Skip Navigation


Europace Advance Access originally published online on August 15, 2008
Europace 2008 10(11):1256-1265; doi:10.1093/europace/eun205
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
10/11/1256    most recent
eun205v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Tzeis, S.
Right arrow Articles by Vardas, P. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tzeis, S.
Right arrow Articles by Vardas, P. E.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org


REVIEWS

Tools and strategies for the reduction of inappropriate implantable cardioverter defibrillator shocks

Stylianos Tzeis1, George Andrikopoulos2, Christof Kolb1 and Panos E. Vardas3,*

1 Deutsches Herzzentrum and 1. Medizinische Klinik, Faculty of Medicine, Technische Universität München, Munich, Germany; 2 First Department of Cardiology, Evagelismos Hospital, Athens, Greece; 3 Cardiology Department, Heraklion University Hospital, PO Box 1352, 71110 Heraklion, Crete, Greece

Manuscript submitted 22 February 2008. Accepted after revision 19 July 2008.

* Corresponding author. Tel: +30 2810 392706/392422; fax: +30 2810 542055. E-mail address: cardio{at}med.uoc.gr or mximeris{at}excite.com


    Abstract
 Top
 Abstract
 Introduction
 Differentiation of...
 Ventricular oversensing
 Optimal therapy programming for...
 Pharmacological and...
 Future possibilities
 Funding
 References
 
Implantable cardioverter defibrillators (ICDs) have been shown to provide a survival benefit in patients at high risk of sudden cardiac death. A major problem associated with ICD therapy is the occurrence of inappropriate shocks which impair patients' quality of life and may also be arrhythmogenic. Despite recent technological advances, the incidence of inappropriate shocks remains high, thus posing a challenge that we have to meet. In the present review we summarise the available tools and the strategies that can be followed in order to reduce inappropriate ICD shocks.

Key Words: Implantable cardioverter defibrillator, Inappropriate shock, ICD programming


    Introduction
 Top
 Abstract
 Introduction
 Differentiation of...
 Ventricular oversensing
 Optimal therapy programming for...
 Pharmacological and...
 Future possibilities
 Funding
 References
 
Implantable cardioverter defibrillators (ICDs) represent the treatment of choice for the secondary prevention of sudden cardiac death, while robust data have also accumulated to validate their mortality benefit in the expanding pool of primary prevention patients.1Go–5Go Despite their undoubted evidence-based credentials, even the newer generation ICDs are plagued by the caveat of inappropriate therapies, which indeed seem to represent their Achilles' heel. Although this major theme has been in the limelight ever since the advent of ICDs, when the use of only rate criteria was associated with a 45% occurrence of inappropriate shocks, the issue still constitutes a major challenge that we have to face.6Go

The incidence of inappropriate therapies remains high, ranging from 15 to 35%, and the related impact on patients' quality of life is paramount.7Go–10Go Implantable cardioverter defibrillator shocks are associated with worse mental health, impaired quality of life,11Go fewer new activities mainly due to an increased incidence of avoidance,12Go depression, and anxiety.13Go,14Go Increased shock frequency has been reported as an independent predictor of anxiety scores14Go as well as low treatment satisfaction.15Go Apart from the detrimental effect on the quality of life, inappropriate therapies may also result in the induction of potentially fatal ventricular tachyarrhythmias (VTs).16Go–19Go Thus, the reduction of the incidence of inappropriate ICD shocks is of primary importance. Moreover, the need to enhance device discrimination is further amplified by the tendency to broaden the eligibility criteria of ICD implantation in primary prevention subpopulations, thus inevitably increasing the relative burden of supraventricular tachycardias (SVTs) falling within the VT/VF detection zone when compared with the proportion of true VT/VF episodes.20Go

The primary criterion for the detection of VTs by ICDs is the ventricular rate, evaluated by sensing electrical signals exceeding the programmable setting of sensitivity and falling within the selected detection zones. If a consecutive number of sensed events or a predetermined percentage of beats in a sliding detection window falls within the detection zone, then the rate criterion is satisfied. This rather crude concept makes the criterion susceptible to misinterpretation and thus false detection of tachyarrhythmias, resulting in inappropriate therapy delivery. The main causes of inappropriate shocks are the erroneous diagnosis of SVT as VT and ventricular oversensing.9Go,20Go–22Go


    Differentiation of supraventricular from ventricular tachycardias
 Top
 Abstract
 Introduction
 Differentiation of...
 Ventricular oversensing
 Optimal therapy programming for...
 Pharmacological and...
 Future possibilities
 Funding
 References
 
In order to enhance the discrimination performance of ICDs, novel discrimination algorithms have been developed with a view to improving the differential diagnosis of SVTs vs. VTs and decreasing the incidence of inappropriate therapies. A key point concerning the use of discrimination algorithms is that the increased specificity (likelihood of appropriately withholding therapies) afforded by the activation or more stringent use of these criteria is achieved in the expense of sensitivity (likelihood of detecting true VT), and vice versa.23Go,24Go Through this prism, the optimal programming should be individualized in relation to the rate of tachyarrhythmia as well as the likelihood of SVT occurrence in a certain tachyarrhythmia zone. The analysis of episodes in ICD recipients has demonstrated that the lower the tachycardia cycle length, the higher the relative percentage of VTs in comparison to SVTs.25Go Thus, the activation of a discrimination algorithm in the VF zone would be of limited value, when one takes into consideration the small odds of rate overlap between VTs and SVTs in this detection zone, and the high likelihood that a considerable proportion of primary prevention ICD recipients will be quite unable to tolerate very rapid tachyarrhythmias because of their underlying compromised ventricular function.

Single-chamber discrimination algorithms
The criteria used by single-chamber ICDs focus on interval regularity, rate onset, and ventricular electrogram morphology (Table 1).


View this table:
[in this window]
[in a new window]

 
Table 1 Single-chamber implantable cardioverter defibrillator detection algorithms from different manufacturers (manufacturers presented in alphabetical order)

 
Onset criterion
The onset criterion is based on the premise that the acceleration of the ventricular rate can differentiate a sinus tachycardia from a VT, since the former usually displays a gradual rate increase, in contrast to the latter, which almost always accelerates rapidly.23Go,26Go–28Go Although the onset criterion is extremely useful in excluding sinus tachycardia, it is also associated with a risk of VT under-detection, for example, in VT initiation during sinus tachycardia, or in VTs which do not exhibit a sudden onset.23Go,27Go,28Go Furthermore, this criterion is not suitable for patients with exercise-induced VTs, since the gradual catecholamine-induced heart rate increase preceding the onset of VT would satisfy the onset criterion, resulting in inappropriate withholding of programmed therapies.

