Europace 2005 7(2):175-180; doi:10.1016/j.eupc.2004.12.010
© 2005 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.
Post-mortem evaluation of 415 pacemakers: in situ measurements and bench tests*
C. Bartscha,*,
W. Irnicha,
M. Jungeb,
W.A. Stertmannc,
M. Rissea and
G. Weilera
aDepartment of Legal Medicine, University Hospital Frankfurter Strasse 58, 35392 Giessen, Germany;
bDepartment of Legal Medicine, University Hospital Hamburg, Germany;
cClinic of Cardiac and Vascular Surgery, University Hospital Giessen, Germany
Manuscript submitted 3 March 2004. Accepted after revision 20 December 2004.
*Corresponding author. Tel.: +49 641 9941421; fax: +49 641 9941419. E-mail address: christine.bartsch{at}forens.med.uni-giessen.de (C. Bartsch).
 |
Abstract
|
|---|
AIM: The hypothesis was that there is more undetected dysfunction
of implanted pacemaker systems than that detected and corrected.
This prompted a research project (sponsored by the German Research
Foundation) to detect pacemaker abnormalities and evaluate their
complications for patients, thus, proving or disproving the
hypothesis.
METHODS AND RESULTS: Four hundred and fifteen pacemakers of deceased patients were analyzed assessing their functionality by in situ measurements and bench tests including five measurements and one telemetric interrogation. Results were divided into four categories and statistically evaluated. Life-threatening abnormalities were found in 3.8%, potentially life-threatening in 3.7%, probably symptomatic, divided into atrial and ventricular problems, 13.3% and 2.8%, respectively, and premature exhaustion in 1.2%. Three of 179 bipolar ventricular leads and 2 of 131 bipolar atrial leads had insulation defects corresponding to 1.7% and 1.5%, respectively. The bipolar complication rate was 2.8 times higher than unipolar.
CONCLUSION: The pacemaker patients investigated, living 4 years with their pacemaker on average, had a post-mortem evaluated complication rate of the category "life-threatening" of 3.8%. This result corresponds to an annual complication rate of 0.94% compared with a rate of only 0.39% in an earlier investigation.
Key Words: implanted pacemaker, post-mortem measurement, in situ measurement, dysfunction, complication rate
 |
Introduction
|
|---|
Due to implantable pacemaker systems, currently approximately
25003000 patients per million in the United States and
Europe live a normal life. As a result of the growing number
of older people in the population, particularly those aged over
75 years, an increasing number of pacemakers will be needed
worldwide [1]

. Possible pacemaker dysfunction in daily clinical
practice has already been described. Lead malfunction, for example,
is still the predominant cause of system dysfunction [2]

. Surgery
related complications during and after implantation are also
well known [3]

, but little is known about technical defects
and dysfunction of generators and electrodes during system life
[4]

.
It was expected that there is more undetected dysfunction of implanted pacemaker systems than that detected and corrected [4]
. This hypothesis challenged us to initiate a research project, sponsored by the German Research Foundation, to detect pacemaker abnormalities and evaluate their complication rate for patients, thus, proving or disproving the hypothesis.
 |
Materials and methods
|
|---|
The population investigated consisted of pacemakers from deceased
patients from different regions in Germany (Giessen, Friedberg,
Hamburg). Between August 2001 and November 2002, 415 pacemakers
together with 556 leads were removed and analyzed in bench tests.
Four hundred and three of these units were also investigated
in situ in the cadaver before cremation.
Table 1 summarizes
all populations in more detail. A measurement method was applied
that is suitable to judge the functionality of a pacemaker system
and guarantees clear results promptly after the patients' death
leaving little room for discussion as to the cause. The functionality
of the pacemaker was assessed by in situ measurements and bench
tests including five measurements and one telemetric interrogation,
carried out at room temperature, in the following sequence [5]

