Europace 2006 8(6):443-448; doi:10.1093/europace/eul041
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
RADIATION EXPOSURE IN EP
Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis
1 2nd Department of Radiology, Medical Physics Unit, University of Athens, Athens, Greece; 2 Department of Cardiology, Athens Euroclinic, 9 Athanassiadou Street, 115 21 Athens, Greece; 3 Department of Medical Physics, University of Athens, Athens, Greece; 4 Quality Assurance Reference Centre, Newcastle General Hospital, Newcastle-upon-Tyne, UK
Aims To perform a comprehensive analysis of all aspects of patient and in-room personnel radiation dosimetry in interventional electrophysiology.
Methods and results Measurements were performed during 19 diagnostic electrophysiology studies and 24 catheter ablations. Kerma-area product and exposure time values were 48.7 (6.4230) Gy cm2 and 25.5 (4.479.2) min for ablation, and 12.5 (4.5117.2) Gy cm2 and 4.5 (1.231) min for diagnostic studies, respectively. Patient effective doses were 15.2 (2.159.6) mSv for ablation and 3.2 (1.323.9) mSv for diagnostic procedures. Radiation risk to the patient was estimated to be up to eight cases of fatal cancer in 10 000 procedures. The risk of development of fatal cancer was less than 3x106 per procedure to the primary operator. The risk for the nurse and technician was much lower. The dose per procedure for the primary operator was 7.1 µGy at the eyes, 0.79 µGy at the chest under the lead apron, 13.68 µGy at the chest over the apron, 3.82 µGy at the thyroid, 17.76 µGy at the left hand, and 12.11 µGy at the left knee.
Conclusion As far as radiation exposure is concerned, electrophysiology studies followed by radiofrequency ablation are safe procedures for both patient and personnel when performed in catheterization laboratories with modern equipment, experienced operators, and standard safety precautions.
Key Words: Electrophysiology, Ablation, Radiation, Dosimetry
* Corresponding author. Tel: +30 210 6416600; fax: +30 210 6819779. E-mail address: dkatritsis{at}euroclinic.gr
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
H. Niinuma, R. T. George, A. Arbab-Zadeh, J. A.C. Lima, and C. A. Henrikson Imaging of pulmonary veins during catheter ablation for atrial fibrillation: the role of multi-slice computed tomography Europace, November 1, 2008; 10(suppl_3): iii14 - iii21. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ector, O. Dragusin, B. Adriaenssens, W. Huybrechts, R. Willems, H. Ector, and H. Heidbuchel Obesity Is a Major Determinant of Radiation Dose in Patients Undergoing Pulmonary Vein Isolation for Atrial Fibrillation J. Am. Coll. Cardiol., July 17, 2007; 50(3): 234 - 242. [Abstract] [Full Text] [PDF] |
||||