Skip Navigation

Europace 2004 6(3):243-247; doi:10.1016/j.eupc.2004.01.007
© 2004 by European Society of Cardiology
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (13)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Bracke, F.A.L.E.
Right arrow Articles by van Gelder, L.M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bracke, F.A.L.E.
Right arrow Articles by van Gelder, L.M.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Lead extraction for device related infections: a single-centre experience

F.A.L.E. Bracke*, A Meijer and L.M. van Gelder

Department of Cardiology, Catharina Hospital Michelangelolaan 2, P.O. Box 1350, 5602 ZA Eindhoven, The Netherlands

Manuscript submitted 18 August 2003. Accepted after revision 26 January 2004.

*Corresponding author. Tel.: +31-40-2397004; fax: +31-40-2447885. E-mail address: frank.bracke{at}cze.nl (F.A.L.E. Bracke).


    Abstract
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
AIMS: We report a single-centre experience of extraction of infected pacing and ICD leads.

METHODS: Retrospective study of consecutive lead extractions for infection. Lead extraction was by traction, or, if unsuccessful, a laser sheath was used. All procedures were performed in the operating room.

RESULTS: Leads were extracted in 82 patients including 66 patients (80%) with infection occurring after replacement of the generator or revision of the leads. Previous treatment, without lead extraction, had failed in 51 patients (62%). Major complications (tamponade or haemothorax) occurred in 6 patients, 2 patients died despite emergency surgery. One patient succumbed to ongoing sepsis. Of the patients alive, a follow-up of at least 6 months (27±17 months) was available in 76 patients. All patients were cured; none had a recurrence. In 31 patients (41%) pacing was abandoned after lead extraction; all remained asymptomatic.

CONCLUSION: Lead extraction is effective in curing pacemaker or ICD related infection, even after failed conservative therapy, but with a significant complication rate. The routine replacement of the generator should be reconsidered in patients in whom the indication for pacing is no longer valid, as the majority of infections occurred after revision of the system.

Key Words: pacemaker, infection, lead extraction


    Introduction
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
According to the NASPE policy statement on recommendations for extraction of chronically implanted transvenous pacing and defibrillator leads, systemic pacemaker related infection is a class 1 indication, localized pocket infections or erosions are a class 2 indication [1]Go. Endovascular extraction techniques including laser sheaths are reported to be effective for lead removal in multi-centre registries and this has stimulated the use of lead extraction in the treatment of device related infections [2]Go. We report a single-centre experience and follow-up of lead extraction in a referral population for device related infections.


    Patients and methods
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
This is a retrospective study of the outcome of lead extraction for pacing or ICD system related infection in consecutive patients treated in a single referral centre. All patients had a follow-up of at least 6 months after lead removal.

There were 82 patients, 80 with a pacemaker and 2 with an ICD. Their mean age was 62±17 years, 52 were male. Twelve patients had clinical signs of systemic infection. Fifty-two patients (62%) were referred after lead-conserving treatment, either with antibiotics or surgery, had failed. Infection had occurred after primary implant in 16 patients (20%) and after generator exchange or revision of the system for lead dysfunction in 66 patients (80%). One, 2, 3 and 4 leads were present in, respectively, 17, 53, 11 and 1 patients. Time from implant of all leads was 72±62 months, range 0.5–342 months. Time from implant of the oldest lead present was 84±68 months.

Lead extraction was performed by 2 operators. It was executed in the operating room under general anaesthesia with the patient prepared for thoracotomy and with a cardiac surgeon and extracorporeal circulation team present in the operating room. Transoesophageal echocardiography throughout the procedure was used to monitor possible complications. Only when all leads were implanted for less than 6 months was the extraction procedure performed in the catheterization laboratory, as in our experience simple traction sufficed to remove the leads [3]Go. A stepwise protocol for extraction was followed. Traction was first applied, if necessary with a locking stylet. If this failed, a laser sheath (SLS I, Spectranetics, Co) was inserted over the lead and locking stylet. Laser energy was provided by an Excimer Xenon chloride laser (Spectranetics, Co). The size of the sheath (12, 14 or 16 F) was chosen according to the recommendations of the manufacturer with a tendency to oversize in case of doubt, or if the chosen size stalled. If the tip of the lead was not yet free by the time the laser sheath reached the myocardium, counter-traction using the outer sheath was applied. In case of laser failure or if a lead dislocated but could not be pulled through the adhesions, a femoral approach with a femoral workstation and a retriever was attempted.

