Implantable cardioverter-defibrillators
Implantable cardioverter-defibrillator therapy in adult patients with tetralogy of Fallot
Departments of Cardiology and Congenital Heart Disease, University of Leeds and Leeds General Infirmary, Leeds, UK
Manuscript submitted 1 December 2007. Accepted after revision 7 April 2008.
* Corresponding author: Division of Cardiovascular Medicine and Diabetes, LIGHT Laboratories, University of Leeds, Clarendon Road, Leeds LS1 9JT, UK. Tel: +44 113 2787206; fax: +44 113 2787206.E-mail address: klauswitte{at}hotmail.com
| Abstract |
|---|
|
|
|---|
Aims: Adults with repaired tetralogy of Fallot (TOF) are at risk of sudden cardiac death (SCD). ESC and AHA guidelines suggest the use of implantable cardioverter defibrillators (ICDs) to protect from this. Few data are available on the benefits of these devices in this population, and there are no randomized studies.
Methods and results: We analysed outcomes with respect to death, ICD therapy delivery, and complications for 20 patients with repaired TOF and 39 dilated cardiomyopathy (DCM) patients followed up at a UK teaching hospital. All TOF patients had clinical ventricular tachycardia (VT), electrophysiological study-inducible VT, or previous arrest due to tachyarrhythmia and received dual-chamber devices with individualized atrial detection algorithms. Tetralogy of Fallot patients were younger than DCM patients, but follow-up duration was not different between the groups. Tetralogy of Fallot patients were more likely to have experienced oversensing (45 vs. 13%; P < 0.02), inappropriate anti-tachycardia pacing delivery (20 vs. 2%; P < 0.05), and inappropriate cardioversion (25 vs. 4%; P = 0.06) than DCM patients and less likely to receive appropriate therapies than DCM patients. The death rate in TOF patients was significantly lower than that in DCM patients (5 vs. 21%; P < 0.05).
Conclusion: Tetralogy of Fallot patients have a higher risk of inappropriate therapies and other complications yet a lower incidence of appropriate therapies from their ICD than DCM patients. Further research into identification of factors predicting SCD in TOF and the benefits of ICD implantation is essential given the potential complications of ICD implantation in young congenital heart disease patients.
Key Words: Implantable defibrillator, Congenital heart disease, Dilated cardiomyopathy
| Introduction |
|---|
|
|
|---|
Advances in the surgical treatment of patients born with congenital heart defects have transformed the prognosis of conditions previously fatal in infancy or childhood. However, annual mortality rate in these patients does not return to normal. The most frequent modes of death include heart failure and sudden cardiac death (SCD) because of arrhythmia. In patients with tetralogy of Fallot (TOF), cardiac dysrhythmia leading to sudden death is thought to account for 30% of late mortality.1
| Aims |
|---|
|
|
|---|
The aim of the current study was to examine the outcomes following ICD implantation in patients with repaired TOF. We chose to compare this group with a group of patients with dilated cardiomyopathy (DCM) because the evidence base for indications for ICD therapy in DCM is most closely allied with the accepted indications for ICD therapy in patients with CHD.
| Methods |
|---|
|
|
|---|
Patients
All patients with TOF or DCM with ICDs implanted between 1 December 1996 and 1 December 2006 at a single centre providing supra-regional congenital cardiology services and regional electrophysiological services were included in the study. Case notes were retrieved and reviewed to identify indication, medical therapy, complications, and outcomes. Each therapy delivered was adjudicated by a physician experienced in device therapy. Therapies were classed as anti-tachycardia pacing (ATP) or shock and described as appropriate if delivered for VT or inappropriate if delivered for an atrial dysrhythmia. Events and mortality were censored at the date of last follow-up.
