Europace Advance Access originally published online on November 24, 2008
Europace 2009 11(1):42-47; doi:10.1093/europace/eun308
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
11/1/42    most recent
eun308v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in Europace
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (2)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Kirchhof, P.
Right arrow Articles by Maintz, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kirchhof, P.
Right arrow Articles by Maintz, D.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org


Ablation: Imaging

Diastolic isthmus length and ‘vertical’ isthmus angulation identify patients with difficult catheter ablation of typical atrial flutter: a pre-procedural MRI study

Paulus Kirchhof1,2,*, Murat Özgün3, Stephan Zellerhoff1, Gerold Mönnig1, Lars Eckardt1, Kristina Wasmer1, Walter Heindel3, Günter Breithardt1,2 and David Maintz3

1 Department of Cardiology and Angiology, Hospital of the University of Münster, Münster, Germany; 2 German Atrial Fibrillation Competence Network (AFNET), Münster, Germany; 3 Institute for Clinical Radiology, Hospital of the University of Münster, Münster, Germany

Manuscript submitted 15 August 2008. Accepted after revision 21 October 2008.

* Corresponding author: Medizinische Klinik und Poliklinik C, Kardiologie und Angiologie, Universitätsklinikum Münster, Albert-Schweitzer-Straße 33, D-48129 Münster, Germany. Tel: +49 251 8345185; fax: +49 251 8347864. E-mail address: kirchhp{at}uni-muenster.de


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Funding
 References
 
Aims: Catheter ablation of isthmus-dependent atrial flutter is technically demanding in some patients and extremely simple in others. The intervention targets a defined anatomical structure, the so-called cavotricuspid isthmus (CTI). We sought to characterize CTI anatomy in vivo in patients with difficult and simple catheter ablation of atrial flutter.

Methods and results: Twenty-six patients were studied. Seven patients with difficult (n = 6) or extremely simple (n = 1) CTI ablation procedures were retrospectively selected from our catheter ablation database. Thereafter, we prospectively studied 19 patients undergoing CTI ablation in our department. We visualized CTI anatomy by ECG- and respiration-gated free precession 1.5 T cardiac magnetic resonance imaging (MRI). Magnetic resonance imaging was analysed for systolic and diastolic CTI length, the angle between the vena cava inferior and CTI, and pouch-like recesses. These parameters were compared between patients with difficult and simple procedures, split by the median number of energy applications. Patients with difficult procedures had a longer diastolic CTI length (diastolic isthmus length 20.3 ± 1.8 mm) than those with simple procedures (diastolic isthmus length 16.6 ± 1.7 mm, all data as mean ± SEM, P < 0.05). Cavotricuspid isthmus angulation with respect to inferior vena cava was closer to 90° in patients with difficult procedures (deviation from 90°: 15 ± 2°) than those with simple procedures (deviation 23 ± 4°, P < 0.05). Systolic CTI length was not different between groups (32 ± 2 mm in both groups, P > 0.2).

Conclusion: Longer diastolic, but not systolic, CTI length and a rectangular angle between CTI and inferior vena cava render CTI catheter ablation difficult. Visualization of isthmus anatomy may help to guide difficult CTI ablation procedures.

Key Words: Catheter ablation, Magnetic resonance imaging, Atrial flutter


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Funding
 References
 
Shortly after the description of its electrophysiological mechanism macro-reentry around the tricuspid valve,1Go–3Go catheter ablation emerged as the treatment of choice for typical, isthmus-dependent atrial flutter.4Go,5Go The ablation procedure creates a linear scar in the myocardium of the cavotricuspid isthmus (CTI), a relatively thin muscular strand that is a critical part of the flutter re-entrant circuit. Catheter ablation in the CTI can cure 90% of the patients with isthmus-dependent atrial flutter,6Go–9Go and re-ablation is often successful in the remaining 10% of the patients. Larger or irrigated-tip radio frequency ablation catheters and the use of three-dimensional catheter localization systems have reduced procedure and fluoroscopy times. The use of an irrigated-tip ablation catheter has increased procedural success rate in a trial,10Go but, in general, larger tip and irrigated tip ablation catheters or the use of cryo-energy rather than radio frequency energy does not markedly improve CTI ablation success.10Go–16Go In contrast to the intuitive expectation of interventional electrophysiologists, the use of mapping systems did not increase CTI ablation success rates in a controlled trial.16Go These observations may be interpreted as a sign that more precise positioning of the ablation catheter or higher energy delivery cannot solve all difficulties in technically demanding CTI ablation procedures.