Stability criterion
The concept of the stability criterion is the differentiation of VTs from episodes of atrial fibrillation with ventricular rates within the detection zones, based on the presence of ventricular rate stability.23Go,26Go–29Go However, the pseudo regularization of ventricular rate in the case of atrial fibrillation with a rapid ventricular response reduces the diagnostic accuracy of this algorithm.27Go,29Go Furthermore, the application of this algorithm poses the risk of under-detection of VTs with an irregular ventricular rate, such as polymorphic VTs and irregular monomorphic VTs, as well as ischaemic VTs with dynamic changes of the re-entrant circuit. In monomorphic VTs, the stability algorithm may rarely result in delayed therapy delivery due to the initial instability of the ventricular tachycardia cycle length.23Go,30Go The combined use of stability and sudden-onset criteria with a time-related override has been associated with a sensitivity of 100% and a specificity of 83%.23Go

Morphology criterion
The morphology criterion was developed in the context of differentiating the origin of detected tachyarrhythmia episodes by evaluating the similarity of the respective electrogram morphology in comparison to a reference electrogram template acquired during supraventricular rhythm without aberration. A similarity index is calculated for each sensed event falling within the tachycardia zone, based on the difference in the areas of compared electrograms. If the estimated similarity index is higher than an operator-programmable limit, then the sensed event is classified to be of supraventricular origin. If a certain proportion of sensed events in the detection window are classified as supraventricular, then the ventricular origin of the arrhythmia is precluded and the programmed therapies are not delivered. When used as a sole discriminator, the Medtronic morphology algorithm results in a specificity (percentage of appropriately classified SVTs) of 78.2% and a sensitivity of 98.6% (percentage of appropriately classified VT/VFs).31Go The use of the St Jude Medical discriminator alone achieved a sensitivity of 77–94.4% and a specificity of 71–95%, while its combined use with stability and onset criteria at the optimal diagnostic setting of two out of three resulted in a sensitivity of 96% and a specificity of 90.9%.32Go–34Go Gold et al.35Go reported that the single-chamber sensitivity of the Guidant morphology algorithm [vector time and correlation (VTC) algorithm] is 99% combined with a specificity of 97%, while in a recent retrospective analysis the single-chamber algorithm achieved a sensitivity of 100% with a specificity of 91%.36Go

Several potential limitations should be taken into account concerning the discrimination performance of morphology algorithms. A key step in the accurate assessment of the similarity index is the proper alignment of the tachycardia electrogram and the reference template. The procedure is applied by aligning the respective positive or negative peaks. Thus, template misalignment may result in overestimation of the non-overlapping electrogram area, underestimation of the matching percentage, and thus to an increased likelihood of event misclassification as VT. Furthermore, the algorithm is prone to misclassification in the case of rate-dependent aberration, since the dissimilarity between the electrogram morphology of the aberrant supraventricular rhythm and the reference template would result in erroneous classification of the index event as VT. The algorithm also displays a poor discrimination performance in the case of VT with a similar morphology to the intrinsic rhythm, as would be the case in certain septal VTs. Caution is also required in order to prevent electrogram truncation, which occurs when the electrogram amplitude exceeds the maximum programmed range, resulting in clipping of some portions of the signal and thus preventing the optimal comparison of the index electrogram with the reference template. Other concerns that need to be addressed are ischaemia-mediated acute changes in electrogram morphology, electrogram maturation, distortion of a sinus tachycardia electrogram due to exercise-induced myopotential interference, especially with low amplitude electrograms, and post-shock electrogram distortion.

Optimal customizing of programming parameters may overcome some of the above-mentioned limitations. Electrogram truncation can be overcome by appropriate adjustment of the electrogram range. For instance, in Medtronic ICDs, it is important to select the far-field configuration (Can to RV coil) as electrogram source, since it is associated with improved detection performance and also minimizes potential posture-related changes in electrogram configuration.37Go–39Go The St Jude Medical morphology algorithm is designed to operate with an RV bipolar electrogram source, while an interesting concept implemented in the Guidant VTC algorithm is that the comparison of the index with the reference electrogram is based on information derived from two different sources: a far-field channel of RV coil to SV coil+can and a near-field channel of RV tip to RV coil. Finally, the activation of automatic electrogram template updating overcomes time-related changes in electrogram morphology, thus averting the need for periodic manual template updating.37Go

Dual-chamber discrimination algorithms
Dual-chamber ICD discriminators are based on information derived from the atrial electrogram pertaining to the timing and pattern of atrial events, while single-chamber ICD differentiation algorithms are also used. These data are incorporated in different flowcharts and algorithms by each manufacturer.40Go A critical step in the hierarchical approach of dual-chamber algorithms is the comparison of the atrial and ventricular rate, since the ventricular rate exceeds the atrial rate in more than 90% of ventricular tachycardias.41Go The dual-chamber discrimination algorithms of the most recent generation ICDs from each manufacturer are briefly presented.

Biotronik SMART detection algorithm42Go,43Go
The initial discrimination test in the SMART algorithm is based on the comparison of the atrial and ventricular rates. If the ventricular rate is higher than the atrial rate, then VT is diagnosed. If the ventricular rate is lower than the atrial rate, the index tachyarrhythmia is differentiated based on atrial and ventricular rate stability, as well as the ratio of atrial to ventricular signals. Finally, if the atrial and ventricular rates are equal, the differentiation is performed using rate stability, and sudden onset (Figure 1, modified from reference42Go). The reported specificity and sensitivity of the SMART detection algorithm are 89 and 100%, respectively.44Go


Figure 1
View larger version (9K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 1 SMART Detection Algorithm (Biotronik).42Go,43Go

 
ELA/Sorin PARAD+ algorithm45Go
The PARAD+ algorithm enables the stepwise discrimination analysis of detected tachyarrhythmias based on the flowchart presented in Figure 2 (modified from reference45Go). If the rate stability criterion is not satisfied, the rhythm is diagnosed as atrial fibrillation. This classification scheme is prone to false detection of polymorphic VTs as atrial fibrillation, although the nominal stability setting has been shown to be associated with high sensitivity in detecting VTs. In the PARAD+ algorithm, all stable dissociated rhythms are further differentiated to atrial fibrillation or VT on the basis of the long cycle criterion (VTLC). The concept of this criterion is that even in fast atrial fibrillation with a rapid ventricular response and pseudo-regularization, a long cycle length exceeding the average interval is present from time to time. If a programmable long cycle length is detected, then atrial fibrillation is diagnosed and therapy delivery is inhibited. Rhythms with N:1 atrioventricular association are classified as atrial flutter. In the next steps, the cycle length acceleration is evaluated for sinus tachycardia discrimination, while a final discrimination between atrial and ventricular tachycardia is performed based on the chamber exhibiting initial acceleration. A recent meta-analysis of prospective trials evaluating the PARAD/PARAD+ algorithms reported that the sensitivity of VT/VF detection was 99.3%, the respective value in slow VT detection was 94%, while the specificity of SVT recognition was 94%.46Go