Statistical analysis
Variables were analysed using a commercially available statistics programme (Statview v5, SAS Institute, Cary, NC, USA). Continuous data are presented as means ± 1SD in text and tables, and the groups were compared using Student's t-test. Categorical data were analysed using
2 tests. We constructed the Kaplan–Meier curves for the two groups for time to first inappropriate therapy, time to any therapy, and death. A P-value of less than 0.05 was taken as significant.
| Results |
|---|
|
|
|---|
Table 1 shows the characteristics of the patient groups. All patients with TOF received dual-chamber devices. Dilated cardiomyopathy patients had dual chamber (n = 19), single-chamber (n = 15), or biventricular ICD (n = 5) devices. All devices were implanted in the pre-pectoral position. Patients with TOF were younger than those with DCM. Significant (more than mild) right ventricular (RV) dilatation or dysfunction was seen in the majority of the TOF patients, whereas left ventricular function was generally well preserved. Patients with DCM were more likely to have had a previous cardiac arrest. More TOF patients had undergone EPS than DCM patients, and only one DCM patient had had an attempted VT ablation procedure. Electrophysiological study in TOF was performed in order to establish the origin of palpitations, attempt VT ablation, or, additionally, in five cases to ablate supra-ventricular dysrhythmias. All TOF patients were symptomatic with palpitations and had documented VT, either on Holter monitor or EPS and most had been in hospital at least once with ventricular dysrhythmia in the 12 months preceding ICD implantation.
|
Patient outcomes are shown in Table 2. Tetralogy of Fallot patients had a higher rate of inappropriate discharges, mostly for misidentified supra-ventricular tachycardias (SVTs), and a lower rate of appropriate discharges, when compared with patients with DCM. Tetralogy of Fallot patients had a slightly higher rate of appropriate ATP therapies than patients with DCM. Only one patient (in the DCM group) had both appropriate ATP and cardioversion. The Kaplan–Meier analysis confirms a higher rate of inappropriate therapy in TOF patients than DCM (Figure 1A), yet no difference between TOF patients and those with DCM for days alive without any therapy (Figure 1B). This is accounted for by a trend towards fewer days alive prior to appropriate therapy for DCM patients. Overall survival in our populations seems to be lower in patients with DCM (Figure 2).
|
|
|
| Discussion |
|---|
|
|
|---|
The present data demonstrate that patients with repaired TOF receiving an ICD have a high rate of inappropriate discharges, mostly for SVTs, and a lower rate of appropriate discharges when compared with patients with DCM. Tetralogy of Fallot patients had a higher rate of appropriate ATP therapies than those with DCM.
Sudden cardiac death in congenital heart disease
Sudden cardiac death is the single most common cause of mortality late after repair of TOF,14
–18
accounting for 30% of deaths in TOF patients over 15 years of follow-up.1
Absolute death rates in this population remain very low, but are still 25–100 times higher than that of an age-matched population. Although the SCD of an otherwise relatively healthy young person is a catastrophic event, such that aggressive preventative therapy might seem proportional, ICD therapy is expensive and carries inherent physical and psychological risks.
For the current adult TOF cohort, the risk factors for ventricular arrhythmias and SCD, (although in all of the studies, SCD makes up very few of the censor events) include absolute QRS duration
180 ms, increasing QRS duration,19
older age at first repair, pulmonary regurgitation and RV dilatation,16
,20
,21
frequent ventricular ectopic beats,21
increased RV systolic pressures,4
,21
,22
complete heart block,21
,23
increased JT dispersion,24
,25
and impaired heart rate variability.26
Other predictors of clinical arrhythmia include late gadolinium enhancement on MRI scan27
and left ventricular dysfunction.28
Many of these variables are merely markers of increased severity of RV dilatation and dysfunction as a result of chronic pulmonary incompetence, and it is too early to tell whether newer, more conservative approaches to RV outflow tract obstruction with less long-term pulmonary regurgitation will improve long-term haemodynamic and electrophysiological outcomes. Some patients with little RV dilatation develop ventricular arrhythmias associated with scar in the RV and may not be identified with these non-specific factors.
