Europace 2003 5(4):361-363; doi:10.1016/S1099-5129(03)00093-X
© 2003 by European Society of Cardiology
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (4)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Srivathsan, K.
Right arrow Articles by Scott, L. R. P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Srivathsan, K.
Right arrow Articles by Scott, L. R. P.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?


CASE REPORT

Pneumopericardium and pneumothorax contralateral to venous access site after permanent pacemaker implantation

K. Srivathsan, R. A. Byrne, C. P. Appleton and L. R. P. Scott

Division of Cardiovascular Diseases, Mayo Clinic Scottsdale Phoenix AZ, USA

Manuscript submitted 18 February 2003. Accepted after revision 22 June 2003.

Correspondence: Luis R. P. Scott, MD, Mayo Clinic Hospital, 5777 East Mayo Boulevard, Phoenix, AZ 85054, U.S.A. Tel.: +1-480-342-1398; Fax: +1-480-342-1388. E-mail: scott.luis{at}mayo.edu


    Abstract
 Top
 Abstract
 Case report
 Discussion
 References
 
A 77-year-old female underwent implantation of a left-sided dual chamber permanent pacemaker for symptomatic bradycardia with active fixation leads. Eight hours after the procedure, the patient complained of shortness of breath and was found to have a 30% right pneumothorax on chest X-ray. Immediately, a chest tube was inserted, promptly relieving the symptoms. A CT scan of the chest revealed extrusion of the helix of the screw-in atrial lead, through the wall of the right atrial appendage. The helix was abutting a bulla in the right lung, the likely cause for pneumothorax and pneumopericardium. The atrial lead was explanted without incident.

Key Words: Contralateral pneumothorax, dual chamber permanent pacemaker, pacemaker insertion complications


    Case report
 Top
 Abstract
 Case report
 Discussion
 References
 
A 77-year-old female with chronic obstructive lung disease and long-standing right bundle branch block was hospitalized for chest pain. A cardiac catheterization was performed and showed no haemodynamically significant epicardial coronary artery disease. During the hospitalization while on telemetry, the patient experienced an episode of loss of consciousness associated with 4.5-s sinus pause. At that time, she was not on negative chronotropic drugs, and no other precipitating factors were identified. A permanent pacemaker implantation was recommended.

The patient underwent a dual chamber permanent pacemaker (Medtronic Kappa, KDR 901-Medtronic, MN, U.S.A.) implantation using left subclavian venous access. Active fixation leads (Medtronic active fixation, Model 5076, Medtronic, MN, U.S.A.) were used for both atrial and ventricular pacing. Atrial lead pacing threshold was 1 V at 0.5 ms and pacing impedance 544 {Omega}. Sensed P wave was 2 mV. Ventricular lead pacing threshold was 0.5 V at 0.5 ms, and pacing impedance 1030 {Omega}. Sensed R wave was 5 mV. The initial chest X-ray after implantation showed adequate lead position and no evidence of pneumothorax.

Eight hours later, the patient complained of sudden onset of shortness of breath. A chest X-ray revealed a right-sided (contralateral) pneumothorax (Fig. 1). A chest tube was inserted, promptly relieving the symptoms. Pacemaker interrogation revealed no change in lead data and a CT scan of the chest (Fig. 2) was done to investigate the aetiology of the contralateral pneumothorax. This revealed extrusion of the helix of the screw-in atrial lead, through the wall of the right atrial appendage, causing a 30% pneumothorax. The helix of the atrial lead was adjacent to an intact bulla of the right lung. Pneumopericardium was also seen. The patient was taken to the electrophysiology laboratory for atrial lead extraction. The atrial lead was uneventfully explanted under continuous intracardiac echocardiography and haemodynamic monitoring with a pulmonary artery catheter. The pneumopericardium resolved within a few days and the right pneumothorax also sealed without complication.



View larger version (127K):
[in this window]
[in a new window]
 
Figure 1 Chest X-ray showing right pneumothorax (arrow) 8 h after placement of a left-sided permanent dual chamber pacemaker for symptomatic bradycardia.

 



View larger version (69K):
[in this window]
[in a new window]
 
Figure 2 CT scan of the chest showing extrusion of the helix of the atrial lead through the right atrial appendage (arrowhead) causing pneumopericardium (arrow). Right pleural effusion (chevron) is also seen.

 

    Discussion
 Top
 Abstract
 Case report
 Discussion
 References
 
Pneumothorax is usually a complication of subclavian venous access[1]Go. Perforation of the right ventricle with or without pericardial effusion is also well recognized[2]Go. Pneumopericardium has been reported from temporary ventricular endocardial lead perforation[3]Go. Pneumothorax, contralateral to venous access site, due to atrial lead perforation is a rare complication[4]Go. Concomitant pneumothorax (contralateral to venous access site) and pneumopericardium secondary to perforation through the wall of the right atrial appendage has to our knowledge not been previously reported.

The benefit on quality of life associated with dual chamber pacing as compared with ventricular pacing for sinus node dysfunction is well documented[5]Go. Physiological pacing may not offer benefit over ventricular pacing for the prevention of stroke or death due to cardiovascular causes[6]Go. Dual chamber systems are widely used in the United States in patients with sinus node disease. The rate of implant related complications between dual and single chamber pacemaker implantation has been reported as either equivalent[7]Go or higher for dual chamber pacing[8]Go.

An atrial lead is essential for dual chamber pacing but dislodgment of this lead is not infrequent[9]Go. In order to reduce the dislodgment rate, active fixation (screw-in) leads have been developed. However, the screw-in atrial leads increase the chance of perforating the thin-walled atrial appendage. Active fixation, polyurethane insulated atrial leads (Accufix) are associated with 2.4% acute lead related complications, (i.e. perforation, microdislodgment and pericarditis)[10]Go. Independent cases of atrial perforation by a screw-in electrode have also been reported[11]Go.