Figure 2
View larger version (18K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 2 PARAD+® diagnostic flowchart (ELA-Sorin group).45Go

 
Guidant Rhythm ID—detection enhancement algorithm47Go
In the Guidant Rhythm ID algorithm (Figure 3, modified from reference47Go), if the average ventricular rate exceeds the average atrial rate by more than 10 bpm then VT is diagnosed. Otherwise, the morphology algorithm is appraised and if it is indicative of SVT then therapy is withheld. Finally, both the criteria of rate stability and atrial rate above an atrial fibrillation threshold need to be met in order for an SVT to be diagnosed. The Rhythm ID algorithm has demonstrated 100% sensitivity for VT/VF detection, while in two studies the reported specificity of SVT discrimination was 92 and 97%.35Go,48Go


Figure 3
View larger version (25K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 3 Guidant Rhythm ID—Dual-chamber Detection Enhancement Algorithm.47Go

 
Medtronic PR Logic49Go
Arrhythmia detection in Medtronic dual-chamber ICDs is enhanced by the use of the PR Logic algorithm, which is optionally programmed in combination with stability onset and morphology criteria. In the PR Logic algorithm, each ventricular coupling interval is divided into zones (junctional, retrograde, and antegrade) (Figure 4, modified from reference49Go). Each atrial sensed event is categorized based on the ventricular zone in which atrial sensing occurred. The recognition of a certain atrial event pattern in two consecutive ventricular intervals enables a pattern code assignment derived from a pool of codes stored in the device. The respective assigned pattern code defines episode classification and SVT or VT rhythm identification. In order to overcome the caveat of sinus tachycardia misinterpretation as VT due to rate-dependent atrioventricular conduction delay and positioning of atrial sensing in the antegrade zone (ECG analogue sinus tachycardia with first-degree AV block classified as short RP tachycardia), the option of operator-dependent customizing of the retrograde–antegrade boundary setting is also provided. Thus, in cases of delayed AV conduction during SVT, the relative size of the antegrade zone should be increased, in contrast to cases of delayed VA conduction, where the relative size of the retrograde zone should be lengthened in order to prevent misinterpretation. The concomitant presence of PR prolongation during SVT with delayed VA conduction during VT renders the tachyarrhythmia differentiation problematic. However, this issue can be addressed by the PR Logic criteria applied in the recent Medtronic ICD models (EnTrust, Virtuoso, and Concerto). The performance of the PR Logic algorithm has been evaluated in several published full-length papers and the reported sensitivity is 100% with a specificity of 66 and 72%.41Go,50Go,51Go


Figure 4
View larger version (24K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 4 PR Logic analysis of AV and VA interval patterns (Medtronic).49Go

 
St Jude Medical dual-chamber algorithm52Go
The first step in the analytical approach of the St Jude Medical algorithm is the comparison of atrial and ventricular rate (rate branch classification). If the ventricular rate is higher than the atrial, then the detected rhythm is classified as VT and therapy is delivered without having to meet further discrimination criteria. When atrial and ventricular rates are equal, then the AV interval discriminator is applied as a pre-qualifier (gate-keeper). In this way, the simultaneous occurrence of sinus tachycardia and VT with coincidentally similar rates can be detected based on the presence of AV dissociation (difference between the second longest and the second shortest AV delay in a programmed detection window). In the next step, sudden-onset and morphology discriminators are implemented. If the ventricular rate is lower than the atrial rate, then the morphology and interval stability criteria are applied with or without the use of the AV interval criterion (assessment of AV dissociation). In the latter two rate branches, the additional applied discriminators can be programmed in ‘if any’ or ‘if all’ diagnostic setting (Figure 5, modified from reference52Go). The sensitivity and specificity of this dual-chamber algorithm are dependent on the selected diagnostic setting and have been reported to vary from 98.7 to 100% and from 73.5 to 90.9%, respectively.52Go,53Go


Figure 5
View larger version (16K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 5 St Jude Medical dual chamber discrimination algorithm flowchart.52Go

 
Special considerations regarding discrimination algorithms
‘Whether or not’ is more important than ‘which one’
Despite the apparent differences, the algorithms introduced by different manufacturers are based on similar paradigms of tachyarrhythmia differentiation. The disparities in the reported specificity of different algorithms for appropriate detection of supraventricular arrhythmias should not be used for direct comparisons, since these differences are at least to a certain degree attributed to variances in the sample sizes of compared studies. Hintringer et al.54Go elegantly demonstrated an inverse correlation between the number of patients and the specificity of dual-chamber ICD detection algorithms. With this in mind, a reliable comparison of the relative efficacy of discrimination algorithms from different manufacturers can be performed only on the basis of a prospective head-to-head randomized clinical trial. The RIGHT study will be the first to evaluate directly the differential efficacy in Guidant vs. Medtronic ICD discrimination algorithms (Rhythm IDTM vs. Enhanced PR LogicTM or WaveletTM).55Go Until the literature acquires new and better data, our main interest should be focused not on ‘which one’, but on ‘whether or not’ the discrimination criteria available in every model have been appropriately utilized in each ICD recipient without complete AV block.

Single- vs. dual-chamber discrimination algorithms
The relative performance of single vs. dual-chamber ICD discriminators is the subject of a long-lasting debate.56Go The implantation of an atrial lead is clearly indicated in the subgroup of ICD recipients with a concomitant indication for permanent pacing or cardiac resynchronization therapy, though without permanent atrial fibrillation.57Go Furthermore, in view of the DAVID-2 results, a dual-chamber ICD can also be selected in order to provide the option of functional AAI pacing, when rate support is considered essential to increase cardiac output, or to titrate the mortality-reducing β-blocker agents in maximally tolerated doses.58Go However, a major issue is whether we have a sufficient amount of evidence to recommend further, on a routine basis, the implantation of a dual-chamber ICD in all patients for the sake of enhanced discrimination performance and avoidance of inappropriate therapies.