Statistical analysis was performed using a two-tailed Fisher's test for contingency tables.


    Results
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
Traction was sufficient for removal of all leads in 27 patients, and a laser sheath was used for at least 1 lead in the remaining 55 patients (Fig. 1). Major complications occurred in 6 patients, only during laser procedures. All these patients had a local infection of the pacemaker pocket and a history of previous unsuccessful conservative therapy. In 5 of these patients use of the laser caused a tear in the superior vena cava, the sixth patient had perforation of the atrium after counter-traction was applied to the atrial wall. All these patients showed haemodynamic instability almost immediately after detection of the complication and had to undergo emergency thoracotomy. Two of the patients with a tear in the superior vena cava died as a result of profound hypovolaemic shock: one of them could not be resuscitated during surgery, the other died post-operatively. Both patients were female and, respectively, 84 and 79 years old. In the other 4 patients surgery was successful without post-operative sequelae. Any leads still present at the time of the complication were removed during surgery. The patients with a tear in the superior vena cava had longer than average implant times: 124±66 months vs. 72±62 months. No other complications were encountered in the study.



View larger version (7K):
[in this window]
[in a new window]
 
Figure 1 Outcome of the lead extraction procedure. FWS: femoral workstation; {dagger} = patients who died.

 
Laser sheath extraction was successful in 43 of the remaining, uncomplicated procedures. The laser failed in 6 patients. The leads were subsequently extracted via a femoral workstation in 5 patients, and in 1 patient a 28-year-old lead was removed during elective surgery. All leads were extracted completely except the tip of 1 electrode that dislocated from the lead body when the lead impacted in the subclavian vein during withdrawal.

Post-operatively, a 90-year-old female patient died from ongoing sepsis 1 month after an uneventful lead removal with traction.

Long-term follow-up could not be obtained in 3 patients (Fig. 2). All of the remaining 76 patients alive were cured from, and remained free of infection during a follow-up of 27±17 months (range 6–70 months). Four patients died during follow-up of causes not related to the extraction procedure or pacemaker therapy (all had new pacemakers implanted and survived the extraction procedure for at least 6 months).



View larger version (11K):
[in this window]
[in a new window]
 
Figure 2 Follow-up of the patients who survived lead extraction. FU: follow-up; AV: atrio-ventricular conduction defect; SSS: sick sinus syndrome; CSS: carotid sinus syndrome; PM: pacemaker; {dagger} = patients who died.

 
In 31 (41%) of these 76 patients it was considered that continuation of pacing therapy was not mandatory, and these patients did not receive a new pacemaker after lead extraction. All these patients remained asymptomatic during follow-up. The original indication for pacemaker implant was determined in 72 patients. Fifteen (32%) of 47 patients with atrio-ventricular conduction disturbances did not receive a new pacemaker in contrast to 15 (60%) of 25 patients with sick sinus or carotid sinus syndrome (p=0.03).


