next 5-days. This observation period was used for sensing and pacing thresholds adjustments. Data collected during this tuning period were erased at day-5. Before patient discharge, only the ATR and the pacemaker mediated tachycardia (PMT) algorithms were turned on and all counters were reset. Then, the patient was discharged on one of the two randomized pacing rates for a 3 month period: either DDD 60 bpm and rate hysteresis of 15 bpm in order to favour continuously spontaneous sinus rhythm, or DDD 80 bpm without rhythm hysteresis. The AV delay was programmed longer than the spontaneous P-R interval (+40 ms) up to 260 ms, to favour spontaneous ventricular conduction. The observation periods were crossed over, either at the end of 3 months if no recurrence occurred, or as soon as symptomatic AF recurred (Fig. 2). To shorten the delay between the recurrence and the cross-over reprogramming, the patient agreed to wear an external loop recorder (Cardiomemo, Instrumedix®, Paris, France), all the time, for recording and sending ECG samples to a remote Holter centre at least weekly and upon any symptomatic recurrence. The loop recorded ECG was faxed to the investigator immediately after the patient's call. At the end of the observation periods, the content of the pacemaker memory was printed out and device function was checked. Every stored event was analyzed by the physician and by an expert technician, checking the EGMs quality and accuracy, channel markers, muscle noise, over- and undersensing, mode switching counters and the documents forwarded by the remote Holter centre.
|
In case of recurrent AF of longer than two days duration (symptomatic or not), the patient was recommended for cardioversion, either by drugs or by external or internal DC shock. Sinus rhythm had to be restored before starting the second period of the trial. Antiarrhythmics and anticoagulants were continued as before the AF recurrence.
| Statistics |
|---|
|
|
|---|
All data are reported as mean±standard deviation. The Student's t-test was used to compare mean atrial pacing percentage, mean number of ATR episodes, mean total time in AF, mean maximum time in AF, and mean minimum time in AF between the 60 bpm and the 80 bpm periods. Kaplan-Meier curves were drawn for the two comparative periods; logrank test was used to compare these two curves; statistical significance was P<0.05.
| Results |
|---|
|
|
|---|
Main objectives
First endpoint
Rapid atrial pacing was shown to prevent AF recurrence longer than 4 s better than slow atrial pacing (Table 1, Fig. 3).
|
|
One pt withdrew prematurely because of severe heart failure precipitated by AF recurrence with rapid ventricular response (ischaemic heart disease; EF 35%) during her first observation period. In the remaining 29 patients, AF recurred in a total of 18 (62%), 16 paced at 60 (15) bpm, after a mean time of 29 days (range 161), and in 9 paced at 80 bpm, after a mean time of 55 days (range 583) (P<0.05) (Fig. 3). AF recurred in both periods in 7 patients (24%).
The first episode of AF recurrence was diagnosed by the stored EGMs during regular follow-up in all 18 patients (100%). Besides the EGM storage, AF recurrence was concomitantly diagnosed by three patients themselves who called the physician immediately at the time of symptoms (17%), and by the remote Holter centre during the weekly systematic checks in five patients (28%) underlining the limited yield of this technique. AF recurrence was asymptomatic in 9 of the 18 patients (50%). Asymptomatic episodes were of variable duration but without fast ventricular rates, probably explaining the lack of symptoms. AF recurrence was converted to sinus rhythm by high dosage of oral amiodarone (30 mg/Kg) in three patients before the cross-over; direct current (DC) shock was not used.
Second endpoint
AF burden estimate. The mean percentage of atrial pacing time was significantly higher in the fast pacing group (79.7 vs 27.7%, P<0.001) (Table 2a). These results remain statistically different (P=0.003) if only the seven patients who experienced AF recurrence in both pacing rates were considered (Table 2b). The parameters directly reflecting the amount of time spent in AF (ATR parameters) were also remarkably better in the fast pacing period and are directly correlated with the mean atrial pacing time.
|
EGM reliability
The main endpoint of that study was achieved by the quality of stored EGMs, by means of the pre-trigger onset recordings and the channel markers capabilities. All stored EGMs have been analyzed separately by two independent experts.