Based on the results of recent prospective trials, as well as recently published meta-analyses, dual-chamber discriminators significantly reduce inappropriately treated episodes (per episode analysis), in comparison to single-chamber discriminators, without significantly reducing the number of patients receiving inappropriate discharges (per patient analysis).25Go,59Go–63Go Furthermore, a cost-effectiveness analysis reported that the least costly strategy for device selection among ICD recipients is the a priori universal implantation of dual-chamber ICDs, since the cost of the additional atrial lead is shown to be offset by the expenses associated with the need for future upgrade, even when the latter rate is as low as 5%. However, this elegantly performed analysis is limited by the retrospective study design, the upgrade indications included, and the low generalizability due to significant variance in device cost among different countries; thus, it certainly cannot be considered adequate for policy setting.64Go

Taking together the reported equivalence of the compared discriminators in per-patient analysis, the superiority of dual-chamber discriminators in per-episode analysis—which is, however, prone to bias associated with the relative high contribution of a small number of patients to the total burden of inappropriate episodes53Go—and the rather underestimated susceptibility of dual-chamber discriminators to tachyarrhythmia misclassification due to atrial sensing errors,52Go,53Go we believe that the decision in favour of single- or dual-chamber ICD discriminators should be individualized. When inappropriate shocks recur despite optimal programming of the available single-chamber differentiation criteria, an attempt to implant an atrial lead for enhancing the discrimination performance should be made.

Atrial sensing and dual-chamber discrimination algorithms
Reliable atrial sensing is a strict prerequisite for accurate performance of discrimination algorithms in dual-chamber ICDs. Two frequent pitfalls that may prevent accurate estimation of the atrial rate and thus impair the discrimination performance of dual chamber algorithms are atrial event under-counting due to prolonged or fixed atrial blanking periods, and atrial event over-counting due to far-field R-wave sensing.65Go Far-field R-wave sensing results in over-counting of atrial events and rhythm misclassification. The dual chamber algorithms have introduced several features in order to address these diagnostic challenges. St Jude Medical and Guidant provide the option of reducing the duration of the programmable post-ventricular atrial blanking (PVAB) period, although this entails the risk of atrial rate overestimation due to cross-chamber sensing. In Medtronic ICDs, the manufacturer-recommended PVAB operation maintains the integration of atrial sensing information during PVAB by the arrhythmia detection features in order to prevent atrial under-counting and impaired SVT–VT discrimination. The associated risk of far-field R-wave oversensing is addressed by the implementation of detection criteria based on the recognition of a consistent short–long pattern of atrial sensed events combined with non-physiological values of AV or VA intervals. In the PARAD algorithm, non-physiological PR intervals shorter than a pre-determined value are rejected as far-field sensing.

Empiric vs. tailored discriminator programming
The continuously expanding ICD features and the resulting cumbersome programming render highly attractive option of a standardized parameter setting. The EMPIRIC trial evaluated the hypothesis whether empiric ICD programming is inferior to physician-tailored programming and demonstrated that the former was associated with a lower percentage of SVT episodes resulting in shock and similar percentages of VT/VF episodes resulting in shock.66Go In our view, the results of this trial should not be posed as an argument in favour of the blind application of a standardized set of criteria, which would undermine the need for gaining detailed familiarity with the ICD parameters available. On the other hand, the reduced percentage of inappropriate shocks in the ‘empiric’ arm reflects the significantly higher percentage of ATP programming and PR Logic criteria activation, further establishing the value of discrimination algorithms as well as the primary importance of consistent ATP programming.66Go Nonetheless, a critical appraisal and customization of programmed settings will inevitably be necessary in the case of inappropriate shock or at some stage of subsequent patient follow-up. Thus, the average implanting physician should be aware of the different features, weak points, and advantages of each discrimination algorithm.


    Ventricular oversensing
 Top
 Abstract
 Introduction
 Differentiation of...
 Ventricular oversensing
 Optimal therapy programming for...
 Pharmacological and...
 Future possibilities
 Funding
 References
 
Apart from false detection of SVTs, another cause of inappropriate shocks is ventricular oversensing. This term pertains to a diverse array of situations (T-wave oversensing, electromagnetic interference, myopotential oversensing, oversensing due to sensing lead dysfunction, and double QRS detection) that result in over-counting of ventricular sensed events, thus satisfying the rate criterion and sometimes leading to inappropriate shocks. Recent retrospective studies have reported that ventricular oversensing resulted in inappropriate shocks in 2.3 and 3.9% of the ICD patient population.67Go,68Go The phenomenon of T-wave oversensing is a rare cause of inappropriate shocks and may be triggered by electrolyte abnormalities and drugs. In order to prevent T-wave oversensing, newer generation ICDs automatically adjust the sensitivity threshold, increasing its value after ventricular events and then allowing a smooth decay with time. Particular caution is necessary in cardiac diseases associated with a low amplitude intracardiac ventricular electrogram (e.g. arrhythmogenic right ventricular cardiomyopathy, and sarcoidosis), tall T-waves (e.g. short QT syndrome),69Go and a long distance of T-wave from R-wave (long QT syndrome), which increase the risk of inappropriate therapy due to T-wave oversensing. Several manufacturers provide the option of specialized programmable parameters in order to overcome these caveats. Biotronik and St Jude Medical enable a programmable time delay before the decay in ventricular sensitivity as well as the possibility of customizing the initial value of ventricular sensitivity as a percentage of the R-wave amplitude.

Inappropriate therapy due to myopotential oversensing may occur in case of programming maximal sensitivity levels (lowest value for the sensitivity setting), although in most cases it is indicative of an insulation defect. When myopotential oversensing is suspected, a provocative test should be implemented, such as shoulder flexion with adduction against resistance.

A tool to prevent the occurrence of inappropriate shocks due to ventricular oversensing is the activation of available alert features triggered when the lead impedance value is out of a programmable range. In a few cases, the immediate patient alert gives the opportunity to the notified physician to proceed with timely lead revision, thus avoiding inappropriate therapy. However, ventricular oversensing is usually an early sign of lead dysfunction, preceding lead fracture and a rise in impedance values. In these cases, inappropriate therapy is the first clue to lead malfunction and its occurrence cannot be prevented by the activation of alert features.


    Optimal therapy programming for shock reduction
 Top
 Abstract
 Introduction
 Differentiation of...
 Ventricular oversensing
 Optimal therapy programming for...
 Pharmacological and...
 Future possibilities
 Funding
 References
 
Since the advent of tiered-therapy, non-committed ICDs, the therapeutic alternative of antitachycardiac pacing (ATP) is available for tachyarrhythmia termination. Several studies have shown that ATP successfully terminates ~90% of VTs having a rate <200 bpm, with an acceptable risk of VT acceleration.70Go–72Go In the PainFREE Rx II trial, a single 8-pulse burst pacing train was effective in terminating 72% of fast VT episodes without impairing patient safety,21Go while in the EMPIRIC trial the consistent ATP application in the ‘empiric’ arm resulted in a significant reduction in the shocked episodes of monomorphic VT when compared with the ‘tailored’ arm, without an increase in the incidence of VT acceleration or syncope.66Go This evidence-based tendency to reposition ICD therapy on a more ATP-oriented basis in order to decrease the number of shocks may also have a favourable impact on the rate of inappropriate shocks. Indeed, the programming of repetitive ATPs, especially in the slow VT zone where VT–SVT rate overlap is maximal, can prolong the time to shock delivery and thus provide the opportunity for some SVTs to self-terminate before shock administration. Furthermore, ventricular ATP may rarely terminate atrial tachycardias, as well as a considerable percentage of junctional tachycardias,46Go although the latter group of SVTs represents a rare cause of inappropriate shocks.