As isthmus ablation targets a well-defined anatomical structure, a specific anatomy of the CTI may be a cause for difficult or non-successful CTI ablation.17Go–19Go Evidence for such anatomical characteristics has been reported by angiographic studies,17Go,19Go,20Go multi-slice computed tomography (MSCT),21Go and one cardiac magnetic resonance imaging (MRI) study.22Go A longer CTI, usually measured during mechanical diastole, and possibly pouches in the CTI region,19Go,22Go may be related to difficult ablation procedures. Magnetic resonance imaging has the potential to visualize cardiac anatomic structures in vivo22Go,23Go and may, in the future, be used to guide ablation procedures.24Go We therefore performed an MRI study of CTI anatomy prior to CTI ablation to characterize anatomic details in ‘difficult’ CTI ablations.


    Methods
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Funding
 References
 
Patients were eligible if they underwent catheter ablation in the CTI for curative treatment of isthmus-dependent atrial flutter. The first seven patients were identified by a retrospective database search of all patients who underwent catheter ablation for atrial flutter at the Department of Cardiology of the University Hospital Münster from 1999 to 2004 (more than 200 patients). In this database, we identified six eligible consenting patients with difficult procedures, i.e. patients with either unsuccessful ablation attempts (n = 2) or those receiving more than 15 point-by-point energy applications to achieve complete CTI block (n = 4), and one patient with a simple procedure. All patients were contacted through telephone and asked whether they were willing to undergo MRI of the isthmus. In a second phase of the study, we prospectively included patients who underwent catheter ablation of atrial flutter in our department from April 2004 to March 2005. The study complied with the standards for research in patients as defined by the WHO and was approved by the local Ethics Committee. All patients gave written informed consent for the MRI study.

Cavotricuspid annulus ablation was performed following standard procedures from a femoral access using a biplane fluoroscopy system. Point-by-point ablation (90 s ablation duration per catheter position) was performed in a temperature-controlled mode (60°C maximal temperature), with a power setting of 70 W (50 W for 4 mm tip electrodes). Complete bidirectional CTI block was assessed at the end of the ablation procedure during pacing in the coronary sinus ostium and low lateral right atrium by a combination of changes in the right atrial activation sequence measured by a multipolar ‘Halo’ catheter, an increase in trans-isthmus conduction times, and double-potential mapping at the CTI.25Go These were measured after the last energy application and after a 30 min waiting period to assess recovery of isthmus conduction. When isthmus conduction recovered, the ablation was continued. The use of a mapping system and the choice of the ablation catheter were driven by the preference of the interventional electrophysiologist in charge of the procedure. All ablation catheters used were available in different curves as supplied by the manufacturer or were equipped with an adjustable curve size. The MRI anatomic information, which was analysed offline, was not used during ablation.

Patients were studied using a 1.5 T magnetic resonance tomography system (Philips® Aptiva, Philips Medical Systems, Hamburg, Germany) by steady-state respiration- and ECG-gated free precession sequences, similar to published protocols for MRI-based image fusion.26Go Scout sequences were used to identify the inferior right atrium and the inferior vena cava. The CTI was visualized in the right and left anterior oblique projections, parallel and perpendicular to the orientation of the interventricular septum. The length and width of cavotricuspid isthmus were measured in diastole and systole as the linear distance between the tricuspid annulus and the end of the right atrial myocardial sleeve. Care was taken to identify the exact extensions of the right atrial myocardium in the MRI for isthmus length determination (Figure 1). On the basis of the unproven hypothesis that a rectangular orientation of the CTI with respect to the most proximal part of the inferior vena cava might prevent adequate tissue contact between the catheter tip and the CTI during the ablation procedure,19Go the angle between the inferior vena cava and the CTI was estimated by the angle between two tangential lines placed parallel to the CTI floor and parallel to the adjacent wall of the inferior vena cava (Figure 1). All MRI analyses were pre-defined and performed by two observers (P.K. and D.M.) blinded to the study group. The complete data set was analysed twice. The first analysis was used to train the analysers. After a waiting period of 6 months, all MRI images were re-analysed by the same observers. When a discrepancy between the initial analysis and the second analysis was detected, the MRI images were jointly analysed for a third time by the two observers to resolve inconsistencies.


