Europace Advance Access originally published online on April 7, 2008
Europace 2008 10(5):528-535; doi:10.1093/europace/eun072
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
ICDs
Dual-chamber defibrillators reduce clinically significant adverse events compared with single-chamber devices: results from the DATAS (Dual chamber and Atrial Tachyarrhythmias Adverse events Study) trial
1 Cardiology Department, Hospital General Universitario Gregorio Maranon, Madrid, Spain; 2 Cardiology Department, Hospital Universitario 12 de Octubre, Madrid, Spain; 3 Cardiology Department, University Hospital of Mannheim, Mannheim, Germany; 4 Heart Diseases Department, Hospital San Filippo Neri, Roma, Italy; 5 Cardiology Department, Hospital Santa Cruz, Carnaxide, Portugal; 6 Estadistica e Investigacion Operativa, Universitat Politecnica de Catalunya, Barcelona, Spain; 7 Scientific and Clinical Department, Medtronic Iberica, Barcelona, Spain; 8 Cardiology Department, Hospital General Universitario de Valencia, Valencia, Spain
Manuscript submitted 8 December 2007. Accepted after revision 22 February 2008.
* Corresponding author: Servicio de Cardiologia (Planta 5), Hospital Gregorio Marañon, Doctor Esquerdo, 46, 28007 Madrid, Spain. Tel: +34 915868281; fax: +34 915868018.E-mail address: almendral{at}secardiologia.es
| Abstract |
|---|
|
|
|---|
Aims: This randomized trial evaluated clinically significant adverse events (CSAEs), in patients implanted with dual-chamber (DC) vs. single-chamber (SC) implantable cardioverter defibrillator (ICD). DC-ICD had atrial tachyarrhythmia (AT) therapy capabilities. Strict programming recommendations were reinforced.
Methods and results: Patients with conventional SC-ICD indication were randomized to DC-ICD, SC-ICD, or a DC-ICD programmed as an SC-ICD (SC-simulated) and followed for 16 months. Patients in the DC and SC-simulated groups crossed over after 8 months. The primary endpoint was a composite of CSAE: all-cause mortality; invasive intervention; hospitalization (>24 h) for cardiovascular causes; inappropriate shocks (two or more episodes); and sustained symptomatic AT lasting >48 h. The outcome variable was a pre-specified score that corrected for clinical severity and follow-up duration. Three hundred and thirty-four patients were analysed (DC-ICD, n = 112; SC-ICD, n = 111; SC-simulated, n = 111). The mean left ventricular ejection fraction was 0.36 ± 0.13, 69% were in functional class
II. CSAE occurred in 65 DC-ICD, 82 SC-ICD, and 84 SC-simulated patients. The outcome variable was 33% lower in the DC-ICD group (OR 0.31; 95% CI 0.14–0.67; P = 0.0028). Mortality was 4% in DC, 9% in SC, and 10% in SC-simulated.
Conclusion: In patients with a standard SC-ICD indication, DC-ICD was associated with less CSAE when compared with SC-ICD.
Key Words: Defibrillation, Tachyarrhythmias, Pacing
| Introduction |
|---|
|
|
|---|
The therapeutic benefits of implantable cardioverter defibrillators (ICDs) have been primarily studied in patients implanted with single-chamber (SC) devices. However, dual-chamber (DC)-ICDs provide atrial pacing, atrioventricular (AV) synchrony, information about the atrial rhythm during tachycardia, and (some DC-ICD) electrical therapy for atrial tachyarrhythmias (ATs). On the other hand, DC-ICDs involve a more complex implant procedure and some of these features could even be harmful.
When the DATAS Trial started its inclusion phase, in year 2000,1
the information regarding DC-ICD was emerging.2
Two additional concerns were: (a) some laboratory findings, available at that time, suggested negative effects of right ventricular (RV) pacing;3
(b) clinical studies had suggested that the ICD electrode could itself be proarrhythmic, at least transiently after implantation,4
raising the concern for some proarrhythmic effect of the atrial electrode.
