Europace Advance Access originally published online on April 22, 2008
Europace 2008 10(6):683-686; doi:10.1093/europace/eun094
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
ABLATION AND ABLATION TECHNIQUES
Impact of cryoablation catheter size on success rates in the treatment of atrioventricular nodal re-entry tachycardia in 160 patients with long-term follow-up
Department of Cardiology, Bristol Royal Infirmary, Level 4, Dolphin House, Bristol Royal Infirmary, Marlborough Street, Bristol BS2 8HW, UK
Manuscript submitted 3 January 2008. Accepted after revision 24 March 2008.
* Corresponding author. Tel: +44 117 342 0489; fax: +44 117 342 0496. E-mail address: tim.cripps{at}ubht.nhs.uk
| Abstract |
|---|
|
|
|---|
Aims: To determine the efficacy and safety of cryoablation for the treatment of atrioventricular nodal re-entry tachycardia (AVNRT).
Methods and results: We analysed the procedural characteristics, acute success, and recurrence rates of 160 consecutive patients treated with cryoablation for the AVNRT and followed up for 18 months. Mean procedural time was 119.1 ± 3.7 min, with an average of 4.6 ± 0.2 Cryo lesions and an acute procedural success rate of 93%. Recurrence rates were 19 (11.9%) cases and were significantly higher in the 4 mm cryocatheter-treated group (12/59, 16.9%), compared with the 6 mm cryocatheter-treated group (9/101, 6.9%, P = 0.01). Recurrence rates were greater where slow pathway block was not achieved 8/12 (66.7%), compared with complete slow pathway block 11/129 (8.5%, P < 0.0001). Recurrence was significantly more likely if atrial echo beats were still present after cryoablation, 12/130 (9.2%) patients with no recurrence vs. 7/19 (36.8%) patients with recurrence (P < 0.0001).
Conclusion: Cryoablation is a safe and efficacious treatment for AVNRT. Complete slow pathway block is associated with long-term success, together with the use of the larger 6 mm cryocatheter. There is always a risk of heart block with radiofrequency ablation, although this experience confirms previous findings that the risk with Cryo is zero.
Key Words: Atrioventricular nodal re-entry tachycardia, Cryoablation
| Introduction |
|---|
|
|
|---|
For several years, radiofrequency (RF) ablation has been the treatment of choice for patients with atrioventricular node re-entry tachycardia (AVNRT) especially, when medical therapy has either failed in the short-term or is not preferred in the long-term.1
| Methods |
|---|
|
|
|---|
A total of 160 consecutive patients undergoing cryoablation for AVNRT from July 2003 to April 2006 were followed up for 18 months. Anti-arrythmic drugs were stopped at least 5 days before the procedure. A standard 4-wire electrophysiology study was performed to diagnose AVNRT, confirmed by the presence of dual AV nodal physiology. Tachycardia was initiated during programmed atrial extrastimulation with or without the administration of isoprenaline. Atrioventricular nodal re-entry tachycardia was confirmed by earliest retrograde activation during tachycardia at the His catheter with failure to reset atrial activation with His synchronous ventricular paced beats. The procedure was deemed success if a complete slow pathway block was achieved or in the presence of single atrial echo beats with no inducible tachycardia with isoprenaline, where the tachycardia was initially easily inducible.
Cryoablation was performed with either 4 or 6 mm tip catheters (FreezorTM and Freezor XtraTM, respectively; Cryocath Technologies, Montreal, Canada). Potential slow pathway sites were mapped initially at a temperature of –30°C for up to 60 s and then ablation performed at –80°C for 2 min if no prolongation of AH interval was seen. Programmed atrial stimulation was then performed to determine the slow pathway block, echo beats, or induction of AVNRT. Cryoablation was then continued at –80°C for 4 min if the slow pathway block or echo beats were evident. This was undertaken with fluoroscopy screening at regular intervals during the freeze time to ensure the stability of the catheter position. Further sites were mapped if slow pathway block was not achieved or if tachycardia was induced. Cryolesions were repeated until either complete slow pathway block or single atrial echo beats only were initiated together with non-inducibility of AVNRT with pacing and isoprenaline. Radiofrequency was then used to complete slow pathway ablation if cryoablation failed and time permitted.