Figure 1
View larger version (188K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 1 Two representative magnetic resonance images taken from a patient with a simple (left panel) and difficult (right panel) CTI ablation procedure. Images depict an approximate equivalent of an approximate 30° RAO projection adjusted to give a perspective view perpendicular to the isthmus length. Images taken during mechanical systole (upper panels) and diastole (lower panels) are shown.

 
Patients were divided into two groups based on the number of point-by-point ablation sites required to achieve complete bidirectional isthmus block (difficult procedures: patients with higher-than-median ablation sites, unsuccessful ablation, or one patient in whom two different energy sources were used to achieve CTI block and simple procedures: patients with lower-than-median ablation sites).


    Results
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Funding
 References
 
Clinical characteristics of the patients are given in Table 1. Blood pressure was not different between patient groups (simple procedures: 129 ± 3 mmHg, difficult procedures: 118 ± 4 mmHg). Coronary artery disease was present in one patient with a simple procedure and in four patients with a difficult procedure (P = ns). Radio frequency ablation was used in 21 patients (10 patients with 8 mm tip catheters and three patients with cooled-tip catheters) and cryo-ablation in six patients. In one patient, both a 4 mm tip radio frequency ablation catheter and subsequently a cryo-ablation catheter were used. Mapping systems and non-fluoroscopic catheter localization systems (one using CARTO®, 10 using LocaLisa®, and two using NavX®) were used in 13 patients to guide the procedure.The mean procedure duration was 195 ± 19 min, and the mean fluoroscopy time was 29 ± 4 min. Usually, fluoroscopy was used to monitor catheter stability during energy application. The clinical and procedural characteristics were not different between the patient groups with the exception of number of energy applications, fluoroscopy duration, and procedure duration, which were longer in patients with difficult procedures (Table 1).


View this table:
[in this window]
[in a new window]

 
Table 1 Clinical and procedural data of the study patients

 
Isthmus MRI could be analysed in 24 of 26 patients (Figure 1). The myocardium usually extended from the tricuspid valve to the Eustachian ridge, but not beyond it. Difficult procedures were associated with a longer diastolic CTI (20.3 ± 1.8 mm, range 7–28, Figure 2A) than simple procedures (CTI length 16.6 ± 1.7 mm, range 9–25, P < 0.05). The angle between inferior vena cava and right atrial floor deviated less from 90° in patients with difficult procedures (15 ± 2°, range 3–30, Figure 2B) than in those with simple procedures (23 ± 4°, range 7–40, P < 0.05). Differences in CTI angulation were caused by either angulation of the proximal part of the inferior vena cava and/or angulation of the CTI within the thorax. Systolic CTI length was markedly shorter than diastolic CTI length in all patients (mean difference 13 ± 1 mm, or 72% of diastolic isthmus length, range 6–25), indicating movement of the CTI. The range of isthmus movement (systolic minus diastolic CTI length) was not different between the groups (simple procedures 15 ± 2 mm, range 8–25 and difficult procedures 12 ± 1 mm, range 6–20). Systolic CTI length was also not different between groups (simple procedures 32 ± 2 mm, range 24–46 and difficult procedures 32 ± 2 mm, range 17–42, P > 0.2). Three patients with incomplete CTI block after the ablation procedure (unsuccessful procedure) also had a long diastolic CTI (22 ± 2 mm, range 19–28) and a rectangular CTI orientation (17 ± 6° angle deviation, range 8–30).


Figure 2
View larger version (11K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 2 (A) Mean diastolic isthmus length in patients with simple (open column) and difficult (filled column) ablation procedures. (B) Mean deviation from 90° of the angle between inferior vena cava and isthmus for simple (open columns) and difficult (filled columns) procedures. All values shown as mean and SEM.

