Europace Advance Access originally published online on March 16, 2006
Europace 2006 8(4):288-292; doi:10.1093/europace/eul009
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
ICD
The RIONI study rationale and design: validation of the first stored electrograms transmitted via home monitoring in patients with implantable defibrillators
1 Department of CardiologyUniversity of BochumBochum Germany; 2 Department of Cardiology and AngiologyMedical School HannoverCarl-Neuberg-Street 1, D-30625 Hannover Germany; 3 Institute of Health Science and TechnologyAalborg UniversityAalborg Denmark; 4 Department of Cardiology, Aalborg SygehusÅrhus Universitu Hospitals Aalborg Denmark; 5 Department of CardiologyHospital Ramón y CajalMadrid Spain; 6 Department of CardiologyCentre Hospitalier Régional Universitaire LilleLille France; 7 Department of Cardiology, Medizinische Klinik und Poliklinik CUniv.-Klinikum MünsterMünster Germany
Appropriate and inappropriate therapies of implantable cardioverter defibrillators (ICDs) have a major impact on morbidity and quality of life in ICD recipients. The recently introduced home monitoring of ICD devices is a promising new technique which remotely offers information about the status of the system. Stored intracardiac electrograms (IEGMs), which are essential for correct classification of appropriate and inappropriate ICD discharges, have until now not been available with ICD home monitoring on a day-by-day basis because of limitations of transferable data. We demonstrate the first compressed IEGMs daily transferable via home monitoring (IEGM-online). Validation of these electrograms will be performed in the Reliability of IEGM-Online Interpretation (RIONI) study. A total of 210 episodes of stored IEGMs will be collected by at least 12 European centres. The primary endpoint of this study is to investigate whether the IEGM-online based evaluation of the appropriateness of the ICDs therapeutic decision following episode detection is equivalent to the evaluation based on the complete ICD episode Holter extracted from the IEGM stored. The evaluation is independently done by an expert board of three experienced ICD investigators. The equivalence of the two methods is accepted if the evaluations yield a different conclusion for <10% of all evaluated IEGMs. The conclusion of the study is expected at the beginning of 2007. If RIONI successfully validates IEGMs transmitted via home monitoring, a strong basis for the use of this promising technique will be established.
Key Words: Implantable defibrillator, Home monitoring, Stored intracardiac electrogram
* Corresponding author. Tel: +49 2323 4991601; fax: +49 2323 499301. E-mail address: christian.perings{at}marienhospital-herne.de
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
J. C. Nielsen, H. Kottkamp, M. Zabel, E. Aliot, U. Kreutzer, A. Bauer, A. Schuchert, H. Neuser, B. Schumacher, H. Schmidinger, et al. Automatic home monitoring of implantable cardioverter defibrillators Europace, June 1, 2008; 10(6): 729 - 735. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stellbrink and H.-J. Trappe The follow-up of cardiac devices: what to expect for the future? Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I113 - I115. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Nagele, S. Hashagen, M. Azizi, S. Behrens, and M.A. Castel Analysis of terminal arrhythmias stored in the memory of pacemakers from patients dying suddenly Europace, June 1, 2007; 9(6): 380 - 384. [Abstract] [Full Text] [PDF] |
||||