The midlands trial of empirical amiodarone versus electrophysiology-guided interventions and implantable cardioverter-defibrillators (MAVERIC): a multi-centre prospective randomised clinical trial on the secondary prevention of sudden cardiac death
aQueen Elizabeth Hospital Birmingham B15 2TH, UK; bGlenfield Hospital Leicester, UK; cClinical Trial Unit, University of Birmingham Birmingham, UK; dManchester Royal Infirmary Manchester, UK
Manuscript submitted 4 August 2003. Accepted after revision 28 March 2004.
*Corresponding author. Tel.: +44-121-627-2043; fax: +44-121-414-1045. E-mail address: michael.griffith{at}uhb.nhs.uk
| Abstract |
|---|
|
|
|---|
AIMS: MAVERIC was a randomised clinical trial designed to test the possibility of prospectively identifying patients who would benefit most from the implantable cardioverter-defibrillator (ICD) by electrophysiology (EP) study in the context of secondary prevention of sudden cardiac death (SCD) through comparing EP-guided interventions (anti-arrhythmic drugs, coronary revascularization, and ICD) against empirical amiodarone therapy.
METHODS: Two hundred and fourteen survivors of sustained ventricular tachycardia (VT), ventricular fibrillation (VF) or SCD were randomized to either treatment strategy, pre-stratified for haemodynamic status at index event, and followed up for a median of 5 years.
RESULTS: Of the 106 amiodarone arm patients, 89 (84%) received the drug and 5 (5%) received an ICD after crossing over. Of the 108 EP arm patients, 31 (29%) received an ICD, 46 (43%) received anti-arrhythmic drugs only (mainly amiodarone or sotalol) and 18 (17%) received coronary revascularization but no ICD. No significant differences in survival or arrhythmia recurrence existed between the two treatment arms after 6 years. However, ICD recipients had a lower mortality than non-ICD recipients, regardless of allocated treatment (hazard ratio = 0.54, p=0.0391).
CONCLUSIONS: Prospective selection of patients to receive the ICD by EP study did not improve survival compared with empirical amiodarone therapy among survivors of VT, VF or SCD, whereas ICD implantation improved survival regardless of allocated treatment. On this basis, routine EP study has no role in the management of such patients, who should be offered empirical ICD therapy according to the results of other secondary prevention ICD trials.
Key Words: ICD, electrophysiology, amiodarone, sudden cardiac death
| Introduction |
|---|
|
|
|---|
In the treatment of patients who are presented with life-threatening ventricular arrhythmias, AVID showed that empirical implantable cardioverter-defibrillator (ICD) therapy was significantly better in preventing death than anti-arrhythmic drugs, principally empirical amiodarone [1]
The MAVERIC trial, initiated before the other major trials (AVID, CASH and CIDS) were completed, sought prospectively to identify patients who would benefit most from ICD therapy by electrophysiology (EP) testing and deliver a mortality benefit over empirical amiodarone therapy comparable to that achieved by empirical ICD implantation. The alternative, to compare EP-guided ICD implantation with empirical ICD implantation, would have been scientifically more appealing but not economically feasible in the United Kingdom. Moreover, it generally takes a larger sample size to prove equivalence than a statistically significant difference between two treatments.
In MAVERIC, the decision to implant an ICD depended on the patient's estimated risk for future life-threatening ventricular arrhythmias, as determined by the presenting arrhythmia, LVEF, scope for coronary revascularization, and baseline EP or Holter abnormalities. Despite the inherent shortcomings of these criteria, they were and probably still are the best methods available for the purpose of risk stratification [6,
7]
. In MAVERIC, the high but not the low risk patients received the ICD.
| Methods |
|---|
|
|
|---|
Study conduction
MAVERIC was conducted in the Midlands region (population 9.1 million) of the UK and involved all the local hospitals with the approval of their local research ethics committees. All survivors of sustained ventricular tachycardia (VT) (i.e. >30 s), ventricular fibrillation (VF) or sudden cardiac death (SCD) in the absence of an acute myocardial infarction in the last 48 h were eligible for inclusion, provided they did not have a likely life expectancy of <6 months from a non-arrhythmic cause or were of child-bearing age. Eligible patients who gave informed consent to join the trial were randomized by sealed envelopes to either empirical amiodarone therapy or EP-guided interventions with pre-stratification for haemodynamic status at index event; otherwise they entered the study's registry (Fig. 1). A clinical event was considered haemodynamically unstable if the patient required resuscitation, was syncopal or had a systolic blood pressure of <80 mmHg. Baseline characteristics including demographics, past medical history and family history were recorded at initial presentation. A baseline estimation of LVEF by either echocardiography or left ventriculography was also obtained.
|
Trial recruitment ran from February 1997 to January 1999. Data were frozen for analysis in January 2003, when the median follow-up was 60 months.