 
Pouch-like recesses in the isthmus have been suspected to render CTI ablation difficult.17Go Usually, the curve of the endocardial surface of the CTI was smooth. Pouch-like recesses were found in three patients with difficult procedures in this study and in three patients with simple procedures (P = ns). Pouches were 3–8 mm deep. The deepest pouch was found in a patient with a simple procedure. Although not part of the pre-specified analyses, we observed that the diastolic movement of the CTI amplified pouch formation during diastole. In extreme cases, the pouch even ‘closed’ to form a secluded area in the CTI by intermittent contact between the Eustachian ridge and the tricuspid annulus. These phenomena were only occasionally observed and not different between study groups.


    Discussion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Funding
 References
 
Main findings
Cardiac MRI can non-invasively visualize CTI anatomy in vivo. We identified several anatomic characteristics associated with difficult CTI ablation procedures:

  1. An angulation of the CTI and the inferior vena cava close to 90° render ablation difficult. This finding suggests that catheter–tissue contact may be difficult in such situations.
  2. The CTI is a moving target that alters its length by 30–50% during the cardiac cycle.
  3. A long diastolic CTI length is associated with difficult ablation procedures. This finding is consistent with prior publications. The (longer) systolic isthmus length was not different between patient groups in this study.
  4. According to MRI criteria, the right atrial myocardium extends approximately to the Eustachian valve, but not further into the vena cava. This relates well to the observation of complete CTI block prior to the ablation of an angiographically complete CTI line.19Go Right atrial angiography may slightly overestimate CTI length.

Magnetic resonance imaging of the cavotricuspid isthmus
Magnetic resonance imaging was feasible using a clinically available 1.5 T MRI system in almost all (92%) study patients. The procedure is non-invasive, was well tolerated, and, in contrast to angiography19Go,20Go or contrast-enhanced computed tomography,21Go does not require exposure to ionizing radiation or potentially nephrotoxic contrast media (~30–50 mL for angiography and ~100 mL for MSCT19Go–21Go). The technique requires validation in larger patient cohorts and in clinical routine, but the integration of MRI into catheter localization and mapping systems26Go,27Go opens the perspective to use real pre-procedural anatomic information to guide CTI ablation in the near future, e.g. for repeat ablations.

Anatomical characteristics of the cavotricuspid isthmus in difficult procedures
Isthmus length and angulation could be readily measured in the study population (Figure 1). Diastolic isthmus length was within the range of published measurements by other techniques.19Go–21Go Interestingly, the systolic (longer) isthmus length did not differ between simple and difficult procedures in this study. Prior publications do not specifically mention the timing of the measurement or only reported measurements for diastolic isthmus length.17Go,19Go,20Go As the catheter was usually kept stable during energy application in this study (in concordance with Da Costa et al.17Go but in contrast to Heidbuchel et al.19Go), this implies that the myocardium under the catheter will move during energy delivery. Our data can be interpreted as a sign that diastolic energy delivery may be more relevant for CTI ablation.

Magnetic resonance imaging allows to delineate CTI extension based on the limits of atrial myocardium in addition to delineation based on the endocardial shape, albeit within the limitations of the spatial resolution of a 1.5 T system. In some patients, we found that the atrial myocardial layer ended before the Eustachian ridge, resulting in a shorter length of the isthmus when measured as the length of atrial myocardium. This may explain that the isthmus lengths measured in this study are slightly shorter than in angiographic studies, which used endocardial border detection (and usually the Eustachian ridge) to delineate the isthmus limit.19Go,20Go Interestingly, one study reported that complete isthmus block occurred before ablation along the entire angiographically determined isthmus length,19Go suggestive of an overestimation of isthmus length by angiography, and consistent with the CTI length measurements in this study.

Prior angiographic studies suggested that pouch-like recesses within the isthmus were associated with difficult procedures,17Go,19Go,22Go although this was not quantified. We could not confirm this suggestion in our study (three patients with pouches in each patient group, Table 1). The pouch depth was also not different between patients with simple and difficult procedures. This difference between our study and prior reports may have several reasons: our patient population, albeit not homogeneously, was enriched with patients who had difficult ablation procedures (n = 6 of the retrospective part of our cohort). Nonetheless, pouches were relatively rare in our study group, suggesting that other factors rendered ablation difficult. Furthermore, we cannot exclude that differences in the MRI technology influenced detection of pouch-like recesses. Occasionally, the cine-mode MRI gave the impression that pouches were ‘closed’ during diastole. This preliminary observation may warrant further evaluation.