Atrial fibrillation and tachycardia episodes were correctly diagnosed in all 133 (100%) EGMs stored in the 1280 pacemaker model which provides the onset of the triggered arrhythmia. The other 38 episodes collected with the 972 pacemaker model were not appropriately classified, corresponding with false positive EGMs (crosstalk phenomena and myopotentials: 2 patients) and to non-diagnostic EGMs (sinus rhythm recovery: 6 patients) related to the absence of the arrhythmia onset capabilities (Table 1).
Other results
In the Pulsar pacemaker models, simultaneous programming of the ATR and the atrial flutter algorithms leads to a potentially wrong classification of the diagnostic triggers.
Due to the principal objective of the study and the limited size of the memory, we did not turn on the ventricular tachycardia and supraventricular tachycardia algorithms in the observation periods. Consequently, retrograde conduction and the incidence of other tachycardias were not assessed.
| Discussion |
|---|
|
|
|---|
Main findings
The main findings from this study are that rapid atrial pacing (80 bpm), delivered at the right atrial appendage, appears to be more advantageous than slow pacing from the same site in preventing recurrence of atrial fibrillation and in reducing AF burden. These findings are validated by accurate intracardiac EGMs, allowing discrimination of true atrial EGMs from myopotentials and ventricular far-field signals.
Previous studies
EGMs storage and reliability
Up to the late 1990's, clinical daily practice showed the limited diagnostic appropriateness of any standard mode switching system when not controlled by EGM recordings. Recent studies have shown the usefulness of reliable EGM information especially when event onset and marker annotations are available to improve diagnosis of arrhythmias [17]
. Endocardial signal filtering represents a critical point of this system where myopotentials may be superimposed on sinus rhythm but are distinguishable from atrial fibrillation.
In the present study, 100% of the signals recorded by the Pulsar 1280 model were appropriately classified. Two patients with the Pulsar 972 model provided 38 endocardial samples (representing 22% of all recorded EGMs) incorrectly classified. So, adding the trigger onset and the channel marker functions seems to be essential for improving the diagnostic capabilities of the pacemaker. The lack of EGMs allowing verification of mode switching counters weakens the diagnostic value of previously reported results.
Atrial overdrive
Pacing modes and atrial pacing rates have been investigated over a long period [17]
. Different overdrive techniques have been proposed: the atria may be paced fast, either at a predefined rate or at a calculated fixed rate above the average sinus rate [6,
7,
12,
18]
. In 1998, the AIDA report suggested that overdrive pacing may be antiarrhythmic [19]
. All available studies showed that, up to 80 bpm, fast DDD pacing is clinically well tolerated and provides a real decrease in AF recurrence, in their total and maximum duration [6]
, but similar results were not found with DDD-R pacing [18]
. Again, all these studies based their approach on events counters, and not on EGM recordings; moreover, patients with possible myopotential oversensing, ventricular cross-talk, atrial undersensing, and endless loop tachycardia possibly interfering with data collection were excluded [6]
.
By using EGMs analysis, this limited pilot study only including brady-tachy syndrome patients in an elderly population fast atrial pacing has been shown to decrease significantly true AF recurrence. Also, we found that myopotentials and crosstalk episodes were equivalent in both pacing periods. Finally, the AV delay had been systematically prolonged by 40 msec over the spontaneous P-R interval in order, as far as possible, to prevent potential left ventricular dysfunction.
Symptoms related to fast atrial pacing
From EGMs or simple markers, more than 50% of AF appears to be asymptomatic [6,
7,
18]
, moreover, patients complain of rate irregularity or palpitations which are not confirmed by intra-atrial recordings. More surprisingly, it has also been reported that AF was confirmed by concomitant Holter recording in only 8% of patients convinced that they were suffering from genuine AF recurrence [22]
. Our study showed an absence of symptoms in 50% of AF recurrence: reliable EGMs might decrease the number of false negative (asymptomatic) patients in future studies.