Another decisive issue for the reduction of inappropriate therapies is the optimized programming of the number of intervals to detect (NID). Increase of the selected NID above the nominal values prolongs the detection window, thus allowing non-sustained VTs or short runs of pseudo- regularized atrial fibrillation to self-terminate before detection criteria are fulfilled and therapies delivered. The conferred benefit in inappropriate shock reduction has been shown not to be accompanied by under-detection of VTs or arrhythmic syncopes.73Go,74Go


    Pharmacological and interventional approaches
 Top
 Abstract
 Introduction
 Differentiation of...
 Ventricular oversensing
 Optimal therapy programming for...
 Pharmacological and...
 Future possibilities
 Funding
 References
 
The cornerstone for avoidance of inappropriate therapies is optimal device programming with the key issues of programming ATP therapies in slow as well as in fast VT zones, increasing the duration of the detection window and activating discrimination algorithms without time overrides. Furthermore, pharmacological and interventional measures can confer an incremental benefit. The dosage of the mortality-reducing β-blockers should be maximized in order to decrease the ventricular rates during supraventricular arrhythmias and thus to minimize the odds of SVT rates falling within VT/VF detection zones. Additionally, antiarrhythmic agents can be used in certain cases in order to prevent supraventricular and non-sustained ventricular tachycardia occurrences that might lead to inappropriate therapies. When the use of antiarrhythmic agents is considered necessary, the first choice should be amiodarone, while sotalol or mexiletine can be a second choice when amiodarone either is ineffective or is discontinued because of adverse effects. However, the increased likelihood of drug-related cardiac and non-cardiac adverse events, especially during long-term administration, and the effect of amiodarone on the defibrillation threshold, which is minor but still clinically meaningful in a minority of cases, represent caveats that should be taken into consideration during clinical decision making.

Catheter ablation constitutes an alternative approach to antiarrhythmic agents, especially for easily amenable arrhythmias such as typical atrial flutter, atrioventricular nodal re-entry tachycardia, or atrioventricular re-entry tachycardias. In the rare case of failure to control ventricular rate with the above-mentioned measures, atrioventricular node ablation may be the ultima ratio.

The strategies that can be implemented for the reduction of inappropriate ICD shocks are summarized in Table 2.


View this table:
[in this window]
[in a new window]

 
Table 2 Strategies for decreasing the likelihood of inappropriate shocks

 

    Future possibilities
 Top
 Abstract
 Introduction
 Differentiation of...
 Ventricular oversensing
 Optimal therapy programming for...
 Pharmacological and...
 Future possibilities
 Funding
 References
 
Novel methods of rhythm discrimination have provided the rationale for the future development of promising algorithms. Proposed parameters that may be used for proper arrhythmia classification in ICD recipients are: the variability of the first post-pacing interval between consecutive bursts of different length,75Go analysis of atrial response patterns during and after ATP,76Go the response of the arrhythmia to simultaneous right atrial and ventricular pacing,77Go and atrial impedance amplitude during the cardiac cycle.78Go The ideal discriminator would be based on the detection of the impact of the tachyarrhythmia on haemodynamic stability and would postpone shock therapy irrespective of the tachycardia rate, as long as an adequate blood pressure level is maintained and pulmonary oedema is avoided. However, taking into consideration the gaping chasm between the ideal ICD and the current reality in everyday clinical practice, as well as the problematic availability of the recently developed technical capabilities, mainly for economic reasons, the above-mentioned discrimination scheme seems unlikely to come to fruition in the near future. Until then, detailed knowledge and use of all available programming options, guided by the special characteristics of each unique patient, provide the only feasible way to achieve the difficult task of reducing inappropriate shocks.


    Funding
 Top
 Abstract
 Introduction
 Differentiation of...
 Ventricular oversensing
 Optimal therapy programming for...
 Pharmacological and...
 Future possibilities
 Funding
 References
 
This work was supported by the Hellenic Cardiovascular Research Society. S.T. is supported by a training fellowship provided by the European Heart Rhythm Association.

Conflict of interest: C.K. has received lecture fees from Biotronik, Medtronic, Sorin, and St Jude Medical. C.K. is an advisor to Biotronik and St Jude Medical.


    References
 Top
 Abstract
 Introduction
 Differentiation of...
 Ventricular oversensing
 Optimal therapy programming for...
 Pharmacological and...
 Future possibilities
 Funding
 References
 
[1] The Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. New Engl J Med (1997) 337:1576–83.[Abstract/Free Full Text]

[2] Connolly SJ, Gent M, Roberts RS, Dorian P, Roy D, Sheldon RS, et al. Canadian Implantable Defibrillator Study (CIDS): a randomized trial of the implantable cardioverter defibrillator against amiodarone. Circulation (2000) 101:1297–302.[Abstract/Free Full Text]

[3] Moss AJ, Hall WJ, Cannom DS, Daubert JP, Higgins SL, Klein H, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia: Multicenter Automatic Defibrillator Implantation Trial Investigators. N Engl J Med (1996) 335:1933–40.[Abstract/Free Full Text]

[4] Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med (2002) 346:877–83.[Abstract/Free Full Text]

[5] Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med (2005) 352:225–37.[Abstract/Free Full Text]

[6] Kenigsberg DN, Ellenbogen KA. ICD arrhythmia detection and discrimination: are we there yet? J Cardiovasc Electrophysiol (2006) 17:1320–2.[CrossRef][Web of Science][Medline]

[7] Glikson M, Lipchenca I, Viskin S, Ballman KV, Trusty JM, Gurevitz OT, et al. Long-term outcome of patients who received implantable cardioverter defibrillators for stable ventricular tachycardia. J Cardiovasc Electrophysiol (2004) 15:658–64.[CrossRef][Web of Science][Medline]

[8] Hreybe H, Ezzeddine R, Barrington W, Bazaz R, Jain S, Ngwu O, et al. Relation of advanced heart failure symptoms to risk of inappropriate defibrillator shocks. Am J Cardiol (2006) 97:544–6.[CrossRef][Web of Science][Medline]

[9] Germano JJ, Reynolds M, Essebag V, Josephson ME. Frequency and causes of implantable cardioverter-defibrillator therapies: is device therapy proarrhythmic? Am J Cardiol (2006) 97:1255–61.[CrossRef][Web of Science][Medline]