    Discussion
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
Lead extraction was highly effective in curing pacemaker or ICD related infection in this study. However, it was accompanied by major complications in 7% of patients including the death of 2 patients directly related to the procedure. In large registries of lead extraction, major complications are reported in 1.9–2.5% and mortality in 0.6–0.8% of procedures [2,Go4]Go. In these studies, complete extraction of the leads was obtained in 86–90% of patients. There may be several factors contributing to the higher complication rate in the present study. First, the implant times of the leads in patients with complications were longer than the rest of the study group. Byrd et al. reported an increased risk of failed or partial extraction with increasing implant duration, effectively doubling every 3 years [5]Go. Also the Accufix Research Institute reported an increase in complication rate of extraction of the Accufix lead (Telectronics Pacing Systems Inc., Englewood, CO, USA) from 2% at 1-year of implant time to 8.3% with implant duration of more than 5 years (http://www.accufix.com). Secondly, although physician's experience has been associated with the risk of complications, it is interesting to note that the majority of complications occurred only late in our experience [4,Go5]Go. Probably, our early results increased the confidence in the laser sheath and we persevered with the laser even when progress stalled. Finally, the first generation of laser sheaths that was used in all study patients was less flexible than the current model, and also lacked the bevelled tip making it more difficult to avoid the vessel wall.

Resulting from our experience we modified the approach when the laser sheath stalls, especially at the level of the superior vena cava–right atrial junction which is vulnerable to perforation. Rather than persevere, we continue extraction with a femoral approach. The proximal isodiametric part of the lead can easily be pulled down from the binding sites in the superior caval vein in most cases. Avoiding extensive traction at this vulnerable point greatly reduces the risk of perforation. If the lead cannot be mobilized from the femoral vein, we try to retrieve it through the internal jugular vein as described by Bongiorni et al. [6]Go. We surgically explore the internal jugular vein to obtain optimal haemostatic control whilst introducing sheaths or retrieving the leads as bleeding in this area can potentially be life threatening. In this situation also the laser can be applied again, now only having to negotiate the distal part of the lead with the laser.

It should be remembered that elective surgery could be an acceptable alternative for complicated endovascular extraction. However, mortality from surgical removal of infected leads is reported in 2.4–17% of patients [7–Go12]Go.

This study again underlines the necessity to perform lead extraction only in the operating room with cardio-pulmonary bypass available on-site and the patient prepared for emergency surgery. General anaesthesia already instituted from the start of the procedure saved valuable time in case of complications. This approach saved the lives of 4 patients, nevertheless bleeding caused by superior vena cava tears can compromise the circulation beyond recovery, as happened in both elderly patients.

Localized pacemaker or ICD related infections are considered a class 2 indication for lead extraction in a NASPE policy statement [1]Go. The results of this study also contribute to a cautious approach in high-risk patients with local infections, and lead extraction may be deferred with an initial attempt at conservative therapy. Implant time, female sex, multiple leads and advanced age increase the risk of lead extraction [2,Go4,Go5,Go13–Go15]Go. An additional consideration is a high operative risk, especially in the event of emergency surgery. Reports of successful conservative treatment have been mostly in relation to low-grade pocket infections or skin erosions [16–Go19]Go. Pacemaker or ICD related sepsis or endocarditis is class 1 indication for lead extraction [1]Go. In these patients and in frank pocket infections we also strongly favour lead extraction as therapy of choice.

Although no standard protocol to reassess the necessity of pacemaker therapy was used in this study, pacing was safely discontinued after lead extraction in 41% of patients, concordant with the 13–52% reported in the literature [12,Go14,Go20]Go. It happened significantly more in patients considered to have either sick sinus or carotid sinus syndrome than in those having atrio-ventricular block (60% vs. 32%).

It is noteworthy that 80% of patients were referred with infection occurring after a generator exchange or lead revision. Also Harcombe et al. reported that the overall rate of late complications, the majority of which consisted of infection or erosion, is significantly higher after elective unit replacement (6.5%) than after a first implant (1.4%) [21]Go.

