Europace 2005 7(2):175-180; doi:10.1016/j.eupc.2004.12.010
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (5)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Bartsch, C.
Right arrow Articles by Weiler, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bartsch, C.
Right arrow Articles by Weiler, G.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© 2005 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.

Post-mortem evaluation of 415 pacemakers: in situ measurements and bench tests*

C. Bartscha,*, W. Irnicha, M. Jungeb, W.A. Stertmannc, M. Rissea and G. Weilera

aDepartment of Legal Medicine, University Hospital Frankfurter Strasse 58, 35392 Giessen, Germany; bDepartment of Legal Medicine, University Hospital Hamburg, Germany; cClinic of Cardiac and Vascular Surgery, University Hospital Giessen, Germany

Manuscript submitted 3 March 2004. Accepted after revision 20 December 2004.

*Corresponding author. Tel.: +49 641 9941421; fax: +49 641 9941419. E-mail address: christine.bartsch{at}forens.med.uni-giessen.de (C. Bartsch).


    Abstract
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusions
 References
 
AIM: The hypothesis was that there is more undetected dysfunction of implanted pacemaker systems than that detected and corrected. This prompted a research project (sponsored by the German Research Foundation) to detect pacemaker abnormalities and evaluate their complications for patients, thus, proving or disproving the hypothesis.

METHODS AND RESULTS: Four hundred and fifteen pacemakers of deceased patients were analyzed assessing their functionality by in situ measurements and bench tests including five measurements and one telemetric interrogation. Results were divided into four categories and statistically evaluated. Life-threatening abnormalities were found in 3.8%, potentially life-threatening in 3.7%, probably symptomatic, divided into atrial and ventricular problems, 13.3% and 2.8%, respectively, and premature exhaustion in 1.2%. Three of 179 bipolar ventricular leads and 2 of 131 bipolar atrial leads had insulation defects corresponding to 1.7% and 1.5%, respectively. The bipolar complication rate was 2.8 times higher than unipolar.

CONCLUSION: The pacemaker patients investigated, living 4 years with their pacemaker on average, had a post-mortem evaluated complication rate of the category "life-threatening" of 3.8%. This result corresponds to an annual complication rate of 0.94% compared with a rate of only 0.39% in an earlier investigation.

Key Words: implanted pacemaker, post-mortem measurement, in situ measurement, dysfunction, complication rate


    Introduction
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusions
 References
 
Due to implantable pacemaker systems, currently approximately 2500–3000 patients per million in the United States and Europe live a normal life. As a result of the growing number of older people in the population, particularly those aged over 75 years, an increasing number of pacemakers will be needed worldwide [1]Go. Possible pacemaker dysfunction in daily clinical practice has already been described. Lead malfunction, for example, is still the predominant cause of system dysfunction [2]Go. Surgery related complications during and after implantation are also well known [3]Go, but little is known about technical defects and dysfunction of generators and electrodes during system life [4]Go.

It was expected that there is more undetected dysfunction of implanted pacemaker systems than that detected and corrected [4]Go. This hypothesis challenged us to initiate a research project, sponsored by the German Research Foundation, to detect pacemaker abnormalities and evaluate their complication rate for patients, thus, proving or disproving the hypothesis.


    Materials and methods
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusions
 References
 
The population investigated consisted of pacemakers from deceased patients from different regions in Germany (Giessen, Friedberg, Hamburg). Between August 2001 and November 2002, 415 pacemakers together with 556 leads were removed and analyzed in bench tests. Four hundred and three of these units were also investigated in situ in the cadaver before cremation. Table 1 summarizes all populations in more detail. A measurement method was applied that is suitable to judge the functionality of a pacemaker system and guarantees clear results promptly after the patients' death leaving little room for discussion as to the cause. The functionality of the pacemaker was assessed by in situ measurements and bench tests including five measurements and one telemetric interrogation, carried out at room temperature, in the following sequence [5]Go:


View this table:
[in this window]
[in a new window]
 
Table 1 Investigated populations

 
    In situ measurements
  1. Signals proportional to pacing pulses were derived with a differential amplifier connected to a digital storage oscilloscope read offline by a personal computer (Fig. 1). The amplifier connected to three cannula electrodes positioned linearly between apex and right clavicle to reduce noise (Fig. 2).
  2. With a test pacemaker (5 V, 1 ms pulse with a rate of 85 min–1) and two additional cannula electrodes, synchronization of the implanted pacemaker could be tested.
  3. The lead impedance was measured with another test pacemaker that could be interrogated by a hand-held programmer.



View larger version (15K):
[in this window]
[in a new window]
 
Figure 1 Block diagram of in situ measurement in pacemaker patients.
 


View larger version (106K):
[in this window]
[in a new window]
 
Figure 2 Arrangement of cannula electrodes.
 