All patients were followed up for a minimum of 18 months, using a postal questionnaire for cryoablation patients at 3, 12, and 18 months enquiring about the recurrence of symptoms with or without outpatient clinical review together with telephone follow-up if required. All patients with recurrence had cardiac monitoring to confirm the presence of AVNRT and were offered a repeat procedure.
Results are expressed as the mean ± SEM except age (mean ± SD). Continuous data were compared using Students t-test or Mann–Whitney U-test. All other categorical data were analysed using the
2 test. Statistical analysis was performed using SPSS for windows version 14.0. P-values less than 0.05 were considered significant.
| Results |
|---|
|
|
|---|
All the 160 patients undergoing ablation for AVNRT had normal structural heart except one patient who had dilated moderately impaired left ventricular systolic function. These patients treated with cryoablation resulted in an acute procedural success rate of 149 (93%), with five patients requiring additional RF ablation (three because of cryocath console failure) to achieve success and there were six failures (tachycardia still inducible). Of the 19 patients with recurrence following cryoablation, 7 underwent successful repeat cryoablation (included in the group), 5 underwent successful RF ablation, with other patients controlled on medication or deciding about repeat procedure. Complications included 1 femoral haematoma, 2 vasovagal episodes, and 13 episodes of transient AV block (initially third degree of no longer than 30 s proceeding to second degree then first degree) of <2 min total duration after re-warming, but no patient required a permanent pacemaker either in the early or late post-procedural phase.
|
In order to determine procedural factors that could affect outcome, we compared patients with and without recurrence. In the cryoablation group, patients were evenly aged and sex matched (Table 2). There were no significant differences in procedure time, freeze time, or number of lesions. Despite increased screening times for patients who suffered recurrence compared with those having no recurrence [16.9 (1.5) vs. 22.6 (2.8) min P = 0.03], there was no difference in fluoroscopy dose [26.4 (2.2) vs. 41.5 (9.1) cGy2, P = 0.243]. Recurrence was significantly more likely to occur without complete slow pathway block, 8/19 (42% recurrence group) vs. 12/130 (9.2% non-recurrence group) (P < 0.0001). Similarly, the presence of echo beats following cryoablation was significantly more greater in the recurrence group when compared with the non-recurrence group [7/19 (36.8%) vs. 12/130 (9.2%), P < 0.0001].
|
Further, we analysed the cryoablation data according to catheter tip size (Table 3). For both the 4 and 6 mm catheter, groups were well matched for age and gender. All other variables were not significantly different: procedure time, fluoroscopy dose and time, the number of lesions together with freeze time, and incidence of slow pathway block/echo beats and non-inducibility. However, there was a significant increase in recurrence rates in the 4 mm catheter tip group when compared with the 6 mm catheter tip group (12/59, 16.9% and 7/101, 7%, respectively, P = 0.01).
|
| Discussion |
|---|
|
|
|---|
Our data confirm that cryoablation is a safe and suitable alternative treatment for AVNRT. Our data are also in keeping with the majority of published literature on cryoablation for AVNRT with high acute success of 93% with 11.9% recurrence rate when compared with other studies ranging from 85 to 99% for acute success together with
10–38% recurrence rate.1
Increased safety is one of the major benefits and differences between Cryo and RF ablation with regard to pacemaker implantation for inadvertent complete heart block, further strengthening our choice of cryoablation. There were no cases of prolonged AV block with cryoablation. Published data suggest a risk of 1–2% for inadvertent complete heart block with RF or marginally higher in our unpublished observations of 3.5%. This may be due to the following factors. The cryocatheter enables increased stability as the tip of the catheter adheres to the endocardium when the lesion develops, compared with the RF catheter that moves with the heartbeat and respiratory movements.12
The predominant rhythm change during RF energy applications includes accelerated junctional rhythm during which the presence of ventriculoatrial block can be a harbinger of AV block such that even if energy is halted promptly, inadvertent block can still occur. This compares with cryoablation where the rhythm can be easily monitored with atrial pacing during junctional rhythm which is not a precursor to the AV block. Also, the development of AV block is more likely to be permanent with RF energy but is completely reversible on re-warming with cryoablation. Therefore, utilizing cryoablation appears to eliminate the risk of permanent pacemaker due to the inadvertent AV block. The rate of the AV block varies between centres; although small, nevertheless, can have a devastating effect when it occurs. As a result, the curative effects of ablation can be opened to a larger group of patients, with results equally good to those achieved with RF.