Electrophysiological studies in congenital heart disease
The use of EPS in patients with DCM has largely been discarded as a prognostic indicator.29
However, in CHD, EPS continues to be used to try to assess patients' arrhythmic risk. No useful mortality data are available from most prospective studies because of small patient numbers and low event rates, but up to 35% of patients have haemodynamically important inducible VT,30
,31
associated with wide QRS, increased RV dimensions, and increased QT dispersion. The only study large enough (n = 252) to provide accurate mortality data based on 15 years of follow-up confirmed inducible VT in 37% and event-free survival in the inducible group of 50% compared with 90% in those without inducible VT. However, there were only 17 deaths (7%) over 15 years, although in those with inducible polymorphic VT (n = 11), the SCD rate at 15 years was 40%.19
Implantable defibrillators in congenital heart disease
Risk–benefit ratio
Although there is a lack of evidence of mortality benefit of ICDs in TOF patients, there is anecdotal suggestion of morbidity benefit. Medical therapy for symptomatic VT is often unsatisfactory in this group even if the side effects of amiodarone in such young patients are accepted,32
many patients continue to have episodes of palpitations because of VT. In patients with TOF with frequent episodes of ATP-responsive VT, device implantation can significantly improve their quality of life by avoiding hospital admission and maintenance of their driving licence.
Despite optimal programming and dual-lead systems in all patients, our TOF group suffered an inappropriate shock rate of 20%. Patients with repaired TOF are generally younger than their DCM counterparts, and RV dysfunction has less effect on physical functioning than left ventricular dysfunction. Hence, such patients are likely to be more active than other groups receiving ICDs which increase the risk of sinus tachycardia at rates falling in the VT detection zones, thereby increasing the risk of inappropriate VT therapies. Atrial dysrhythmias, as a result of the surgical atriotomy,33
are also a frequent cause of inappropriate therapy, and the hypertrophied and dilated RV in TOF increases the risk of far-field sensing of the QRS complex. Consistent with these frequently occurring factors in TOF patients, the largest retrospective study so far published suggests a 30% inappropriate ATP or shock rate in 121 patients.34
Other complications include a higher lead failure rate and the ongoing risk of infection due to numerous generator replacements in these young patients. Finally, the onset of ventricular dysrhythmias and ICD implantation has important lifestyle implications. Some patients (particularly thin young women) find even current ICDs unacceptably bulky. There are stringent licensing authority restrictions about driving and particular types of employment may be threatened. Patients with ICDs are more likely to report ICD-related lower quality of life and depression, with a greater prevalence in young patients.35
,36
This problem may be exacerbated in patients with CHD as a consequence of their lifelong cardiac condition and childhood experiences.
Outcomes
Our data demonstrate a low frequency of events with only one appropriate cardioversion and one death after a mean of 2 years of follow-up. This contrasts with the higher appropriate discharge rate in patients with DCM. Although our cohort is small, and limited by a modest period of follow-up, the indications for implantation and period of follow-up were similar between the groups, yet the mortality difference between the groups was significantly different. In the absence of randomized trials in patients with TOF, a comparison with the DCM population is relevant. The mean 1 year mortality for DCM patients recruited to the control arms of the primary prevention arms of ICD studies was 9%.37
In contrast, the SCD mortality in TOF is <2% per year in most studies and increases to 6% in patients followed for more than 30 years. In patients with DCM, no single trial has demonstrated a clear benefit of ICD therapy on mortality.38
It required a large meta-analysis with over 1800 patients to identify a statistically significant benefit.37
Data on secondary prevention remains neutral, despite meta-analysis, probably hampered by small numbers. Our results and data from other published reports suggest that the arrhythmic mortality from SCD in TOF patients is even lower than that of DCM patients.