Angulation of the isthmus with respect to the inferior vena cava
A rectangular orientation of the CTI with respect to the inferior vena cava was associated with difficult procedures in this study. Only one angiographic study reported that measurements of CTI angulation were performed,19Go but there are no results for simple and difficult procedures in that paper. This is the first study that identified a rectangular orientation of the CTI with respect to the proximal part of the inferior vena cava as a factor that renders isthmus ablation difficult. Hypothetically, a rectangular angle between the inferior vena cava and the isthmus area could hinder adequate contact of the tip of the steerable catheter to parts of the myocardial sleeves within the isthmus, especially in the caval part of the CTI. It is also conceivable that a rectangular position of the CTI could facilitate intermittent loss of contact of the catheter tip to the CTI during the cardiac cycle. These factors could impede completion of the ablation line. Ideally, this hypothesis could be tested in a system that allows MRI during the ablation procedure, whereas the catheter is in contact with the CTI.28Go

Limitations
This study is hypothesis-generating, and larger, prospective, multi-centre series are required to confirm the observations reported and to determine the clinical usefulness of CTI visualization prior to catheter ablation. The patient population is heterogeneous, partially prospectively and partially retrospectively selected, and relatively small. The majority of patients in this study underwent MRI during sinus rhythm. As MRI, comparable with other ‘signal-averaged’ imaging modalities such as computed tomography or positron emission tomography, relies on ECG and respiration triggering, MRI during atrial fibrillation or flutter may be more difficult to interpret. The present study used a 1.5 T MRI system available in clinical routine at many centres. The spatial resolution of this system has prevented reliable measurements of myocardial thickness in this study, and thin or isolated myocardial sleeves may extend beyond the myocardial boundaries identified in this study (i.e. beyond the Eustachian ridge). Finally, the MRI system and sequences used in this study did not allow to delineate small coronary vessels that could at times be associated with failure of isthmus ablation18Go and may have limited our ability to identify pouch-like recesses. Future studies, possibly using MRI systems with higher spatial resolution, may help better delineate these details.

Conflict of interest: C.K. and S.P. have received travel grants and honoraria from Siemens Medical Systems. P.S. served on advisory boards for Siemens. P.K. has received honoraria or travel grants from or served on advisory boards for Medtronic, St. Jude Medical, Siemens, and various pharmaceutical companies. P.S. and P.K. have been involved in research on image integration which was partially supported by TomTec Imaging systems. D.M. has received honoraria from Philips Medical Systems and from Siemens.


    Funding
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Funding
 References
 
This study is supported by the German Federal Ministry of Education and Research (BMBF) through the Atrial Fibrillation Competence NETwork (AFNET, projects B7, B11 and C1, grant no. 01Gi0204).


    References
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Funding
 References
 
[1] Olshansky B, Okumura K, Hess PG, Waldo AL. Demonstration of an area of slow conduction in human atrial flutter. J Am Coll Cardiol (1990) 16:1639–48.[Abstract]

[2] Klein G, Guiraudon G, Sharma A, Milstein S. Demonstration of macro-reentry and feasibility of operative therapy in the common type of atrial flutter. Am J Cardiol (1986) 57:587–91.[CrossRef][Web of Science][Medline]

[3] Cosio F, Arribas F, Barbero JE, et al. Validation of double-spike electrograms as markers of conduction delay or block in atrial flutter. J Am Coll Cardiol (1988) 61:775–80.