Cryoablation catheter tip size was the most interesting predictive factor for long-term recurrence. Our data showed that recurrence was more likely when using the smaller 4 mm Cryotip catheter when compared with the 6 mm catheter. This strengthens the results of De Sisti et al.,11
as we compared significantly more patients in each group (59 vs. 101, compared with their 8 vs. 61). It has been suggested that the size and number of lesions may be predictive of recurrence.12
Although the number of lesions and total freeze time were similar between the two catheters, it is likely that the 6 mm tip produces a larger lesion that is deeper and, therefore, more effective. This has been confirmed in recent animal studies, in which 4 and 8 mm cryocatheters delivered cryoablation to porcine ventricular myocardium in a saline bath. Lesion sizes and tissue temperatures were related to electrode size and also to electrode orientation and contact pressure. Electrode temperature was a poor predictor of lesion size, whereas time to cryoablation electrode rewarming was a better predictor.13
,14
The success of a larger lesion size may be suggested in our results by the higher, but not significant, elimination of slow pathway conduction in the 6 mm group together with the presence of more atrial echoes in the 4 mm group that again does not reach statistical significance. This may require further investigation to provide an insight into the mechanism of relative failure of the 4 mm cryocatheter to include other variables utilized in the animal studies.
The importance of attaining complete slow pathway block is suggested by the significantly lower recurrence rates where complete slow pathway block was achieved. Loss of slow pathway conduction, therefore, seems to predict long-term outcome compared with either sustained single atrial echo beats or non-inducible tachycardia similar to that seen by Gupta et al.7
We also confirmed higher recurrence rates in patients with persistent atrial echo beats after cryoablation, which is a new finding and different from that which occurs with RF ablation. This may be due to differences in lesion characteristics where RF lesions develop further after RF energy is delivered, with fibrosis and vascular damage, hence the risk of delayed AV block, while Cryo lesions contract with time.12
,15
During cryoablation, therefore, the abolition of any residual slow pathway would be the most important to achieve long-term success.
This study has its limitations though. Patients treated with the 4 and 6 mm cryocatheter were not randomized to treatment, even though both groups were evenly age- and sex-matched with a potential for time lag and bias. Although the 4 mm cryocatheter was introduced first, we therefore cannot exclude a learning curve effect that could have contributed to this difference. However, each of the operators has >10 years of experience for ablation for AVNRT with many RF cases performed before the Cryo study period commenced. Also, the technique used in identifying potential slow pathway ablation sites did not differ for both forms of ablation.
In conclusion, our single centre results are the largest series of patients undergoing cryoablation with long-term follow-up. We therefore suggest that cryoablation is a safe and efficacious alternative to RF ablation for the treatment of AVNRT. These results also suggest the importance of achieving slow pathway block to improve success together with the larger 6 mm cryocatheter. The failure to cause inadvertent complete AV block in our patients compared with published data for RF energy suggests that cryoablation is a safer alternative that has formed our choice for ablation for AVNRT.
Conflict of interest: none declared.
| References |
|---|
|
|
|---|
[1] Calkins H, Yong P, Miller JM, Olshansky B, Carlson M, Saul JP, et al. Catheter ablation of accessory pathways, atrioventricular nodal re-entrant tachycardia, and the atrioventricular junction: final results of a prospective multicentre clinical trial: the Atakr Multicentre Investigators Group. Circulation (1999) 99:262–70.
[2] Hindricks G. The Multicentre European Radiofrequency Survey (MERFS): complications of radiofrequency catheter ablation of arrhythmias: the Multicentre European Radiofrequency Survey Investigators of the working group on arrhythmias of the European Society of Cardiology. Eur Heart J (1993) 14:1644–53.