:
In situ measurements- Signals proportional to pacing pulses were derived with a differential amplifier connected to a digital storage oscilloscope read offline by a personal computer (Fig. 1). The amplifier connected to three cannula electrodes positioned linearly between apex and right clavicle to reduce noise (Fig. 2).
- With a test pacemaker (5 V, 1 ms pulse with a rate of 85 min1) and two additional cannula electrodes, synchronization of the implanted pacemaker could be tested.
- The lead impedance was measured with another test pacemaker that could be interrogated by a hand-held programmer.
Bench tests- 4. The parameters: pulse period, pulse amplitude, pulse duration, sensing threshold, and magnet influence were measured in the laboratory. The connector pins were checked for notches, which are a prerequisite for good contact between pin and screw(s). The bipolar lead insulation was checked with an ohm-meter.
- 5. Programmed pacemaker parameters and historical data were interrogated by telemetry (implantation date, date of the last follow-up, date of attainment of the electric replacement indicator (ERI) state).
- 6. If no pacing pulses were delivered by the device, the can was opened. The battery voltage was an indication of whether exhaustion and/or defect was present. In cases of battery exhaustion, the circuitry was connected to a 2.8 V power source to assess its function. Without a pacing pulse, the pacemaker was judged to be defective.
To evaluate the measurements of all 415 pacemakers the results were divided into different subgroups with respect to the severity of their impact (Tables 18).
 |
Results
|
|---|
Three hundred and seventy six cases had dates of birth and death
available allowing mean age at death to be calculated; 84.7
years (
Table 2). In 117 deceased pacemaker patients the date
of implantation could be interrogated. Mean age at implantation
of the last pacemaker was 80.7 years, in the other 298 cases
the implantation age was not available. Of the 411 ventricular
leads (as four generators were received by us without leads),
387 leads were checked for notches while 24 pacemakers were
from Intermedics, which do not possess screws (
Table 1Table 5).
Fig. 4Fig. 6 demonstrate that abnormal pulses usually
show characteristic deviations from normal (
Fig. 3), which have
been described earlier [5

7]

. A shunting resistor within
the pacemaker circuitry parallel to the electrode decreases
the output rapidly and below zero (
Fig. 4). If the insulation
of a bipolar lead degrades with fluid between the conductors,
a shunting capacitance is created that rounds leading and trailing
edges (
Fig. 5). Breakage or bad contact produces sharp spikes
followed by a low and flattened curve (
Fig. 6).
The discovered abnormalities were divided into four categories:
"life-threatening", "potentially life-threatening", "probably
symptomatic" and "premature exhaustion" (
Table 3). If the elective
replacement indicator (ERI) was reached within warranty period
(48 months), exhaustion was termed "premature" which was found
in 1.2%.
Table 4 shows the results for the category "life-threatening", from which a complication rate of 3.8% can be derived. The leading problem was battery exhaustion (1.7%), followed by failures of ventricular leads (1.0%). Pacemaker defects were about 0.5% and lastly infection, 0.2%. Infection as cause of death was confirmed by autopsy.
The category "potentially life-threatening" included 3.7% (Table 5) which is restricted only to ventricular problems, consisted of complications "infection" (diagnosis without autopsy), "missing notch" and "indifferent screw not tightened", in which missing notches on ventricular lead pins were most frequent (2.8%). A notch is a prerequisite for a good contact. Missing notches are potentially dangerous because of corrosion between different metals, an effect which was confirmed in some cases by high output voltages (5 V or higher) and long duration (1 ms).
In the category "probably symptomatic" (Table 6) the complication rate for atrial problems was 13.3% with deactivated atrial channels being the leading problem. Ventricular problems were found in 2.8% with insulation defects being most prominent (0.9%). There was a discrepancy in one case between the magnet test resulting in "ERI" while the programmer interrogation claimed battery to be "good". This uncertainty is surely of clinical relevance. "Magnet function unreliable" meant that several trials were necessary to open the reed switch. A lead that could be withdrawn by low extraction force was "not tightened".
Three of 179 bipolar ventricular leads and two of 131 bipolar atrial leads had an insulation defect corresponding to 1.7% and 1.5%, respectively (Table 7). Table 8 shows that the failure rate of ventricular bipolar leads is 8.9% versus 2.6% for unipolar yielding a defect ratio bipolar versus unipolar of 3.5%. If the defect ratio of all bipolar leads (6.8%) versus unipolar (2.4%) is calculated, the defect ratio drops to 2.8% (Table 8).
 |
Discussion
|
|---|
We analyzed and evaluated the functionality of 415 implanted
pacemaker systems post mortem to find out possible complications
and to derive complication rates. The applied post mortem in
situ measurement method seems to be unique worldwide. To our
knowledge, no data comparable to our results are available.
In the category "life-threatening" we found a complication rate
of 3.8% of patients that lived on average 4 years with their
pacemaker which means that those patients died with or because
of a dysfunction of their pacemaker. Thus, 3.8% divided by 4
years yields an annual complication rate of 0.95% which can
be compared with the 0.39% of an earlier investigation [4]