Study protocol
The MAVERIC protocol was as shown in Fig. 1. Investigation for possible myocardial ischaemia with exercise stress testing and coronary angiography was mandatory for patients in the EP arm but discretionary for patients in the amiodarone arm. Coronary revascularization by either percutaneous transluminal coronary angioplasty (PTCA) or coronary artery bypass grafting surgery (CABG) was administered according to current guidelines [8]
. Left ventricular aneurysmectomy was performed if indicated. After any appropriate coronary revascularization had been completed, patients in the EP arm underwent serial programmed ventricular stimulation by a protocol modified from Wellens et al. [9]
. The protocol comprised six stages, with drive cycles of 600 ms and 400 ms and up to three extra-stimuli from either the right ventricular apex or outflow tract. Positive end-points were defined as the induction of either a sustained monomorphic VT of >30 s duration or VF. When VT/VF was non-inducible, Holter monitoring could be used as an alternative means of assessing arrhythmia suppression if the patient had >30 ventricular ectopic beats per hour on average [10]
. If neither programmed ventricular stimulation nor Holter monitoring could be used for assessing arrhythmia suppression, an ICD was implanted. The definitions of successful arrhythmia suppression were as stated in Fig. 1, and the sequence of anti-arrhythmic drugs tried was determined by the patient's LVEF. If arrhythmia suppression with either sotalol or amiodarone was unsuccessful, further treatment depended on haemodynamic status of the index event: if haemodynamically unstable, the patient received an ICD; otherwise, various combinations of anti-arrhythmic drugs (such as the combined use of sotalol or mexiletine with amiodarone) were tried.
In the EP arm, when specific forms of VT particularly amenable to ablation (Fig. 1) or misdiagnosis of supraventricular tachycardia with aberrant conduction as VT was suspected, a full EP study was performed, with radiofrequency ablation if appropriate.
End-points
The trial's primary end-point was death; the secondary end-points were VT/VF/SCD recurrence and cross-over in treatment. Follow-up for death was through the Office of National Statistics of the United Kingdom, and for hospitalization and arrhythmia recurrence through scrutiny of study patients' case notes every 3 months. The cause and mode of death were decided by two adjudicators independent of all other aspects of study conduction through scrutiny of the deceased's case notes. Deaths which were sudden (i.e. within 1 h of onset of symptom) but without any documentary evidence of arrhythmia were classified as sudden cardiac rather than arrhythmic deaths.
Statistical analyses and power calculation
Baseline characteristics were compared with the unpaired t-test or contingency table. Survival and recurrence-free survival were depicted as KaplanMeier curves and compared with the log-rank statistic by the intention-to-treat principle. A p-value of <0.05 was regarded as significant. Multivariate analysis on factors that might affect survival was conducted with logistic regression. A sample size of 200 was estimated to have a 90% power to detect a 50% relative risk reduction in mortality.
| Results |
|---|
|
|
|---|
Out of 689 patients meeting the inclusion criteria, 214 joined the trial. Of the 122 trial patients haemodynamically stable at index event, 60 were in the EP arm and 62 in the amiodarone arm. Of the 92 trial patients haemodynamically unstable at index event, 48 were in the EP arm and 44 in the amiodarone arm.
Baseline characteristics
The two arms were comparable for all characteristics examined except for age, which was lower for the EP arm than for the amiodarone arm (65.9±10.3 years versus 68.5±9.4 years, p=0.051) (Table 1). By virtue of pre-stratified randomization, the two arms were also comparable in the proportion of patients haemodynamically stable at index event.
|
Treatments received
Of the 62 patients haemodynamically stable at index event in the amiodarone arm, 54 (87%) were maintained on amiodarone only. Six patients (10%) crossed over to the EP arm: 4 received an ICD; 1 received radiofrequency ablation of VT substrate and 1 received combined mexiletine and amiodarone therapy. Two patients (3%) died of cardiogenic shock shortly after randomization before completion of amiodarone loading.