Limitations
Atrial pacing site
As there is no definitive consensus about the optimal pacing site(s) in the right atrium, we did not try to test different pacing sites either solely in the right atrium or by combining right and left atrial sites using the coronary sinus [23]
. If the low interatrial septal and the Bachmann bundle area seem to be more effective than the right atrial appendage simultaneously to depolarize both atria, validation and routine application of these techniques remains under evaluation [20,
21]
.
Study follow-up
The 6 month observation period may appear to be too short even in such a homogeneous cohort of stable brady-tachy syndrome patients with a high AF burden. A larger study designed with longer comparative periods and quality of life (QoL) assessments would probably provide more details about the circumstances of AF recurrence, the protective effect of pacemaker implantation, the association with other arrhythmias, the impact of different antiarrhythmics and their relationship with the underlying cardiac disease.
The design of this preliminary study did not cover the evaluation of specific AF prevention algorithms. These algorithms could be tested in a larger prospective multicentre study.
With the exception of one patient in New York Heart Association class 3 who developed heart failure during an AF recurrence with fast ventricular response when programmed on DDD 80 bpm, none of the others complained of side effects due to 80 bpm atrial pacing.
The external loop recorder was underused in this study but its limited diagnostic yield would not suggest systematic use in larger studies.
| Conclusions |
|---|
|
|
|---|
The results of this study concerning patients suffering from brady-tachy syndrome show that atrial pacing at 80 bpm significantly decreased the rate of AF recurrence and is haemodynamically well tolerated, over a 3-month period. Larger studies in this patient population should assess QoL associated with fast atrial pacing and the use of AF prevention algorithms.
| References |
|---|
|
|
|---|
[1] Wyse DG, Waldo AL, DiMarco JP, Domanski MJ, Rosenberg Y, Schron EB, et al. Atrial fibrillation follow-up investigation of rhythm management (AFFIRM) investigators. N Engl J Med 2002; 347: 18251833.
[2] Taylor FC, Cohen H, Ebrahim S. Systematic review of long term anticoagulation or antiplatelet treatment in patients with non-rheumatic atrial fibrillation. Br Med J 2001; 322: 321326.
[3] Wellens HJ, Lau CP, Lüderitz B, Akhtar M, Waldo AL, Camm AJ, et al. Atrioverter: an implantable device for the treatment of atrial fibrillation. Circulation 1998; 98: 16511656.
[4] Daubert JC, Mabo P, Berder V, Gras D, Leclercq C. Atrial tachyarrhythmia associated with high degree interatrial conduction block: prevention by permanent atrial resynchronisation. Eur J Card Pacing Electrophysiol 1994; 4:Suppl. 1 3544.
[5] Savelieva I and Camm AJ. The results of pacing trials for the prevention and termination of atrial tachyarrhythmias: is there any evidence of therapeutic breakthrough? J Intervent Card Electrophysiol 2003; 8: 103115.[CrossRef][ISI][Medline]
[6] Garrigue S, Barold S, Cazeau S, Gencel L, Jaïs P, Haïssaguerre M, et al. Prevention of atrial arrhythmias during DDD pacing by atrial overdrive. Pacing Clin Electrophysiol 1998; 21: 17511759.[CrossRef][Medline]
[7] Funck RC, Adamec R, Lurje L, Capucci A, Ritter P, Skehan D, et al. Atrial overdriving is beneficial in patients with atrial arrhythmias: first results of the PROVE study. Pacing Clin Electrophysiol 2000; 23: 18911893.[Medline]
[8] Ogawa M, Suyama K, Kurita T, Shimuzu W, Matsuo K, Taguchi A, et al. Acute effects of different atrial pacing sites in patients with atrial fibrillation: comparison of single site and biatrial pacing. Pacing Clin Electrophysiol 2001; 24: 14701478.[Medline]
[9] Mirza I, James S, Holt P. Biatrial pacing for paroxysmal atrial fibrillation. A randomized prospective study into the suppression of paroxysmal atrial fibrillation using biatrial pacing. J Am Coll Cardiol 2002; 40: 457463.