Several risk factors may be responsible for the increased complication rate of screw-in atrial leads. Variation in patient anatomy such as an extremely thin-walled or multilobed atrial appendage may play a role. Lead factors such as design and stiffness of the helix may differ between manufacturers and could be important. The experience of the operator is equally important to avoid over-screwing during atrial lead fixation.

Usually the lead parameters, in particular the pacing threshold, will show a significant change on lead perforation[12]Go. The helix may offer mechanical support by anchoring the lead but may not be part of the electrode. Thus the lead parameters may not change in some cases after perforation[4]Go. In this case, the helix was part of the electrode (active helix), but the lead parameters did not change. The reason for the lack of change in lead parameters is unclear. However, a large part of the electrode may have been in contact with the atrial myocardium resulting in lack of change in the lead parameters.

Pneumothorax secondary to subclavian venous access is uncommon and occurs between 1 and 2% with experienced operators [1,Go13]Go. Pneumothorax may be detected during the procedure or delayed until 48 h after implantation. Pneumothorax secondary to perforation of the atrial lead through the wall of the atrial appendage has been reported once previously[4]Go. Concomitant pneumothorax and pneumopericardium as happened in this case has not been previously reported.

This patient developed pneumopericardium, pleural effusion but no discernible pericardial effusion. The lead may have plugged the small defect, although no pericardial effusion developed on explantation of the lead. The presence of a bulla abutting the lead helix as seen on the CT scan suggests that bullous lung disease likely played a role in the development of this complication.

After lead extraction, implantation of a new atrial lead was deferred to allow time for the atrial appendage to heal.

This case illustrates that if a right pneumothorax develops in a patient where the pacemaker was placed from a left subclavian approach, perforation of the atrial appendage should be considered and immediately investigated. In haemodynamically stable patients, CT scan of the chest is the investigation of choice. In unstable patients emergency echocardiography may be useful to identify pericardial effusion. Extra post-procedural vigilance should be considered when implanting atrial leads in patients with bullous emphysema.


    References
 Top
 Abstract
 Case report
 Discussion
 References
 
[1] Aggarwal RK, Connelly DT, Ray SG, Ball J, Charles RG. Early complications of permanent pacemaker implantation: no difference between dual and single chamber systems. Br Heart J 1995; 73: 571–575.[Abstract/Free Full Text]

[2] Gondi B and Nanda NC. Real-time, two-dimensional echocardiographic features of pacemaker perforation. Circulation 1981; 64: 97–106.[Abstract/Free Full Text]

[3] Deanfield J, Jonathan A, Fox K. Pericardial complications of endocardial and epicardial pacing. Br Med J (Clin Res Ed) 1981; 283: 635–636.

[4] Ho WJ, Kuo CT, Lin KH. Right pneumothorax resulting from an endocardial screw-in atrial lead. Chest 1999; 116: 1133–1134.[Abstract/Free Full Text]

[5] Lamas GA, Orav EJ, Stambler BS, et al. Quality of life and clinical outcomes in elderly patients treated with ventricular pacing as compared with dual-chamber pacing. Pacemaker Selection in the Elderly Investigators. N Engl J Med 1998; 338: 1097–1104.[Abstract/Free Full Text]

[6] Connolly SJ, Kerr CR, Gent M, et al. Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes. Canadian Trial of Physiologic Pacing Investigators. N Engl J Med 2000; 342: 1385–1391.[Abstract/Free Full Text]

[7] Mueller X, Sadeghi H, Kappenberger L. Complications after single versus dual chamber pacemaker implantation. Pacing Clin Electrophysiol 1990; 13: 711–714.[CrossRef][Medline]

[8] Chauhan A, Grace AA, Newell SA, et al. Early complications after dual chamber versus single chamber pacemaker implantation. Pacing Clin Electrophysiol 1994; 17: 2012–2015.[CrossRef][Medline]

[9] Markewitz A, Wenke K, Weinhold C. Reliability of atrial screw-in leads. Pacing Clin Electrophysiol 1988; 11: 1777–1783.[Medline]

[10] Glikson M, von Feldt LK, Suman VJ, Hayes DL. Clinical surveillance of an active fixation, bipolar, polyurethane insulated pacing lead, Part I: The atrial lead. Pacing Clin Electrophysiol 1994; 17: 1399–1404.[CrossRef][Medline]

[11] Van Nooten G, Verbeet T, Deuvaert FE. Atrial perforation by a screw-in electrode via a left superior vena cava. Am Heart J 1990; 119: 1439–1440.[Medline]

[12] Ellenbogen KA, Wood MA, Shepard RK. Delayed complications following pacemaker implantation. Pacing Clin Electrophysiol 2002; 25: 1155–1158.[CrossRef][Medline]

[13] Link MS, Estes NA 3rd, Griffin JJ, et al. Complications of dual chamber pacemaker implantation in the elderly. Pacemaker Selection in the Elderly (PASE) Investigators. J Interv Card Electrophysiol 1998; 2: 175–179.[CrossRef][Web of Science][Medline]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
CMAJHome page
A. Sestito, G. A. Sgueglia, F. Infusino, P. Zecchi, F. Crea, and G. A. Lanza
A 60-year-old man with chest pain following pacemaker implantation
Can. Med. Assoc. J., March 29, 2005; 172(7): 874 - 874.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (4)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Srivathsan, K.
Right arrow Articles by Scott, L. R. P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Srivathsan, K.
Right arrow Articles by Scott, L. R. P.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?