Skip Navigation



Europace Advance Access published online on September 5, 2008

Europace, doi:10.1093/europace/eun257
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
10/12/1375    most recent
eun257v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Nägele, H.
Right arrow Articles by Eisermann, C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nägele, H.
Right arrow Articles by Eisermann, C.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2008. For permissions please email: journals.permissions@oxfordjournals.org

Cardiac contractility modulation in non-responders to cardiac resynchronization therapy

H. Nägele*, S. Behrens and C. Eisermann

St Adolfstift, Medical Clinic, Hamburger Str. 41, Reinbek D-21465, Germany

Aims: Cardiac resynchronization therapy (CRT) has become a standard therapy in cases of heart failure and asynchrony. Unfortunately, 20–30% of patients were non-responsive (NR) to CRT. In this report we used cardiac contractility modulation (CCM) as an adjunctive measure in NR patients.

Methods and results: Sixteen NR patients, mean age 65 ± 9 years, mean ejection fraction 27.3 ± 7.4%, and New York Heart Association (NYHA) class III (n = 9) or IV (n = 7) despite CRT plus optimized medical therapy, received an additional CCM-implantation contra-lateral to the existing CRT system (OPTIMIZER III, Impulse Dynamics, Orangeburg, NY, USA). Cardiac contractility modulation delivers non-excitatory high-energy stimulatory impulses during the absolute refractory period, thus improving contractility [left ventricular (LV) dp/dt)] by stimulating the septum with two screw-in leads and one additional atrial lead for triggering the impulses. Acute LV dp/dt changes induced by CCM stimulation were measured by 5F Millar catheters placed in the LV during the implantation procedure in 14 of 16 cases. Patients were followed prospectively. Left ventricular dp/dt increased from a mean of 568 ± 153 to 646 ± 147 mmHg/s (+14%, P < 0.001) in the acute intraoperative testing. We noted the following complications and events during a follow-up of an average of 147 ± 80 days (range 68–326) after CCM: intraoperative ventricular flutter needing cardioversion (n = 1), atrial lead dislocation (n = 1), coronary sinus (CS) lead dislocation (n = 1), painful stimulation requiring repositioning of septal leads (n = 1), true defibrillator shocks (n = 3), cardiac decompensations (n = 3), atrial fibrillation (n = 4), renal failure (n = 1), and pneumonia (n = 2). NYHA class improved from 3.4 to 2.8 (P < 0.01), and the ejection fraction increased from 27.3 ± 5 to 31.1 ± 6 (P < 0.01). Three patients (19%) died suddenly presumably due to electromechanical dissociation after 318, 104, and 81 days. No electrical interference was observed between the CCM and CRT systems, and in particular, at no time was the CRT–implantable cardioverter-defibrillator found to be delivering inadequate shocks.

Conclusion: The CCM method is feasible and could be applied with calculated risks as a possible useful adjunct in CRT-NR when no other options are available; however, mortality and event rates are high in this very sick population.

Key Words: Heart failure, Cardiac resynchronization therapy, Non-responder, Non-excitatory stimulation


* Corresponding author. Tel: +49 40 7280 5158; fax: +49 40 7280 2729. E-mail address: herbert.naegele{at}krankenhaus-reinbek.de

Manuscript submitted 9 June 2008. Accepted after revision 15 August 2008.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.