Bench tests
4. The parameters: pulse period, pulse amplitude, pulse duration, sensing threshold, and magnet influence were measured in the laboratory. The connector pins were checked for notches, which are a prerequisite for good contact between pin and screw(s). The bipolar lead insulation was checked with an ohm-meter.
5. Programmed pacemaker parameters and historical data were interrogated by telemetry (implantation date, date of the last follow-up, date of attainment of the electric replacement indicator (ERI) state).
6. If no pacing pulses were delivered by the device, the can was opened. The battery voltage was an indication of whether exhaustion and/or defect was present. In cases of battery exhaustion, the circuitry was connected to a 2.8 V power source to assess its function. Without a pacing pulse, the pacemaker was judged to be defective.

To evaluate the measurements of all 415 pacemakers the results were divided into different subgroups with respect to the severity of their impact (Tables 18).


View this table:
[in this window]
[in a new window]
 
Table 2 Age distribution

 


View this table:
[in this window]
[in a new window]
 
Table 3 Conspicuous aspects of evaluation

 


View this table:
[in this window]
[in a new window]
 
Table 4 Category "life-threatening"

 


View this table:
[in this window]
[in a new window]
 
Table 5 Category "potentially life-threatening"

 


View this table:
[in this window]
[in a new window]
 
Table 6 Category "probably symptomatic" with respect to atrial and ventricular problems

 


View this table:
[in this window]
[in a new window]
 
Table 7 Bipolar insulation defects

 


View this table:
[in this window]
[in a new window]
 
Table 8 Lead complication statistics

 

    Results
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusions
 References
 
Three hundred and seventy six cases had dates of birth and death available allowing mean age at death to be calculated; 84.7 years (Table 2). In 117 deceased pacemaker patients the date of implantation could be interrogated. Mean age at implantation of the last pacemaker was 80.7 years, in the other 298 cases the implantation age was not available. Of the 411 ventricular leads (as four generators were received by us without leads), 387 leads were checked for notches while 24 pacemakers were from Intermedics, which do not possess screws (Table 1Table 5). Fig. 4–Fig. 6 demonstrate that abnormal pulses usually show characteristic deviations from normal (Fig. 3), which have been described earlier [5–Go7]Go. A shunting resistor within the pacemaker circuitry parallel to the electrode decreases the output rapidly and below zero (Fig. 4). If the insulation of a bipolar lead degrades with fluid between the conductors, a shunting capacitance is created that rounds leading and trailing edges (Fig. 5). Breakage or bad contact produces sharp spikes followed by a low and flattened curve (Fig. 6).



View larger version (15K):
[in this window]
[in a new window]
 
Figure 3 Normal ventricular, unipolar pulse derived from the thorax surface.

 



View larger version (16K):
[in this window]
[in a new window]
 
Figure 4 Unipolar pulse of a defective pacemaker with 131 {Omega} shunt in parallel to output.

 



View larger version (20K):
[in this window]
[in a new window]
 
Figure 5 Pulse of a bipolar lead with insulation defect.

 



View larger version (20K):
[in this window]
[in a new window]
 
Figure 6 Bipolar pulse of a broken lead with capacitive coupling.

 
The discovered abnormalities were divided into four categories: "life-threatening", "potentially life-threatening", "probably symptomatic" and "premature exhaustion" (Table 3). If the elective replacement indicator (ERI) was reached within warranty period (48 months), exhaustion was termed "premature" which was found in 1.2%.

Table 4 shows the results for the category "life-threatening", from which a complication rate of 3.8% can be derived. The leading problem was battery exhaustion (1.7%), followed by failures of ventricular leads (1.0%). Pacemaker defects were about 0.5% and lastly infection, 0.2%. Infection as cause of death was confirmed by autopsy.

The category "potentially life-threatening" included 3.7% (Table 5) which is restricted only to ventricular problems, consisted of complications "infection" (diagnosis without autopsy), "missing notch" and "indifferent screw not tightened", in which missing notches on ventricular lead pins were most frequent (2.8%). A notch is a prerequisite for a good contact. Missing notches are potentially dangerous because of corrosion between different metals, an effect which was confirmed in some cases by high output voltages (5 V or higher) and long duration (1 ms).

In the category "probably symptomatic" (Table 6) the complication rate for atrial problems was 13.3% with deactivated atrial channels being the leading problem. Ventricular problems were found in 2.8% with insulation defects being most prominent (0.9%). There was a discrepancy in one case between the magnet test resulting in "ERI" while the programmer interrogation claimed battery to be "good". This uncertainty is surely of clinical relevance. "Magnet function unreliable" meant that several trials were necessary to open the reed switch. A lead that could be withdrawn by low extraction force was "not tightened".

Three of 179 bipolar ventricular leads and two of 131 bipolar atrial leads had an insulation defect corresponding to 1.7% and 1.5%, respectively (Table 7). Table 8 shows that the failure rate of ventricular bipolar leads is 8.9% versus 2.6% for unipolar yielding a defect ratio bipolar versus unipolar of 3.5%. If the defect ratio of all bipolar leads (6.8%) versus unipolar (2.4%) is calculated, the defect ratio drops to 2.8% (Table 8).