Of the 60 patients haemodynamically stable at index event in the EP arm, 33 (55%) were maintained on anti-arrhythmic drugs only (20 on amiodarone alone; 9 on sotalol alone; 1 on amiodarone and mexiletine; 1 on sotalol and mexiletine; 2 on mexiletine alone). Seven patients (12%) received coronary revascularization but no ICD (4 had CABG only; 1 had CABG and left ventricular aneurysmectomy; 1 was maintained on amiodarone after CABG; 1 was maintained on amiodarone after PTCA). Fourteen patients (23%) received an ICD (4 received an ICD only; 6 were maintained on amiodarone after receiving an ICD; 2 were maintained on sotalol after receiving an ICD; 1 was maintained on mexiletine after receiving an ICD; 1 received an ICD after PTCA). Three patients (5%) were successfully treated with radiofrequency ablation (1 for VT substrate; 2 for supraventricular tachycardia substrate). Three patients (5%) died shortly after randomization before receiving full EP-guided interventions.
Of the 44 patients haemodynamically unstable at index event in the amiodarone arm, 35 (80%) were maintained on amiodarone only. Six patients (14%) were intolerant of amiodarone: 2 were switched over to sotalol but 4 were not maintained on any other anti-arrhythmic therapy. Three patients (7%) had recurrences while on amiodarone and crossed over to the EP arm (1 received CABG and remained on amiodarone; 1 received an ICD, and 1 received a permanent pacemaker as the syncopal episode was deemed to be due to heart block induced/aggravated by amiodarone).
Of the 48 patients haemodynamically unstable at index event in the EP arm, 13 (27%) were maintained on anti-arrhythmic drugs only (10 on amiodarone alone; 2 on sotalol alone and 1 on sotalol and mexiletine). Eleven patients (23%) received coronary revascularization but no ICD (9 received CABG alone; 1 was maintained on amiodarone after CABG; 1 was maintained on amiodarone and sotalol after PTCA). Seventeen patients (35%) received an ICD (8 received an ICD only; 4 received an ICD after CABG; 5 were maintained on amiodarone after receiving an ICD). One patient (2%) had sotalol-induced long QT syndrome precipitating polymorphic VT and was managed simply by withdrawal of sotalol. Two patients (4%) had co-existent uncontrolled atrial fibrillation and were treated with atrioventricular nodal ablation and permanent pacemaker implantation. Two patients (4%) died shortly after randomization before receiving full EP-guided interventions. Two patients (4%) withdrew from the study shortly after randomization.
Overall, of the 106 amiodarone arm patients, 89 (84%) received the drug and 5 (5%) received an ICD after crossing over. Of the 108 EP arm patients, 31 (29%) received an ICD, 46 (43%) received anti-arrhythmic drugs only (mainly amiodarone or sotalol) and 18 (17%) received coronary revascularization but no ICD.
Death
After a maximum of 6 years follow-up (median 5 years), there was no significant difference in survival between the two treatment arms, with or without pre-stratification for haemodynamic stability at index event (Fig. 2). However, there was a statistically non-significant trend for patients randomized to EP-guided interventions to have an initially worse but subsequently better survival experience than patients randomised to empirical amiodarone therapy, especially when the index event was associated with haemodynamic compromise.
|
Event-free survival
For the composite end-point of death and recurrence of arrhythmia, the amiodarone arm had a slightly better experience than the EP arm throughout follow-up, but the difference did not reach statistical significance (log-rank = 1.42, p=0.2335).
Effects of ICD implantation on survival
When the survival experiences of all trial patients were compared according to whether they received an ICD or not rather than their allocated treatments, ICD recipients consistently did better than non-ICD recipients, and the difference reached statistical significance (Fig. 3(a)). However, the survival benefit of ICD implantation was more marked for patients haemodynamically unstable at index event than those haemodynamically stable at index event (Fig. 3(b) and (c)).
|
In terms of baseline characteristics (Table 2), ICD recipients were significantly younger and also less likely to have diabetes than non-ICD recipients, but the two groups were otherwise comparable in other respects.
|
Multivariate analysis on factors affecting survival
Age, LVEF<35%, diabetes and congestive cardiac failure were independently associated with an increased risk for death (Table 3). ICD implantation was associated with a reduced risk for death but the association did not reach statistical significance (p=0.080).
|
| Discussion |
|---|
|
|
|---|
The overall neutral results of MAVERIC imply that empirical amiodarone therapy is just as effective as EP-guided interventions in the secondary prevention of SCD. The sample size in this trial was relatively small but comparable to those in CASH [3]
MAVERIC was different from other secondary prevention trials such as AVID, CIDS and CASH in two crucial respects. First, patients haemodynamically stable at index event were included in MAVERIC (>50% of the study population) but excluded from the other studies. Second, MAVERIC compared empirical amiodarone therapy against EP-guided interventions rather than against empirical ICD therapy. In MAVERIC, ICD therapy was guided by programmed ventricular stimulation rather than routinely indicated in the EP arm, which accounted for its relatively low utilization rate (29%). These two factors may explain why EP-guided ICD implantation did not achieve the same survival benefit over empirical amiodarone therapy in MAVERIC as empirical ICD therapy in other secondary prevention trials [1
4]
.