.
Thus, life-threatening dysfunction of implanted pacemaker systems
are 2.4 times higher than anticipated. Moreover, the category
"probably symptomatic" including atrial and ventricular problems
reveals a surprisingly high percentage of 13% respectively,
3% of pacemakers which do not work correctly or optimally with
atrial problems being predominant. Lead defects are more frequent
(ventricular 1%, atrial 3.4%) than generator defects (0.5%)
(
Table 4 and
Table 6) which support the statement that lead
malfunction is still the predominant cause of system dysfunction
[2]

. The defect ratio of ventricular bipolar leads is 3.5 times
higher compared with unipolar leads. Indeed, bipolar leads are
advantageous with respect to electromagnetic interference, but
this is counterbalanced by a higher defect rate.
The discrepancy in one case between magnet test resulting in "ERI" while the programmer interrogation claimed battery to be "good" raises concern because such an uncertainty provokes explantation which is surely of clinical relevance.
The age at last implant, which was 80.7 years in our study, is not representative of the German pacemaker population which is characterized by mean age at initial implantation for men of 73.3 years and for women 77.1 years (mean value 74.9 years) [8]
. Thus, our population of deceased patients undergoing cremation is approximately 6 years older than the general population. It has been shown [9,
10]
that pacemaker treatment is less generous with respect to mode selection with increasing age. Whether this can also be extrapolated to post implant care of older patients cannot be derived from our data.
 |
Conclusions
|
|---|
Our data reveal that there is much undetected pacemaker dysfunction
and possibly more than that detected and corrected. A post mortem
evaluated the complication rate in the category "life-threatening"
of 3.8% of patients that lived on average 4 years with their
pacemaker and the high rate of deviations from normal of 13%
should be taken into account and lead to improvements. Perhaps
follow-up intervals for pacemakers should be shortened. Incorrect
handling during or after implantation produces an astonishingly
high proportion of atrial (1.5%) and ventricular lead (3.1%)
dysfunction that must be reduced. More data will reveal whether
these results are also valid for other regions in Germany.
 |
Footnotes
|
|---|
*The paper was presented in part at the 24th Scientific Session
of the North American Society of Pacing and Electrophysiology
(NASPE), DC May 15, 2003.

 |
References
|
|---|
[1] Irnich W. Pacemaker-related patient mortality. Pacing Clin Electrophysiol 1999; 22: 12791283.
[Medline][2] Daley WR and Kaczmarek RG. The epidemiology of cardiac pacemakers in the older US population. J Am Geriatr Soc 1998; 46: 10161019.[Medline]
[3] Geppert A and Rauscha F. Pacemaker dysfunction in daily clinical practice. Wien Klin Wochenschr 2001; 113: 1526.[Medline]
[4] Da Costa A, Kirkorian G, Isaaz K, Touboul P. Secondary infections after pacemaker implantation. Rev Med Intern 2000; 21: 256265.
[5] Bartsch C, Irnich W, Risse M, Junge M, Weiler G. Postmortem in situ diagnosis of pacemakers and electrodes to detect dysfunction. Leg Med 2003; 5: S397S400.[CrossRef]
[6] Irnich W, Stertmann WA, Müller R, Rempe M, Hehrlein FW. How to detect defective bipolar leads. In Vardas PE (Ed.). Proceedings EUROPACE '97, Athens (Greece), June 811 1997; Bologna Monduzzi Editore pp. 815819.
[7] Irnich W, Batz L, Müller R, Stertmann WA. Physics of failing bipolar leads. In Vardas PE (Ed.). Proceedings EUROPACE '97, Athens (Greece), June 811 1997; Bologna Monduzzi Editore pp. 11071111.
[8] Markewitz A. Annual report 2000/2001 of the German central register of pacemakers. Herzschr Elektrophys 2002; 13: 242257.[CrossRef]
[9] Batz L and Irnich W. Demographic aspects of pacemaker therapy. Herzschr Elektrophys 1995; 6: 410.
[10] Batz L and Irnich W. Are there specific gender differences in pacemaker selection? Biomed Tech 1997; 42:Suppl 7778.

CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:

|
 |

|
 |
 
M. Novak, P. Dvorak, P. Kamaryt, B. Slana, and J. Lipoldova
Autopsy and clinical context in deceased patients with implanted pacemakers and defibrillators: intracardiac findings near their leads and electrodes
Europace,
November 1, 2009;
11(11):
1510 - 1516.
[Abstract]
[Full Text]
[PDF]
|
 |
|