[4] Feld GK, Fleck RP, Chen PS, Boyce K, Bahnson TD, Stein JB, et al. Radiofrequency catheter ablation for the treatment of human type 1 atrial flutter. Identification of a critical zone in the reentrant circuit by endocardial mapping techniques. Circulation (1992) 86:1233–40.[Abstract/Free Full Text]

[5] Saoudi N, Atallah G, Kirkorian G, Touboul P. Catheter ablation of the atrial myocardium in human type I atrial flutter. Circulation (1990) 81:762–71.[Abstract/Free Full Text]

[6] Lesh MD. Catheter ablation of atrial flutter and tachycardia. In: Cardiac Electrophysiology—from Cell to Bedside—Zipes DP, Jalife J, eds. (2000) Vol. 1. Philadelphia: W.B. Saunders. 1009–27.

[7] Natale A, Newby KH, Pisano E, Leonelli F, Fanelli R, Potenza D, et al. Prospective randomized comparison of antiarrhythmic therapy versus first-line radiofrequency ablation in patients with atrial flutter. J Am Coll Cardiol (2000) 35:1898–904.[Abstract/Free Full Text]

[8] Anselme F, Saoudi N, Poty H, Douillet R, Cribier A. Radiofrequency catheter ablation of common atrial flutter: significance of palpitations and quality-of-life evaluation in patients with proven isthmus block. Circulation (1999) 99:534–40.[Abstract/Free Full Text]

[9] Fischer B, Jais P, Shah D, Chouairi S, Haissaguerre M, Garrigues S, et al. Radiofrequency catheter ablation of common atrial flutter in 200 patients. J Cardiovasc Electrophysiol (1996) 7:1225–33.[Web of Science][Medline]

[10] Jais P, Shah DC, Haissaguerre M, Hocini M, Garrigue S, Le Metayer P, et al. Prospective randomized comparison of irrigated-tip versus conventional-tip catheters for ablation of common flutter. Circulation (2000) 101:772–6.[Abstract/Free Full Text]

[11] Kottkamp H, Hugl B, Krauss B, Wetzel U, Fleck A, Schuler G, et al. Electromagnetic versus fluoroscopic mapping of the inferior isthmus for ablation of typical atrial flutter: a prospective randomized study. Circulation (2000) 102:2082–6.[Abstract/Free Full Text]

[12] Da Costa A, Cucherat M, Pichon N, Messier M, Laporte S, Romeyer-Bouchard C, et al. Comparison of the efficacy of cooled-tip and 8-mm-tip catheters for radiofrequency catheter ablation of the cavotricuspid isthmus: a meta-analysis. Pacing Clin Electrophysiol (2005) 28:1081–7.[CrossRef][Medline]

[13] Montenero AS, Bruno N, Zumbo F, Antonelli A, Fiocca L, Barbieri L, et al. Cryothermal ablation treatment of atrial flutter—experience with a new 9 French 8 mm tip catheter. J Interv Card Electrophysiol (2005) 12:45–54.[Medline]

[14] Passman RS, Kadish AH, Dibs SR, Engelstein ED, Goldberger JJ. Radiofrequency ablation of atrial flutter: a randomized controlled study of two anatomic approaches. Pacing Clin Electrophysiol (2004) 27:83–8.[Medline]

[15] Schneider MA, Ndrepepa G, Dobran I, Schreieck J, Weber S, Plewan A, et al. LocaLisa catheter navigation reduces fluoroscopy time and dosage in ablation of atrial flutter: a prospective randomized study. J Cardiovasc Electrophysiol (2003) 14:587–90.[CrossRef][Web of Science][Medline]

[16] Earley MJ, Showkathali R, Alzetani M, Kistler PM, Gupta D, Abrams DJ, et al. Radiofrequency ablation of arrhythmias guided by non-fluoroscopic catheter location: a prospective randomized trial. Eur Heart J (2006) 27:1223–9.[Abstract/Free Full Text]

[17] Da Costa A, Faure E, Thevenin J, Messier M, Bernard S, Abdel K, et al. Effect of isthmus anatomy and ablation catheter on radiofrequency catheter ablation of the cavotricuspid isthmus. Circulation (2004) 110:1030–5.[Abstract/Free Full Text]

[18] Igawa O, Adachi M, Hisatome I, Matsui Y. Histopathologic background for resistance to conventional catheter ablation of common atrial flutter. J Cardiovasc Electrophysiol (2004) 15:829–32.[CrossRef][Web of Science][Medline]