The DATAS Trial was designed to compare DC-ICD (capable of electrical therapies for AT) and SC-ICD, the outcome being clinically significant adverse events (CSAEs). We hypothesized that both the potential for device-related complications and the potential to reduce or increase spontaneous deleterious effects of the primary disease process would translate into relatively simple and clinically measurable consequences, i.e. CSAE. In order to reduce the potentially negative effects of RV pacing and homogenize the population, strict programming recommendations were made to reduce RV pacing, with the tools available in the ICD at that time. Finally, in order to assess the potential for proarrhythmia (and for a better look at AT in SC devices), a third arm of patients with a DC hardware but a SC programming was added to the usual parallel design. The main outcome of the trial is presented here.
| Methods |
|---|
|
|
|---|
Study design, randomization, and data collection
The design of the trial has been published.1
|
Follow-up started immediately after randomization. A 1-month wash out period was implemented after programmed crossover.
Primary endpoint and main outcome
The primary endpoint was a composite of five pre-determined CSAE: (i) all-cause mortality, (ii) invasive intervention due to cardiovascular cause, (iii) hospitalization (longer than 24 h) or prolongation of hospitalization due to cardiovascular cause, (iv) inappropriate shocks: two or more episodes with inappropriate shocks, and (v) sustained symptomatic ATs that (a) require urgent termination or (b) lasted more than 48 h leading to therapeutic intervention.
Since this composite endpoint was specifically designed to assess the global impact of clinically relevant events, a CSAE-score was created in order to rank clinical severity. Death was obviously the worst outcome; we assigned a high score to a premature crossover because it represented a basic failure of the assigned therapy. Point assignment was as follows: each CSAE: 1 point; death: maximum number of CSAE points in any individual patient in the entire study plus one; premature crossover: maximum number of CSAE points in any individual patient in that study period. Main outcome was defined as the CSAE-score over length of follow-up, i.e. CSAE-score rate.
Patient eligibility criteria
Patients were eligible for the study if they met a standard Class I indication for an SC-ICD according to the 1998 ACC/AHA guidelines for implantation of cardiac pacemakers and antiarrhythmia devices. After November 2001, patients as those enrolled in the MUSTT study5
were also accepted for inclusion. Every patient eligible for ICD therapy was screened at each centre.
Exclusion criteria were the following: (i) permanent AT; (ii) absence of structural heart disease; (iii) implantation criteria for DC pacing: symptomatic sinus node disease, second degree AV block (except asymptomatic Mobitz I), and complete AV block; (iv) a previously implanted pacemaker or ICD; (v) mechanical right heart valve; (vi) any medical condition that would preclude the testing required by the protocol; (vii) any medical condition that would limit study participation; (viii) the patient is unwilling or unable to cooperate or give written informed consent; (ix) legal guardians (of a minor) refuse to give informed consent; (x) inaccessibility for follow-up at the study centre; (xi) indication for cardiac resynchronization therapy; (xii) enrolment or planning to be enrolled in another clinical trial.
Implantable cardioverter defibrillator devices and programming
All devices were commercially available from Medtronic Inc. (Minneapolis, MN, USA). The DC devices had automatic atrial tiered antitachycardia therapies.
There were strong programming recommendations as previously reported1
and summarized as follows: (i) for all devices, a zone with antitachycardia pacing for cycle length <360 ms; (ii) for SC (or SC-simulated) devices: (a) stability criteria of 50 ms for arrhythmia discrimination; (b) lower pacing rate of
50 bpm or less; (iii) for DC devices: (a) arrhythmia discrimination via the PR logic function; (b) lower pacing rate of 70 bpm, in the DDD mode; (c) paced AV of
230 ms, sensed AV of
200 ms, or even longer values to reduce RV pacing; (d) atrial tachycardia detection zone: atrial cycle length
320 ms; (e) atrial fibrillation detection: atrial cycle length <150 ms; (f) tiered therapies for AT (antitachycardia pacing, 50 Hz, cardioversion).
Statistical analysis
It should be noted that: (a) the DC group included twice the number of patients since it pooled together patients from the two crossover arms (first period of DC/SC-simulated group and second period of SC-simulated/DC true group) and (b) every patient in the SC group would be followed for a longer period of time (17 months) than patients in DC (8 month); (c) in order to make groups as comparable as possible, a 1-month window (absence of adverse events counting) was opened at the 9th month in the SC arm. We accounted for this disparity in follow-up duration by calculating a CSAE-score rate; defined as the CSAE-score value divided by months of follow-up. The intention-to-treat principle was used.