Finally, and particularly relevant to the TOF population, in the ICD arm of DEFINITE, the largest randomized trial of primary prevention in DCM, 33 of 229 patients received 70 appropriate therapies, but there were only 15 sudden deaths in the 229 patients receiving medical therapy. This implies that therapies delivered for VT are not always life saving, because much VT will terminate spontaneously.39
This is likely to be more common in the TOF patients than in DCM patients.
| Conclusion |
|---|
|
|
|---|
Full implementation of current guidelines suggesting that ICD implantation as a therapy for patients with TOF (or any CHD) would lead to a large increase in ICD implants in low risk, young individuals with no proven mortality benefit, and great potential for physical and psychological damage. Before the NICE guidelines can be adopted for all individuals with TOF and other potentially arrhythmogenic congenital heart defects, the mortality and morbidity benefits and psychological aspects of ICD therapy in this group should be evaluated more closely.
| Limitations |
|---|
|
|
|---|
Our study is limited by its retrospective nature, small numbers, and short duration of follow-up. However, our data were collected in a consecutive fashion and no patients were excluded. Furthermore, given the high rate of mortality in patients with DCM, it is likely that a longer duration of follow-up would further expose the differences between the populations in terms of total mortality and sudden death. Our analysis using DCM and TOF patients is not without its limitations, but DCM patients represent the nearest appropriate comparator group for whom randomized controlled data are available.
| Acknowledgements |
|---|
|
|
|---|
The authors are grateful to Ms Rachel Mundell and Mrs Eve Butterfield for secretarial assistance. The data relating to patients with Tetralogy of Fallot have also been included in a multicentre analysis of event rates in patients with congenital heart disease and ICDs.
Conflict of interest: none declared.
| References |
|---|
|
|
|---|
[1] Oechslin EN, Harrison DA, Connelly MS, Webb GD, Siu SC. Mode of death in adults with congenital heart disease. Am J Cardiol (2000) 86:1111–6.[CrossRef][Web of Science][Medline]
[2] Silka MJ, Hardy BG, Menashe VD, Morris CD. A population-based prospective evaluation of risk of sudden cardiac death after operation for common congenital heart defects. J Am Coll Cardiol (1998) 32:245–51.
[3] Gillette PC, Yeoman MA, Mullins CE, McNamara DG. Sudden death after repair of tetralogy of Fallot. Electrocardiographic and electrophysiologic abnormalities. Circulation (1977) 56:566–71.
[4] Garson A Jr, Nihill MR, McNamara DG, Cooley DA. Status of the adult and adolescent after repair of tetralogy of Fallot. Circulation (1979) 59:1232–40.
[5] Rosenthal A, Behrendt D, Sloan H, Ferguson P, Snedecor SM, Schork A. Long-term prognosis (15–26 years) after repair of tetralogy of Fallot: I. Survival and symptomatic status. Ann Thorac Surg (1984) 38:151–6.[Abstract]
[6] Chandar JS, Wolff GS, Garson A Jr, Bell TJ, Beder SD, Bink-Boelkens M, et al. Ventricular arrhythmias in postoperative tetralogy of Fallot. Am J Cardiol (1990) 65:655–61.[CrossRef][Web of Science][Medline]
[7] Garson A Jr. Ventricular arrhythmias after repair of congenital heart disease: who needs treatment? Cardiol Young (1991) 1:177–81.[Medline]
[8] Murphy JG, Gersh BJ, Mair DD, Fuster V, McGoon MD, Ilstrup DM, et al. Long-term outcome in patients undergoing surgical repair of tetralogy of Fallot. N Engl J Med (1993) 329:593–9.
[9] Morris CD, Menashe VD. 25-year mortality after surgical repair of congenital heart defect in childhood. A population-based cohort study. JAMA (1991) 266:3447–52.
[10] Garson A Jr, McNamara DG. Sudden death in a pediatric cardiology population, 1958 to 1983: relation to prior arrhythmias. J Am Coll Cardiol (1985) 5:134B–137B.[Medline]
[11] Alexander ME, Walsh EP, Saul JP, Epstein MR, Triedman JK. Value of programmed ventricular stimulation in patients with congenital heart disease. J Cardiovasc Electrophysiol (1999) 10:1033–44.[Web of Science][Medline]
[12] Zipes DP, Camm AJ, Borggrefe M, Buxton AE, Chaitman B, Fromer M, et al, American College of Cardiology/American Heart Association Task Force, European Society of Cardiology Committee for Practice Guidelines, European Heart Rhythm Association the Heart Rhythm Society. ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death—executive summary: a report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (writing committee to develop guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death) developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society. Eur Heart J (2006) 27:2099–140.