    Discussion
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusions
 References
 
We analyzed and evaluated the functionality of 415 implanted pacemaker systems post mortem to find out possible complications and to derive complication rates. The applied post mortem in situ measurement method seems to be unique worldwide. To our knowledge, no data comparable to our results are available. In the category "life-threatening" we found a complication rate of 3.8% of patients that lived on average 4 years with their pacemaker which means that those patients died with or because of a dysfunction of their pacemaker. Thus, 3.8% divided by 4 years yields an annual complication rate of 0.95% which can be compared with the 0.39% of an earlier investigation [4]Go. Thus, life-threatening dysfunction of implanted pacemaker systems are 2.4 times higher than anticipated. Moreover, the category "probably symptomatic" including atrial and ventricular problems reveals a surprisingly high percentage of 13% respectively, 3% of pacemakers which do not work correctly or optimally with atrial problems being predominant. Lead defects are more frequent (ventricular 1%, atrial 3.4%) than generator defects (0.5%) (Table 4 and Table 6) which support the statement that lead malfunction is still the predominant cause of system dysfunction [2]Go. The defect ratio of ventricular bipolar leads is 3.5 times higher compared with unipolar leads. Indeed, bipolar leads are advantageous with respect to electromagnetic interference, but this is counterbalanced by a higher defect rate.

The discrepancy in one case between magnet test resulting in "ERI" while the programmer interrogation claimed battery to be "good" raises concern because such an uncertainty provokes explantation which is surely of clinical relevance.

The age at last implant, which was 80.7 years in our study, is not representative of the German pacemaker population which is characterized by mean age at initial implantation for men of 73.3 years and for women 77.1 years (mean value 74.9 years) [8]Go. Thus, our population of deceased patients undergoing cremation is approximately 6 years older than the general population. It has been shown [9,Go10]Go that pacemaker treatment is less generous with respect to mode selection with increasing age. Whether this can also be extrapolated to post implant care of older patients cannot be derived from our data.


    Conclusions
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusions
 References
 
Our data reveal that there is much undetected pacemaker dysfunction and possibly more than that detected and corrected. A post mortem evaluated the complication rate in the category "life-threatening" of 3.8% of patients that lived on average 4 years with their pacemaker and the high rate of deviations from normal of 13% should be taken into account and lead to improvements. Perhaps follow-up intervals for pacemakers should be shortened. Incorrect handling during or after implantation produces an astonishingly high proportion of atrial (1.5%) and ventricular lead (3.1%) dysfunction that must be reduced. More data will reveal whether these results are also valid for other regions in Germany.


    Footnotes
 
*The paper was presented in part at the 24th Scientific Session of the North American Society of Pacing and Electrophysiology (NASPE), DC May 15, 2003. Back


    References
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 Conclusions
 References
 
[1] Irnich W. Pacemaker-related patient mortality. Pacing Clin Electrophysiol 1999; 22: 1279–1283.[Medline]

[2] Daley WR and Kaczmarek RG. The epidemiology of cardiac pacemakers in the older US population. J Am Geriatr Soc 1998; 46: 1016–1019.[Medline]

[3] Geppert A and Rauscha F. Pacemaker dysfunction in daily clinical practice. Wien Klin Wochenschr 2001; 113: 15–26.[Medline]

[4] Da Costa A, Kirkorian G, Isaaz K, Touboul P. Secondary infections after pacemaker implantation. Rev Med Intern 2000; 21: 256–265.

[5] Bartsch C, Irnich W, Risse M, Junge M, Weiler G. Postmortem in situ diagnosis of pacemakers and electrodes to detect dysfunction. Leg Med 2003; 5: S397–S400.[CrossRef]

[6] Irnich W, Stertmann WA, Müller R, Rempe M, Hehrlein FW. How to detect defective bipolar leads. In Vardas PE (Ed.). Proceedings EUROPACE '97, Athens (Greece), June 8–11 1997; Bologna Monduzzi Editore pp. 815–819.

[7] Irnich W, Batz L, Müller R, Stertmann WA. Physics of failing bipolar leads. In Vardas PE (Ed.). Proceedings EUROPACE '97, Athens (Greece), June 8–11 1997; Bologna Monduzzi Editore pp. 1107–1111.

[8] Markewitz A. Annual report 2000/2001 of the German central register of pacemakers. Herzschr Elektrophys 2002; 13: 242–257.[CrossRef]

[9] Batz L and Irnich W. Demographic aspects of pacemaker therapy. Herzschr Elektrophys 1995; 6: 4–10.

[10] Batz L and Irnich W. Are there specific gender differences in pacemaker selection? Biomed Tech 1997; 42:Suppl 77–78.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
EuropaceHome page
M. Novak, P. Dvorak, P. Kamaryt, B. Slana, and J. Lipoldova
Autopsy and clinical context in deceased patients with implanted pacemakers and defibrillators: intracardiac findings near their leads and electrodes
Europace, November 1, 2009; 11(11): 1510 - 1516.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (5)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Bartsch, C.
Right arrow Articles by Weiler, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bartsch, C.
Right arrow Articles by Weiler, G.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?