Two statistical analyses were pre-planned before and during the inclusion phase in the absence of any information about allocated group. The protocol, as well as the sample size calculation, was based on the distribution free Mann–Whitney–Wilcoxon statistic. In order to obtain a more standard effect size measure, an additional primary analysis was changed to the odds ratio (OR) of CSAE-score between DC and SC obtained from SAS GENMOD procedure with the length of follow-up as an offset variable and making the assumption of a negative binomial distribution in order to allow the over-dispersion introduced by the highest values imputed to deaths. Both statistics would be able to observe a combination of effects both in the reduction of the proportion of patients with tendency to develop a CSAE, as well as in the mean of CSAE over all patients.
Sample size and power for the main analysis
The assumed effect of the DC treatment was a reduction from 30 to 15% in the proportion of patients who develop a CSAE, as well as a 15% reduction in the mean of CSAE (from 6 to 5.1). The estimated sample size1
was 200 (DC true) vs. 100 (SC true) patients followed for 8 months, with a two-sided
= 0.05 and a power of 88.8%. The sample size was set up to 360 patients (120 patients per arm), considering loses in follow-up.
| Results |
|---|
|
|
|---|
Study population
Patients were enrolled between November 2000 and December 2003. Figure 2 depicts a flow diagram of patients entering the trial with a final number of 334 patients for analysis. Their relevant clinical data are provided in Table 1.
|
|
Device implantation and follow-up
Lead- and procedure-related complications are depicted in Table 2. There were no statistically significant differences among groups, with a non-significant trend towards a higher procedure-related complication rate in DC-ICD, even if considering DC and SC-simulated (identical hardware, see fourth column in Table 2) as a single group [14/223 (6.3%) vs. 3/111 (2.7%), P = NS].
|
Of the 334 patients, 12 (3.6%) were lost to follow-up (Figure 2). The mean follow-up duration was 16 months in DC, 15.6 in SC, and 15.5 in SC-simulated.
There were no significant differences among groups in the percentage of patients on each cardioactive medication at hospital discharge, and at 8- or 17-month follow-up, except for digoxin at the 8-month follow-up (12% in DC, 3% in SC-simulated, P = 0.03).
There were 20 premature crossovers during follow-up with only one of those being in the SC-ICD arm (Table 3, Figure 2).
|
The percentage of ventricular pacing, obtained from the device counters at each follow-up visit, is shown in Table 3.
Primary endpoint
The absolute number of CSAEs was 65 in the DC group when compared with 82 in the SC and 84 in the SC-simulated groups (Figure 3A). The correspondent CSAE-score is depicted in Figure 3B. The rate of CSAE-score (Figure 3C) was 33% lower in the DC group, a difference that was statistically significant (P = 0.0028; OR 0.31; 95% CI 0.14–0.67). The proportion of patients performing better with DC was also statistically significant (P = 0.030; OR 0.56; 95% CI 0.51–0.62). A secondary sensitivity analysis was performed, including also the 20 patients randomized but not included in the analysis, showing similar results.
|
Although the study was not powered to make statistical comparisons for each of the components of the primary endpoint, these are shown in Table 4 and Figure 4. Of note, total mortality was similar in SC (9%) and SC-simulated (10%) groups, but tended to be lower in the DC group (4%).
|
|
| Discussion |
|---|
|
|
|---|
Main findings
The main finding of this study is that in a setting where the excess hardware-related complications of DC devices are limited and the RV pacing effects are reduced by strict programming, the impact of CSAEs was lower in patients with a DC-ICD than in patients with an SC-ICD device.
Primary outcome
Since efficacy for termination of ventricular tachyarrhythmias is expected to be identical in SC and DC devices, but common sense suggests that DC-ICD may have more complications but also some advantages and risks, we addressed the question of whether adverse events differ in these two types of ICD. Some DC-ICDs are also able to sense and treat AT,6
,7
with the ability to potentially improve the clinical status of patients by reducing AT burden, an additional element of the study. Not all adverse events have the same clinical implications, and what finally matters is the likelihood of those that (whether produced or reduced by ICD) are clinically relevant, i.e. resulting in death, an invasive intervention, hospitalization, undesired shocks, or prolonged ATs (what we have considered as CSAE). For these reasons, our study focused on CSAE and a score was developed for further tuning by assigning a weight to each CSAE.