In MAVERIC, ICDs were implanted only in patients considered at high risk either by their presentation or risk stratification by EP study. Yet, ICD implantation was associated with a greatly improved survival in the trial. This suggests that the prognosis of even the "high-risk" patients was so much improved by ICD therapy that it actually became better than that of the "low-risk" patients. This benefit just became non-significant with multivariate analysis, but this is likely to be due to the sample size involved. However, even though the majority of ICD recipients were in the EP arm, the survival benefit conferred by ICD implantation did not translate into an overall survival advantage for the whole treatment arm. This most probably reflects a dilution effect on the survival benefit by the poor prognosis of other "low-risk" patients who did not receive ICDs.
MAVERIC did not directly compare EP-guided against empirical ICD implantation, and it may be argued that such a direct head-to-head comparison would have been better. However, such a comparison will have to be an equivalence trial, which will require a large sample size. If MAVERIC had demonstrated survival benefit from EP-guided ICD implantation over empirical amiodarone therapy comparable to that observed for empirical ICD implantation over empirical amiodarone therapy in other secondary prevention trials, then such a direct head-to-head comparison would have been indicated. The neutral results of MAVERIC obviate the need for such a comparison. However, this does not mean that more precise targeting of ICD therapy is unnecessary or impossible. There is emerging evidence that empirical ICD therapy is better than anti-arrhythmic (amiodarone) therapy in only certain subgroups of survivors of VT, VF or SCD [5
,12
14]
.
MAVERIC shows that EP study has a minimal impact on the diagnosis of patients presented with VT, VF or SCD. The substrate of nearly all broad complex tachycardias can be identified from the surface ECG. In MAVERIC, only 2 of 108 patients in the EP arm were found to have a supraventricular cause for their broad complex tachycardias. However, for both these patients, a supraventricular cause for their tachycardias was strongly suspected from the presentation ECGs. Thus if the diagnosis of VT is clear from the surface ECG, diagnostic EP study is unnecessary. As this trial does not support a role for EP testing in risk stratification either, EP study should be reserved for patients with an uncertain diagnosis.
In MAVERIC, coronary revascularization was not part of the randomized treatments. In the amiodarone arm, it was left to the discretion of the attending physician. In the EP arm, an aggressive revascularization policy was pursued, with coronary angiograms performed in all patients and coronary revascularization attempted as far as it was technically feasible. Thus MAVERIC could not address the role of coronary revascularization in the management of survivors of VT, VF or SCD. The CABG Patch trial showed that prophylactic ICD implantation in patients with coronary artery disease, a depressed LVEF and an abnormal signal-averaged ECG at the time of elective CABG did not confer any survival benefit [15]
, but that was in the context of primary prevention of SCD. In contrast, the AVID registry showed that coronary revascularization produced a survival benefit of its own and did not diminish that of ICD therapy in the secondary prevention of SCD [16]
. Consequently, coronary revascularization should be recommended to survivors of VT, VF or SCD according to current guidelines [8]
.
In conclusion, MAVERIC shows that prospective selection of patients to receive in ICD by EP study combined with aggressive coronary revascularization did not improve survival compared with empirical amiodarone therapy among survivors of VT, VF or SCD. On the other hand, ICD recipients had a significantly better survival experience than non-ICD recipients, even though this observation was based on non-randomized data. The overall results of MAVERIC suggest that the methods used to select patients for ICD implantation in this trial were not sensitive enough to identify all high-risk patients who would benefit from the therapy, and routine EP testing and aggressive coronary revascularization do not add value in the management of such patients. On this basis, patients presented with VT, VF or SCD should be offered empirical ICD therapy according to the results of other secondary prevention ICD trials [1
4]
.
| References |
|---|
|
|
|---|
[1] The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med 1997; 337: 15761583.
[2] Connolly SJ, Gent M, Roberts RS, et al. Canadian implantable defibrillator study (CIDS): a randomized trial of the implantable cardioverter defibrillator against amiodarone. Circulation 2000; 101: 12971302 [For the CIDS Investigators.].