[13] http://guidance.nice.org.uk/TA95/guidance/pdf/English.
[14] Gillette PC, Garson A Jr. Sudden cardiac death in the pediatric population. Circulation (1992) 85:I64–I69.[Medline]
[15] Nollert G, Fischlein T, Bouterwek S, Bohmer C, Klinner W, Reichart B. Long-term survival in patients with repair of tetralogy of Fallot: 36-year follow-up of 490 survivors of the first year after surgical repair. J Am Coll Cardiol (1997) 30:1374–83.[Abstract]
[16] Gatzoulis MA, Balaji S, Webber SA, Siu SC, Hokanson JS, Poile C, et al. Risk factors for arrhythmia and sudden cardiac death late after repair of tetralogy of Fallot: a multicentre study. Lancet (2000) 356:975–81.[CrossRef][Web of Science][Medline]
[17] Bricker JT. Sudden death and tetralogy of Fallot. Risks, markers, and causes. Circulation (1995) 92:158–9.
[18] Jonsson H, Ivert T, Brodin LA, Jonasson R. Late sudden deaths after repair of tetralogy of Fallot. Electrocardiographic findings associated with survival. Scand J Thorac Cardiovasc Surg (1995) 29:131–9.[Web of Science][Medline]
[19] Khairy P, Landzberg MJ, Gatzoulis MA, Lucron H, Lambert J, Marçon F, et al. Value of programmed ventricular stimulation after tetralogy of fallot repair: a multicenter study. Circulation (2004) 109:1994–2000.
[20] Harrison DA, Harris L, Siu SC, MacLoghlin CJ, Connelly MS, Webb GD, et al. Sustained ventricular tachycardia in adult patients late after repair of tetralogy of Fallot. J Am Coll Cardiol (1997) 30:1368–73.[Abstract]
[21] Gatzoulis MA, Balaji S, Webber SA, Siu SC, Hokanson JS, Shinohara T. Risk stratification for arrhythmia and sudden cardiac death late after repair of tetralogy of Fallot. In: The Adult with Tetralogy of Fallot.—Gatzoulis MA, Murphy DJ, eds. (2001) Armonk, New York: Futura Publishing Company, Inc. 1–23.
[22] Katz NM, Blackstone EH, Kirklin JW, Pacifico AD. Late survival and symptoms after repair of tetralogy of Fallot. Circulation (1982) 65:403–10.
[23] Hokanson JS, Moller JH. Significance of early transient complete heart block as a predictor of sudden death late after operative correction of tetralogy of Fallot. Am J Cardiol (2001) 87:1271–7.[CrossRef][Web of Science][Medline]
[24] Berul CI, Hill SL, Geggel RL, Hijazi ZM, Marx GR, Rhodes J, et al. Electrocardiographic markers of late sudden death risk in postoperative tetralogy of Fallot children. J Cardiovasc Electrophysiol (1997) 8:1349–56.[Web of Science][Medline]
[25] Vogel M, Sponring J, Cullen S, Deanfield JE, Redington AN. Regional wall motion and abnormalities of electrical depolarization and repolarization in patients after surgical repair of tetralogy of Fallot. Circulation (2001) 103:1669–73.
[26] Davos CH, Davlouros PA, Wensel R, Francis D, Davies LC, Kilner PJ, et al. Global impairment of cardiac autonomic nervous activity late after repair of tetralogy of Fallot. Circulation (2002) 106:I69–I75.[Web of Science][Medline]
[27] Babu-Narayan SV, Kilner PJ, Li W, Moon JC, Goktekin O, Davlouros PA, et al. Ventricular fibrosis suggested by cardiovascular magnetic resonance in adults with repaired tetralogy of fallot and its relationship to adverse markers of clinical outcome. Circulation (2006) 113:405–13.