Several studies have now been published comparing DC and SC-ICD,8
–14
but to the best of our knowledge, this is the first study with a pooled CSAE as the primary endpoint. The results of the primary outcome of the DATAS study are, in this sense, unique, as they show that this outcome variable is better with a DC-ICD than with an SC-ICD device.
Procedure/lead-related complications
The overall incidence of lead- and device-related complications of 9% compares favourably with that of most recent ICD trials of SC- and DC-ICD that provided a detailed report of complications.8
,10
Most of the complications resulted in a CSAE, indicating that the parameters selected for our composite CSAE were sensitive enough to pick up procedure- and lead-related complications.
Ventricular pacing
The issue of the negative effects of RV pacing on ventricular function facilitating heart failure has been of increasing interest,15
and this deleterious effect shown to be clinically relevant in DAVID.9
More recently, subanalyses from randomized clinical trials (RCTs) have described the effect of the proportion of paced beats on clinical outcome.16
–18
In our study, this effect was not so obvious. The number of heart failure-related hospitalizations was higher in DAVID than in DATAS (13.3 and 22.6% in SC and DC, respectively, at 1 year in DAVID, 5 and 8% at 8 months in DATAS). DAVID selected patients with lower left ventricular ejection fraction (LVEF) (mean LVEF 27% in DAVID and 36% in DATAS) and had more ventricular pacing (
60% in DAVID and 40% in DATAS).
The recent INTRINSIC RV trial showed if ventricular pacing is minimized with an algorithm, DC-ICD is not inferior to SC-ICD.13
Our results, with an intermediate decrease in %RV pacing producing less deleterious effects, could be consistent with these findings.
Inappropriate shocks
The extent to which DC-ICD devices decrease inappropriate therapies has been controversial. Some studies found no benefit of DC devices.8
,14
More recently, both in a population with slow VT10
and in a more general population,11
a modest reduction was found. Since frequent ICD shocks are detrimental for patient's well-being, in our clinically focused approach, it was considered that two or more inappropriate shocks (first inappropriate shock could just mean a need for reprogramming, then preventing further inappropriate shocks) represented a significant problem. Our results point in a similar direction as these recent studies: it is infrequent to have two or more inappropriate shocks, but seems to be even more so in patients with DC-ICD.
Atrial tachyarrhythmias
Several reports have substantiated the efficacy of some DC-ICD to detect and terminate AT.6
,7
However, since most of these ATs lasted only minutes to hours19
and the symptomatic status was largely ignored (device retrieved arrhythmias), their clinical significance is unclear. In contrast, our study considered AT as part of the primary endpoint only if the episodes were symptomatic and/or long-lasting requiring therapeutic intervention, thus with undisputable clinical significance. As such, they were observed in a minority of patients, but mostly in patients with SC devices.
Total mortality
Two RCTs reported a non-significant trend towards a higher mortality in DC-ICD.9
,12
A recent RCT,13
including only patients in whom a DC algorithm minimized ventricular pacing to <20%, reported non-inferiority of DC-ICD with a favourable mortality trend. In DATAS, with a strict but relatively simple programming and no patient exclusion as to the amount of ventricular pacing, mortality trended similarly. Moreover, the mortality of the SC-simulated group (10%) was similar to that of the SC group, showing inherent consistency of our findings. It can be argued that the short follow-up in DATAS (before crossover) minimizes the deleterious effects of RV pacing. However, an increase in mortality was an early effect in DAVID.9
In contrast to the other reported RCT, our study included electrical therapy for AT. ATs in patients with decreased LVEF are associated with an increase in total mortality, and the composite of death or hospitalization.20
AT left untreated could promote ventricular tachyarrhythmias.21
Inappropriate recognition and therapy of AT could even lead to a vicious life-threatening cycle.22
Limitations
The follow-up might have been too short to reveal differences in heart failure-related hospitalizations and related mortality, since by the crossover design of DATAS, each DC patient was not followed for >8 months. However, hospitalizations in the second 8-month period tended to decrease in all arms.