[10] Mabo P, Funck R, De Roy L, Poezevara Y, Anselme F, Di Donna P. Impact of ventricular pacing on AF prevention. Pacing Clin Electrophysiol 2002; 25: 621.
[11] Carlson MD, Ip J, Messenger J, Beau S, Kalbfleisch S, Gervais P, et al. A new pacemaker algorithm for the treatment of atrial fibrillation. Results of the Atrial Dynamic Overdrive Pacing Trial (ADOPT). J Am Coll Cardiol 2003; 42: 627633.
[12] Camm AJ. Results of AF Therapy. The hotline sessions of the 23rd European Congress of Cardiology. Eur Heart J 2001; 22: 20332037.
[13] Lee M, Weachter R, Pollack S, Kremers M, Naik AM, Silverman R, et al. The effect of atrial pacing therapies on atrial tachyarrhythmia burden and frequency. Results of a randomized trial in patients with bradycardia and atrial tachyarrhythmias. J Am Coll Cardiol 2003; 41: 19261932.
[14] Israel CW, Hugl B, Unterberg C, Lawo T, Kennis I, Hettrick D, et al. Pace-terminating and pacing for prevention of atrial tachyarrhythmias: results from a multicenter study with an implantable device for atrial therapy. J Cardiovasc Electrophysiol 2001; 12: 11211128.[CrossRef][ISI][Medline]
[15] Gregoratos G, Abrams J, Epstein AE, Freedman RA, Hayes DL, Hlatky MA, et al. American College of Cardiology/American Heart Association Task Force on Practice Guidelines/North American Society for Pacing and Electrophysiology Committee to Update the 1998 Pacemaker Guidelines. ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). Circulation 2002; 106: 21452161.
[16] Israel CW and Böckenförde JB. Pacemaker event counters: possible sources of error in calculation of AV synchrony in VDD single lead systems as an example for present limitations. Pacing Clin Electrophysiol 1998; 21: 489493.[Medline]
[17] Nowak B, Sperzel J, Rauscha F, Mont L, Alzueta J, Senden PJ, et al. Diagnostic value of onset-recordings and marker annotations in dual chamber pacemaker stored electrograms. Europace 2003; 5: 103109.
[18] Levy T, Walker S, Rex S, Paul V. Does atrial overdrive pacing prevent paroxysmal atrial fibrillation in paced patients? Int J Cardiol 2000; 75: 9197.[CrossRef][ISI][Medline]
[19] Defaye P, Dournaux F, Mouton E. Prevalence of supraventricular arrhythmias from the automated analysis of data stored in the DDD pacemakers of 617 patients: the AIDA study. Pacing Clin Electrophysiol 1998; 21: 250255.[CrossRef][Medline]
[20] Padeletti L, Botto G, Spampinato A. Prevention of paroxysmal atrial fibrillation in patients with sinus bradycardia: role of right atrial linear ablation and pacing site. J Cardiovasc Electrophysiology 2003; 14: 733738.
[21] de Voogt W, Van den Bos A, De Vusser P, Lau CP, Koistinen Y, Mairesse G, et al. OASES trial: overdrive atrial septum stimulation in patients with paroxysmal atrial fibrillation and class I and II pacemaker indication. Europace 2003; 4:Suppl. A A12.
[22] Saksena S and Madan N. Hybrid therapy of atrial fibrillation algorithms and outcomes. J Interventional Cardiac Electrophysiology 2003; 9: 235247.[CrossRef]
[23] Saksena S, Prakash A, Ziegler P, Hummel JD, Friedman P, Plumb VJ, et al. Improved suppression of atrial fibrillation with dual-site atrial pacing and antiarrhythmic drug therapy. J Am Coll Cardiol 2002; 40: 11401150.
![]()
CiteULike
Connotea
Del.icio.us What's this?
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||



2 = 4248, P<0.05.