[10] Sweeney MO, Wathen MS, Volosin K, Abdalla I, DeGroot PJ, Otterness MF, et al. Appropriate and inappropriate ventricular therapies, quality of life, and mortality among primary and secondary prevention implantable cardioverter defibrillator patients: results from the Pacing Fast VT REduces Shock ThErapies (PainFREE Rx II) trial. Circulation (2005) 111:2898–905.[Abstract/Free Full Text]

[11] Heller SS, Ormont MA, Lidagoster L, Sciacca RR, Steinberg S. Psychosocial outcome after ICD implantation: a current perspective. Pacing Clin Electrophysiol (1998) 21:1207–15.[CrossRef][Medline]

[12] Lemon J, Edelman S, Kirkness A. Avoidance behaviors in patients with implantable cardioverter defibrillators. Heart Lung (2004) 33:176–82.[CrossRef][Web of Science][Medline]

[13] Kamphuis HC, de Leeuw JR, Derksen R, Hauer RN, Winnubst JA. Implantable cardioverter defibrillator recipients: quality of life in recipients with and without ICD shock delivery: a prospective study. Europace (2003) 5:381–9.[Abstract/Free Full Text]

[14] Bilge AK, Ozben B, Demircan S, Cinar M, Yilmaz E, Adalet K. Depression and anxiety status of patients with implantable cardioverter defibrillator and precipitating factors. Pacing Clin Electrophysiol (2006) 29:619–26.[CrossRef][Medline]

[15] Ladwig KH, Deisenhofer I, Simon H, Schmitt C, Baumert JJ. Characteristics associated with low treatment satisfaction in patients with implanted cardioverter defibrillators: results from the LICAD study. Pacing Clin Electrophysiol (2005) 28:506–13.[CrossRef][Medline]

[16] Vollmann D, Luthje L, Vonhof S, Unterberg C. Inappropriate therapy and fatal proarrhythmia by an implantable cardioverter-defibrillator. Heart Rhythm (2005) 2:307–9.[CrossRef][Web of Science][Medline]

[17] Healy E, Goyal S, Browning C, Robotis D, Ramaswamy K, Rofino-Nadoworny K, et al. Inappropriate ICD therapy due to proarrhythmic ICD shocks and hyperpolarization. Pacing Clin Electrophysiol (2004) 27:415–6.[CrossRef][Medline]

[18] Veltmann C, Borggrefe M, Schimpf R, Wolpert C. Fatal inappropriate ICD shock. J Cardiovasc Electrophysiol (2007) 18:326–8.[CrossRef][Web of Science][Medline]

[19] Rosenqvist M, Beyer T, Block M, den Dulk K, Minten J, Lindemans F. Adverse events with transvenous implantable cardioverterdefibrillators: a prospective multicenter study. European 7219 Jewel ICD investigators. Circulation (1998) 98:663–70.[Abstract/Free Full Text]

[20] Wilkoff BL, Hess M, Young J, Abraham WT. Differences in tachyarrhythmia detection and implantable cardioverter defibrillator therapy by primary or secondary prevention indication in cardiac resynchronization therapy patients. J Cardiovasc Electrophysiol (2004) 15:1002–9.[CrossRef][Web of Science][Medline]

[21] Wathen MS, DeGroot PJ, Sweeney MO, Stark AJ, Otterness MF, Adkisson WO, et al. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (Pain-FREE Rx II) trial results. Circulation (2004) 110:2591–6.[Abstract/Free Full Text]

[22] Kolb C, Zrenner B, Schmitt C. Incidence of electromagnetic interference in implantable cardioverter defibrillators. Pacing Clin Electrophysiol (2001) 24:465–8.[CrossRef][Medline]

[23] Brugada J, Mont L, Figueiredo M, Valentino M, Matas M, Navarro-López F. Enhanced detection criteria in implantable defibrillators. J Cardiovasc Electrophysiol (1998) 9:261–8.[Web of Science][Medline]

[24] Swerdlow CD, Ahern T, Chen PS, Hwang C, Gang E, Mandel W, et al. Underdetection of ventricular tachycardia by algorithms to enhance specificity in a tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol (1994) 24:416–24.[Abstract]

[25] Bänsch D, Steffgen F, Grönefeld G, Wolpert C, Böcker D, Mletzko RU, et al. The 1+1 trial: a prospective trial of a dual- versus a single-chamber implantable defibrillator in patients with slow ventricular tachycardias. Circulation (2004) 110:1022–9.[Abstract/Free Full Text]

[26] Schaumann A, von zur Mühlen F, Gonska BD, Kreuzer H. Enhanced detection criteria in implantable cardioverter-defibrillators to avoid inappropriate therapy. Am J Cardiol (1996) 78:42–50.[Web of Science][Medline]

[27] Swerdlow CD, Chen PS, Kass RM, Allard JR, Peter CT. Discrimination of ventricular tachycardia from sinus tachycardia and atrial fibrillation in a tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol (1994) 23:1342–55.[Abstract]

[28] Weber M, Böcker D, Bänsch D, Brunn J, Castrucci M, Gradaus R, et al. Efficacy and safety of the initial use of stability and onset criteria in implantable cardioverter defibrillators. J Cardiovasc Electrophysiol (1999) 10:145–53.[Web of Science][Medline]

[29] Kettering K, Dörnberger V, Lang R, Vonthein R, Suchalla R, Bosch RF, et al. Enhanced detection criteria in implantable cardioverter defibrillators: sensitivity and specificity of the stability algorithm at different heart rates. Pacing Clin Electrophysiol (2001) 24:1325–33.[CrossRef][Medline]

[30] García-Alberola A, Yli-Mäyry S, Block M, Haverkamp W, Martínez-Rubio A, Kottkamp H, et al. RR interval variability in irregular monomorphic ventricular tachycardia and atrial fibrillation. Circulation (1996) 93:295–300.[Abstract/Free Full Text]

[31] Klein GJ, Gillberg JM, Tang A, Inbar S, Sharma A, Unterberg-Buchwald C, et al. Improving SVT discrimination in single chamber ICDs—a new electrogram morphology based algorithm. J Cardiovasc Electrophysiol (2006) 17:1310–9.[CrossRef][Web of Science][Medline]

[32] Boriani G, Occhetta E, Pistis G, Menozzi C, Jorfida M, Sermasi S, et al. Combined use of morphology discrimination, sudden onset, and stability as discriminating algorithms in single chamber cardioverter defibrillators. Pacing Clin Electrophysiol (2002) 25:1357–66.[CrossRef][Medline]

[33] Theuns DA, Rivero-Ayerza M, Goedhart DM, van der Perk R, Jordaens LJ. Evaluation of morphology discrimination for ventricular tachycardia diagnosis in implantable cardioverter-defibrillators. Heart Rhythm (2006) 3:1332–8.[CrossRef][Web of Science][Medline]