It is possible that recently introduced sophisticated algorithms designed to minimize RV pacing could have further decreased RV pacing.13
,23
However, they were introduced after the enrolment of patients in DATAS.
Premature crossovers are a limitation of RCT. Despite the need for AEAC authorization,
10% of the patients implanted with a DC-ICD hardware crossed over as opposed to 1% in the SC-ICD. This was likely to have reflected the ease for software crossover once the DC hardware is in place (in the DC and SC-simulated arms) when compared with a hardware upgrade (in the SC-ICD arm). However, our scoring system severely penalized the premature crossover, thus correcting for this difference.
Clinical implications
The characteristics of our patient population seem to be similar to those of other recently published ICD series in which patients were not selected based on LV function10
–12
and thus represents an unselected ICD population, with mean LVEF over 0.30. To this extent, DATAS can be considered representative of the daily clinical ICD practice in typical European countries.
In this context, the DATAS study demonstrates a reduction in pooled CSAE with the use of DC-ICD. The study's conclusions could lead to the development of trials to explore if CSAE can be reduced even further with algorithms intended to minimize RV pacing.
| Funding |
|---|
|
|
|---|
This study was funded in full by Medtronic.
| Appendix |
|---|
|
|
|---|
DATAS Investigators and study centres:
B. Lüderitz, J. Schwab, T. Lewalter, R. Schimpf, University Hospital, Bonn, Germany; M. Santini, R. Ricci, C. Pignalberi, M. Russo, San Filippo Neri, Rome, Italy; P. Hanrath, Ch. Stellbrink, K. Mischke, R. Koos, University Hospital RWTH, Aachen, Germany; J. Brugada, L. Mont, M. Matas, H. Clinic i Provincial, Barcelona, Spain; J. Gill, R. Simon, A. Rinaldi, N. Gall, St Thomas' Hospital, London, UK; M. Glikson, Sheba Medical Center, Tel-Hashomer, Israel; J. Roda, S. Villalba, V. Palanca, J. Belchi, H. General Universitario, Valencia, Spain; C. Muto, M. Canciello, G. Carreras, B. Tuccillo, Loreto Mare Hospital, Naples, Italy; A. Arenal, E. Gonzalez-Torrecillas, F. Atienza, H. Gregorio Marañon, Madrid, Spain; M. Borggrefe, S. Spehl, 1st Department of Medicine Cardiology, University Hospital Mannheim, Mannheim, Germany; J.L. Merino, R. Peinado, H. La Paz, Madrid, Spain; J.C. Rodriguez, O. Medina, J. García, H. Insular de Gran Canaria, Las Palmas, Spain; F. Morgado, Santa Cruz, Lisbon, Portugal; I. Lozano, J. Toquero, R. Arroyo, H. Puerta de Hierro, Madrid, Spain; J.M. Ormaetxe, M. Arkotxa, H. de Basurto, Bilbao, Spain; G. Steinbeck, E. Hoffman, S. Janko, U. Dorwarth, Ludwig-Maximilian-University Hospital, München, Germany; M. Geist, V. Turkisher, Wolfson Medical Center, Holon, Israel; P. Della Bella, G. Fassini, C. Carbucicchio, F. Giraldi, Centro Cardiologico Monzino, Milano, Italy; P. Golino, M. Viscusi, F. Mascia, Hospedale Civile, Caserta, Italy; L. Tercedor, M. Alvarez, H. Virgen de las Nieves, Granada, Spain; J.G. Martinez, A. Ibañez, H. General Universitario, Alicante, Spain; A. Moya, E. Rodriguez, C. Alonso, H. Valle Hebron, Barcelona, Spain; M. Lopez Gil, J. Sanz, H. 12 Octubre, Madrid, Spain; R. Garcia-Civera, R.Ruiz, S. Morell, R. SanJuan, H. Clinico Universitario, Valencia, Spain; A. García-Alberola, J. Martinez, J.J. Sanchez, H. Virgen de la Arrixaca, Murcia, Spain; M. Manz, D.Burkhardt, A. Markewitz, Krankenhaus Marienhof, Koblenz, Germany; E. Castellanos, L. Rodriguez-Padial, H. Virgen de la Salud, Toledo, Spain; M. Sassara, A. Achilli, E. Scabbia, Civile Hospital, Viterbo, Italy; J. Olagüe, J.E. Pareja, M.J. Sancho-Tello, H. La Fe, Valencia, Spain; S. Hohnloser, G. Grönefeld, Johann Wolfgang Goethe University, Frankfurt, Germany; T Fuchs, Assaf Harofe Medical Center, Tzerifin, Israel; W. Jung, N. Schwick, B. Roggenbuck-Schwilk, Klinikum Villingen-Schwenningen, Villingen, Germany; B. Lemke, T. Lawo, T. Deneke, S. Holt, BG Kliniken Bergmannsheil, Bochum, Germany; G.Baumann, H. Bondke, M. Claus, Campus Charite Mitte, Berlin, Germany; A. Maresta, S. Silvani, D. Cornacchia, E. Tampieri, Civile Hospital, Ravenna, Italy; J.J. Manzano, A. Medina, E. Caballero, F. Wangüemert, H. General Dr Negrín, Las Palmas, Spain.