[3] Lowe MD, Meara M, Mason J, Grace AA, Murgatroyd FD. Catheter cryoablation of supraventricular arrhythmias: a painless alternative to radiofrequency energy. Pacing Clin Electrophysiol (2003) 26:500–3.[CrossRef][Medline]
[4] Skanes AC, Klein G, Krahn A, Yee R. Cryoablation: potentials and pitfalls. J Cardiovasc Electrophysiol (2004) 15:S28–S34.[CrossRef][Web of Science][Medline]
[5] Friedman PL, Dubuc M, Green MS, Jackman WM, Keane DTJ, Marinchak RA, et al. Catheter cryoablation of supraventricular tachycardia: results of the multicentre prospective Frosty trial. Heart Rhythm (2004) 1:129–38.[CrossRef][Web of Science][Medline]
[6] Zrenner B, Dong J, Schreieck J, Deisenhofer L, Estner H, Luani B, et al. Transvenous cryoablation versus radiofrequency ablation of the slow pathway for the treatment of atrioventricular nodal re-entrant tachycardia: a prospective randomised pilot study. Eur Heart J (2004) 25:2226–31.
[7] Gupta D, Al-Lamee RK, Earley MJ, Kistler P, Harris SJ, Nathan AW, et al. Cryoablation compared with radiofrequency ablation for atrioventricular nodal re-entrant tachycardia: analysis of factors contributing to acute and follow-up outcome. Europace (2006) 8:1022–6.
[8] Kimman GP, Theuns DA, Janse PA, Rivero-Ayerza M, Scholten MF, Szili-Tarok T, et al. CRAVT: a prospective randomised study comparing transvenous cryothermal and radiofrequency ablation in atrioventricular nodal re-entrant tachycardia. Eur Heart J (2004) 25:2232–7.
[9] Greiss I, Novak PG, Khaily P, Guerra PG, Macle L, Roy D, et al. Slow pathway ablation for AVNRT: a comparison between cryoablation and radiofrequency energy in a 5 year experience. Heart Rhythm (2005) 2:S270.
[10] Kimman GP, Theuns DA, Janse PA, Maximo R-A, Scholten MF, Szili-Torok T, et al. One year follow up in a prospective randomised study comparing radiofrequency and cryoablation of arrhythmias in Kochs triangle: clinical symptoms and event recording. Europace (2006) 8:595–5.
[11] De Sisti A, Tonet J, Barakett N, Lacotte J, Leclercq JF, Frank R. Transvenous cryoablation of the slow pathway for the treatment of atrioventricular nodal re-entrant tachycardia: a single-centre initial experience study. Europace (2007) 9:401–6.
[12] Friedman PL. How to ablate atrioventricular nodal re-entry using cryoenergy. Heart Rhythm (2005) 2:893–6.[CrossRef][Web of Science][Medline]
[13] Parvez B, Goldberg SM, Pathak V, Schubert CM, Wood MA. Time to electrode rewarming after cryoablation predicts lesion size. J Cardiovasc Electrophysiol (2007) 18:845–8.[CrossRef][Web of Science][Medline]
[14] Wood MA, Parvez B, Ellenbogen AL, Shaffer KM, Goldberg SM, Gaspar MP, et al. Determinants of lesion size and tissue temperatures during catheter cryoablation. Pace (2007) 30:644–54.[Medline]
[15] Lustgarten DC, Keane D, Ruskin J. Cryothermal ablation: mechanism of tissue injury and current experience in the treatment of tachyarrythmia. Prog Cardiovasc Dis (1999) 41:481–98.[CrossRef][Web of Science][Medline]
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
H. Bastani, J. Schwieler, P. Insulander, F. Tabrizi, F. Braunschweig, G. Kenneback, N. Drca, B. Sadigh, and M. Jensen-Urstad Acute and long-term outcome of cryoablation therapy of typical atrioventricular nodal reentrant tachycardia Europace, July 4, 2009; (2009) eup177v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Y. Chan, N. S. Mok, C. L. Lau, Y. K. Lo, C. C. Choy, S. T. Lau, and Y. C. Choi Treatment of atrioventricular nodal re-entrant tachycardia by cryoablation with a 6 mm-tip catheter vs. radiofrequency ablation Europace, May 17, 2009; (2009) eup121v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. De Sisti, J. Tonet, F. Gueffaf, F. Touil, J.-F. Leclercq, P. Aouate, J. Lacotte, F. Hidden-Lucet, and R. Frank Effects of inadvertent atrioventricular block on clinical outcomes during cryoablation of the slow pathway in the treatment of atrioventricular nodal re-entrant tachycardia Europace, December 1, 2008; 10(12): 1421 - 1427. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||