Even though the implant rate of pacemakers in the Netherlands is among the lowest in Western Europe (±350 implants/million inhabitants), the above findings stress the importance of careful assessment of pacing indications not only at implant but also at the time of generator replacement or lead revision. Therefore, reassessment of the original indication should be made in any patient, when elective generator replacement approaches, to judge if pacing has to be continued. It can then be decided to leave the generator and leads in situ rather than replacing them with the risk of complications. However, leaving a generator with a depleted battery raises some concerns. Firstly, as the end of battery life approaches, one has to be sure that the pacemaker does not revert to unwanted pacing. This should be confirmed with the pacemaker manufacturer. Secondly, although runaway pacemaker behaviour, with high stimulation rates of more than 2000 bpm, has been described following battery depletion this is a rare condition in modern pacemakers and has resulted only in subthreshold stimuli [22]Go. Thirdly, there is the possibility of battery leakage, but this is highly unlikely in modern pacemakers. As an alternative, only the generator may be removed, abandoning the leads properly insulated and fixed: this diminishes the risk of skin erosion, which is a main cause of infection after generator replacement. Although extraction of the leads has been advocated when pacing is abandoned, this should be heavily discouraged as the risk outweighs any potential benefit [19,Go23]Go.

In conclusion, although lead extraction is highly effective for eradicating pacemaker or ICD lead related infection, it is not without complications and the application of it as first line therapy should be tailored to the risk profile of both the extraction procedure and infection in a given patient. The automatic replacement of the generator when depleted has to be reconsidered given the frequent possibility of abandoning pacing therapy after lead extraction and the high incidence of infection occurring after generator replacement.


    References
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
[1] Love CJ, Wilkoff BL, Byrd CL, Belott PH, Brinker JA, Fearnot NE, et al. Recommendations for extraction of chronically implanted transvenous pacing and defibrillator leads: indications, facilities, training. North American Society of Pacing and Electrophysiology Lead Extraction Conference Faculty. Pacing Clin Electrophysiol 2000; 23: 544–551.[CrossRef][Medline]

[2] Byrd CL, Wilkoff BL, Love CJ, Sellers TD, Reiser C. Clinical study of the laser sheath for lead extraction: the total experience in the United States. Pacing Clin Electrophysiol 2002; 25: 804–808.[CrossRef][Medline]

[3] Bracke F, Meijer A, van Gelder B. Extraction of pacemaker and implantable cardioverter defibrillator leads: patient and lead characteristics in relation to the requirement of extraction tools. Pacing Clin Electrophysiol 2002; 25: 1037–1040.[Medline]

[4] Smith HJ, Fearnot NE, Byrd CL, Wilkoff BL, Love CJ, Sellers TD. Five-years experience with intravascular lead extraction. U.S. Lead Extraction Database. Pacing Clin Electrophysiol 1994; 17: 2016–2020.[CrossRef][Medline]

[5] Byrd CL, Wilkoff BL, Love CJ, Sellers TD, Turk KT, Reeves R, et al. Intravascular extraction of problematic or infected permanent pacemaker leads:1994–1996. U.S. Extraction Database, MED Institute. Pacing Clin Electrophysiol 1999; 22: 1348–1357.[CrossRef][Medline]

[6] Bongiorni MG, Soldati E, Arena G, Gherarducci G, Ratti M, Gianessi C, et al. Transvenous removal of difficult pacing and ICD leads; a new technique through the internal jugular vein. Pacing Clin Electrophysiol 2000; 23: 696.

[7] Brodman R, Frame R, Andrews C, Furman S. Removal of infected transvenous leads requiring cardiopulmonary bypass or inflow occlusion. J Thorac Cardiovasc Surg 1992; 103: 649–654.[Abstract]

[8] Vogt PR, Sagdic K, Lachat M, Candinas R, von Segesser LK, Turina MI. Surgical management of infected permanent transvenous pacemaker systems: ten year experience. J Card Surg 1996; 11: 180–186.[Web of Science][Medline]

[9] Wilhelm MJ, Schmid C, Hammel D, Kerber S, Loick HM, Herrmann M, et al. Cardiac pacemaker infection: surgical management with and without extracorporeal circulation. Ann Thorac Surg 1997; 64: 1707–1712.[Abstract/Free Full Text]

[10] Byrd CL, Schwartz SJ, Sivina M, Yahr WZ, Greenberg JJ. Technique for the surgical extraction of permanent pacing leads and electrodes. J Thorac Cardiovasc Surg 1985; 89: 142–144.[Abstract]