| Acknowledgements |
|---|
|
|
|---|
This study would not have been possible without Mercedes Ortiz, from Hospital Gregorio Marañon (Madrid, Spain), our study data coordinator.
Author contributions: original idea and launching of the study: Aurelio Quesada. Members of the Steering Committee: Aurelio Quesada, Jesus Almendral, Fernando Arribas, Christian Wolpert, Renato Ricci, Pedro Adragao, Xavier Navarro. Statistical design and analysis: Erik Cobo. Funding to pay the Open Access publication charges for this article was provided by Medtronic.
www.clinicaltrials.gov Identifier: NCT00157820 [ClinicalTrials.gov] .
Conflict of interest: J.A. has received honoraria from Medtronic, Guidant (now Boston Scientific), Johnson & Johnson and St Jude Medical for lectures, and has served as a consultant for Johnson & Johnson. C.W. has received honoraria from Medtronic and St Jude Medical for lectures. E.C. is a consultant for Ferrer International and Medtronic, and receives teaching grants from Instituto de Formación Novartis. X.N. is an employee of Medtronic. A.Q. is currently conducting research sponsored by Medtronic and has served as a paid consultant for Medtronic and Boston Scientific.
| References |
|---|
|
|
|---|
[1] Quesada A, Almendral J, Arribas F, Ricci R, Wolpert C, Adragao P, et al. The DATAS rationale and design: a controlled, randomized trial to assess the clinical benefit of dual chamber (DDED) defibrillator. Europace (2004) 6:142–50.
[2] Trappe HJ, Achtelik M, Pfitzner P, Voigt B, Weismuller P. Single-chamber versus dual-chamber implantable cardioverter defibrillators: indications and clinical results. Am J Cardiol (1999) 83:8–16D.[CrossRef]
[3] Tse HF, Lau CP. Long-term effect of right ventricular pacing on myocardial perfusion and function. J Am Coll Cardiol (1997) 29:744–9.[Abstract]
[4] Kim SG, Ling J, Fisher JD, Wang G, Rameneni A, Roth JA, et al. Comparison and frequency of ventricular arrhythmias after defibrillator implantation by thoracotomy versus nonthoracotomy approaches. Am J Cardiol (1994) 74:1245–8.[CrossRef][ISI][Medline]
[5] Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. N Engl J Med (1999) 341:1882–90.
[6] Ricci R, Pignalberi C, Disertori M, Capucci A, Padeletti L, Botto G, et al. Efficacy of a dual chamber defibrillator with atrial antitachycardia functions in treating spontaneous atrial tachyarrhythmias in patients with life-threatening ventricular tachyarrhythmias. Eur Heart J (2002) 23:1471–9.
[7] Adler SW, Wolpert C, Warman EN, Musley SK, Koehler JL, Euler DE, et al. Efficacy of pacing therapies for treating atrial tachyarrhythmias in patients with ventricular arrhythmias receiving a dual-chamber implantable cardioverter defibrillator. Circulation (2001) 104:887–92.