[34] Grönefeld GC, Schulte B, Hohnloser SH, Trappe HJ, Korte T, Stellbrink C, et al. Morphology discrimination: a beat-to-beat algorithm for the discrimination of ventricular from supraventricular tachycardia by implantable cardioverter defibrillators. Pacing Clin Electrophysiol (2001) 24:1519–24.[CrossRef][Medline]

[35] Gold MR, Shorofsky SR, Thompson JA, Kim J, Schwartz M, Bocek J, et al. Advanced rhythm discrimination for implantable cardioverter defibrillators using electrogram vector timing and correlation. J Cardiovasc Electrophysiol (2002) 13:1092–7.[CrossRef][Web of Science][Medline]

[36] Corbisiero R, Lee MA, Nabert DR, Coman JA, Breiter DJ, Schwartz M, et al. Performance of a new single-chamber ICD algorithm: discrimination of supraventricular and ventricular tachycardia based on vector timing and correlation. Europace (2006) 8:1057–61.[Abstract/Free Full Text]

[37] Compton SJ, Merrill JJ, Dorian P, Cao J, Zhou D, Gillberg JM. Continuous template collection and updating for electrogram morphology discrimination in implantable cardioverter defibrillators. Pacing Clin Electrophysiol (2006) 29:244–54.[CrossRef][Medline]

[38] Lüthje L, Vollmann D, Rosenfeld M, Unterberg-Buchwald C. Electrogram configuration and detection of supraventricular tachycardias by a morphology discrimination algorithm in single chamber ICDs. Pacing Clin Electrophysiol (2005) 28:555–60.[CrossRef][Medline]

[39] Wolber T, Binggeli C, Holzmeister J, Brunckhorst C, Strobel U, Boes C, et al. Wavelet-based tachycardia discrimination in ICDs: impact of posture and electrogram configuration. Pacing Clin Electrophysiol (2006) 29:1255–60.[CrossRef][Medline]

[40] Aliot E, Nitzsché R, Ripart A. Arrhythmia detection by dual-chamber implantable cardioverter defibrillators. A review of current algorithms. Europace (2004) 6:273–86.[Abstract/Free Full Text]

[41] Wilkoff BL, Kuhlkamp V, Volosin K, Ellenbogen K, Waldecker B, Kacet S, et al. Critical analysis of dual chamber implantable cardioverter-defibrillator arrhythmia detection: Results and technical considerations. Circulation (2001) 103:381–6.[Abstract/Free Full Text]

[42] Smart Detection algorithm functioning. Biotronik Lumax Function Manual.

[43] Theuns D, Klootwijk AP, Kimman GP, Szili-Török T, Roelandt JR, Jordaens L. Initial clinical experience with a new arrhythmia detection algorithm in dual chamber implantable cardioverter defibrillators. Europace (2001) 3:181–6.[Abstract/Free Full Text]

[44] Sinha AM, Stellbrink C, Schuchert A, Mox B, Jordaens L, Lamaison D, et al. Clinical experience with a new detection algorithm for differentiation of supraventricular from ventricular tachycardia in a dual-chamber defibrillator. J Cardiovasc Electrophysiol (2004) 15:646–52.[CrossRef][Web of Science][Medline]

[45] PARAD/PARAD+:. Cardiac Rhythm Management: Sorin group. http://www.sorin-crm.com/sorin_elamedical/read/intl-medical-professionalstachyarrhythmia-management-features-parad-parad-.

[46] Anselme F, Mletzko R, Bowes R, Mabo P, Sadoul N, Schoels W, et al. Prevention of inappropriate shocks in ICD recipients: a review of 10,000 tachycardia episodes. Pacing Clin Electrophysiol (2007) 30:S128–S133.[Medline]

[47] The Boston Scientific Cardiac Rhythm Management Compass. (2007) Rhythm ID—DR Detection Enhancement Mode.

[48] Lee MA, Corbisiero R, Nabert DR, Coman JA, Giudici MC, Tomassoni GF, et al. Clinical results of an advanced SVT detection enhancement algorithm. Pacing Clin Electrophysiol (2005) 28:1032–40.[CrossRef][Medline]

[49] Medtronic Virtuoso DR Manual. http://manuals.medtronic.com/OrderManagement.

[50] Dijkman B, Wellens HJ. Dual chamber arrhythmia detection in the implantable cardioverter defibrillator. J Cardiovasc Electrophysiol (2000) 11:1105–15.[Web of Science][Medline]

[51] Kühlkamp V, Wilkoff BL, Brown AB, Volosin KJ, Hügl BJ, Stafford W, et al. Experience with a dual chamber implantable defibrillator. Pacing Clin Electrophysiol (2002) 25:1041–8.[CrossRef][Medline]

[52] Boriani G, Biffi M, Dall'Acqua A, Martignani C, Frabetti L, Zannoli R, et al. Rhythm discrimination by rate branch and QRS morphology in dual chamber implantable cardioverter defibrillators. Pacing Clin Electrophysiol (2003) 26:466–70.[CrossRef][Medline]

[53] Bailin SJ, Niebauer M, Tomassoni G, Leman R, The Photon Investigators. Clinical investigation of a new dual-chamber implantable cardioverter defibrillator with improved rhythm discrimination capabilities. J Cardiovasc Electrophysiol (2003) 14:144–9.[Web of Science][Medline]

[54] Hintringer F, Deibl M, Berger T, Pachinger O, Roithinger FX. Comparison of the specificity of implantable dual chamber defibrillator detection algorithms. Pacing Clin Electrophysiol (2004) 27:976–82.[CrossRef][Medline]

[55] Berger RD, Lerew DR, Smith JM, Pulling C, Gold MR. The Rhythm ID Going Head to Head Trial (RIGHT): design of a randomized trial comparing competitive rhythm discrimination algorithms in implantable cardioverter defibrillators. J Cardiovasc Electrophysiol (2006) 17:749–53.[CrossRef][Web of Science][Medline]

[56] Gatzoulis KA, Gialafos IE. Dual chamber antitachycardia cardioverter defibrillators. Another step for the treatment of high-risk patients. Hellenic J Cardiol (2003) 44:71–9.