[19] Heidbuchel H, Willems R, van Rensburg H, Adams J, Ector H, Van de Werf F. Right atrial angiographic evaluation of the posterior isthmus: relevance for ablation of typical atrial flutter. Circulation (2000) 101:2178–84.[Abstract/Free Full Text]

[20] Cabrera JA, Sanchez-Quintana D, Ho SY, Medina A, Wanguemert F, Gross E, et al. Angiographic anatomy of the inferior right atrial isthmus in patients with and without history of common atrial flutter. Circulation (1999) 99:3017–23.[Abstract/Free Full Text]

[21] Komatsu S, Okuyama Y, Omori Y, Oka T, Mizuno H, Honda T, et al. Evaluation of the cavotricuspid isthmus and right atrium by multidetector-row computed tomography in patients with common atrial flutter. Heart Vessels (2005) 20:264–70.[Medline]

[22] Lim KT, Murray C, Liu H, Weerasooriya R. Pre-ablation magnetic resonance imaging of the cavotricuspid isthmus. Europace (2007) 9:149–53.[Abstract/Free Full Text]

[23] Tops LF, van der Wall EE, Schalij MJ, Bax JJ. Multi-modality imaging to assess left atrial size, anatomy and function. Heart (2007) 93:1461–70.[Abstract/Free Full Text]

[24] Nazarian S, Kolandaivelu A, Zviman MM, Meininger GR, Kato R, Susil RC, et al. Feasibility of real-time magnetic resonance imaging for catheter guidance in electrophysiology studies. Circulation (2008) 118:223–9.[Abstract/Free Full Text]

[25] Johna R, Eckardt L, Fetsch T, Breithardt G, Borggrefe M. A new algorithm to determine complete isthmus conduction block after radiofrequency catheter ablation for typical atrial flutter. Am J Cardiol (1999) 83:1666–8. A1666–7.[Medline]

[26] Ector J, De Buck S, Adams J, Dymarkowski S, Bogaert J, Maes F, et al. Cardiac three-dimensional magnetic resonance imaging and fluoroscopy merging: a new approach for electroanatomic mapping to assist catheter ablation. Circulation (2005) 112:3769–76.[Abstract/Free Full Text]

[27] Knackstedt C, Schauerte P, Kirchhof P. Electro-anatomic mapping systems in arrhythmias. Europace (2008) 10:iii28–iii34.[Abstract/Free Full Text]

[28] Dickfeld T, Calkins H, Zviman M, Kato R, Meininger G, Lickfett L, et al. Anatomic stereotactic catheter ablation on three-dimensional magnetic resonance images in real time. Circulation (2003) 108:2407–13.[Abstract/Free Full Text]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?

Related articles in Europace:

How anatomy can guide ablation in isthmic atrial flutter
José-Angel Cabrera, Siew Yen Ho, and Damián Sánchez-Quintana
Europace 2009 11: 4-6. [Full Text]  



This article has been cited by other articles:


Home page
Circ Arrhythm ElectrophysiolHome page
D. Vollmann, L. Luthje, J. Seegers, G. Hasenfuss, and M. Zabel
Remote Magnetic Catheter Navigation for Cavotricuspid Isthmus Ablation in Patients With Common-Type Atrial Flutter
Circ Arrhythm Electrophysiol, December 1, 2009; 2(6): 603 - 610.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
P. Kirchhof, J. Bax, C. Blomstrom-Lundquist, H. Calkins, A. J. Camm, R. Cappato, F. Cosio, H. Crijns, H.-C. Diener, A. Goette, et al.
Early and comprehensive management of atrial fibrillation: Proceedings from the 2nd AFNET/EHRA consensus conference on atrial fibrillation entitled 'research perspectives in atrial fibrillation'
Europace, July 1, 2009; 11(7): 860 - 885.
[Full Text] [PDF]


Home page
EuropaceHome page
J.-A. Cabrera, S. Y. Ho, and D. Sanchez-Quintana
How anatomy can guide ablation in isthmic atrial flutter
Europace, January 1, 2009; 11(1): 4 - 6.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
11/1/42    most recent
eun308v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in Europace
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (2)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Kirchhof, P.
Right arrow Articles by Maintz, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kirchhof, P.
Right arrow Articles by Maintz, D.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?