Europace Advance Access originally published online on May 16, 2007
Europace 2007 9(7):523-527; doi:10.1093/europace/eum082
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
BIVENTRICULAR PACING
Single-centre experience with coronary sinus lead stability and long-term pacing parameters
Department of Internal Medicine, University Hospital, Olomouc, Czech Republic
Aims To compare both short- and long-term stability and pacing parameters of CS leads. Long-term maintenance of coronary sinus (CS) pacing lead position has not yet been assessed by serial X-ray digital imaging.
Methods and results Using X-ray digital imaging in anteroposterior and left lateral views, we performed analysis of the CS lead position at the end of the implantation procedure, 48 h following implantation and after 12 months in 93 consecutive patients. Attain OTW leads, QuickSite leads and EasyTrak leads were implanted in 44, 12 and 37 cases, respectively. X-ray detectable late CS lead dislodgement was observed in 8 cases (8.6%) during 12-months follow-up. Dislodgment was once accompanied by loss of capture and need for repositioning. Another two major dislodgements occurred without any threshold change. Five (5.4%) minor dislocations occurred with either a decrease or an increase in pacing threshold. Left ventricular (LV) pacing threshold was changed by
1 V after 12 months in 14 patients (15%) when compared to values 48 h after implantation. All but one case of
1 V threshold change (either positive or negative) occurred without any radiological evidence of lead displacement. LV pacing threshold was significantly higher than RV pacing threshold at both baseline and during follow-up. Leads from all three manufacturers had comparable pacing thresholds 12 months following implantation.
Conclusion The position of CS leads may change during 1-year follow-up. However, it rarely required reoperation in the selected group of patients. Long-term pacing parameters between different manufacturers are comparable. Increase in pacing threshold is not a reliable marker of CS lead dislodgement.
Key Words: CS lead stability, Long-term performance, Biventricular pacing
* Corresponding author: Cardiocenter, Hospital
eské Bud
jovice, B. N
mcové 585/54, 370 87
eské Bud
jovice, Czech Republic. Tel: +420 588443201. E-mail address: alanbulava{at}seznam.czManuscript submitted 13 December 2006. Accepted after revision 30 March 2007.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
G. Z. Duray, J. Schmitt, S. Cicek-Hartvig, S. H. Hohnloser, and C. W. Israel Complications leading to surgical revision in implantable cardioverter defibrillator patients: comparison of patients with single-chamber, dual-chamber, and biventricular devices Europace, March 1, 2009; 11(3): 297 - 302. [Abstract] [Full Text] [PDF] |
||||