[11] Klug D, Lacroix D, Savoye C, Goullard L, Grandmougin D, Hennequin JL, et al. Systemic infection related to endocarditis on pacemaker leads: clinical presentation and management. Circulation 1997; 95: 2098–2107.[Abstract/Free Full Text]

[12] Cacoub P, Leprince P, Nataf P, Hausfater P, Dorent R, Wechsler B, et al. Pacemaker infective endocarditis. Am J Cardiol 1998; 82: 480–484.[CrossRef][Web of Science][Medline]

[13] Choo MH, Holmes DR Jr., Gersh BJ, Maloney JD, Merideth J, Pluth JR, et al. Permanent pacemaker infections: characterization and management. Am J Cardiol 1981; 48: 559–564.[CrossRef][Web of Science][Medline]

[14] Lewis AB, Hayes DL, Holmes DRJ, Vlietstra RE, Pluth JR, Osborn MJ. Update on infections involving permanent pacemakers. Characterization and management. J Thorac Cardiovasc Surg 1985; 89: 758–763.[Abstract]

[15] Molina JE. Undertreatment and overtreatment of patients with infected antiarrhythmic implantable devices. Ann Thorac Surg 1997; 63: 504–509.[Abstract/Free Full Text]

[16] Yamada M, Takeuchi S, Shiojiri Y, Maruta K, Oki A, Iyano K, et al. Surgical lead-preserving procedures for pacemaker pocket infection. Ann Thorac Surg 2002; 74: 1494–1499.[Abstract/Free Full Text]

[17] Hurst LN, Evans HB, Windle B, Klein GJ. The salvage of infected cardiac pacemaker pockets using a closed irrigation system. Pacing Clin Electrophysiol 1986; 9: 785–792.[CrossRef][Medline]

[18] Garcia-Rinaldi R, Revuelta JM, Bonnington L, Soltero-Harrington L. The exposed cardiac pacemaker. Treatment by subfascial pocket relocation. J Thorac Cardiovasc Surg 1985; 89: 136–141.[Abstract]

[19] Bracke FA, Meijer A, van Gelder LM. Pacemaker lead complications: when is extraction appropriate and what can we learn from published data? Heart 2001; 85: 254–259.[Free Full Text]

[20] Bohm A, Banyai F, Preda I, Zamolyi K. The treatment of septicemia in pacemaker patients. Pacing Clin Electrophysiol 1996; 19: 1105–1111.[CrossRef][Medline]

[21] Harcombe AA, Newell SA, Ludman PF, Wistow TE, Sharples LD, Schofield PM, et al. Late complications following permanent pacemaker implantation or elective unit replacement. Heart 1998; 80: 240–244.[Abstract/Free Full Text]

[22] Vijayaraman P, Vaidya K, Kim SG, Gross JN, Ferrick KJ, Palma E, et al. Runaway pulse generator malfunction resulting from undetected battery depletion. Pacing Clin Electrophysiol 2002; 25: 220–222.[Medline]

[23] Bohm A, Pinter A, Duray G, Lehoczky D, Dudas G, Tomcsanyi I, et al. Complications due to abandoned noninfected pacemaker leads. Pacing Clin Electrophysiol 2001; 24: 1721–1724.[CrossRef][Medline]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
EuropaceHome page
S. Hamid, A. Arujna, S. Khan, A. Ladwiniec, M. McPhail, J. Bostock, M. Mobb, N. Patel, C. Bucknall, and C. A. Rinaldi
Extraction of chronic pacemaker and defibrillator leads from the coronary sinus: laser infrequently used but required
Europace, February 1, 2009; 11(2): 213 - 215.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
R. G. Hauser
The Growing Mismatch Between Patient Longevity and the Service Life of Implantable Cardioverter-Defibrillators
J. Am. Coll. Cardiol., June 21, 2005; 45(12): 2022 - 2025.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (13)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Bracke, F.A.L.E.
Right arrow Articles by van Gelder, L.M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bracke, F.A.L.E.
Right arrow Articles by van Gelder, L.M.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?