[28] Ghai A, Silversides C, Harris L, Webb GD, Siu SC, Therrien J. Left ventricular dysfunction is a risk factor for sudden cardiac death in adults late after repair of tetralogy of Fallot. J Am Coll Cardiol (2002) 40:1675–80.
[29] Grimm W, Christ M, Bach J, Muller HH, Maisch B. Noninvasive arrhythmia risk stratification in idiopathic dilated cardiomyopathy: results of the Marburg Cardiomyopathy Study. Circulation (2003) 108:2883–91.
[30] Lucron H, Marcon F, Bosser G, Lethor JP, Marie PY, Brembilla-Perrot B. Induction of sustained ventricular tachycardia after surgical repair of tetralogy of Fallot. Am J Cardiol (1999) 83:1369–73.[CrossRef][Web of Science][Medline]
[31] Marie PY, Marcon F, Brunotte F, Briancon S, Danchin N, Worms AM, et al. Right ventricular overload and induced sustained ventricular tachycardia in operatively repaired tetralogy of Fallot. Am J Cardiol (1992) 69:785–9.[CrossRef][Web of Science][Medline]
[32] Bongard V, Marc D, Philippe V, Jean-Louis M, Maryse LM. Incidence rate of adverse drug reactions during long-term follow-up of patients newly treated with amiodarone. Am J Ther (2006) 13:315–9.[CrossRef][Medline]
[33] Magnin-Poull I, De Chillou C, Miljoen H, Andronache M, Aliot E. Mechanisms of right atrial tachycardia occurring late after surgical closure of atrial septal defects. J Cardiovasc Electrophysiol (2005) 16:681–7.[CrossRef][Web of Science][Medline]
[34] Khairy P, Harris L, Landzberg MJ, Viswanathan S, Barlow A, Gatzoulis MA, et al. Implantable cardioverter-defibrillators in tetralogy of Fallot. Circulation (2008) 117:363–70.
[35] Thomas SA, Friedmann E, Kao CW, Inguito P, Metcalf M, Kelley FJ, et al. Quality of life and psychological status of patients with implantable cardioverter defibrillators. Am J Crit Care (2006) 15:389–98.
[36] Friedmann E, Thomas SA, Inguito P, Kao CW, Metcalf M, Kelley FJ, et al. Quality of life and psychological status of patients with implantable cardioverter defibrillators. J Interv Card Electrophysiol (2006) 17:65–72.[CrossRef][Web of Science][Medline]
[37] Desai AS, Fang JC, Maisel WH, Baughman KL. Implantable defibrillators for the prevention of mortality in patients with nonischemic cardiomyopathy: a meta-analysis of randomized controlled trials. JAMA (2004) 292:2874–9.
[38] Kadish A, Dyer A, Daubert JP, Quigg R, Estes NA, Anderson KP, et al, Defibrillators in Non-Ischemic Cardiomyopathy Treatment Evaluation (DEFINITE) Investigators. Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy. N Engl J Med (2004) 350:2151–8.
[39] Ellenbogen KA, Levine JH, Berger RD, Daubert JP, Winters SL, Greenstein E, et al, Defibrillators in Non-Ischemic Cardiomyopathy Treatment Evaluation (DEFINITE) Investigators. Are implantable cardioverter defibrillator shocks a surrogate for sudden cardiac death in patients with nonischemic cardiomyopathy? Circulation (2006) 113:776–82.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
M. J. Silka and Y. Bar-Cohen Should patients with congenital heart disease and a systemic ventricular ejection fraction less than 30% undergo prophylactic implantation of an ICD?: Patients With Congenital Heart Disease and a Systemic Ventricular Ejection Fraction Less Than 30% Should Undergo Prophylactic Implantation of an Implantable Cardioverter Defibrillator Circ Arrhythm Electrophysiol, October 1, 2008; 1(4): 298 - 306. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||