[3] Kuck KH, Cappato R, Siebels J, Rüppel R. Randomized comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from cardiac arrest. The Cardiac Arrest Study Hamburg (CASH). Circulation 2000; 102: 748754.
[4] Connolly SJ, Hallstrom AP, Cappato R, et al. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Antiarrhythmics vs implantable defibrillator study. Cardiac Arrest Study Hamburg. Canadian Implantable Defibrillator Study. Eur Heart J 2000; 21: 20712078.
[5] Sheldon R, Connolly S, Krahn A, et al. On behalf of the CIDS Investigators. Identification of patients most likely to benefit from implantable cardioverter-defibrillator therapy. The Canadian Implantable Defibrillator Study. Circulation 2000; 101: 16601664.
[6] Stevenson WG, Brugada P, Waldecker B, Zehender M, Wellens HJ. Clinical, angiographic, and electrophysiologic findings in patients with aborted sudden death as compared with patients with sustained ventricular tachycardia after myocardial infarction. Circulation 1985; 71: 11461152.
[7] Schmitt C, Barthel P, Ndrepepa G, et al. Value of programmed ventricular stimulation for prophylactic internal cardioverter-defibrillator implantation in postinfarction patients preselected by noninvasive risk stratifiers. J Am Coll Cardiol 2001; 37: 19011907.
[8] Eagle KA, Guyton RA, Davidoff R, et al. ACC/AHA guidelines for coronary artery bypass graft surgery: executive summary and recommendations: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to revise the 1991 guidelines for coronary artery bypass graft surgery). Circulation 1999; 100: 14641480.
[9] Zehender M, Brugada P, Geibel A, Waldecker B, Stevenson W, Wellens HJ. Programmed electrical stimulation in healed myocardial infarction using a standardized ventricular stimulation protocol. Am J Cardiol 1987; 59: 578585.[CrossRef][ISI][Medline]
[10] Hedblad B, Janzon L, Johansson BW, Juul-Moller S. Survival and incidence of myocardial infarction in men with ambulatory ECG-detected frequent and complex ventricular arrhythmias. 10 year follow-up of the Men born 1914 study in Malmo, Sweden. Eur Heart J 1997; 18: 17871795.
[11] Moss AJ, Hall WJ, Cannom DS, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. N Engl J Med 1996; 335: 19331940 [For the Multicenter Automatic Defibrillator Implantation Trial Investigators.].
[12] Hallstrom AP, McAnulty JH, Wilkoff BL, et al. Patients at lower risk of arrhythmia recurrence: a subgroup in whom implantable defibrillators may not offer benefit. Antiarrhythmics Versus Implantable Defibrillator (AVID) Trial Investigators. J Am Coll Cardiol 2001; 37: 10931099.
[13] Exner DV, Sheldon RS, Pinski SL, Kron J, Hallstrom A. Do baseline characteristics accurately discriminate between patients likely versus unlikely to benefit from implantable defibrillator therapy? Evaluation of the Canadian implantable defibrillator study implantable cardioverter defibrillatory efficacy score in the antiarrhythmics versus implantable defibrillators trial. Am Heart J 2001; 141: 99104.[CrossRef][ISI][Medline]
[14] Domanski MJ, Epstein A, Hallstrom A, Saksena S, Zipes DP. Survival of antiarrhythmic or implantable cardioverter defibrillator treated patients with varying degrees of left ventricular dysfunction who survived malignant ventricular arrhythmias. J Cardiovasc Electrophysiol 2002; 13: 580583.[CrossRef][ISI][Medline]
[15] Bigger JT Jr. Prophylactic use of implanted cardiac defibrillators in patients at high risk for ventricular arrhythmias after coronary-artery bypass graft surgery. N Engl J Med 1997; 337: 15691575 [For the Coronary Artery Bypass Graft (CABG) Patch Trial Investigators.].
[16] Cook JR, Rizo-Patron C, Curtis AB, et al. Effect of surgical revascularization in patients with coronary artery disease and ventricular tachycardia or fibrillation in the antiarrhythmics versus implantable defibrillators (AVID) registry. Am Heart J 2002; 143: 821826.[CrossRef][ISI][Medline]
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
F. S. Luyster, J. W. Hughes, D. Waechter, and R. Josephson Resource loss predicts depression and anxiety among patients treated with an implantable cardioverter defibrillator. Psychosom Med, September 1, 2006; 68(5): 794 - 800. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Goldberger and R. Lampert Implantable Cardioverter-Defibrillators: Expanding Indications and Technologies JAMA, February 15, 2006; 295(7): 809 - 818. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||