[8] Deisenhofer I, Kolb C, Ndrepepa G, Schreieck J, Karch M, Schmeider S, et al. Do current dual chamber cardioverter defibrillators have advantages over conventional single chamber cardioverter defibrillators in reducing inappropriate therapies? A randomized, prospective study. J Cardiovasc Electrophysiol (2001) 12:134–42.[CrossRef][ISI][Medline]
[9] The DAVID Trial Investigators. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) trial. JAMA (2002) 288:3115–23.
[10] Bansch D, Steffgen F, Gronefeld G, Wolpert C, Bocker D, Mletzko RU, et al. The 1+1 trial: a prospective trial of a dual- versus a single-chamber implantable defibrillator in patients with slow ventricular tachycardias. Circulation (2004) 110:1022–9.
[11] Friedman PA, McClelland RL, Bamlet WR, Acosta H, Kessler D, Munger TM, et al. Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis. The Detect Supraventricular Tachycardia Study. Circulation (2006) 113:2871–9.
[12] Kolb C, Deisenhofer I, Schmieder S, Barthel P, Zrenner B, Karch MR, et al. Long-term follow-up of patients supplied with single-chamber or dual-chamber cardioverter defibrillators. Pacing Clin Electrophysiol (2006) 29:946–52.[CrossRef][Medline]
[13] Olshansky B, Day JD, Moore S, Gering L, Rosenbaum M, McGuire M, et al. Is dual-chamber programming inferior to single-chamber programming in an implantable cardioverter-defibrillator? Results of the INTRINSIC RV (Inhibition of Unnecessary RV Pacing With AVSH in ICDs) Study. Circulation (2007) 115:9–16.
[14] Theuns DA, Klootwijk AP, Goedhart DM, Jordaens LJ. Prevention of inappropriate therapy in implantable cardioverter-defibrillators: results of a prospective, randomized study of tachyarrhythmia detection algorithms. J Am Coll Cardiol (2004) 44:2362–7.
[15] Sweeney MO, Prinzen FW. A new paradigm for physiologic ventricular pacing. J Am Coll Cardiol (2006) 47:282–8.
[16] Sweeney MO, Hellkamp AS, Ellenbogen KA, Greenspon AJ, Freedman RA, Lee KL, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation (2003) 107:2932–7.
[17] Steinberg JS, Fischer A, Wang P, Schuger C, Daubert J, McNitt S, et al. The clinical implications of cumulative right ventricular pacing in the Multicenter Automatic Defibrillator Trial II. J Cardiovasc Electrophysiol (2005) 16:359–65.[ISI][Medline]
[18] Sharma AD, Rizo-Patron C, Hallstrom AP, O'Neill GP, Rothbart S, Martins JB, et al. Percent right ventricular pacing predicts outcomes in the DAVID trial. Heart Rhythm (2005) 8:830–4.[Medline]
[19] Darbar D, Warman EN, Hammill SC, Friedman PA, Worldwide Jewel AF Investigators. Recurrence of atrial tachyarrhythmias in implantable cardioverter-defibrillator recipients. Pacing Clin Electrophysiol (2005) 28:1047–51.[CrossRef][Medline]
[20] Dries DL, Exner DV, Gersh BJ, Domanski MJ, Waclawiw MA, Stevenson LW. Atrial fibrillation is associated with an increased risk for mortality and heart failure progression in patients with asymptomatic and symptomatic left ventricular systolic dysfunction: a retrospective analysis of the SOLVD trials. Studies of Left Ventricular Dysfunction. J Am Coll Cardiol (1998) 32:695–703.
[21] Stein KM, Euler DE, Mehra R, Seidl K, Slotwiner DJ, Mittal S, et al. Do atrial tachyarrhythmias beget ventricular tachyarrhythmias in defibrillator recipients? J Am Coll Cardiol (2002) 40:335–40.
[22] Birgersdotter-Green U, Rosenqvist M, Lindemans FW, Ryden L, Radegran K. Holter documented sudden death in a patient with an implanted defibrillator. Pacing Clin Electrophysiol (1992) 15:1008–14.[CrossRef][Medline]
[23] Sweeney MO, Shea JB, Fox V, Adler S, Nelson L, Mullen TJ, et al. Randomized pilot study of a new atrial-based minimal ventricular pacing mode in dual-chamber implantable cardioverter-defibrillators. Heart Rhythm (2004) 1:160–7.[CrossRef][ISI][Medline]
![]()