[57] Vardas PE, Auricchio A, Blanc JJ, Daubert JC, Drexler H, Ector H, et al. European Society of Cardiology; European Heart Rhythm Association. Guidelines for cardiac pacing and cardiac resynchronization therapy. The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Europace (2007) 9:959–98.[Free Full Text]

[58] Cleland JG, Coletta AP, Abdellah AT, Witte KK, Hobson N, Clark AL. Clinical trials update from Heart Rhythm 2007 and Heart Failure 2007: CARISMA, PREPARE, DAVID II, SAVE-PACE, PROTECT and AREA-IN-CHF. Eur J Heart Fail (2007) 9:850–3.[Abstract/Free Full Text]

[59] Theuns DA, Rivero-Ayerza M, Boersma E, Jordaens L. Prevention of inappropriate therapy in implantable defibrillators: a meta-analysis of clinical trials comparing single-chamber and dual-chamber arrhythmia discrimination algorithms. Int J Cardiol (2008) 125:352–7.[CrossRef][Web of Science][Medline]

[60] Deisenhofer I, Kolb C, Ndrepepa G, Schreieck J, Karch M, Schmieder S, et al. Do current dual chamber cardioverter defibrillators have advantages over conventional single chamber cardioverter defibrillators in reducing inappropriate therapies? A randomized, prospective study. J Cardiovasc Electrophysiol (2001) 12:134–42.[CrossRef][Web of Science][Medline]

[61] Kuhlkamp V, Dornberger V, Mewis C, Suchalla R, Bosch RF, Seipel L. Clinical experience with the new detection algorithms for atrial fibrillation of a defibrillator with dual chamber sensing and pacing. J Cardiovasc Electrophysiol (1999) 10:905–15.[Web of Science][Medline]

[62] Friedman PA, McClelland RL, Bamlet WR, Acosta H, Kessler D, Munger TM, et al. Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis: the Detect SupraVentricular Tachycardia study. Circulation (2006) 113:2871–9.[Abstract/Free Full Text]

[63] Theuns DA, Klootwijk AP, Goedhart DM, Jordaens LJ. Prevention of inappropriate therapy in implantable cardioverter-defibrillators: results of a prospective, randomized study of tachyarrhythmia detection algorithms. J Am Coll Cardiol (2004) 44:2362–7.[Abstract/Free Full Text]

[64] Goldberger Z, Elbel B, McPherson CA, Paltiel AD, Lampert R. Cost advantage of dual-chamber versus single-chamber cardioverter-defibrillator implantation. J Am Coll Cardiol (2005) 46:850–7.[Abstract/Free Full Text]

[65] Kolb C, Wille B, Maurer D, Schuchert A, Weber R, Schibgilla V, et al. Management of far-field R wave sensing for the avoidance of inappropriate mode switch in dual chamber pacemakers: results of the FFS-test study. J Cardiovasc Electrophysiol (2006) 17:992–7.[CrossRef][Web of Science][Medline]

[66] Wilkoff BL, Ousdigian KT, Sterns LD, Wang ZJ, Wilson RD, Morgan JM. A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicenter EMPIRIC trial. J Am Coll Cardiol (2006) 48:330–9.[Abstract/Free Full Text]

[67] Rauwolf T, Guenther M, Hass N, Schnabel A, Bock M, Braun MU, et al. Ventricular oversensing in 518 patients with implanted cardiac defibrillators: incidence, complications, and solutions. Europace (2007) 9:1041–7.[Abstract/Free Full Text]

[68] Occhetta E, Bortnik M, Magnani A, Francalacci G, Marino P. Inappropriate implantable cardioverter-defibrillator discharges unrelated to supraventricular tachyarrhythmias. Europace (2006) 8:863–9.[Abstract/Free Full Text]

[69] Schimpf R, Wolpert C, Gaita F, Giustetto C, Borggrefe M. Short QT syndrome. Cardiovasc Res (2005) 67:357–66.[Abstract/Free Full Text]

[70] Wathen MS, Sweeney M, DeGroot PJ, Stark AJ, Koehler JL, Chrisner MB, et al. Shock reduction using antitachycardia pacing for spontaneous rapid ventricular tachycardia in patients with coronary artery disease. Circulation (2001) 104:796–801.[Abstract/Free Full Text]

[71] Yee R, Klein G, Guiradon GM, Jones DL, Sharma Ad, Norris C. Initial clinical experience with the pacemaker cardioverter-defibrillator. Can J Cardiol (1990) 6147–56.

[72] Luceri R, Habal SM, David IB, Puchferran RL, Murtore C. Changing trends in therapy delivery with a third generation noncommitted implantable cardioverter defibrillator: results of a large single center clinical trial. Pacing Clin Electrophysiol (1993) 16:159–64.[CrossRef][Medline]

[73] Wilkoff BL, Stern R, Williamson B, Wathen M, Holloman K, Fieberg A, et al. Aggressively programmed implantable defibrillator detection and therapy reduces shocks compared to physician-tailored programming in primary prevention patients: results from the PREPARE trial. Heart Rhythm Society 2007 Scientific Sessions, 10 May 2007: Denver, CO.

[74] Gunderson BD, Abeyratne AI, Olson WH, Swerdlow CD. Effect of programmed number of intervals to detect ventricular fibrillation on implantable cardioverter-defibrillator aborted and unnecessary shocks. Pacing Clin Electrophysiol (2007) 30:157–65.[CrossRef][Medline]

[75] Arenal A, Ortiz M, Peinado R, Merino JL, Quesada A, Atienza F, et al. Differentiation of ventricular and supraventricular tachycardias based on the analysis of the first postpacing interval after sequential anti-tachycardia pacing in implantable cardioverter-defibrillator patients. Heart Rhythm (2007) 4:316–22.[CrossRef][Web of Science][Medline]

[76] Ridley DP, Gula LJ, Krahn AD, Skanes AC, Yee R, Brown ML, et al. Atrial response to ventricular antitachycardia pacing discriminates mechanism of 1:1 atrioventricular tachycardia. J Cardiovasc Electrophysiol (2005) 16:601–5.[CrossRef][Web of Science][Medline]

[77] Saba S, Baker L, Ganz L, Barrington W, Jain S, Ngwu O, et al. Simultaneous atrial and ventricular anti-tachycardia pacing as a novel method of rhythm discrimination. J Cardiovasc Electrophysiol (2006) 17:695–701.[CrossRef][Web of Science][Medline]

[78] Schmidt B, Asbach S, Schweika O, Zehender M, Bode C, Faber TS. Atrial fibrillation reduces the atrial impedance amplitude during cardiac cycle: a novel detection algorithm to improve recognition of atrial fibrillation in pacemaker patients. Europace. (2007) 9:812–6.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
EuropaceHome page
Task force members, J. Vijgen, G. Botto, J. Camm, C.-J. Hoijer, W. Jung, J.-Y. Le Heuzey, A. Lubinski, T. M. Norekval, M. Santomauro, et al.
Consensus statement of the European Heart Rhythm Association: updated recommendations for driving by patients with implantable cardioverter defibrillators
Europace, August 1, 2009; 11(8): 1097 - 1107.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
10/11/1256    most recent
eun205v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Tzeis, S.
Right arrow Articles by Vardas, P. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tzeis, S.
Right arrow Articles by Vardas, P. E.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?